Skip to main content
. 2024 Dec 18;13(1):A0161. doi: 10.5702/massspectrometry.A0161

Table 1. Optimized parameters for HPLC and ICP-MS conditions.

HPLC
Column Agilent Bio WAX
(4.6 mm i.d. × 50 mm)
Mobile phase 50 mmol L–1 CH3COONH4,
2 mmol L–1 PDCA (pH 6.80)
Flow rate 0.6 mL min–1
Column temperature Room temperature
Injection volume 10 μL
ICP-MS
RF power 1200 W
Plasma gas 15 L min–1
Auxiliary gas 1.0 L min–1
Carrier gas 1.0 L min–1
Sampling depth 5.5 mm
Spray chamber temperature 2°C
Injector 2.5 mm i.d.
CRC gas He 4.3 mL min–1
KED +3 V
Monitored isotope (m/z) 52Cr, 35Cl
Integration time 1 s for 52Cr, 0.001 s for 35Cl

CRC, collision/reaction cell; HPLC, high-performance liquid chromatography; ICP-MS, inductively coupled plasma-mass spectrometry; KED, kinetic energy discrimination; PDCA, 2,6-pyridinedicarboxylic acid.