Skip to main content
. 2024 Nov 26;19(1):194. doi: 10.1186/s11671-024-04127-0

Table 2.

The performance comparison of electrochemical biosensors for detecting miRNA

Type Material Capture probe Detection limit Linear range References
Labeled biosensor AuNPs Ferrocene/methylene blue/cp-DNA 33 fM 100 fM–100 nM [37]
AuNPs/Graphitic carbon nitride nanosheets Ferrocene/DNAzyme walker 4.3/78.6 aM 100 aM–100 nM [38]
Magnetic Nanoparticles/ Biotin/cp-DNA 0.27 aM 1 aM–1 pM [39]
Label-free biosensor MXene Tetrahedral DNA nanostructures 5 pM 5 pM–10 nM [40]
poly(L-cysteine)/MoS2 probe DNA 78 fM 100 fM–1 nM [41]
ZIF-8/AuNPs 3D DNA walker 29 pM 0.1 nM–10 μM [42]
AuNPs/NSF cp-DNA 46.3 fM 0.1 pM–1 μM This work