Skip to main content
. Author manuscript; available in PMC: 2024 Dec 27.
Published in final edited form as: IEEE J Sel Top Quantum Electron. 2009 Dec 15;16(4):863–869. doi: 10.1109/jstqe.2009.2035362

Fig. 1.

Fig. 1.

(a) Design schematic of the robust OCT imaging needle. The fiber-optic GRIN lens and microreflector were encased in a small glass tube and then assembled in a hypodermic tube. (b) Spot size and working distance as a function of the pitch number of the GRIN fiber lens with a spacer of 450 μm long. The simulation is based on ABCD matrix calculation. The dashed lines on the plots indicate our target parameters.