Skip to main content
. 2024 Jan 22;10(3):e24952. doi: 10.1016/j.heliyon.2024.e24952

Table 2.

Experimental results of the pull-out tests where CoV is the Coefficient of Variation.

Pull-out T = 25 Ncm

Label Fmax [N] CoV ksec [N/mm] CoV

BENEfit4plusV_2_7 303.65 0.09 134.26 0.09
BENEfit4plusV_2_9 383.08 0.07 199.56 0.10
BENEfit4plusV_2_11 576.19 0.09 203.25 0.10
MSE_OAS_T1511_1.5_13.1 446.94 0.05 144.41 0.09
MSE_OAS_T1511_1.5_15.1 609.12 0.06 134.10 0.12
MSE_OAS_T1511_1.8_13.3 649.54 0.12 169.50 0.05
MSE_OAS_T1511_1.8_15.1 632.40 0.14 197.46 0.08
Pull-out T = 30 Ncm
Label Fmax [N] CoV ksec [N/mm] CoV

BENEfit4plusV_2_7 423.64 0.14 96.78 0.07
BENEfit4plusV_2_9 533.80 0.12 189.56 0.08
BENEfit4plusV_2_11 597.90 0.08 172.52 0.08
MSE_OAS_T1511_1.5_13.1 546.41 0.14 193.23 0.15
MSE_OAS_T1511_1.5_15.1 648.51 0.14 232.24 0.15
MSE_OAS_T1511_1.8_13.3 631.39 0.09 275.29 0.06
MSE_OAS_T1511_1.8_15.1 683.27 0.12 245.94 0.10
Pull-out T = 35 Ncm
Label Fmax [N] CoV ksec [N/mm] CoV

BENEfit4plusV_2_7 390.35 0.09 130.17 0.08
BENEfit4plusV_2_9 546.19 0.10 172.44 0.05
BENEfit4plusV_2_11 684.58 0.12 164.40 0.11
MSE_OAS_T1511_1.5_13.1 629.22 0.12 235.55 0.14
MSE_OAS_T1511_1.5_15.1 741.01 0.08 200.67 0.12
MSE_OAS_T1511_1.8_13.3 791.19 0.11 290.55 0.14
MSE_OAS_T1511_1.8_15.1 736.38 0.10 248.78 0.11