Table 3.

HU consistency for HyperSight CBCT. The top table is for 125 kV CBCT, and the bottom table is for 140 kV CBCT

125 kV
Catphan position
Material Catphan HU Center 10 cm superior 10 cm inferior 10 cm up 10 cm down 10 cm left 10 cm right Max diff.
Air −1046:−986 −1000 −1000 −994 −1000 −1000 −1000 −985 6
PMP −220: −172 −197 −190 −192 −188 −191 −190 −176 21
LDPE −121:−87 −105 −99 −112 −92 −92 −111 −93 20
Polystyrene −65:−29 −50 −47 −54 −56 −44 −48 −33 23
Acrylic 92:137 109 108 102 109 112 121 109 19
Delrin 344:387 335 340 326 349 329 338 334 23
Bone 50% 667:783 663 671 635 625 643 674 576 98
Bone 20% 211:263 215 227 206 221 197 204 217 30
Teflon 941:1060 939 939 914 916 934 932 933 25
140 kV
Catphan position
Material Catphan HU Center 10 cm superior 10 cm inferior 10 cm up 10 cm down 10 cm left 10 cm right Max diff.
Air −1046:−986 −999 −1000 −999 −1000 −1000 −1000 −986 14
PMP −220: −172 −188 −187 −191 −179 −189 −192 −174 18
LDPE −121:−87 −96 −94 −107 −84 −93 −107 −88 23
Polystyrene −65:−29 −44 −45 −52 −51 −43 −46 −35 17
Acrylic 92:137 118 113 102 111 110 115 114 16
Delrin 344:387 346 344 331 358 338 340 338 27
Bone 50% 667:783 644 646 614 598 621 648 558 90
Bone 20% 211:263 214 216 196 216 189 187 205 29
Teflon 941:1060 940 938 911 930 938 930 926 29

Abbreviations: CT = computed tomography; CBCT = cone-beam computed tomography; HU = Hounsfield unit.