Reliability, biological variability, and accuracy of multi-frequency bioelectrical impedance analysis for measuring body composition components

David P Looney 1,2,3, Erica A Schafer 1,4, Christopher L Chapman 1,4, Riana R Pryor 5, Adam W Potter 1,*, Brandon M Roberts 1, Karl E Friedl 1
PMCID: PMC11649400  PMID: 39691170

Abstract

Introduction

Bioelectrical impedance analysis (BIA) systems are gaining popularity for use in research and fitness assessments as the technology improves and becomes more affordable and easier to use. Multifrequency BIA (MF-BIA) may improve accuracy and precision using octopolar contacts for segmental analyses.

Purpose

Evaluate reliability, biological variability, and accuracy of component measures (total body water, mass, and composition) of commercially available MF-BIA system (InBody 770, Cerritos, California, USA).

Methods

Fourteen healthy military-age adults were assessed by MF-BIA in duplicate on five laboratory visits across 3 weeks (10 measures each). Participants were evaluated at the same time of day after refraining from strenuous exercise (> 48 h), alcohol consumption (> 24 h), and caffeine, nicotine, and food (> 10 h). Systematic error (test–retest reliability) and biological variability (day-to-day reliability) were summarized by intraclass correlation coefficient (ICC) values determined for body mass (fat, fat-free, total) and body water (extracellular, intracellular, total). Body composition measurements derived from BIA on the second visit were also tested for accuracy compared to dual-energy x-ray absorptiometry (DXA).

Results

Test–retest reliability was very high for all measurements of whole-body water and mass (ICC ≥ 0.999) and high for regional body water and mass (ICC 0.973–1.000). Biological variability was observable with very minor differences between tests (same day) for total and regional body water (0.0–0.2 L) and total and regional body mass measurements (0.0–0.2 kg); while between day differences were slightly higher (0.0–0.5 L and 0.1–0.7 kg). Compared to DXA, the MF-BIA whole-body measures showed an offset in %BF (Bias −4.0 ± 2.8%; Standard error of the estimate (SEE), 2.6%), an overprediction for total body fat-free mass (Bias 2.8 ± 2.1 kg; SEE 2.2 kg) and an underprediction of total body fat mass (Bias −2.9 ± 2.0 kg; SEE 1.9 kg).

Conclusion

Under controlled conditions with fit and healthy men and women, this MF-BIA system has high methodological reliability and demonstrates stable day-to-day measurements of major body composition components. Previously reported ~3% body fat offset compared to criterion methods was again confirmed. Precision of the InBody 770 shows consistency and supports further testing of this specific device as a new military standards method and suitability across a wider range of %BF.

Keywords: body composition, BMI, FMI, DXA, body water, BIA

Introduction

Bioelectrical impedance (BIA) is an appealing technology for practical and affordable everyday assessment of body composition across many domains, including preventive medicine, athletics, and assessing compliance with military body fat standards. Interest in electrical impedance technologies has soared in the past two decades as evidenced by the number of publications in this field (1). The cost and time associated with measuring body composition with BIA is considerably less than the criterion methods such as dual-energy x-ray absorptiometry (DXA) or the previously used hydrodensitometry (2). In 1994, the National Institutes of Health convened a Technology Assessment meeting on BIA that was reviewed by an expert extramural panel. The panel concluded that “BIA provides a reliable estimate of total body water (TBW) under most conditions [and] it can be a useful technique for body composition analysis in healthy individuals…” (3). However, the leading researchers involved in this early consensus also noted that BIA is affected by factors such as body position, hydration and prandial state, skin temperature, and recent physical activity (4, 5). Since this time, considerable improvements in the BIA technology have been made, which increases the interest in using BIA to replace current reliance on anthropometric estimates of adiposity (e.g., body mass index (BMI), skinfold thicknesses, body circumferences) with a method that is equitable across sexes and has better accuracy and precision at the individual level. Compared to anthropometric methods such as body circumferences and skinfold thicknesses, a method such as BIA which does not involve assessment of different fat sites in men and women can also be viewed as more equitable, and the ability of a user to obtain their own reliable measurements without contact with an observer increases user acceptability. Biophysically-based strategies such as BIA also hold out the promise of greater precision free from observer error; the US Army briefly used the valid and accurate Durnin & Womersley equations for skinfold thicknesses but then abandoned the method because of the substantial interobserver variation when applied at scale (i.e., the entire US Army) (6).

The original concept of BIA was to use stature2/resistance of a cylindrical model of the human body to predict body water and fat-free mass. These assessments were based on measurement of resistance to 50 kHz of current applied across the length of the connected series of cylinders (i.e., between hand and foot on one side of the body). This frequency passed through cell membranes and assessed total body water (7). Subsequent advances include octopolar assessments with two sensors on each hand and foot, and analysis across multiple frequencies (MF-BIA) or frequency sweep (bioelectrical impedance spectroscopy, BIS) (1, 8). Most devices are standardized for standing or prone body positions and studies have generally incorporated a standardized period of adjustment with postural changes (e.g., >15 min). Further improvements such as segmental analyses may reduce error from limb length assumptions, and most commercially available systems standardize posture (usually standing) and limb placement. The estimations of regional fluid and cell mass is still largely empirical and Ward has highlighted the need for a mechanistic understanding of impedance and phase angle measurement associations (9).

As part of a large-scale body composition survey of the U.S. Marine Corps, Potter et al. (10) determined that a commercially available BIA system provided similar results to DXA when a + 3% body fat correction was applied across a range of adiposity in free living conditions. However, this +3% body fat correction has yet to be cross-validated in another population of healthy, physically active, military-age women and men. Recently, the U.S. Army, Army National Guard, and Marine Corps have acquired and distributed BIA devices to be used by service members and reservists (11, 12). Given the importance of body composition standards in these occupations to physical readiness, the reliability and accuracy of these modern technologies must be evaluated for mass use beyond research applications. Furthermore, while BMI has historically been relied upon for identifying overweight and obese individuals, its usefulness compared to other methods such as the use of percent body fat has been called into question (13). Therefore, understanding the reliability and accuracy of modern BIA devices serves to help a broad range of the population and, with further testing, could replace BMI in clinical weight management.

In the present study, we assessed “best performance” using reliability and accuracy metrics of the technology under optimal controlled laboratory conditions to estimate percent body fat (%BF) and body cell mass (intracellular fluid and solids). Subsequent assessments can then consider the most important factors that must be controlled outside of the laboratory to ensure meaningful measurements. In this study, we tested a MF-BIA device that has been widely used in field research for %BF estimations. Furthermore, we examined accuracy of the MF-BIA compared to full body DXA, including segmental estimations of body composition, and we considered device and biological variability with repeated measurements on five separate days for each participant.

Current technology and prediction models may have improved enough to be able to justify the use of BIA for practical and affordable everyday assessments in preventive medicine, obesity prevention, fitness guidance, and for assessment of compliance with military body fat standards. This could replace current reliance on anthropometric estimates of adiposity (e.g., BMI, % body fat predictive equations, circumference tape methods) with a method that is equitable across sexes and has better accuracy and precision at the individual level. Accurate estimates of body measurements are also valuable for calculating derived metrics such as the various body mass indices. While BMI has historically been relied upon for identifying overweight and obese individuals, its usefulness compared to other methods such as the use of percent body fat has been called into question (13).

Methods

Design

We performed a repeated measures design study to evaluate the test–retest reliability and day-to-day variability of measurements of regional and whole-body water, mass, and composition from a commercially available system (InBody 770, Cerritos, California, USA). Study participants completed BIA tests in duplicate on five laboratory visits scheduled over 19 ± 6 days. Test–retest reliability was examined by comparing measurements between duplicate tests on each visit while day-to-day variability was assessed using data collected across the five visits. Additionally, BIA-derived body mass and composition measurements collected on the second visit were compared against criterion data measured by DXA. We also evaluated whether adding the +3% correction to BIA-derived %BF from Potter et al. (10) improved accuracy and concordance with DXA measurements.

Participants

Fourteen healthy military-age adults (age, 24 ± 6 years old; height, 172 ± 8 cm; body mass, 72.4 ± 15.9 kg) participated in the current study including 3 women (age, 21 ± 3 years old; height, 167 ± 7 cm; body mass, 62.4 ± 9.9 kg) and 11 men (age, 25 ± 6 years old; height, 173 ± 8 cm; body mass, 75.2 ± 16.4 kg). The fourteen participants included 11 active-duty U.S. Army Soldiers and 3 civilian men. Participants were recruited from the Natick Human Research Volunteer (HRV) Pool, U.S. Army Natick Soldier System Center (NSSC) active-duty military personnel, active-duty population located at other military organizations, as well as the federally and non-federally employed civilian population. Participants were required to be between 18 and 44 years old, at least recreationally active (aerobic or resistance exercise for at least 30 min on at least 2 days per week), and free of any musculoskeletal injuries, illnesses, or medical conditions that compromise the ability to exercise. Participants gave their voluntary written informed consent after a briefing on the purpose of the study and potential risks. This study was approved by the Institutional Review Board (IRB) at the U.S. Army Medical Research and Development Command (MRDC; Fort Detrick, MD, USA). Investigators adhered to Department of Defense Instruction 3216.02 and 32 CFR 219 on the use of volunteers in research.

Procedures

Participants arrived at the laboratory in the same physical training uniform (e.g., shorts, t-shirt, socks, compression garment) at the same time of day (0700) after refraining from strenuous exercise (> 48 h) and alcohol consumption (> 24 h) as well as avoid caffeine, nicotine, food, over the counter medications, and dietary supplements (> 10 h). All metallic items (jewelry, etc.) were removed prior to testing. There was no attempt to control for menstrual cycle or contraceptive hormone use in the women. To ensure proper hydration, participants were asked to drink at least 500 mL water the night before as well as the morning of each visit. Participants provided a urine sample that was assessed for hydration status via urine specific gravity (USG) (<1.030) using a standard refractometer. No additional water was consumed within 1 h of the start of body composition testing. Height was measured using a laboratory stadiometer on the first visit.

Bioelectrical impedance analysis (BIA) was conducted using a commercially available system (InBody 770, Cerritos, California, USA). The InBody 770 system uses a multi-frequency assessment method (1, 5, 50, 250, 500 and 1,000 kHz) with proprietary algorithms apparently based on a sum of segments approach (14). For each test, the participant stood barefoot on the footplate with heels on the rear sole electrodes for an initial body mass measurement. Participants stood still for the duration of the BIA assessment with arms positioned straight and away from the body, thighs not in contact with one another, and hands gripping the hand electrode so that the thumb was placed on the oval electrode and the four fingers wrapped the surface of the bottom hand electrode. In addition to whole body measurements, we recorded the regional measurements reported by the InBody 770 system, which include arms (left and right arms combined), legs (left and right legs combined), and trunk. We also reported the remainder or difference between each whole-body measurement and the sum of its associated arm, leg, and trunk measurements.

On the second visit only, study participants were assessed via DXA (iDXA, GE Lunar, Madison, Wisconsin, USA). Participants laid still in the supine position within the outlined assessment area of the DXA table for a whole-body scan that lasted ~10 min. Similar to BIA, we recorded whole body and regional measurements (arms, legs, trunk, and remainder) for the DXA; these regional assessments were based on the default regions of interest in the software. All DXA scans were performed by the same cross-trained team members and required DXA training and certification to meet our rigorous laboratory credentialing standard. The coefficient of variation for same day repeated measures of percent body fat with the GE iDXA has been shown to be 1.06% (15).

Statistical analyses

All data were analyzed using R (Version 4.3.3; R Foundation for Statistical Computing; Vienna, Austria) (16) and reported as mean ± standard deviation (SD) unless specified otherwise. For each outcome, we determined the intraclass correlation coefficient (ICC) (17) for test–retest reliability by partitioning out variance components using a linear mixed-effects model (18) that included random effects of visit within participant on intercepts. We compared body composition measurements from both BIA tests on the second visit against the criterion DXA using a linear mixed-effects model that included random effects of participant on intercepts. The agreement between BIA and DXA was evaluated quantitatively based on the bias (mean difference, BIA − DXA), standard deviation (SD) of differences, standard error of the estimate (SEE), Pearson correlation coefficient (r), and concordance correlation coefficient (CCC). We also included Bland–Altman plots with 95% limits of agreement (95% LoA) to visually depict the agreement between the two systems with the x-axes represented as the mean of paired measurements (19, 20).

Results

Regional and whole-body water components measured by BIA had exceptional test–retest reliability (ICC ≥ 0.973) (Table 1). Test–retest reliability of body mass measurements was nearly perfect for all compartments and regions (ICC ≥ 0.995) except for fat mass of the remainder, which was still excellent (ICC = 0.949) (Table 2). Test–retest reliability was also nearly perfect for BIA-measured percent body fat as well as BMI, FMI, and fat free mass index (FFMI) (ICC ≥ 0.998) (Table 3).

Table 1.

Test–retest reliability and variability of regional and whole-body water measurements (L) by bioelectrical impedance analysis (BIA).

Variability between
Compartment Region Mean SD Participant Day Test Range ICC
Intracellular Arms 3.2 0.8 0.8 0.0 0.0 0.2 ± 0.1 0.999
Legs 8.7 1.6 1.6 0.1 0.0 0.2 ± 0.1 1.000
Trunk 12.8 2.5 2.5 0.1 0.1 0.4 ± 0.1 0.999
Remainder 2.9 0.7 0.7 0.1 0.1 0.4 ± 0.2 0.985
Whole 27.5 5.3 5.3 0.2 0.1 0.7 ± 0.3 1.000
Extracellular Arms 1.9 0.5 0.5 0.0 0.0 0.1 ± 0.0 0.999
Legs 5.0 0.9 0.9 0.1 0.0 0.2 ± 0.1 1.000
Trunk 7.4 1.3 1.3 0.1 0.1 0.3 ± 0.1 0.998
Remainder 1.7 0.4 0.4 0.1 0.1 0.3 ± 0.1 0.973
Whole 16.0 2.9 2.9 0.2 0.1 0.6 ± 0.2 0.999
Total Arms 5.1 1.3 1.3 0.1 0.0 0.3 ± 0.1 0.999
Legs 13.7 2.5 2.5 0.1 0.0 0.4 ± 0.1 1.000
Trunk 20.2 3.8 3.8 0.2 0.1 0.6 ± 0.2 0.999
Remainder 4.6 1.0 1.0 0.2 0.1 0.6 ± 0.3 0.991
Whole 43.6 8.1 8.1 0.5 0.2 1.3 ± 0.5 1.000

ICC, intraclass correlation coefficient for test–retest reliability within day; Range, difference between maximum and minimum measurement within each individual across the ten tests; Remainder, Whole – (Arms + Legs + Trunk); SD, standard deviation.

Table 2.

Test–retest reliability and variability of regional and whole-body mass measurements (kg) by bioelectrical impedance analysis (BIA).

Variability between
Compartment Region Mean SD Participant Day Test Range ICC
Fat Arms 1.4 1.1 1.1 0.1 0.1 0.3 ± 0.1 0.997
Legs 3.8 1.8 1.8 0.2 0.1 0.5 ± 0.3 0.998
Trunk 6.9 4.5 4.5 0.3 0.1 0.9 ± 0.3 0.999
Remainder 1.2 0.3 0.2 0.2 0.1 0.4 ± 0.5 0.949
Whole 13.3 7.4 7.3 0.5 0.2 1.5 ± 0.6 0.999
Fat-Free Arms 6.6 1.7 1.7 0.1 0.0 0.3 ± 0.1 0.999
Legs 17.6 3.2 3.2 0.2 0.0 0.5 ± 0.2 1.000
Trunk 26.1 4.9 4.9 0.3 0.1 0.8 ± 0.3 0.999
Remainder 9.3 1.9 1.9 0.2 0.1 0.7 ± 0.4 0.996
Whole 59.6 11.1 11.1 0.6 0.2 1.8 ± 0.7 1.000
Total Arms 8.0 2.3 2.2 0.1 0.1 0.3 ± 0.1 1.000
Legs 21.5 4.1 4.1 0.3 0.1 0.8 ± 0.4 1.000
Trunk 32.9 8.1 8.1 0.3 0.1 0.9 ± 0.3 1.000
Remainder 10.4 2.0 2.0 0.3 0.1 0.9 ± 0.4 0.995
Whole 72.9 15.9 15.9 0.7 0.0 1.7 ± 0.7 1.000

ICC, intraclass correlation coefficient for test–retest reliability within day; Range, difference between maximum and minimum measurement within each individual across the ten tests; Remainder, Whole – (Arms + Legs + Trunk); SD, standard deviation.

Table 3.

Test–retest reliability and variability of key bioelectrical impedance analysis (BIA) measurements.

Variability between
Measurement Mean SD Participant Day Test Range ICC
Body fat (% body mass) 17.6 7.4 7.3 0.6 0.3 1.9 ± 0.9 0.998
Body mass index (kg/m2) 24.5 4.0 4.0 0.2 0.0 0.6 ± 0.2 1.000
Dry lean mass (kg) 16.0 3.0 3.0 0.2 0.1 0.5 ± 0.2 0.999
Extracellular water (% TBW) 36.8 0.6 0.5 0.2 0.0 0.5 ± 0.2 0.994
Fat mass index (kg/m2) 4.5 2.3 2.3 0.2 0.1 0.5 ± 0.2 0.999
Fat-free mass index (kg/m2) 20.0 2.6 2.6 0.2 0.1 0.6 ± 0.3 0.999
Skeletal muscle mass (kg) 33.9 6.9 6.9 0.3 0.1 1.0 ± 0.4 1.000
Visceral Fat Area (cm2) 52.7 34.2 34.1 2.5 1.1 7.0 ± 2.9 0.999

ICC, intraclass correlation coefficient for test–retest reliability within day; Range, difference between maximum and minimum measurement within each individual across the ten tests; Remainder, Whole – (Arms + Legs + Trunk); SD, standard deviation; TBW, total body water.

When compared to DXA, BIA typically measured lower regional fat, fat-free, and total mass except for fat-free mass measured at the trunk (Bias, 0.4 kg) and all mass measured for the remainder (Bias ≥0.3 kg) (Table 4). Percent body fat was consistently lower when measured by BIA versus DXA but strongly correlated (Bias ± SD, −4.0 ± 2.8%; SEE, 2.6%; r, 0.932; CCC, 0.819) (Figure 1). Adding the 3% body fat correction to BIA measurements from Potter et al. (10) increased both accuracy and concordance (Bias ± SD, −1.0 ± 2.8%; SEE, 2.6%; r, 0.932; CCC, 0.920). Agreement between BIA and DXA was good for FMI (Bias ± SD, −1.0 ± 0.7 kg∙m2; SEE, 0.6 kg∙m2; r, 0.961; p < 0.001; CCC, 0.886) and FFMI (Bias ± SD, 0.9 ± 0.7 kg∙m2; SEE, 0.7 kg∙m2; r, 0.957; p < 0.001; CCC, 0.896) but nearly perfect for body mass index (Bias ± SD, 0.0 ± 0.1 kg∙m2; SEE, 0.1 kg∙m2; r, 1.000; p < 0.001; CCC, 1.000) (Figure 2).

Table 4.

Accuracy of bioelectrical impedance analysis (BIA) mass measurements (kg) relative to dual-energy x-ray absorptiometry (DXA).

Compartment Region Bias ± SD
(kg)
SEE
(kg)
Bias ± SD
(%)
r p CCC
Fat Arms −0.4 ± 0.6 0.7 −44.9 ± 46.8 0.805 < 0.001 0.731
Legs −1.5 ± 1.3 1.0 −33.0 ± 26.5 0.838 < 0.001 0.644
Trunk −1.3 ± 1.1 1.0 −25.3 ± 36.1 0.973 < 0.001 0.935
Remainder 0.3 ± 0.3 0.3 24.2 ± 21.6 0.169 0.534 0.063
Whole −2.9 ± 2.0 1.9 −22.3 ± 18.4 0.966 < 0.001 0.896
Fat-Free Arms −0.4 ± 0.5 0.4 −6.0 ± 7.1 0.968 < 0.001 0.930
Legs −2.5 ± 1.8 1.2 −12.6 ± 7.0 0.922 < 0.001 0.730
Trunk 0.4 ± 2.3 2.4 1.4 ± 8.8 0.876 < 0.001 0.863
Remainder 5.3 ± 1.5 1.4 80.2 ± 11.7 0.676 0.006 0.033
Whole 2.8 ± 2.1 2.2 4.9 ± 3.7 0.980 < 0.001 0.947
Total Arms −0.9 ± 0.9 0.8 −10.6 ± 8.3 0.923 < 0.001 0.859
Legs −4.0 ± 2.0 1.3 −16.6 ± 5.8 0.949 < 0.001 0.681
Trunk −0.8 ± 2.3 2.3 −3.0 ± 6.8 0.957 < 0.001 0.950
Remainder 5.6 ± 1.6 1.4 72.7 ± 11.2 0.686 0.005 0.039
Whole −0.1 ± 0.3 0.3 −0.1 ± 0.4 1.000 < 0.001 1.000

CCC, concordance correlation coefficient; p, p-value for Pearson’s correlation coefficient; r, Pearson’s correlation coefficient; SD, standard deviation; SEE, standard error of the estimate.

Figure 1.

Figure 1

Bland–Altman plots of agreement between percent body fat measured by MF-BIA and DXA with no correction and + 3% correction (10) to MF-BIA. Black dashed line, linear regression line of best fit; Blue solid line, bias; DXA, dual-energy x-ray absorptiometry; MF-BIA, Multi-frequency Bioelectrical Impedance Analysis; red solid lines, 95% limits of agreement.

Figure 2.

Figure 2

Bland–Altman plots of agreement between fat mass index (FMI), fat-free mass index (FFMI), and body mass index (BMI) measured by BIA and DXA. Black dashed line, linear regression line of best fit; Blue solid line, bias; DXA, dual-energy x-ray absorptiometry; MF-BIA, Multi-frequency Bioelectrical Impedance Analysis; red solid lines, 95% limits of agreement.

Discussion

Our study demonstrates the high test-rest reliability and relatively low day-to-day variability of the InBody 770 BIA system when measuring common regional and whole-body water, mass, and composition measurements in healthy military-age adults across five controlled laboratory visits. When compared to DXA, BIA tended to measure lower values for arm, leg, and trunk mass as well as lower fat mass and higher fat-free mass for the whole body. Although BIA-derived percent body fat was systematically lower than when measured by DXA, adding the +3% body fat correction from Potter et al. (10) improved accuracy and concordance. Esco et al. (21) found that a closely related system from the same manufacturer (InBody 720) also underestimated percent body fat by 3.33% and overestimated fat free mass (FFM) by 2.12 kg in college female athletes when compared with DXA (21). Mecherques-Carini et al. (22) replicated our findings in a recent study by also finding an approximate 3% bias in body fat percentage as measured by BIA with a different commercially-available, segmental multifrequency BIA device with eight electrodes (MC-780-MA, Tanita Corporation, Tokyo, Japan) compared to DXA in young adults following an overnight fast. Another study comparing a SECA device (SECA 515) to iDXA also demonstrated an underestimation of fat mass and overestimation of fat-free mass (23).

Previous studies have also shown the InBody 770 tends to underestimate %BF (3.0 to 3.4%) while overestimating fat-free mass (2.2–2.7 kg) (10, 24–27). This appears to be a systematic error related to the predictive software. The lower arm, leg, and trunk masses measured by BIA versus DXA could be attributed to the different methodology used by each system for segmentation of body regions. Full-body DXA scans are generally separated into regions of interest automatically based on positioning of anatomical landmarks on a two-dimensional reconstructed image of the individual (28) whereas BIA estimates regional measurements based on empirical equations that incorporate sex, height, and body mass (29). Consequently, the proprietary algorithms employed by the systems examined in this study may also use different anatomical landmarks for separating body regions. Additionally, it is possible that the empirical equations used by the InBody 770 rely on inaccurate assumptions of body proportions for a modern population of healthy, physically active, military-age adults. This could be determined by a follow-up study comparing disagreements in BIA and DXA-derived body mass measurements between groups of individuals with extreme body types (high vs. low armspan-to-height ratio, etc.). Ultimately, what is made abundantly clear by the consistently higher remainder measured by the InBody 770 for each mass compartment is that this BIA system attributes considerably less mass to the arms, legs, and trunk than the DXA system. BMI and waist circumference were important predictors of the variance between MF-BIA and DXA with an earlier version of the InBody octopolar system (8).

There are several methodological considerations of the present study that warrant discussion. First, this study was designed to maximize internal validity and experimental control. Depending on the measurement, we demonstrated excellent reliability with BIA measurement in this tightly controlled experiment. These results may not be applicable to other types of BIA systems. Further research is required to assess how reliability is impacted in free-living applications such as for military standards and in fitness facilities, or by circadian rhythm influences over the time of day. While we recruited both women and men for the present study, we are limited by the small number of women (n = 3) who participated and who are also underrepresented in the military services.

To our knowledge, this is the first study to evaluate test–retest reliability of the InBody 770 BIA system over five visits within a relatively short period of time (19 ± 6 days). The consistency of the body water, mass, and composition measurements within individuals supports the use of the InBody 770 for longitudinal monitoring applications. An uncontrolled field study with Inbody measurements on separate days will provide a snapshot of biological variability produced by prandial, exercise, and hydration factors which were tightly controlled in this study. In this study, there was no attempt to control for phase of the menstrual cycle with potential changes in total body water; however, the biological variability in this limited sample of women was not markedly different than the ICC for the men.

Conclusion

In tightly controlled conditions InBody 770 BIA system provides exceptionally reliable measurements of common regional and whole-body water, mass, and composition measurements in healthy miliary-age adults under controlled laboratory conditions. Users can add the +3% body fat correction from Potter et al. to their InBody measurements to obtain closer results to DXA.

Acknowledgments

This research was supported in part by appointments to the Department of Defense (DOD) Research Participation Program administered by the Oak Ridge Institute for Science and Education (ORISE) through an interagency agreement between the US Department of Energy (DOE) and the DOD. ORISE is managed by ORAU under DOE contract number DE-SC0014664. All opinions expressed in this paper are the author’s and do not necessarily reflect the policies and views of DOD, DOE, or ORAU/ORISE.

Funding Statement

The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. Funding for this work has been provided by US Army Medical Research and Development Command (USAMRDC), Military Operational Medicine Research Program (MOMRP).

Data availability statement

The datasets presented in this article are available by the authors upon reasonable request.

Ethics statement

The studies involving humans were approved by Institutional Review Board (IRB) at the U.S. Army Medical Research and Development Command (MRDC; Fort Detrick, MD, USA). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

DL: Writing – review & editing, Writing – original draft, Visualization, Validation, Supervision, Resources, Project administration, Methodology, Investigation, Funding acquisition, Formal analysis, Data curation. ES: Writing – review & editing, Writing – original draft, Methodology, Formal analysis, Data curation. CC: Writing – review & editing, Writing – original draft, Methodology, Investigation, Formal analysis, Data curation. RP: Writing – review & editing, Writing – original draft, Methodology, Formal analysis. AP: Writing – review & editing, Writing – original draft, Project administration, Investigation, Formal analysis, Data curation, Conceptualization. BR: Writing – review & editing, Writing – original draft, Methodology, Investigation, Formal analysis, Data curation. KF: Writing – review & editing, Writing – original draft, Supervision, Methodology, Investigation, Formal analysis, Conceptualization.

Conflict of interest

DL was employed by Maximize Human Performance, LLC and CoachMePlus.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Author disclaimer

The views expressed in this work are those of the authors and do not reflect the official policy of the Department of Army, Department of Defense, US Government, or CoachMePlus.

References

  • 1.Ward LC. Electrical bioimpedance: from the past to the future. J Electr Bioimpedance. (2021) 12:1–2. doi: 10.2478/joeb-2021-0001, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Buskirk ER. Underwater weighing and body density: a review of procedures In: Brozek J, Henschel A, editors. Techniques for measuring body composition. National Academy of Sciences: Washington DC; (1961). 90–106. [Google Scholar]
  • 3.National Institutes of Health . Bioelectrical impedance analysis in body composition measurement: National Institutes of Health technology assessment conference statement. Washington DC: US Department of Health and Human Services; (1994). [Google Scholar]
  • 4.Ellis KJ, Bell SJ, Chertow GM, Chumlea WC, Knox TA, Kotler DP, et al. Bioelectrical impedance methods in clinical research: a follow-up to the NIH technology assessment conference. Nutr. (1999) 15:874–80. doi: 10.1016/S0899-9007(99)00147-1 [DOI] [PubMed] [Google Scholar]
  • 5.Chumlea WC, Guo SS. Bioelectrical impedance: a history, research issues, and recent consensus In: Emerging technologies for nutrition research - potential for assessing military performance capability. Committee on military nutrition research. eds. Carlson-Newberry SJ, Costello RB. Washington DC: National Academy Press; (1997). 169–92. [PubMed] [Google Scholar]
  • 6.Friedl KE, DeWinne CM, Taylor RL. The use of the Durnin-Womersley generalized equations for body fat estimation and their impact on the Army weight control program. Mil Med. (1987) 152:150–5. PMID: [PubMed] [Google Scholar]
  • 7.Lukaski HC, Johnson PE, Bolonchuk WW, Lykken GI. Assessment of fat-free mass using bioelectrical impedance measurements of the human body. Am J Clin Nutr. (1985) 41:810–7. doi: 10.1093/ajcn/41.4.810 [DOI] [PubMed] [Google Scholar]
  • 8.Shafer KJ, Siders WA, Johnson LK, Lukaski HC. Validity of segmental multiple-frequency bioelectrical impedance analysis to estimate body composition of adults across a range of body mass indexes. Nutr. (2009) 25:25–32. doi: 10.1016/j.nut.2008.07.004 [DOI] [PubMed] [Google Scholar]
  • 9.Ward LC, Brantlov S. Bioimpedance basics and phase angle fundamentals. Rev Endocr Metab Disorders. (2023) 24:381–91. doi: 10.1007/s11154-022-09780-3, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Potter AW, Nindl LJ, Soto LD, Pazmino A, Looney DP, Tharion WJ, et al. High precision but systematic offset in a standing bioelectrical impedance analysis (BIA) compared with dual-energy X-ray absorptiometry (DXA). BMJ Nutr Prev Health. (2022) 5:254–62. doi: 10.1136/bmjnph-2022-000512 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.United States Marine Corps . Marine corps order 6110.3A W/CH-3, body composition and military appearance program. Washington DC: Headquarters United States Marine Corps; (2019). [Google Scholar]
  • 12.Department of the Army . All Army activities memorandum (ALARACT). 2023;046/2023. Notification of new Army body fat assessment for Army body composition program. Washington DC: Secretary of the Army; (2023). [Google Scholar]
  • 13.Potter AW, Chin GC, Looney DP, Friedl KE. Defining overweight and obesity by percent body fat instead of body mass index. J Clin Endocrinol Metab. (2024) 15:dgae341. doi: 10.1210/clinem/dgae341, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ward LC. Segmental bioelectrical impedance analysis: an update. Curr Opin Clin Nutr Metab Care. (2012) 15:424–9. doi: 10.1097/MCO.0b013e328356b944 [DOI] [PubMed] [Google Scholar]
  • 15.Bailey BW, LeCheminant G, Hope T, Bell M, Tucker LA. A comparison of the agreement, internal consistency, and 2-day test stability of the InBody 720, GE iDXA, and BOD POD® gold standard for assessing body composition. Measurement Phys Educ Exerc Sci. (2018) 22:231–8. doi: 10.1080/1091367X.2017.1422129 [DOI] [Google Scholar]
  • 16.R Core Team . R: A language and environment for statistical computing. Foundation for statistical computing. Vienna, Austria: (2014). [Google Scholar]
  • 17.Nakagawa S, Johnson PCD, Schielzeth H. The coefficient of determination R(2) and intra-class correlation coefficient from generalized linear mixed-effects models revisited and expanded. J Royal Soc Interface. (2017) 14:20170213–1. doi: 10.1098/rsif.2017.0213 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Bates D, Mächler M, Bolker B, Walker S. Fitting linear mixed-effects models using lme4. J Stat Softw. (2015) 67:1–48. doi: 10.18637/jss.v067.i01 [DOI] [Google Scholar]
  • 19.Gerke O. Reporting standards for a bland-Altman agreement analysis. Diagnostics. (2020) 10:334. doi: 10.3390/diagnostics10050334 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Parker RA, Weir CJ, Rubio N, Rabinovich R, Pinnock H, Hanley J, et al. Application of mixed effects limits of agreement in the presence of multiple sources of variability: exemplar from the comparison of several devices to measure respiratory rate in COPD patients. PLoS One. (2016) 11:e0168321. doi: 10.1371/journal.pone.0168321, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Esco MR, Snarr RL, Leatherwood MD, Chamberlain NA, Redding ML, Flatt AA, et al. Comparison of total and segmental body composition using DXA and multifrequency bioimpedance in collegiate female athletes. J Strength Cond Res. (2015) 29:918–25. doi: 10.1519/JSC.0000000000000732 [DOI] [PubMed] [Google Scholar]
  • 22.Mecherques-Carini M, Albaladejo-Saura M, Vaquero-Cristóbal R, Baglietto N, Esparza-Ros F. Validity and agreement between dual-energy X-ray absorptiometry, anthropometry and bioelectrical impedance in the estimation of fat mass in young adults. Front Nutr. (2024) 11:1421950. doi: 10.3389/fnut.2024.1421950, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Day K, Kwok A, Evans A, Mata F, Verdejo-Garcia A, Hart K, et al. Comparison of a bioelectrical impedance device against the reference method dual energy X-ray absorptiometry and anthropometry for the evaluation of body composition in adults. Nutrients. (2018) 10:1–11. doi: 10.3390/nu10101469, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Antonio J, Kenyon M, Ellerbroek A, Carson C, Burgess V, Tyler-Palmer D, et al. Comparison of dual-energy X-ray absorptiometry (DXA) versus a multi-frequency bioelectrical impedance (InBody 770) device for body composition assessment after a 4-week Hypoenergetic diet. J Funct Morphol Kinesiol. (2019) 4:1–9. doi: 10.3390/jfmk4020023 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Lahav Y, Goldstein N, Gepner Y. Comparison of body composition assessment across body mass index categories by two multifrequency bioelectrical impedance analysis devices and dual-energy X-ray absorptiometry in clinical settings. Eur J Clin Nutr. (2021) 75:1275–82. doi: 10.1038/s41430-020-00839-5, PMID: [DOI] [PubMed] [Google Scholar]
  • 26.McLester CN, Nickerson BS, Kliszczewicz BM, McLester JR. Reliability and agreement of various InBody body composition analyzers as compared to dual-energy X-ray absorptiometry in healthy men and women. J Clin Densitom. (2020) 23:443–50. doi: 10.1016/j.jocd.2018.10.008, PMID: [DOI] [PubMed] [Google Scholar]
  • 27.Raymond CJ, Dengel DR, Bosch TA. Total and segmental body composition examination in collegiate football players using multifrequency bioelectrical impedance analysis and dual X-ray absorptiometry. J Strength Cond Res. (2018) 32:772–82. doi: 10.1519/JSC.0000000000002320, PMID: [DOI] [PubMed] [Google Scholar]
  • 28.Hart NH, Nimphius S, Spiteri T, Cochrane JL, Newton RU. Segmental musculoskeletal examinations using dual-energy X-ray absorptiometry (DXA): positioning and analysis considerations. J Sports Sci Med. (2015) 14:620–6. [PMC free article] [PubMed] [Google Scholar]
  • 29.Moonen H, van Zanten FJL, Driessen L, de Smet V, Slingerland-Boot R, Mensink M, et al. Association of bioelectric impedance analysis body composition and disease severity in COVID-19 hospital ward and ICU patients: the BIAC-19 study. Clin Nutr. (2021) 40:2328–36. doi: 10.1016/j.clnu.2020.10.023, PMID: [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets presented in this article are available by the authors upon reasonable request.


Articles from Frontiers in Nutrition are provided here courtesy of Frontiers Media SA