Table 1.

Nominal instrument parameters.

GC Parameter Setting
GC System Agilent 7890A GC/Agilent 6890 GC (Section 3.7, analyst 2 only)
Column Agilent DB-1, 60m × 0.32 mm, 3.0 μm film thickness, P/N: 123-1064
Carrier Flow 18 psi (constant pressure)
Carrier Gas Helium
Inlet Temperature 180 °C
Split Ratio 5:1
Oven Program 5 °C hold for 6 min,
25 °C/min to 250 °C, hold 5.6 min (variable)
Detector FID
Detector Temperature 300 °C
Detector Gasses Hydrogen: 40 mL/min.
Air: 400 mL/min.
Make up (He): 30 mL/min (Constant)
Analysis Time 20 min (variable)
Data Collection Rate 10 Hz

HS Parameter Setting

Headspace Autosampler Perkin Elmer Turbomatrix 40
Headspace Oven Temperature 110 °C (variable)
Needle Temperature 130 °C
Transfer Line Temperature 140 °C
GC Cycle Time 30 min
Vial Equilibration Time 10 min (variable)
Pressurization Time 2.0 min
Inject Time 0.05 min
Withdraw Time 0.2 min
Shaker On
Vial Vent On
Track Oven Off
Carrier Pressure 24 psi