Skip to main content
. 2024 Nov 17;25(22):12332. doi: 10.3390/ijms252212332

Table 1.

Validation parameters. Low-, medium- and high-QCs were 3, 8 and 16 ng/mL, respectively.

Analyte Linear Range
(ng/mL)
r2 LOD
(ng/mL)
LOQ
(ng/mL)
Accuracy
(%CV)
Inter-Day
Precision
(%CV)
Intra-Day
Precision
(%CV)
Recovery
(%)
Matrix Effect
(%)
lQC mQC hQC lQC mQC hQC lQC mQC hQC lQC mQC hQC lQC mQC hQC
Etazene 1–20 0.995 0.25 0.5 8 21 11 12 18 4 15 7 13 99 89 84 88 108 97
Flunitazene 1–20 0.997 0.25 0.5 10 9 8 15 12 9 12 9 18 95 110 105 80 110 95
Isotonitazene 1–20 0.993 0.35 0.5 17 20 12 8 16 11 18 11 12 91 117 92 108 93 96
Protonitazene 1–20 0.992 0.35 0.5 18 15 16 3 5 2 11 12 16 93 92 99 112 108 101

Abbreviations: %CV, percent coefficient of variation; hQC, high-QC; LOD, limit of detection; LOQ, limit of quantification; lQC, low-QC; mQC, medium-QC; r2, coefficient of determination.