Table 2. Diagnostic performance of the point-of-care and laboratory high-sensitivity cardiac troponin I assays performed using the initial blood samples.

Population Assay Positive Negative MI Non-MI Sensitivity(95% CI) Specificity(95% CI) P value PPV(95% CI) NPV(95% CI)
Overall(N=585) POC 56 529 14 571 100% (76.8 to 100.0) 92.6% (90.2 to 94.7) < 0.001 25.0% (20.0 to 30.8) 100% (99.3 to 100.0)
Laboratory 82 503 14 571 100% (76.8 to 100.0) 88.1% (85.2 to 90.6) 17.1% (14.1 to 20.5) 100% (99.3 to 100.0)
Male(N=304) POC 35 269 9 295 100% (66.4 to 100.0) 91.2% (87.4 to 94.2) = 0.001 25.7% (19.3 to 33.3) 100% (98.6 to 100.0)
Laboratory 50 254 9 295 100% (66.4 to 100.0) 86.1% (81.6 to 89.8) 18.0% (14.2 to 22.6) 100% (98.6 to 100.0)
Female(N=281) POC 19 262 5 276 100% (47.8 to 100.0) 94.9% (91.6 to 97.2) < 0.001 26.3% (17.7 to 37.3) 100% (98.6 to 100.0)
Laboratory 43 238 5 276 100% (47.8 to 100.0) 86.2% (81.6 to 90.1) 11.6% (8.9 to 15.0) 100% (98.5 to 100.0)
Younger(<65 years)(N=248) POC 13 235 4 244 100% (39.8 to 100.0) 96.3% (93.1 to 98.3) = 0.289 30.8% (20.0 to 45.8) 100% (98.4 to 100.0)
Laboratory 17 231 4 244 100% (39.8 to 100.0) 94.7% (91.1 to 97.1) 23.5% (15.4 to 34.3) 100% (98.4 to 100.0)
Older(≥65 years)(N=337) POC 41 296 10 327 100% (69.2 to 100.0) 90.5% (86.8 to 93.5) < 0.001 24.4% (18.8 to 31.1) 100% (98.8 to 100.0)
Laboratory 76 261 10 327 100% (69.2 to 100.0) 79.8% (75.1 to 84.0) 13.2% (10.9 to 15.8) 100% (98.6 to 100.0)

The McNemar test was used to produce the P values given.

MI, myocardial infarction; N, number; NPV, negative predictive value; POC, point-of-care; PPV, positive predictive value