Total count and misreported count of event narratives containing key death search terms. Total count includes all event narratives classified as deaths, injuries, or malfunctions. Misreported count includes all event narratives classified as injuries or malfunctions. Only BMS and DES reports were included in the analysis.

Year Combined DES and BMS
BMS
DES
Misreport Total % Misreport Total % Misreport Total %
2011 38 616 6.17 3 62 4.84 35 554 5.96
2012 49 676 7.25 3 58 5.17 46 618 7.44
2013 35 515 6.80 3 53 5.66 32 462 6.93
2014 20 435 4.60 5 46 10.87 15 389 3.86
2015 18 372 4.84 4 37 10.81 14 334 4.19
2016 20 456 4.39 4 52 7.69 16 382 4.19
2017 34 405 8.40 22 76 28.95 11 315 3.49
2018 38 376 10.11 24 71 33.80 14 303 4.62
2019 41 329 12.46 18 58 31.03 23 270 8.52
Total 293 4180 7.01 86 513 16.76 206 3627 5.68

∗ BMS bare metal stents; DES drug eluting stents.