Skip to main content
. 2024 Dec 2;24(23):7704. doi: 10.3390/s24237704

Figure 7.

Figure 7

Actuation pad dimensions. The solenoid plunger diameter dictated the actuation pad geometry. Three designs were tested to determine the geometry allowing the best channel closure when not in use. Design (A) provided a dead space and permitted fluidic flow even when closed. Design (B) did not allow the closure of the valve. Design (C) was chosen due to its superior valve closure capabilities.