Injection precision, linearity, and QL data from various solvent systems.

Solvent System Injection Precision (% RSDall of Standard Peak Area) Correlation of Determination (R) Slope (pA∗min.∗mL/mg) Y-intercept (pA∗min.) Std. Dev. of Residuals Residual Sum of Squares % RSD of RFs Across the Range QL S/N at 0.5 μg/mL (Each Injection) QL Injection Precision (% RSD6 of Peak Area)
1 mM HCl in DMAc 1.1 % 1.0000 17.248 −0.00047 0.00100 4.0 × 10−6 3.0 % 17, 19, 19, 20, 19, 16 4.4 %
1 mM HCL in DMF 2.8 % 1.0000 19.873 −0.00097 0.00276 3.0 × 10−5 3.2 % 19, 19, 21, 22, 22, 21 2.5 %
1 mM HCl in DMI 0.5 % 0.9999 18.768 0.00132 0.00437 7.6 × 10−5 5.5 % 19, 15, 20, 19, 20, 20 2.9 %
1 mM HCl in NMP 1.6 % 1.0000 18.380 0.00010 0.00207 1.7 × 10−5 3.3 % 15, 17, 12, 17, 14, 15 3.6 %
1 mM HCl in Cyrene® 5.7 % 0.9988 9.611 0.01010 0.00889 3.2 × 10−4 22.7 % 8, 13, 10, 10,
12, 12
5.8 %
10 mM Tetrabutylammonium Chloride in DMAc 0.8 % 1.0000 17.294 −0.00320 0.00241 2.3 × 10−5 1.0 % 10, 11, 12, 11, 13, 11 5.8 %