Skip to main content
. 2024 Oct 29;15(11):321. doi: 10.3390/jfb15110321

Table 2.

Summary of contact lens parameters and property testing.

Sample Control
Parameters Diameter (mm) 13.90
(±0.033)
14.03
(±0.106)
Back Curve (mm) 8.30
(±0.091)
8.29
(±0.059)
Central Thickness (mm) 0.071 0.072
Properties Water content (%) 55 55
Luminous transmittance
(2°,%) 95 98
(10°,%) 95 98
UV transmittance
(UV-A,%) 14 16.8
(UV-B,%) 2.8 4.7
Refraction 1.404 1.406
Oxygen permeability Dk 61 64
Total Organic Carbon (µg/lens) 5.4 4.2