Table 2.

Optimized conditions for LC-MS quantification of PFOS and PFOA

LC MS
Instrument Agilent 1100 Series Instrument: atmospheric pressure electrospray ionization (ESI) MS detector (Agilent Technologies, USA)
Column ODS-C18 ((2.1 × 150 mm, 5 mm), Waters, USA
Mobile phase

A: CH3CN

B: CH3COONH4/H2O (10mM)

Polarity mode Negative
Elution Linear gradient: started at 35% of A and then increased to 45% of A at 2%/min for 5 min and maintained at 45% untile the end of run time Nebulizer N2 (70 psi)
Drying gas N2 (10 L/min, 350 °C)
Fragmenter voltage 130 V
capillary voltage 4000 V
Flow rate 0.2 ml/min m/z (SIM mode)

499 for PFOS

413 for PFOA

Oven temperature Room temperature
Injection volume 20 µL Data processing ChemStation software
Run time 20 min