Table 3.

The outcomes of new‐generation devices

Outcomes NGDs On‐label devices Off‐label devices P **
ES (95% CI) I 2 (%) P * ES (95% CI) I 2 (%) P * ES (95% CI) I 2 (%) P *
30 day procedural outcomes
Device success

0.945

(0.913–0.971)

76.8 0.000

0.978

(0.964–0.989)

8.2 0.358

0.899

(0.848–0.941)

10.3 0.350 <0.001
Permanent pacemaker implantation

0.088

(0.061–0.119)

57.0 0.000

0.069

(0.046–0.095)

40.0 0.041

0.184

(0.132–0.242)

0.0 0.928 <0.001
Conversion to SAVR

0.022

(0.009–0.038)

0.0 0.981

0.025

(0.012–0.042)

0.0 0.957 NA NA NA NA
Annulus rupture

0.002

(0.000–0.017)

0.0 0.941 NA NA NA NA NA NA NA
Reintervention

0.023

(0.007–0.045)

13.9 0.324 NA NA NA

0.028

(0.000–0.114)

54.6 0.111 NA
Greater‐than‐mild PVL

0.012

(0.004–0.022)

0.0 0.713

0.009

(0.002–0.019)

0.0 0.942

0.038

(0.012–0.074)

0.0 0.611 0.003
Mild PVL

0.209

(0.176–0.244)

12.8 0.304

0.203

(0.165–0.243)

19.0 0.241 NA NA NA NA
None/trace PVL

0.774

(0.708–0.835)

71.3 0.000

0.780

(0.705–0.847)

73.4 0.000 NA NA NA NA
30 day clinical outcomes
30 day mortality

0.042

(0.027–0.059)

43.8 0.008

0.026

(0.013–0.043)

22.1 0.192

0.051

(0.016–0.102)

30.4 0.219 0.006
1 year clinical outcomes
1 year mortality

0.081

(0.051–0.117)

67.3 0.001

0.059

(0.035–0.087)

23.3 0.251 NA NA NA NA

CI, confidence interval; ES, effect size; I 2, the variation attributable to heterogeneity; NA, not available; NGDs, new‐generation devices including on‐label devices and off‐label devices; PVL, paravalvular leak; SAVR, surgical aortic valve replacement.

*

P value for significance of I 2.

**

P value for significance of the χ 2 or Fisher's exact test between off‐label and on‐label devices.