Fig. 1. Modified Bland–Altman plots of WatchBP central systolic pressure compared against multiple reference methods.
Plots compare the difference between the cSBP estimates of the WatchBP and the reference cSBP values (Y-axis) against the reference cSBP (X-axis). Bias (blue dashed lines) and limits of agreement (black dotted lines) are shown. n = 11 patients with 25 individual comparisons for the left panel, n = 27 with 57 comparisons for the middle panel, and n = 39 with 78 comparisons for the right panel. The blue shaded box represents the range of acceptable bias values. The black bracket on the right-hand side of each plot represents the acceptable limits of agreement (LOA, given the measured bias). In the third panel, the filled black dots represent data from data from Group 1 and the unfilled black dots represent data from Group 2.