Table 2. Validation Results of Accuracy and Precision for CP-III Concentrations in Human Plasmaa.

  nominal CP-III concentrations (ng/mL)
endogenous LLOQ sample low QC sample mid QC sample high QC sample
  0.01 + endo 0.03 + endo 15 + endo 40 + endo
within-run
1 mean (ng/mL) 0.0355 0.0438 0.0606 14.2 35.9
accuracy (%)   96.3 [81.3–103.3] 92.5 [87.0–97.7] 94.5 [89.5–101.0] 90.0 [87.6–92.8]
precision (%CV) 14.8 8.05 3.99 4.44 2.06
2 mean (ng/mL) 0.0422 0.0527 0.0678 13.7 37.1
accuracy (%)   101.0 [88.2–116.9] 94.0 [85.9–103.9] 91.3 [89.5–94.8] 92.7 [87.4–98.9]
precision (%CV) 9.43 9.22 6.22 1.90 4.32
3 mean (ng/mL) 0.0500 0.0522 0.0695 14.6 38.4
accuracy (%)   89.3 [81.7–91.7] 106.8 [87.5–96.3] 96.9 [90.0–107.0] 95.8 [86.5–108.4]
precision (%CV) 8.08 3.90 4.03 6.69 8.57
run-to-run
  accuracy (%)   95.5 92.1 94.2 92.9
precision (%CV) 17.4 11.0 8.21 5.49 6.64
a

CP-III, coproporphyrin-III; LLOQ, lower limit of quantitation; QC, quality control; Endo, endogenous level; CV, coefficient of variation.