Skip to main content
. 2024;25(9):3283–3291. doi: 10.31557/APJCP.2024.25.9.3283

Table 2.

Asymmetric Factor for “Z”Patient (10% Isodose volume) and Entire Patient (0% Isodose Volume) for 3%3mm and 3%2mm Gamma Passing Criteria.

Sr No "Z"patient(Zk3D) Patient(Zk3D)
3%3mm 3%2mm 3%3mm 3%2mm
1 1.63 2.1 0.54 0.68
2 3.26 4.15 0.73 0.9
3 2.25 2.8 0.49 0.59
4 3.12 3.75 1.04 1.24
5 1.35 1.94 0.29 0.41
6 2.54 3.45 0.58 0.78
7 2.33 3.2 0.76 1.02
8 1.57 2.2 0.47 0.66
9 1.75 2.12 0.54 0.65
10 1.89 2.5 0.7 0.93
11 1.23 1.57 0.27 0.36
12 3.33 3.86 0.96 1.13
13 3.01 3.3 1.09 1.18
14 2.49 3.28 0.73 0.94
15 1.63 1.98 0.68 0.81
16 1.34 1.73 0.39 0.49
17 0.76 0.87 0.25 0.33
18 2.58 2.79 0.79 0.91
19 2.82 3.1 0.82 0.9
20 3.47 3.52 1.06 1.08
Avg ± Std deviation 2.22±0.79 2.71±0.87 0.66±0.25 0.80±0.27
p value < 0.01 < 0.01 < 0.01 < 0.01