Table 3. The most studied dynamic fluid responsiveness tests.

Method Limitations Maneuver or clinical condition Indicates fluidresponsiveness Sensitivity, % (95% CI) Specificity, % (95% CI)
Central venous pressure Right ventricular infarction or failurePulmonary hypertensionconstrictive pericarditisHypertensive pneumothorax Decreased CVP with spontaneous inspiration ↓ ≥1 mmHg 94 (NA) 81 (NA)
Pulse pressure variation Spontaneous ventilationCardiac arrhythmiaAcute respiratory distress syndromeRight ventricle failure Controlled ventilation, Vt ≥ 7 mL/kg PBW ∆ ≥11% 84 (75-90) 84 (77-90)
Increase in PPV from Vt 6-8 mL/kg PBW ­↑ ≥3.5% 94 (NA) 100 (NA)
Variation in systolic volume Spontaneous ventilation Cardiac arrhythmia Controlled ventilation, Vt ≥ 7 mL/kg PBW ∆ ≥13% 79 (67-87) 84 (74-90)
Acute respiratory distress syndromeRight ventricle failure Increase in SV variation from Vt 6-8 mL/kg PBW ­↑ ≥2.5% 88 (NA) 100 (NA)
Respiratory variation of inferior vena cava diameter Spontaneous ventilationCardiac arrhythmiaAcute respiratory distress syndrome Controlled ventilation, Vt ≥ 8 mL/kg PBW ∆ ≥15% 77 (44-94) 85 (49-97)
Cardiac output/cardiac index Right ventricular infarction or failureIntrathoracic or intra-abdominal pressure >12 mmHgCardiac tamponade or tension pneumothoraxAdrenergic stimulation (pain, cough, or discomfort) Passive lower limb raising ↑ cardiac output ≥11% 88 (80-93) 92 (89-95)
Not intubated Unable to stop spontaneous breathing for at least 15s End-expiratory occlusion test ↑ cardiac index ≥5% 91 (72-99) 100 (72-100)
*

If the result is negative in ventilation with low tidal volume, pulse pressure and stroke volume variations must be reassessed after a transient increase (1 min) in tidal volume to 8mL/kg. Absolute increases ≥3.5% in pulse pressure variation and ≥2.5% in stroke volume variation indicate fluid responsiveness.

∆: variation; CVP: central venous pressure; NA: not applicable; PBW: predicted body weight; SV: stroke volume; Vt: tidal volume; 95%CI: 95% confidence interval.