Abstract
Introduction
The anterior cruciate ligament is the most commonly injured ligament in the knee. Its injury is often evaluated with orthopedic tests during physical examination, but this turns out to be a subjective assessment. A knee arthrometer is a mechanical device developed in the 1970s to improve the diagnostic accuracy of anterior cruciate ligament injury. However, due to the high cost and low availability of the devices, they are rarely used in clinical practice and training. In this scenario, the objective of this study was to develop and validate a knee arthrometer to be fabricated by additive manufacturing (3D printing), thus creating a device that is easily accessible to all.
Methods
This experimental cross-sectional study involved the Orthopedic Skills Laboratory, Department of Surgery, at the Federal University of Paraná and 3D Printing Laboratory of the Professor Leide Parolin Marinoni Simulation Center of Parana Clinical Hospital, in partnership with the Engineering Department at the Federal University of Technology - Paraná, Brazil. Additionally, face validity was conducted at the Workers' Hospital and Parana Clinical Hospital. Ten orthopedic professionals with experience in the use of arthrometers were invited to assemble, evaluate the developed equipment, and answer a questionnaire regarding the assembly, appearance, and usability of the 3D arthrometer.
Results
At the end of the project, the arthrometer was successfully assembled from 3D-printed and unprinted parts, with an approximate cost of US$30 per device. All participants were able to assemble the 3D arthrometer with the instructions provided, and only one participant requested the instructor's assistance once during assembly. Regarding use, all participants were able to perform the arthrometry successfully and without the need for any assistance. Regarding face validity, we obtained a high degree of agreement among participants regarding the ease of assembly, good appearance, ease of use, and practicality of the equipment.
Conclusion
Using 3D printing technology to fabricate a knee arthrometer is feasible, resulting in an inexpensive and accessible tool.
Keywords: 3d printing, additive manufacturing, anterior cruciate ligament (acl), anterior cruciate ligament tear, arthrometer
Introduction
The anterior cruciate ligament (ACL) is the most commonly studied ligament in the knee and one of the most researched of all the entire musculoskeletal system [1]. Furthermore, this ligament has a large socioeconomic impact, since ACL injury usually requires surgical treatment, with approximately 350,000 reconstructions per year in the United States, at an estimated cost of 1 billion dollars [2].
ACL ruptures are often evaluated with orthopedic tests, such as the Lachman test, the anterior drawer sign, and the pivot shift test, but they are subjective assessments [3]. The diagnostic sensitivity and specificity of the tests vary widely depending on the examiner’s experience, the patient’s body type, and the delay from injury to examination [4]. Although physical examination is the first step in the assessment of knee injuries, it has a low diagnostic value in the acute setting [5]. Furthermore, manual tests cannot quantify or detect minor changes in laxity (e.g., in the face of partial ACL tears) [6]. In a cadaveric study, Hole et al. found that the Lachman test was reliable only after sectioning at least 75% of the ligament [7].
To improve the accuracy of ACL injury diagnosis, the arthrometer was developed in the 1970s, which is a mechanical device capable of objectively measuring ligament laxity [8]. It is very useful to aid in the diagnosis of partial ligament ruptures and also to evaluate and monitor post-operative results [9]. Several arthrometer models have been developed covering a wide range of complexity and availability [10,11]. However, the use of arthrometers in the routine evaluation of patients with ACL injuries is still limited by the high cost, difficult use, and low availability of devices, as some models have been discontinued and are therefore difficult to find.
In this scenario, the purpose of this study was to develop and validate a knee arthrometer to be fabricated by additive manufacturing (3D printing), thus creating a device that is easily accessible to all orthopedists.
Materials and methods
This experimental cross-sectional study involved the Orthopedic Skills Laboratory, Department of Surgery, at the Federal University of Paraná and 3D Printing Laboratory of the Professor Leide Parolin Marinoni Simulation Center of Parana Clinical Hospital, in partnership with the Engineering Department at the Federal University of Technology - Paraná, Brazil. Additionally, face validity was conducted at the Workers' Hospital and Parana Clinical Hospital. The ethics committee approval number is 81340424.8.0000.5225.
A 3D arthrometer was developed and modeled using SolidWorks® 3D CAD software (Dassault Systèmes, SolidWorks Corporation, Waltham, USA). The 3D printable components of the device were then fabricated by additive manufacturing (3D printing) with a Prusa i3 MK3S® printer using polyethylene terephthalate glycol (PETG) filament (Prusa, Prague, Czech Republic). The 3D printed and unprinted parts of the arthrometer are shown in Figure 1 and Table 1.
Table 1. Materials used and their respective costs.
Pieces | Quantity | Cost |
Caliper | 1 | US $9 |
Nylon Cable Ties 3,6mm | 2 | US $2 |
M4x16 screws - Hex Socket Cap Head | 6 | US $2.5 |
M4 nut | 6 | US $1.5 |
M3x30 screws - Hex Socket Cap Head | 2 | US $1 |
M3 nut | 2 | US $1 |
Butterfly M6 screw | 1 | US $2 |
M6 nut | 1 | US $1 |
Printed parts | 7 | US $10 |
Total Cost | US $30 |
To perform face validity, 10 orthopedic professionals with experience in the use of arthrometers were invited to assemble, evaluate the developed equipment, and answer a questionnaire regarding the assembly, appearance, and usability of the 3D arthrometer. Before assembly, participants watched a video with instructions, and during assembly, they had access to an instruction manual and a trained evaluator who could help if necessary. After assembly, participants watched a video with information on the use of the equipment and then had to perform arthrometry on a healthy participant. In the end, they were asked to answer a questionnaire with a Likert scale adapted for this study, with five questions formulated, which were graded on a spectrum of five descriptors between “Completely disagree” and “Completely agree”.
Descriptive and exploratory statistics were used, in addition to the degree of agreement of the items being assessed by the Content Validity Index (CVI), in which the sum of agreement of the items “Completely agree” and “Partially agree” marked by the participants was considered. An acceptable content validity index should be at least 0.80 and preferably greater than 0.90 [13].
Step-by-step assembly of the arthrometer
The first step is coupling the distal curve to the distal arch, with two M4x16 mm screws and two M4 nuts. The same procedure is performed for the proximal arch to the proximal curve. The set, distal arch, and distal curve are then coupled to the main bar of the arthrometer with two M4x16 screws and two M4 nuts (Figure 2).
Then, an M6 nut is then placed inside the pre-countered hole of the proximal angle bracket, and an M6x25 knurled thumb screw is tightened from bottom to top to fix this region of the arthrometer to the main bar (Figure 3). The thumb screw allows changing the length of the arthrometer, being easily adaptable to the patient’s characteristics.
Subsequently, the measuring instrument (digital caliper) is inserted in the predetermined hole in the main bar on the upper surface of the device. The caliper is fixed to the main bar by passing two nylon cable ties through four pre-drilled holes in the main bar, looping the caliper jaws, and fixing the caliper to the main bar of the arthrometer (Figure 4).
Lastly, two caliper covers (right and left) are then fixed to the caliper jaws with two M3x30 screws (Figure 5).
The entire process of assembling the arthrometer is demonstrated in a video attached as supplementary data of this paper (Video 1), as well as a video with instructions for using a 3D arthrometer (Video 2). These two videos were used as guides for research participants during the validation process.
Results
The arthrometer was successfully fabricated by 3D printing. Figure 6 shows the complete ready-to-use assembly of the device, which was accomplished at an approximate cost of US$30.
The participants evaluated were all men, orthopedists, and knee surgeons recognized by the SBCJ (Brazilian Society of Knee Surgery). All participants were able to assemble the 3D arthrometer with the instructions provided, and only one participant requested the instructor's assistance once during assembly. Regarding use, all participants were able to perform the arthrometry successfully and without the need for any assistance.
As a result of the validation, in the graph in Figure 7, we have all the questions that were evaluated and their respective answers. In general, all the questions asked had a Content Validity Index above 0.9. The equipment obtained 100% agreement regarding the usability and ease of use of the equipment. Regarding assembly, 80% of the participants strongly agreed that the instructions provided allowed for easy assembly. Regarding the appearance of the 3D arthrometer and its practicality for use during a consultation procedure, 90% of the participants strongly agreed with the statements. The greatest disparity occurred in the question regarding the comparison of the 3D arthrometer with others available on the market, where 10% of participants disagreed with the statement, but the total IVC of the question remained at 0.9.
Discussion
The rate of ACL injuries is increasing despite improved knowledge of ACL injury and prevention [1]. However, less than 10% of patients with an ACL injury have the diagnosis made by the first physician to see them [14]. The arthrometer was developed to improve the detection of ACL injuries and to quantify the results [5, 8, 15]. In the current study, we present an alternative arthrometer that is affordable, reproducible, easily portable, and easy to handle. We fabricated the arthrometer by additive manufacturing (3D printing) to obtain an affordable and high-strength tool.
The KT-1000 (MEDmetric Corp., San Diego, California) is the most widely used knee arthrometer and is currently considered the standard instrumented testing device for measuring anterior knee translation [16]. Van Eck et al. reported that arthrometer testing with the KT-1000 performed with maximum manual force had the highest sensitivity, specificity, accuracy, and positive predictive value for the diagnosis of ACL rupture [17]. Although considered the gold standard for assessing ACL injuries, the KT-1000 arthrometer was discontinued a few years ago, making it difficult to find and limiting access to the equipment. Furthermore, the device itself is considerably large and heavy, which makes it difficult to use during examination. Given these issues, several studies have attempted to develop a simpler and more portable device [4, 11]. Among them, the Rolimeter (Aircast Europa, Neubeuern, Germany) stands out, which is a simpler, more compact, and easier-to-handle device than the KT-1000 and is also easily portable [10, 11]. Several studies have compared the Rolimeter with the KT-1000, with a good correlation between the two devices; the main advantages of the Rolimeter are its low acquisition cost, simple learning curve, and lightness [11, 18-21]. However, the Rolimeter is no longer manufactured.
The device developed in this work can be used similarly to the Rolimeter to measure the anterior translation of the tibia in relation to the femur by applying maximum manual force. Because it is similar in size to the Rolimeter, it offers the same advantages: ease of handling, quick measurement procedure, and ease of transportation. We replaced the analog gauge with a digital caliper to simplify the assessment and improve precision [19]. A major advantage of our 3D printable arthrometer is its open-source license (Creative Commons 4.0 International License Attribution-ShareAlike) and its significantly lower cost compared to industrial arthrometers. Our device can be manufactured at a cost of approximately US$30, as shown in Table 1, while similar commercial arthrometers used to be sold for hundreds of euros. The 3D printable knee arthrometer project is already available for download and public use [12]. Furthermore, to facilitate access and use, the authors have made videos available on the assembly and use of the arthrometer. A “how to use” video proved necessary in our journey with the arthrometer, as we found that few professionals had experience using the device, precisely due to the difficulty in acquiring an arthrometer (Videos 1, 2).
Face validity aims to assess the degree to which the object in question appears effective in achieving its stated objectives [22]. Regarding this issue, the 3D arthrometer performed excellently in the questionnaire results. All questions asked had a Content Validity Index above 0.9. In other words, we obtained a high degree of agreement among participants regarding the usability, ease of assembly, appearance, and practicality of the equipment.
A simple measurement procedure and low acquisition cost are decisive for the successful use of a device in the doctor's daily routine and currently, orthopedic surgery training lacks arthrometers in daily practice. This device is essential both for the diagnosis of ACL injury and for research in the field of medical education. Ideally, each orthopedic surgeon in training should have an arthrometer available for individual use whenever necessary for progressive training and familiarity with the device. This ideal scenario is not currently attainable with the devices available in the market. We hope that the arthrometer model presented here will gain widespread use in clinical practice.
Limitations
The study has some limitations that should be considered. First, the sample size for face validation was limited, which may affect the generalizability of the findings to broader populations. Furthermore, the study did not include direct comparisons with existing arthrometers, which may limit the understanding of the device’s efficacy. Moreover, the lack of clinical validation also represents a significant limitation, as the arthrometer’s performance in real-world medical settings has not yet been fully evaluated. With these limitations in mind, we aim to conduct future studies to validate our 3D printable knee arthrometer using a simultaneous validation design, comparing it with other arthrometers and evaluating its clinical performance.
Conclusions
Using 3D printing technology to manufacture a knee arthrometer is feasible, resulting in a cheap and accessible tool. During validation, all participants were able to assemble the 3D arthrometer with the instructions provided, and successfully perform arthrometry. Overall, participants expressed agreement on the device's usability, ease of assembly, the appearance of the 3D arthrometer and its practicality for use. We hope that this equipment will enable more orthopedic specialists to have easier access to their own personal arthrometers for individual use.
Disclosures
Human subjects: Consent for treatment and open access publication was obtained or waived by all participants in this study. Comitê de Ética do Hospital do Trabalhador - Curitiba (Paraná) - Brasil issued approval 6.944.989.
Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Carolline P. Nunes, Edmar Stieven Filho, Ayrton A. Martins Neto, Isabela P. Nascimento, Caue O. Maia, Jose A. Foggiatto, Fernando M. Rosa, Paul A. Milcent, Maisa S. Namba, Lucas Santiago
Acquisition, analysis, or interpretation of data: Carolline P. Nunes, Edmar Stieven Filho, Ayrton A. Martins Neto, Isabela P. Nascimento, Caue O. Maia, Jose A. Foggiatto, Fernando M. Rosa, Paul A. Milcent, Maisa S. Namba, Lucas Santiago
Drafting of the manuscript: Carolline P. Nunes, Edmar Stieven Filho, Ayrton A. Martins Neto, Isabela P. Nascimento, Caue O. Maia, Jose A. Foggiatto, Fernando M. Rosa, Paul A. Milcent, Maisa S. Namba, Lucas Santiago
Critical review of the manuscript for important intellectual content: Carolline P. Nunes, Edmar Stieven Filho, Ayrton A. Martins Neto, Isabela P. Nascimento, Caue O. Maia, Jose A. Foggiatto, Fernando M. Rosa, Paul A. Milcent, Maisa S. Namba, Lucas Santiago
Supervision: Carolline P. Nunes, Edmar Stieven Filho, Ayrton A. Martins Neto, Isabela P. Nascimento, Caue O. Maia, Jose A. Foggiatto, Fernando M. Rosa, Paul A. Milcent, Maisa S. Namba, Lucas Santiago
References
- 1.Trends in publications on the anterior cruciate ligament over the past 40 years on PubMed. Kambhampati SB, Vaishya R. Orthop J Sports Med. 2019;7:2325967119856883. doi: 10.1177/2325967119856883. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.The effectiveness of a functional knee brace on joint-position sense in anterior cruciate ligament-reconstructed individuals. Sugimoto D, LeBlanc JC, Wooley SE, Micheli LJ, Kramer DE. J Sport Rehabil. 2016;25:190–194. doi: 10.1123/jsr.2014-0226. [DOI] [PubMed] [Google Scholar]
- 3.The diagnostic accuracy of ruptures of the anterior cruciate ligament comparing the Lachman test, the anterior drawer sign, and the pivot shift test in acute and chronic knee injuries. Katz JW, Fingeroth RJ. Am J Sports Med. 1986;14:88–91. doi: 10.1177/036354658601400115. [DOI] [PubMed] [Google Scholar]
- 4.Objective evaluation of anterior knee laxity; comparison of the KT-1000 and GNRB® arthrometers. Collette M, Courville J, Forton M, Gagnière B. https://doi.org/10.1007/s00167-011-1869-2. Knee Surg Sports Traumatol Arthrosc. 2012;20:2233–2238. doi: 10.1007/s00167-011-1869-2. [DOI] [PubMed] [Google Scholar]
- 5.Test-retest reliability of repeated knee laxity measurements in the acute phase following a knee trauma using a Rolimeter. Ericsson D, Östenberg AH, Andersson E, Alricsson M. https://doi.org/10.12965/jer.1735104.552. J Exerc Rehabil. 2017;13:550–558. doi: 10.12965/jer.1735104.552. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.A novel non-invasive method for measuring knee joint laxity in four dof: In vitro proof-of-concept and validation. Pedersen D, Vanheule V, Wirix-Speetjens R, Taylan O, Delport HP, Scheys L, Andersen MS. https://doi.org/10.1016/j.jbiomech.2018.10.016. J Biomech. 2019;82:62–69. doi: 10.1016/j.jbiomech.2018.10.016. [DOI] [PubMed] [Google Scholar]
- 7.Increased tibial translation after partial sectioning of the anterior cruciate ligament. The posterolateral bundle. Hole RL, Lintner DM, Kamaric E, Moseley JB. Am J Sports Med. 1996;24:556–560. doi: 10.1177/036354659602400423. [DOI] [PubMed] [Google Scholar]
- 8.Measurements of anterior laxity in normal and anterior cruciate absent knees with two instrumented test devices. Sherman OH, Markolf KL, Ferkel RD. http://www.ncbi.nlm.nih.gov/pubmed/3802633. Clin Orthop Relat Res. 1987:156–161. [PubMed] [Google Scholar]
- 9.Right and left knee laxity measurements: a prospective study of patients with anterior cruciate ligament injuries and normal control subjects. Sernert N, Kartus JT Jr, Ejerhed L, Karlsson J. https://doi.org/10.1016/j.arthro.2004.04.056. Arthroscopy. 2004;20:564–571. doi: 10.1016/j.arthro.2004.04.056. [DOI] [PubMed] [Google Scholar]
- 10.An investigation to examine the inter-tester and intra-tester reliability of the Rolimeter® knee tester, and its sensitivity in identifying knee joint laxity. Hatcher J, Hatcher A, Arbuthnot J, McNicholas M. J Orthop Res. 2005;23:1399–1403. doi: 10.1016/j.orthres.2005.06.003.1100230623. [DOI] [PubMed] [Google Scholar]
- 11.The rolimeter: a new arthrometer compared with the KT-1000. Ganko A, Engebretsen L, Ozer H. https://doi.org/10.1007/s001670050008. Knee Surg Sports Traumatol Arthrosc. 2000;8:36–39. doi: 10.1007/s001670050008. [DOI] [PubMed] [Google Scholar]
- 12.The 3D printable Knee Arthrometer by edmar | Download free STL model | Printables.com. [ Oct; 20214 ]. 2023. https://www.printables.com/model/292733-the-3d-printable-knee-arthrometer https://www.printables.com/model/292733-the-3d-printable-knee-arthrometer
- 13.ABC of content validation and content validity index calculation. Yusoff MSB. Educ Med J. 2019;11:49–54. [Google Scholar]
- 14.Anterior cruciate ligament injury: a persistently difficult diagnosis. Parwaiz H, Teo AQ, Servant C. https://doi.org/10.1016/j.knee.2015.09.016. Knee. 2016;23:116–120. doi: 10.1016/j.knee.2015.09.016. [DOI] [PubMed] [Google Scholar]
- 15.Use of arthrometers to assess knee laxity and outcomes. Cannon WD. Sports Med Arthrosc. 2002;10:191–200. [Google Scholar]
- 16.Instrumented measurement of anterior laxity of the knee. Daniel DM, Malcom LL, Losse G, Stone ML, Sachs R, Burks R. http://www.ncbi.nlm.nih.gov/pubmed/3997924. J Bone Joint Surg Am. 1985;67:720–726. [PubMed] [Google Scholar]
- 17.Methods to diagnose acute anterior cruciate ligament rupture: a meta-analysis of instrumented knee laxity tests. van Eck CF, Loopik M, van den Bekerom MP, Fu FH, Kerkhoffs GM. https://doi.org/10.1007/s00167-012-2246-5. Knee Surg Sports Traumatol Arthrosc. 2013;21:1989–1997. doi: 10.1007/s00167-012-2246-5. [DOI] [PubMed] [Google Scholar]
- 18.Evaluation of anterior knee joint instability with the Rolimeter. A test in comparison with manual assessment and measuring with the KT-1000 arthrometer. Balasch H, Schiller M, Friebel H, Hoffmann F. https://doi.org/10.1007/s001670050149. Knee Surg Sports Traumatol Arthrosc. 1999;7:204–208. doi: 10.1007/s001670050149. [DOI] [PubMed] [Google Scholar]
- 19.Instrumented Arthrometry of the anterior cruciate ligament. A comparison. Krautter A, Rohnstock L, Moeller K, et al. Biomed Tech. 2012;57:889–892. [Google Scholar]
- 20.A new mechanical testing device for measuring anteroposterior knee laxity. Schuster AJ, McNicholas MJ, Wachtl SW, McGurty DW, Jakob RP. https://doi.org/10.1177/0363546504267050. Am J Sports Med. 2004;32:1731–1735. doi: 10.1177/0363546504267050. [DOI] [PubMed] [Google Scholar]
- 21.Estudio comparativo de dos sistemas de medición de la laxitud del ligamento cruzado anterior. Ladero F, Maestro A. Rev Ortop Traumatol. 2006;50:263–267. [Google Scholar]
- 22.Principles and methods of validity and reliability testing of questionnaires used in social and health science researches. Bolarinwa OA. Niger Postgrad Med J. 2015;22:195–201. doi: 10.4103/1117-1936.173959. [DOI] [PubMed] [Google Scholar]