Table 4.

Comparative result of various methods using the ECG or BCG as the proximal reference, and IPG or PPG as the distal reference for the cuffless blood pressure measurement.

Reference PTT Signals (Sensor Placement) PCC
[39] BCG (foot) and BPW (finger) SBP: 0.70
DBP: 0.66
NA
[40] BCG (chair) SBP: 0.755
DBP: 0.532
ME
SBP: 0.93 ± 6.24 mmHg
DBP: 0.21 ± 5.42 mmHg
[41] BCG (foot) and PPG (foot) NA RMSE
SBP: 11.8 ± −1.6 mmHg
DBP: 7.6 ± −0.5 mmHg,
[42] ECG and IPG (arm) SBP: 0.700
DBP: 0.450
NA
[15] BCG (foot) and PPG (foot) SBP: 0.775
DBP: 0.532
RMSE
SBP: 6.7 ± 1.6 mmHg
DBP: 4.8 ± 1.47 mmHg
[16] BCG (foot) and IPG (foot) SBP: 0.754
DBP: 0.533
RMSE
SBP: 7.3 ± 2.1 mmHg
DBP: 4.5 ± 1.8 mmHg
Proposed method:
Server-based Computing
BCG (foot) and IPG (foot) SBP: 0.80
DBP: 0.81
ME
SBP: 2.64 ± 9.71 mmHg
DBP: 1.52 ± 6.32 mmHg
Proposed method:
Edge Computing
BCG (foot) and IPG (foot) SBP: 0.73
DBP: 0.78
ME
SBP: 2.2 ± 10.9 mmHg
DBP: 1.87 ± 6.79 mmHg