Skip to main content
. 2024 Dec 23;96(12):e70129. doi: 10.1002/jmv.70129

Table 1.

Percent agreement and Cohen's kappa score for the comparison of BioFire‐RP2.1 and QIAstat‐Dx‐RP2.0 multiplex panel tests in 269 respiratory clinical specimens.

QIAstat‐Dx‐RP2.0 Agreement Kappa
positive negative
BioFire‐RP2.1 Positive 148 33 86.3% 0.71
Negative 4 84