TABLE 3.
Summary of QA recommendations for the Iris collimator.
Parameter | Frequency | Tolerance | Method |
---|---|---|---|
Iris beam CAX / Laser coincidence | Quarterly / Commissioning / Path Calibration |
Preferred: <0.5 mm Acceptable: <1.0 mm |
Film, water tank |
Beam QA (OCR, TPR/PDD, OF and FWHM) (Section 7.3) | Commissioning, annual | Consistent with commissioning (Table 10) | Water tank |
Field size consistency | Monthly/Recalibration |
Preferred: ±0.2 mm from baseline Acceptable: ±0.5 for FS ≥ 10 mm |
Film, camera, dose area product, EPID |
End to end targeting | Commissioning/Annual/Monthly |
Static E2E: <0.95 mm Synchrony E2E: <0.95 mm preferred, <1.5 mm acceptable |
Film |
AQA targeting | Daily | <0.95 mm from baseline | Film |
DQA (Section 3.5.4 and 7.2) | Commissioning/Monthly | 2%/2 mm (>90%) for static and 3%/3 mm (>90%) for motion tracking | Depends on equipment availability |