Table 5.
Pairwise comparisons between Area under the ROC curve of index assessment methods.
At Enamel threshold | Visual-tactile (FDI) and VistaCam | |
Difference between areas | 0.0654 | |
SE | 0.0237 | |
95% CI | 0.0190 to 0.112 | |
z-statistics | 2.761 | |
Significance level | P = 0.0058* | |
Visual-tactile (FDI) and DIAGNOdent | ||
Difference between areas | 0.0422 | |
SE | 0.0267 | |
95% CI | −0.0102 to 0.0945 | |
z-statistics | 1.578 | |
Significance level | P = 0.1145 | |
VistaCam and DIAGNOdent | ||
Difference between areas | 0.0232 | |
SE | 0.0181 | |
95% CI | −0.0123 to 0.0587 | |
z-statistics | 1.279 | |
Significance level | P = 0.2008 | |
At Dentin threshold | Visual-tactile (FDI) and VistaCam | |
Difference between areas | 0.171 | |
SE | 0.0262 | |
95% CI | 0.119 to 0.222 | |
z-statistics | 6.521 | |
Significance level | P < 0.0001* | |
Visual-tactile (FDI) and DIAGNOdent | ||
Difference between areas | 0.0851 | |
SE | 0.0227 | |
95% CI | 0.0406 to 0.130 | |
z-statistics | 3.745 | |
Significance level | P = 0.0002* | |
VistaCam and DIAGNOdent | ||
Difference between areas | 0.256 | |
SE | 0.0245 | |
95% CI | 0.208 to 0.304 | |
z-statistics | 10.456 | |
Significance level | P < 0.0001* |
*Significant, SE Standard Error, 95% CI 95% confidence interval.