Table 2. Diagnostic performance of the point-of-care and laboratory high-sensitivity cardiac troponin I assays performed using the initial blood samples.
Population | Assay | Positive | Negative | MI | Non-MI | Sensitivity(95% CI) | Specificity(95% CI) | P value | PPV(95% CI) | NPV(95% CI) |
Overall(N=585) | POC | 56 | 529 | 14 | 571 | 100% (76.8 to 100.0) | 92.6% (90.2 to 94.7) | < 0.001 | 25.0% (20.0 to 30.8) | 100% (99.3 to 100.0) |
Laboratory | 82 | 503 | 14 | 571 | 100% (76.8 to 100.0) | 88.1% (85.2 to 90.6) | 17.1% (14.1 to 20.5) | 100% (99.3 to 100.0) | ||
Male(N=304) | POC | 35 | 269 | 9 | 295 | 100% (66.4 to 100.0) | 91.2% (87.4 to 94.2) | = 0.001 | 25.7% (19.3 to 33.3) | 100% (98.6 to 100.0) |
Laboratory | 50 | 254 | 9 | 295 | 100% (66.4 to 100.0) | 86.1% (81.6 to 89.8) | 18.0% (14.2 to 22.6) | 100% (98.6 to 100.0) | ||
Female(N=281) | POC | 19 | 262 | 5 | 276 | 100% (47.8 to 100.0) | 94.9% (91.6 to 97.2) | < 0.001 | 26.3% (17.7 to 37.3) | 100% (98.6 to 100.0) |
Laboratory | 43 | 238 | 5 | 276 | 100% (47.8 to 100.0) | 86.2% (81.6 to 90.1) | 11.6% (8.9 to 15.0) | 100% (98.5 to 100.0) | ||
Younger(<65 years)(N=248) | POC | 13 | 235 | 4 | 244 | 100% (39.8 to 100.0) | 96.3% (93.1 to 98.3) | = 0.289 | 30.8% (20.0 to 45.8) | 100% (98.4 to 100.0) |
Laboratory | 17 | 231 | 4 | 244 | 100% (39.8 to 100.0) | 94.7% (91.1 to 97.1) | 23.5% (15.4 to 34.3) | 100% (98.4 to 100.0) | ||
Older(≥65 years)(N=337) | POC | 41 | 296 | 10 | 327 | 100% (69.2 to 100.0) | 90.5% (86.8 to 93.5) | < 0.001 | 24.4% (18.8 to 31.1) | 100% (98.8 to 100.0) |
Laboratory | 76 | 261 | 10 | 327 | 100% (69.2 to 100.0) | 79.8% (75.1 to 84.0) | 13.2% (10.9 to 15.8) | 100% (98.6 to 100.0) |
The McNemar test was used to produce the P values given.
MI, myocardial infarction; N, number; NPV, negative predictive value; POC, point-of-care; PPV, positive predictive value