Abstract
Recent progress in materials chemistry has resulted in the development of several ceramic materials that are now being used in dental implants. The advantages of ceramic materials over conventional metallic materials are that they do not induce allergic reactions in individuals with metal allergies, they do not interfere with magnetic resonance imaging, and they provide improved esthetics. In addition, some ceramic materials are tougher than metallic materials and less brittle. However, despite these advantages, few ceramic dental implant materials are currently approved for use in Japan. In FY2022, the Ministry of Health, Labour and Welfare of Japan commissioned a project called the “Project for the Development of a Guideline for the Evaluation of Ceramic Dental Implants,” the goal of which was to consider how best to facilitate swift clinical development and approval of emerging ceramic dental implant materials. At a meeting of experts from professional societies, related industry organizations, and government agencies, the issues related to evaluation of the efficacy and safety of ceramic implant were discussed. Here, we summarize the outcomes of that meeting as a set of principles for the premarketing evaluation of ceramic dental implant materials in Japan.
Keywords: Dental Implant, Ceramics, Safety, Efficacy, Material
Introduction
Dental caries and periodontal disease are non-communicable oral diseases and risk factors for tooth loss. The burden of these oral diseases increases with aging [1]. Dental implants are a therapeutic option to restore functional loss of mastication and improve esthetics in individuals without teeth. The current gold-standard material for dental implants is titanium and its alloys; however, these materials can induce allergic reactions in individuals with metal allergies and can have poor esthetics. To address these issues, ceramic implant materials such as zirconia have been developed [2]. Zirconia is superior to titanium with respect to osteoblastic adhesion and proliferation; moreover, it does not induce allergic reactions in individuals with metal allergy and it has better material fatigue and esthetic properties, although it has lower ductility and a lower elastic modulus than does titanium [2, 3].
Despite the advantages of ceramic implant materials, few such materials are currently approved for use in Japan. In FY2022, the Ministry of Health, Labour and Welfare of Japan commissioned a project called the “Project for the Development of a Guideline for the Evaluation of Ceramic Dental Implants,” the goal of which was to consider how best to facilitate swift clinical development and approval of emerging ceramic dental implant materials. At a meeting of experts from professional societies, related industry organizations, and government agencies, the issues related to the evaluation of efficacy and safety of ceramic implant materials were discussed. In this document, we summarize the outcomes of this meeting as a set of principles for the premarketing evaluation of ceramic dental implant materials in Japan.
General Matters
In this document, the term “dental implants” is used to refer to medical devices with the generic names of “dental endosseous implant” “dental implant fixtures,” “dental implant systems,” and “dental implant abutments,” as specified in the Japanese Act on Securing Quality, Efficacy and Safety of Products Including Pharmaceuticals and Medical Devices (The Pharmaceutical and Medical Device Act) [4]. Definitions of the other terms used in this document are provided in Table 1. In addition, it should be noted that this document covers only dental implants made from ceramic materials containing zirconia as the main ingredient. For dental implants made of ceramic materials other than zirconia, the applicability of the principles contained herein should be considered with respect to the individual properties of these other ceramic materials.
Table 1.
Definitions of terms used in this document
| Term | Definition |
|---|---|
| Dental implant | A medical device made of biocompatible materials that is surgically implanted in, or directly connected to, the maxilla or mandible to restore masticatory function. In this document, this term refers to dental endosseous implants, dental implant fixtures, dental implant systems, and dental implant abutments. |
| Dental endosseous implant | A dental implant that is partially or totally implanted in the jaw bone. In this document, this term includes dental implant fixtures and dental implant abutments. |
| Dental implant fixture | The part of a dental implant that is surgically implanted into bone. Screw-type and cylinder-type fixtures are available. |
| Dental implant abutment | An element that is fixed to the dental implant fixture to form an abutment of the superstructure or used temporarily until the gingiva heals. In this document, this term includes the abutment screw used for fixing the dental implant abutment. |
| Dental implant system | A system consisting of dental implants, instruments for implant surgery, and laboratory instruments used to fabricate the superstructure. In this document, this term refers only to a system composed of a dental implant fixture and a dental implant abutment. |
| Immediate loading | A procedure involving implantation of a dental endosseous implant material or dental implant fixture and subsequent occlusal loading of a prosthesis within the following 48 h. |
| Early loading | A procedure involving implantation of a dental endosseous implant material or dental implant fixture and subsequent occlusal loading of a prosthesis onto the dental implant before bone healing (which requires 4 months for the maxilla and 3 months for the mandible). Immediate loading is excluded. |
| Temporary implant | A dental endosseous implant or dental implant fixture that is not intended for permanent use. |
| Surface treatment | A process applied to a part of, or the entire surface of, a dental implant. In this document, this term is used in reference to a product that has been manufactured and has undergone some surface processing on the manufactured surface. |
The premarketing evaluation of ceramic dental implants (hereafter, “the product”) should address three overarching points. First, the uniqueness, improvements, and equivalence of the product compared with current approved products should be clarified. For example, it should be clarified that the product is equivalent to currently approved titanium dental implants, or that it is superior to currently approved titanium dental implants in terms of immediate loading and early loading, or that is addresses an unmet medical need. If an esthetics claim related to the color tone is made, the applicant should clarify the clinical significance of the color tone and then evaluate the esthetic property of the product. Similarly, if a claim of a low risk of inducing allergic symptoms to metal is made, the applicant should confirm the absence of metal in the material or the degree to which any metal is present.
Secondly, the method of using the product to ensure adequate safety should be established, particularly when the safety concerns differ from those for currently approved titanium dental implants. For example, for dental implant fixtures, the method of forming an insertion socket and inserting the product should be clarified. When dental implant abutments are connected to dental implant fixtures, and when dental implant superstructure materials are connected to dental implant abutments, the tightening method and the recommended torque should be established. The points to consider in cases in which removal of the dental implant fixture is required should also be established.
Thirdly, the concomitant use of any medical devices should be clarified. For example, superstructures should be identified by information such as the brand name and certification number. When a dedicated instrument is used for implanting a dental fixture or fastening an abutment, the instrument should be identified by information such as the brand name, generic name, and certification/notification number. If the instrument to be used is not identified as a dedicated instrument, it is acceptable to specify the conditions required for the medical device to be used and to provide information such as any critical specifications and the generic name of the instrument.
Evaluation of Quality, Safety, and Performance
Within the categories of quality, safety, and performance, there are a number of individual items that should be evaluated. These items are summarized below and in Table 2. The evaluations should be performed by using raw materials or a test sample prepared under the same conditions as the final product [5, 6].
Table 2.
Items for premarketing evaluation of ceramic dental implant materials in Japan
| Category | Item |
|---|---|
| Safety |
• Biological safety [6, 14, 15] • Magnetic resonance safety [16] |
| Quality |
• Raw materials • Chemical composition • Fine structure (mean crystal grain size) • Surface treatment • Hot-water stability • Assurance of sterility • Residual ethylene oxide testing [11] • Packaging |
| Performance |
• Surface roughness [22] • Bending strength [7] • X-ray detectability [24, 25] • Solubility and degradability [19] • Evaluation required according to the characteristics of the product |
| Other | • Simulated-use test [15] |
Quality
Raw Materials
Such materials should fulfill the standards of either ISO 13,356 or ASTM F1843 [7, 8]. If a raw material does not fulfill one of those standards, its physical and chemical properties, including its bulk density, chemical composition, and radioactivity, should be evaluated [7]. In addition, clarification that the material does not undergo chemical change during the manufacturing process should be obtained.
Chemical Composition
If a ceramic material contains a metal, either as part of its chemical composition or as an additive material, the purpose and chemical composition of the metal should be clarified.
Fine Structure
The fine structure (mean crystal grain size) should be clarified in either a sample of the final product or a test sample prepared under the same conditions as the final product.
Surface Treatment
For surface-treated dental implant fixtures or dental implant abutments, the surface treatment method and treatment conditions should be clarified. Surface treatment other than roughening may need to be evaluated separately in consideration of its intended clinical effect. Under visual observation, no abnormality in shape, burr on the surface, scratch, adhesion of foreign matter, coating with foreign matter, or any other visual abnormality should be seen.
Stability
The stability of the product should be evaluated [9], and subsequently an appropriate storage method and shelf life should be established. For products that do not require any specific storage method to ensure their quality, or for products that do not deteriorate over time, stability does not need to be evaluated [10].
Assurance of Sterility
The sterility of dental implants supplied as sterilized shall be assured on the basis of the appropriate sterilization validation standards or other equivalent or superior standards.
Residual Ethylene Oxide Test
The residual gas concentration of dental implants sterilized with ethylene oxide gas should be evaluated [11].
Risk Assessment
A risk analysis and assessment, including of any surface treatment residues on dental implant fixtures or dental implant abutments and of the fatigue strength of dental implants, should be performed [12, 13].
Safety
Biological Safety
The biological safety of the product should be evaluated [6, 14, 15]. The notification categorizes the contact site of dental implants as “implantable” and the contact time as “long-term (permanent),” and evaluation based on these categories is required [6]. In this case, the method of using the product should be fully considered. If the packaging material comes into direct or indirect contact with dental implants via the filling fluid in the package, the evaluation should also examine the influence of such contact.
Magnetic Resonance Safety
Magnetic resonance safety should be evaluated as described in the notification of PSEHB/ELD Notification No. 0801-1/PSEHB/SD Notification No. 0801-4 dated August 1, 2019 [16].
Physical Performance and Other Requirements
Ceramic materials show elastic, not plastic, deformation, which eliminates the potential for instantaneous brittle fracture at high applied stress levels. However, the physical strength and toughness of the product should still be evaluated. For instance, the bending strength should be 500 MPa or more (2axis bending) or 800 MPa or more (4-point bending) for dental implant abutments and 800 MPa or more (4-point bending) for dental implant fixtures [7].
The fatigue strength of the assembly should be equivalent, or superior, to that of current approved products or otherwise be clinically acceptable. Each final product of a dental implant fixture, dental implant abutment, dedicated abutment screw, and dental implant superstructure material should be analyzed. In addition, a risk assessment should be performed using the whole system constructed from the final products. The final products should have the highest fracture risk in terms of fatigue strength when they are assembled by tightening at the specified torque [17, 18]. For final products assembled from several components individually sterilized or processed/adjusted before use, the final product should be evaluated by using the individual components with the highest fracture risk. In particular, if part of the assembly is an approved product from another manufacturer, the combination shall be subjected to risk assessment by fatigue test. The material should have sufficient impact resistance as a dental implant and should be equivalent, or superior, to the raw material used for the approved product.
Solubility and degradability should be evaluated [19]. The amount dissolved within 16 h should not exceed 50 µg/cm2. Osteointegration capacity and the status of adverse events should be confirmed in the canine jaw bone [15].
Other Matters
For the product, if it is difficult to confirm their equivalence to an approved product for the evaluations explained above, evaluation by supplementary testing, such as by animal testing, should be considered, depending on the characteristics of the product.
Matters Related to Nonclinical Studies
Clinical evaluation to demonstrate clinical efficacy and safety is not required if there is no novelty to the approved product in terms of shape, dimension, method of use, etc. However, the safety and efficacy of the product should still be appropriately evaluated, with a focus on equivalence to an approved product (Table 2). If there are medical devices used concomitantly, evaluation in the presence of such devices is necessary.
Matters Related to Clinical Studies
If there is novelty in the shape, size, method of use, etc., and evaluation of the efficacy and safety is insufficient or difficult to achieve by nonclinical studies alone, clinical evaluation will be necessary to demonstrate clinical efficacy and safety. If there are medical devices used concomitantly, evaluation in the presence of such devices is also necessary.
Safety should be evaluated by performing medical examinations and tests to confirm that there is no pain, discomfort, change in perception, or sign of infection caused by the dental implant. It should also be evaluated whether the safety of the dental implant is equivalent to that of approved titanium dental implants.
Mobility status, occlusal pain, and amount of bone resorption should be evaluated over time after attachment of the superstructure. With the dental implant fixture in place, it may be acceptable to evaluate the implant stability quotient by resonance frequency analysis and the clinical mobility by using a dynamic periodontal tissue examination/diagnostic device to confirm changes over time in osteointegration. However, evaluation of these parameters alone should not constitute the primary evaluation, and the success or failure of the dental implant treatment should also be evaluated.
Conclusions
Here, we present a set of principles for premarketing evaluation of the safety and efficacy of zirconia-based ceramic dental implants used to restore functional defects and improve esthetics in individuals with missing teeth. These principles arose from a meeting of experts under the “Project for the Development of a Guideline for the Evaluation of Ceramic Dental Implants” of the Ministry of Health, Labour and Welfare of Japan, and they are the principles that are considered most important at present. There are currently few types of ceramic dental implants approved for use in Japan, and it is hoped that these principles will facilitate the clinical development and approval of emerging ceramic dental implant materials.
We intend to revise this set of principles in the future on the basis of technological innovations and the accumulation of knowledge, and they should not be considered binding in regard to the contents of applications for approval. The evaluations mentioned in this document should be performed on the basis of scientific rationality after fully understanding the characteristics of the product, and with reference to other relevant domestic and international guidelines.
Acknowledgements
We thank all members of the committee of the development of a guideline for the evaluation of ceramic dental implants for their useful comments. The views mentioned here do not necessarily represent the views and findings of the Ministry of Health, Labour and Welfare or the Pharmaceuticals and Medical Devices Agency of Japan. This work was supported by a grant from the Ministry of Health, Labour and Welfare of Japan.
Author Contributions
Participated in research design: Hara, Sato, Tanishiro, Tamaki, Baba, Hirose, Yoshida, Watanabe, Nishikawa, Okuda, Murakami, and KondohConducted experiments: Hara, Sato, Tanishiro, Tamaki, Baba, Hirose, Yoshida, Watanabe, Okuda, and KondohPerformed data analysis: Hara, Sato, Tanishiro, Tamaki, Baba, Hirose, Yoshida, Watanabe, Nishikawa, Okuda, Murakami, and KondohWrote, or contributed to the writing of, the manuscript: Hara, Sato, Tanishiro, Tamaki, Baba, Hirose, Yoshida, Watanabe, Nishikawa, Okuda, Murakami, Niwa and Kondoh.
Funding
Open Access funding provided by Osaka University.
Data Availability
No datasets were generated or analysed during the current study.
Declarations
Competing Interests
The authors declare no competing interests.
Conflict of interest
Eiichi Hirose and Kiyoshi Watanabe are employees of PLATON JAPAN and GC CORPORATION, respectively. The other authors declare no conflicts of interest.
References
- 1.Gavriilidou NN, Belibasakis GN. Root caries: the intersection between periodontal disease and dental caries in the course of ageing. Br Dent J. 2019;227:1063–7. [DOI] [PubMed] [Google Scholar]
- 2.Yoshinari M. Future prospects of zirconia for oral implants-A review. Dent Mater J. 2020;39:37–45. [DOI] [PubMed] [Google Scholar]
- 3.Hempel U, Hefti T, Kalbacova M, Wolf-Brandstetter C, Dieter P, Schlottig F. Response of osteoblast-like SAOS-2 cells to zirconia ceramics with different surface topographies. Clin Oral Implants Res. 2010;21:174–81. [DOI] [PubMed] [Google Scholar]
- 4.The Act on Securing Quality. Efficacy and Safety of Products Including Pharmaceuticals and Medical devices (the Pharmaceutical and Medical device act). Article 2, Act 63 of 2019.
- 5.Partial Amendment of Basic Principles of Biological Safety. Evaluation Required for Application for Marketing Approval of Dental Medical Devices. PSEHB/ELD notification 0612-4 dated June 12, 2018.
- 6.Partial Revision of Basic Principles of Biological Safety Evaluation of Dental Medical Devices. PSEHB/ELD Notification No. 0531-5 dated May 31, 2021.
- 7.ISO 13356. Implants for surgery - ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP).
- 8.ASTM F 1873. Standard specification for high-purity dense Yttria Tetragonal Zirconium Oxide Polycrystal (Y-TZP) for Surgical Implant Applications.
- 9.Establishment of Shelf Life of Medical Devices and, Tests S. PFSB/ELD/OMDE Notification No. 0905001 dated September 5, 2008.
- 10.Handling of Stability Tests. Related to establishment of the Shelf Life for Application for Marketing Approval (certification) of Medical devices. PFSB/ELD/OMDE notification 1227-5 dated December 27, 2012.
- 11.Handling of Acceptable Limits of Sterilization Residues after Ethylene Oxide Sterilization. PFSB/ELD/OMDE notification 1012-2 dated October 12, 2010.
- 12.JIS T 14971. Medical devices - Application of risk management to medical devices.
- 13.ISO 14971. Medical devices - Application of risk management to medical devices.
- 14.JIS T 0993-1. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
- 15.JIS T 6001. Evaluation of biocompatibility of medical devices used in dentistry.
- 16.Actions on MR Safety of Implantable Medical. Devices, etc. PSEHB/ELD Notification No. 0801-1/PSEHB/SD Notification No. 0801-4 dated August 1, 2019.
- 17.JIS T 6005. Dynamic fatigue test methods for endosseous dental implants.
- 18.ISO 14801. Dentistry-implants - Dynamic loading test for endosseous dental implants.
- 19.ISO 6872. Dentistry - Ceramic materials.
- 20.Determination of Shelf Life and Stability Studies of Medical Devices. PFSB/ELD/OMDE Notification No. 0905001 dated September 5, 2008.
- 21.Regarding the handling. Of stability tests related to setting the validity period when applying for medical device manufacturing and marketing approval (certification). PFSB/ELD/OMDE notification 1227-5 dated December 27, 2012.
- 22.Revision of the Standards for Approval of Dental Implants. PSEHB Notification No. 1102-2 dated November 2, 2022.
- 23.JIS T. 6526, Dental ceramic materials.
- 24.JIS T 6006, Dentistry --. Test Method for Determining Radio-Opacity of Materials.
- 25.ISO 13116. Dentistry -Test method for determining radio-opacity of materials.