Table 2.
Qualities of dosimeters suitable for IVD.
Dosimeter | Thermoluminescent Dosimeters (TLDs) | Semiconductor Diodes | Metal oxide semiconductor field effect transistor (MOSET) | Radioluminescence (RL) | Plastic Scintillator Detector (PSD) | Optically Stimulated Luminescent Detectors (OSL) |
---|---|---|---|---|---|---|
Advantages | − Small − High Sensitivity − No angular dependence − Dynamic range − Low cost − Well studied system |
− Variable size − High sensitivity − Dynamic range − Simple calibration − Reasonable cost − Online dosimetry |
− Very small size − High sensitivity − Dynamic range −Low cost − Online dosimetry |
− Very small size − Very sensitive − No angular dependent − No energy dependence − Dynamic range − Online dosimetry |
− Very small size − No angular dependence − No energy dependence − Online dosimetry − Simple calibration |
− Stable − No directional effects − Dose rate independent |
Disadvantages | − No on-line dosimetry − Energy dependent − Requires calibration and preparation for read-out |
− Energy dependent − Angular dependence |
− Energy dependent − Angular dependence (due to epoxy resin bulb) − Limited lifespan (20,000 mV) |
− Frequent recalibration − Stem effect − Not commercially available |
− Stem/Cherenkov effect* | − Stem/Cherenkov effect* − Need a laser reading |
Stem effect = unwanted light generated in the fibre cable upon irradiation.