Table 1.

Comparison of repeatability estimates of the PCR measurements using different reaction mixes and platforms over three days (using standard deviations of the Ct value).

Repeatability Estimates (standard deviations of the Ct value) based on Individual Plasmid Copy Number
Plasmid Copy Number Platform and PCR Reaction Mix

ABI 7700 LightCycler RotorGene

Excite FastStart Excite FastStart Excite FastStart

102 0.543 0.182 0.568 0.233 0.888 0.186
103 0.452 0.204 1.177 0.190 0.184 0.445
104 0.223 0.305 1.123 0.100 0.256 0.592
105 0.243 0.168 0.947 0.224 0.209 0.221
106 0.362 0.183 1.378 0.281 0.772 0.333
107 0.473 0.189 0.919 0.156 0.319 0.247
108 0.236 0.184 0.454 0.352 0.322 0.166
Degrees of Freedom 6–8 6 6–8 5–6 6–8 6
Mean 0.362 0.202 0.938 0.219 0.421 0.313
Standard Deviation 0.130 0.047 0.331 0.082 0.286 0.156

(Note that one replicate is missing for the LC platform coupled with the Fast Start reaction mix – the capillary broke).