Abstract
Purpose
To develop a 3D distortion‐free reduced‐FOV diffusion‐prepared gradient‐echo sequence and demonstrate its application in vivo for diffusion imaging of the spinal cord in healthy volunteers.
Methods
A 3D multi‐shot reduced‐FOV diffusion‐prepared gradient‐echo acquisition is achieved using a slice‐selective tip‐down pulse in the phase‐encoding direction in the diffusion preparation, combined with magnitude stabilizers, centric k‐space encoding, and 2D phase navigators to correct for intershot phase errors. The accuracy of the ADC values obtained using the proposed approach was evaluated in a diffusion phantom and compared to the tabulated reference ADC values and to the ADC values obtained using a standard spin echo diffusion‐weighted single‐shot EPI sequence (DW‐SS‐EPI). Five healthy volunteers were scanned at 3 T using the proposed sequence, DW‐SS‐EPI, and a clinical diffusion‐weighted multi‐shot readout‐segmented EPI sequence (RESOLVE) for cervical spinal cord imaging. Image quality, perceived SNR, and image distortion were assessed by two expert radiologists. ADC maps were calculated, and ADC values obtained with the proposed sequence were compared to those obtained using DW‐SS‐EPI and RESOLVE.
Results
Consistent ADC estimates were measured in the diffusion phantom with the proposed sequence and the conventional DW‐SS‐EPI sequence, and the ADC values were in close agreement with the reference values provided by the manufacturer of the phantom. In vivo, the proposed sequence demonstrated improved image quality, improved perceived SNR, and reduced perceived distortion compared to DW‐SS‐EPI, whereas all measures were comparable against RESOLVE. There were no significant differences in ADC values estimated in vivo for each of the sequences.
Conclusion
3D distortion‐free diffusion‐prepared imaging can be achieved using the proposed sequence.
Keywords: diffusion imaging, diffusion‐prepared imaging, distortion‐free diffusion, reduced FOV diffusion, spinal cord DWI
1. INTRODUCTION
DWI is a powerful tool for probing the microstructure to assess a range of pathologies, including stroke, tumors, and inflammatory conditions. 1 , 2
Conventional DWI employs a Stejskal–Tanner pulsed gradient spin echo 3 in combination with a single‐shot EPI (SS‐EPI) readout, which is efficient and insensitive to motion. 4 However, SS‐EPI is prone to geometrical distortion, signal loss, and blurring, which are exacerbated at higher field strengths due to increased off‐resonance and shorter relaxation times. Signal loss and image blurring are also increased with the longer readout durations required for higher spatial resolution, which is a key benefit of imaging at higher field strengths.
Scan acceleration using parallel imaging 5 , 6 to reduce the readout duration and the effective echo spacing can help to reduce 7 , 8 , 9 but cannot fully eliminate these effects, especially for imaging regions with high susceptibility gradients 10 such as the frontal lobes, skull base, cervical spine, and near metal implants where shimming of the static magnetic field can be very challenging.
A further reduction in the readout duration and effective echo spacing can be achieved using multi‐shot EPI, including interleaved EPI 11 , 12 or readout‐segmented EPI. 13 , 14 The main challenge introduced with these multi‐shot approaches is the intershot nonlinear phase errors due to physiological motion in the presence of strong diffusion gradients. It is therefore necessary to correct for these phase offsets before combining the acquired shots. Solutions for phase correction include phase navigators, with initial methods proposing a 1D linear phase correction using 1D navigator echoes 15 , 16 and the use of 2D navigators for correcting linear and constant phase offsets in k‐space. 12 2D nonlinear phase correction was proposed in DW‐PROPELLER 17 and was later formulated in a general refocusing reconstruction framework. 18 Other approaches that do not involve an explicit acquisition of a phase navigator have also been proposed. These include self‐navigated techniques that derive shot‐dependent phase maps from the acquired data, 19 , 20 , 21 , 22 , 23 , 24 , 25 , 26 navigator‐free reconstruction techniques that recover uncorrupted data via a regularization prior such as a low rank constraint 27 , 28 , 29 , 30 , 31 , 32 , 33 or a deep learning–based prior, 34 , 35 , 36 as well as a combination of these approaches. 37
Another approach to reduce distortions in diffusion imaging is to use reduced FOV (RFOV) EPI techniques 38 , 39 , 40 by restricting the FOV in the phase‐encoding direction with, for example, outer volume suppression or 2D RF excitation pulses. RFOV approaches offer some advantages over parallel imaging, including a simpler implementation of the phase correction (no parallel imaging reconstruction is required for the navigator shots, which would also be undersampled to obtain a phase map with sufficient resolution), no g‐factor SNR penalty, and avoiding artifacts and unwanted signals from short T1, high‐signal producing regions (e.g., chest and/or back fat for spinal cord imaging) outside of the anatomical region of interest (ROI).
However, distortion may still be present in areas of high susceptibility gradients. Other EPI‐based approaches 41 , 42 , 43 achieve distortion‐free diffusion imaging using techniques such as point‐spread‐function EPI encoding, 44 , 45 which combines EPI with additional phase encoding steps; or echo planar time‐resolved imaging, 46 which enables high acceleration of a ky‐t acquisition to reconstruct distortion‐free images at each time point. However, these methods are based on additional encoding, therefore prolonging scan times.
Other techniques have proposed using an alternative imaging readout to EPI. For example, diffusion‐preparation (DP) modules have been proposed, which can be combined with any imaging readout to completely avoid distortion even in regions of high‐susceptibility gradients. 47 The DP module typically consists of a nonselective tip‐down pulse, followed by diffusion‐sensitizing gradients, refocusing pulses, and a nonselective tip‐up pulse, which stores the diffusion‐prepared magnetization in the longitudinal direction prior to readout. One of the main challenges with diffusion‐prepared imaging is that phase errors due to motion and eddy currents will result in spatially variable magnitude errors. 48 Several solutions have been proposed, including preparatory gradients, 49 motion‐compensated diffusion gradients, 50 phase cycling of the tip‐up pulse, 48 , 51 and magnitude stabilizers. 52 , 53 Recent work suggests that magnitude stabilizers are essential for correction of motion‐related phase errors. 54 The use of magnitude stabilizers in the DP combined with rewinder gradients in the imaging module leads to the generation of stimulated echoes, and is therefore closely related to the diffusion‐weighted STEAM (DW STEAM) 55 and particularly to the DW turbo‐STEAM sequence, 56 , 57 , 58 , 59 which extends DW STEAM using a DP with slice‐selective 90° and 180° pulses.
DP has been combined with turbo spin echo (TSE), 60 , 61 , 62 , 63 balanced steady‐state free precession, 50 , 64 , 65 and gradient‐echo (GRE) readouts. 48 , 51 , 53 Each of these readouts have advantages and limitations, which need to be considered for a particular application and field strength. Whereas TSE and balanced steady‐state free precession offer higher relative SNR, GRE uses RF pulses with lower flip angles and therefore has a lower specific absorption rate (SAR) burden, 66 which is advantageous at higher field strengths. One potential disadvantage of the GRE readout for diffusion‐prepared imaging is the introduction of T1 weighting, which can be minimized by employing a centric encoding. 51 T1 effects can also impact on accuracy of calculated ADC values. These biases can be avoided by employing magnitude stabilizers 67 or by employing phase cycling and ensuring the delay time before each DP is slightly greater than the T1, which is also the condition for optimal SNR efficiency. 48
In this work, we present a novel technique for distortion‐free diffusion‐prepared imaging, which extends the DW turbo‐STEAM and the diffusion‐prepared stimulated echo turboFLASH 53 sequences to a 3D multi‐shot, reduced FOV acquisition. The proposed sequence uses a multi‐shot 3D GRE readout, combined with centric encoding, magnitude stabilizers, and reduced FOV. The technique is demonstrated in a diffusion phantom and for 3 T sagittal imaging of the cervical spinal cord, which is particularly prone to distortion using a conventional SS‐EPI acquisition. Part of this work has been presented at the International Society for Magnetic Resonance in Medicine 2024 annual meeting, 68 where a parallel and independent work pursued a similar idea. 69
2. METHODS
2.1. Sequence design and image reconstruction
The sequence is similar in its design and principles to the DW turbo‐STEAM 56 and diffusion‐prepared stimulated echo turboFLASH 53 sequences and enables a 3D multi‐shot diffusion‐prepared acquisition with reduced FOV and explicit 2D phase navigation. The sequence (Figure 1) consists of a twice‐refocused DP scheme, paired with a centric‐encoded 3D GRE readout, followed by a linearly encoded 2D phase correction navigator. Each imaging shot acquires a single partition of the 3D acquisition, that is, in each imaging shot all the in‐plane phase encoding steps are acquired for one partition‐encoding step. The 2D phase navigator is acquired with the same excitation pulse but without any partition encoding and with lower resolution in the phase‐encoding direction. The k‐space acquisition for 3D diffusion imaging and 2D phase navigation is illustrated in Figure 2. The DP is preceded by a chemical shift‐selective fat‐saturation pulse. The tip‐down pulse in the DP is applied simultaneously with a slab‐selective gradient in the phase‐encoding direction, thereby restricting the tip‐down excitation in the DP module to water protons within a reduced FOV. The tip‐down excitation is followed by a pair of bipolar diffusion gradients, each surrounding a 180° adiabatic hyperbolic secant refocusing pulse. A magnitude stabilizer approach is used to avoid phase‐offset–induced magnitude errors and convert these into conventional intershot phase errors. Finally, a nonselective 90° tip‐up pulse is applied. It is worth noting that whereas this acts as a tip‐up pulse for the inner‐volume magnetization, it acts as a tip‐down pulse for the unprepared outer‐volume magnetization and any recovered fat magnetization. A spoiler gradient is executed immediately after the DP to dephase any remaining transverse magnetization, including outer volume and/or fat protons. The original phase of the spins is restored by a magnitude stabilizer rewinder gradient, which is played out after each RF readout pulse and importantly also acts as a spoiler gradient for the outer volume and fat magnetization.
FIGURE 1.

Pulse sequence diagram for 3D diffusion‐prepared gradient echo acquisition. A chemical shift‐selective fat saturation pulse (orange) precedes the tip‐down pulse (yellow), which is applied in combination with a slab‐selective gradient in the phase‐encoding direction. This is followed by a pair of bipolar diffusion gradients, each surrounding a 180° adiabatic hyperbolic secant refocusing pulse (purple). A magnitude stabilizer dephasing gradient (blue) is applied before the nonselective tip‐up pulse (green). The magnitude stabilizer re‐phasing gradient (red) is incorporated in the slice rewinder gradient as an additional moment after each readout excitation pulse.
FIGURE 2.

K‐space sampling for the 3D diffusion acquisition and 2D phase navigator. Each colored circle represents the acquisition of a frequency‐encoded line (kx) in the 3D diffusion acquisition, with centric ordering in the phase‐encoding (ky) direction (for each shot, blue‐colored k‐space points are acquired earlier than red‐colored k‐space points). Each shot of the 3D diffusion acquisition acquires all ky samples for a single partition‐encoding step (i.e., one row). The rectangle at the centre of k‐space indicates the k‐space samples acquired with the 2D navigator, which are acquired with linear encoding and are identical for every shot.
The remaining inner‐volume intershot phase variations are corrected with the 2D phase navigator in the reconstruction: a Hamming filter is applied to each phase navigator in k‐space for each receiver coil to reduce Gibbs ringing. A 2D inverse Fourier transform is then applied to both the Hamming‐filtered phase navigator data and to each diffusion‐prepared imaging shot for each receiver coil to obtain 2D (complex) images, N i (x,y) and S i (x,y), for the i‐th navigator and diffusion‐prepared shot, respectively. The Hamming‐filtered phase navigator data are then normalized to have a magnitude equal to 1. A complex conjugate multiplication of the two respective datasets is then performed in image space. 13 The navigator normalization and phase correction of S i (x,y) are summarized in the following equation:
where S i , corr (x, y) is the result of the phase correction of S i (x, y), and N* i (x, y) and |N i (x, y)| are the complex conjugate and the magnitude of N i (x, y), respectively.
Finally, a Fourier transform is applied to produce the phase‐corrected image. 18 The reconstruction was implemented inline in the vendor reconstruction software (Syngo MR E11P, Siemens Healthineers AG, Erlangen, Germany).
2.2. Imaging studies
All imaging was performed at 3 T (Biograph mMR, Siemens Healthineers AG). Phantom imaging was acquired with a 16‐channel head and neck coil, and imaging in vivo was acquired with a 32‐channel spine array and posterior section of a 16‐channel head and neck coil.
2.2.1. Phantom study
A National Institute of Standards and Technology (NIST)–traceable diffusion phantom (Caliber MRI, Boulder, CO) with six vials containing varying concentrations of polyvinylpyrrolidone (PVP) and water was used in the phantom experiments.
The phantom was imaged with the proposed sequence and a standard 2D DW‐SS‐EPI sequence to evaluate the accuracy of ADC estimates. The diffusion sequences were acquired with matched parameters where possible, as detailed in Table 1. Additional parameters for the proposed sequence include number of navigator lines: 14, one dummy scan for magnetization preparation, and 4π intravoxel phase dispersion for the magnitude stabilizer, and the tip‐down pulse in the DP is a 4 ms Hamming‐filtered sinc pulse with time‐bandwidth product = 12.
TABLE 1.
Acquisition parameters for phantom and in vivo studies.
| Phantom | In Vivo | ||||
|---|---|---|---|---|---|
| Proposed | SS‐EPI | Proposed | SS‐EPI | RESOLVE | |
| b‐values (s/mm2) | 0, 500 | 0, 500 | 0, 500 | 0, 500 | |
| Averages | 1, 1 | 1, 5 | 1, 5 | ||
| In‐plane resolution (mm × mm) | 2.1 × 2.1 | 2.1 × 2.1 | |||
| Slice thickness (mm) | 4 | 4 | |||
| Number of slices | 40 | 8 | |||
| Partial Fourier | None | None | |||
| FOV read × FOV phase (mm2) | 200 × 75 | 430 × 263 | 200 × 50 | 430 × 263 | 220 × 112 |
| TR (ms) | 2000 a | 15600 b | 2000 a | 4200 b | 2800 b |
| Echo spacing (ms) | 4.21 | 0.66 | 4.21 | 0.66 | 0.38 |
| TE (ms) | 2.36 | 79 | 2.36 | 79 | 62 |
| Diffusion prep time (ms) | 65 | N/A | 65 | N/A | N/A |
| Excitation flip angle (°) | 10 | 90 | 10 | 90 | 90 |
| Phase‐encode lines per shot | 41 | 63 | 27 | 63 | 48 |
| Readout segments | 5 | ||||
| Acceleration | None | GRAPPA 2 | None | GRAPPA 2 | GRAPPA 2 |
| Fat suppression | FatSat 95 | STIR 96 | FatSat 95 | STIR 96 | FatSat with gradient reversal 97 |
| Receiver bandwidth (Hz/Px) | 410 | 2025 | 410 | 2025 | 945 |
| Phase‐encode bandwidth (Hz/Px) | 24 | 24 | 53 | ||
| Phase oversampling | 10% | 0% | 10% | 0% | 75% |
| Slice oversampling | 0% | 50% | |||
| Acquisition time (min:s) | 5:22 | 5:12 | 6:56 | 1:26 | 3:54 |
Abbreviations: FatSat, fat saturation; STIR, short tau inversion recovery; RESOLVE, readout‐segmented EPI diffusion sequence; SS‐EPI, single shot EPI.
For the proposed sequence, TR is the time between successive shots (partition encodings) of the 3D acquisition.
For EPI and RESOLVE sequences, TR is the time between successive excitations of the same slice.
ADC maps were calculated for both sequences using mono‐exponential fitting of the signal from the b0 and b500 trace‐weighted images.
2.2.2. In vivo evaluation
Five subjects (four male, one female, mean age 30 ± 4 years) were prospectively recruited for the study. This study was approved by the institutional ethics committee (REMAS 8700), and written informed consent was obtained from all participants for the scan and inclusion in this study. The cervical spinal cord was imaged with the proposed sequence, a standard 2D DW‐SS‐EPI sequence, clinical 2D multi‐shot readout‐segmented EPI diffusion sequence (RESOLVE), 13 and 2D T2‐weighted TSE sequence as anatomical reference. The sequence parameters for all diffusion sequences are detailed in Table 1. Additional parameters for the proposed sequence were matched to the phantom study.
ADC maps were calculated for all sequences using mono‐exponential fitting of the signal from the b0 and b500 trace‐weighted images. For all diffusion sequences, a ROI in the upper cervical spinal cord was manually delineated on the ADC map on the imaging slice that captured the centre of the spinal cord. For each volunteer and each diffusion sequence, the ADC values were estimated as the mean ADC in the ROI. For statistical analysis, a Wilcoxon signed‐rank test was performed between the ADC values of the proposed sequence and the ADC values of DW‐SS‐EPI and RESOLVE, with a p‐value <0.05 indicating a significant difference.
2.3. Qualitative assessment
The images were assessed qualitatively independently by two radiologists with >20 and >5 years of experience in MRI. Overall image quality (1 = nondiagnostic, 2 = poor, 3 = fair, 4 = good, 5 = excellent), perceived distortion (1 = major distortion resulting in nondiagnostic image quality, 2 = major distortion but not limiting diagnosis, 3 = minor distortion but not limiting diagnosis, 4 = no distortion), and perceived SNR (1 = poor, 2 = fair, 3 = good, 4 = very good, 5 = excellent) were assessed for each subject/acquisition.
3. RESULTS
3.1. Phantom study
ADC maps in the diffusion phantom calculated for the proposed sequence and the conventional DW‐SS‐EPI sequence are shown in Figure 3A, and a plot of the obtained ADC values using both sequences against the reference ADC values provided by the manufacturer of the phantom is shown in Figure 3B. There was a good correlation between the ADC values obtained from the proposed sequence and the reference ADC values (slope and intercept of linear fit: 0.98 and 0.06 × 10−3 mm2/s R2 = 1.0, RMS error = 0.05 × 10−3 mm2/s). A similar correlation was observed between the ADC values obtained from the DW‐SS‐EPI sequence and the reference ADC values (slope and intercept of linear fit: 1.01 and 0.03 × 10−3 mm2/s, R2 = 1.0, RMS error = 0.04 × 10−3 mm2/s).
FIGURE 3.

Diffusion phantom measurements. (A) ADC maps calculated using the conventional SS‐EPI diffusion sequence (top) and the proposed sequence (bottom). (B) Plot comparing the mean and SDs of ADC values in six vials measured with the conventional SS‐EPI sequence (green) and the proposed sequence (blue) against the reference values (purple) provided by the manufacturer. Proposed and conventional measurements are slightly offset in the x‐direction for clarity. SS‐EPI, single shot EPI.
3.2. In vivo evaluation
The left column of Figure 4 shows one slice from the T2‐weighted TSE and b0 images for the proposed sequence, DW‐SS‐EPI, and RESOLVE for all five volunteers. The middle and right columns of Figure 4 show the same slice from the b500 trace‐weighted images and calculated ADC maps, respectively. Contour plots of the CSF and cord, delineated on the TSE images, are shown overlaid on the b0 and b500 images in Supporting Information Figure S1 to better highlight the distortion properties of each sequence. Figure 5 shows the estimated ADC values in the ROI for each volunteer and diffusion sequence. There were no significant differences in ADC estimates measured with the proposed sequence (1.06 ± 0.12 × 10−3 mm2/s) compared with DW‐SS‐EPI (0.90 ± 0.11 × 10−3 mm2/s, p = 0.19) or RESOLVE (0.86 ± 0.06 × 10−3 mm2/s, p = 0.13). The mean image quality, perceived distortion, and perceived SNR scores from the qualitative assessment are shown in Table 2.
FIGURE 4.

Example images acquired in vivo: T2‐weighted TSE (top row), b0 (left column); b500 trace‐weighted images (middle column); and ADC maps (right column) for the proposed sequence, DW‐SS‐EPI, and RESOLVE for all five volunteers. It is noted that the contrast between the bone marrow and intervertebral discs is lower for the proposed sequence compared to the reference sequences, which may be attributed to complex averaging of b500 images with the proposed sequence, leading to elevated ADC values in low SNR regions (such as bone marrow). DW, diffusion‐weighted; SS‐EPI, single shot EPI; RESOLVE, readout‐segmented EPI diffusion sequence; TSE, turbo spin echo.
FIGURE 5.

Mean and SDs of measured ADC values in ROIs in the upper cervical spine for the proposed sequence, DW‐SS‐EPI, and RESOLVE in the five healthy volunteers. DW, diffusion‐weighted; SS‐EPI, single shot EPI; RESOLVE, readout‐segmented EPI diffusion sequence; ROI, region of interest.
TABLE 2.
Mean scores from qualitative assessment by two expert readers of b500 images acquired in vivo in five healthy volunteers.
| DW‐SS‐EPI | RESOLVE | Proposed | |
|---|---|---|---|
| Image quality | 1.5 ± 0.4 | 3.8 ± 0.3 | 3.3 ± 0.3 |
| Perceived distortion | 1.5 ± 0.4 | 3.3 ± 0.4 | 3.6 ± 0.2 |
| Perceived SNR | 1.8 ± 0.4 | 3.7 ± 0.3 | 3.3 ± 0.3 |
4. DISCUSSION
In this study, a 3D RFOV diffusion‐prepared GRE sequence is presented for distortion‐free diffusion imaging. The sequence was tested in vivo for spinal cord DWI and compared to standard DW‐SS‐EPI and to a clinical multi‐shot DWI sequence with reduced distortions (readout‐segmented EPI, RESOLVE 13 ), which has been successfully applied for spine DWI (among other applications) in previous studies. 70 , 71 The proposed sequence demonstrated improved image quality, improved perceived SNR, and reduced perceived distortion compared to DW‐SS‐EPI, whereas all measures were comparable against RESOLVE. Furthermore, all images acquired in vivo demonstrated effective suppression of outer‐volume and fat signal. The obtained ADC values in a phantom and in the upper cervical spine in healthy volunteers were consistent with those obtained using the reference sequences (DW‐SS‐EPI for the phantom scans; DW‐SS‐EPI and RESOLVE for the study in vivo).
The RFOV approach is a key element of the sequence design because this reduces the readout duration and enables acquisition of all the lines of a partition within a single shot while maintaining sufficient signal for the navigator acquisition, which is acquired after the imaging readout and permits a straightforward phase correction.
The efficacy of the novel RFOV method introduced in this study can be attributed to the outer‐volume signal suppression achieved using a slice‐selective/nonselective tip‐up/tip‐down DP, combined with a magnitude stabilizer approach, and this combination is the main contribution of this paper. The outer volume is not excited by the tip‐down pulse of the DP. Therefore, the “tip‐up” pulse and subsequent spoiler gradient serve to saturate the outer volume magnetization before the readout. On its own, this outer‐volume suppression may be insufficient because small inefficiencies in the outer‐volume saturation or regrowth of outer‐volume magnetization over the course of the readout may still significantly contaminate the low‐signal diffusion‐prepared magnetization. 54 Therefore, it is important to also recognize the contribution of the magnitude stabilizer rewinder gradient to the suppression of outer volume signal: The dephasing gradient of the magnitude stabilizer in the DP, applied before the tip‐up pulse, does not influence magnetization in the outer volume because it has not been excited by the tip‐down pulse, and thus the magnitude stabilizer rewinder gradient serves to spoil any signal from outer‐volume magnetization that recovers during the course of the readout.
ADC mapping for the proposed sequence was performed using mono‐exponential fitting, as proposed in the stimulated echo turboFLASH sequence. 53 This is possible because the magnitude stabilizers spoil the transverse magnetization, and therefore the net longitudinal magnetization is zero after the tip‐up pulse. Therefore, after one dummy shot, the steady‐state longitudinal magnetization available before the tip‐down pulse is independent of the diffusion weighting, that is, it will be the same independently of the used b‐value (but still dependent on T1 recovery). In addition, the regrowing longitudinal magnetization does not contribute to the stimulated echo pathway. These factors make the ADC estimation more straightforward for this sequence than methods using analytical signal fitting or dictionary matching with a forward simulation of the sequence in which prior knowledge of T1 and T2 relaxation times is required. 48 , 51
Fat suppression is critical in EPI‐based diffusion due to chemical shift–related displacement in the low‐bandwidth phase‐encoding direction. 72 In the proposed sequence, effective fat suppression is also desirable to achieve optimal image quality. Chemical shift–related displacement of fat in the phase‐encoding direction when applying the slab‐selective tip‐down pulse may lead to fat signal outside of the reduced FOV being excited by the tip‐down pulse and folding back in the reduced FOV. This effect may be exacerbated due to the tip‐down pulse being a relatively long 90° pulse exciting a larger slab thickness than typical 2D slices, leading to a low slab‐selection gradient strength. Unsuppressed fat signal may also be detrimental for the 2D phase navigator because the 2D navigator is essentially a projection of a 3D slab and the projection of surrounding fat tissue might overlap with the ROI, resulting in a less reliable phase estimation. Such contamination in the diffusion‐weighted images and in the phase navigator is particularly problematic due to the low ADC of fat, resulting in hyperintense signal compared to the tissue of interest. Furthermore, radiologists are more familiar with reading fat‐suppressed DWI because this is the standard in the clinic. Fat signal was successfully suppressed in our spinal cord experiments in a similar manner, as described above for outer‐volume signal because the fat saturation pulse, when effective, ensures that fat signal is also not excited by the tip‐down pulse of the preparation module. Whereas the fat signal was successfully suppressed in all images acquired in vivo in this study, there may be cases/anatomies where the chemical shift‐selective saturation fails, especially in regions with high B0 inhomogeneity and magnetic susceptibility gradients. However, this sequence could be combined with a multi‐echo Dixon readout for fat–water separation to achieve a more robust fat suppression. 33 , 73 , 74
From a sequence design perspective, the proposed approach builds on principles proposed in DW turbo‐STEAM 56 , 58 and diffusion‐prepared stimulated echo turboFLASH 53 to enable a 3D multi‐shot reduced FOV acquisition. A reduced FOV version of the DW turbo‐STEAM sequence was previously proposed and applied for diffusion imaging of the spine 75 ; however, this sequence was restricted to a single‐shot 2D acquisition and used slice‐selective 90° pulses in orthogonal directions to achieve FOV reduction, whereas the proposed sequence uses a combination of slice‐selective and nonselective tip‐down/‐up pulses. A 2D multi‐shot self‐navigated extension to DW turbo‐STEAM was proposed using a radial trajectory and a nonlinear reconstruction algorithm 58 ; however, the proposed sequence combines a 3D acquisition with a Cartesian trajectory, 2D phase navigators, and reduced FOV DP.
Compared to diffusion‐prepared TSE sequences, 60 , 61 , 62 , 63 diffusion‐prepared GRE sequences offer an alternative for non‐EPI diffusion imaging with a much lower SAR deposition burden. This is because SAR is proportional to the square of the RF flip angles, 76 and the proposed sequence uses low flip angle RF excitation pulses (if we consider the oversimplified situation where pulse shapes and durations are the same, and for 180° TSE refocusing pulses, the energy of the 10° excitation pulse in the proposed sequence amounts to 0.3% of the energy of the TSE refocusing pulse). This could be beneficial especially given that the DP requires the use of a pair of SAR‐intensive adiabatic refocusing pulses at 3 T to achieve robustness to B1 (and B0) inhomogeneity effects. 77
A further difference between the proposed sequence and diffusion‐prepared TSE is that whereas TSE readouts are affected by T2 decay along the echo train, the proposed sequence with a centric trajectory is affected by the T1 decay of the stimulated echoes. 75 , 78 Whereas both may lead to blurring artifacts, T1 relaxation times are typically longer than T2 relaxation times, and the achieved echo spacings for GRE readouts are usually shorter than for TSE readouts. Thus, it is expected that the blurring effects related to T1 decay are less pronounced than those due to T2 decay. However, T1‐related blurring may be observed for species with short T1 relaxation times such as unsuppressed fat or tissue after contrast agent injection.
Despite the advantages of increased Fourier averages due to partition encoding with a 3D acquisition and the lower receiver bandwidths per pixel, the SNR efficiency of the proposed sequence is relatively low compared to the DW‐SS‐EPI and RESOLVE sequences. This is in part due to the inherent 50% signal loss incurred by all diffusion sequences, which employ magnitude stabilizers due to the creation of a stimulated echo. 79 In addition, the use of a low flip angle GRE readout further reduces the expected signal compared to a spin echo EPI sequence. Two‐dimensional diffusion‐weighted sequences also benefit from slice interleaving, which increases the scan efficiency compared to the proposed 3D sequence, where time is needed for magnetization recovery after each GRE shot, as evidenced by the shorter scan times of the 2D DW‐SS‐EPI and RESOLVE sequences. However, other factors impacting the SNR efficiency of the proposed sequence are application‐specific. For example, application of this sequence to regions requiring an increased number of partitions could offer an increase in SNR efficiency because signal averaging of the high b‐value acquisitions could be traded for Fourier averaging by increasing the number of partition‐encoding steps, resulting in preserved SNR and scan time with increased coverage. Another application‐specific factor impacting SNR is the shot‐to‐shot TR. A relatively long shot‐to‐shot TR (2 s) was used for the proposed sequence in this study to ensure sufficient longitudinal magnetization recovery and a high steady‐state signal in the spinal cord. However shorter shot‐to‐shot TRs could be achieved and could be beneficial for the study of regions with short T1 relaxation times, resulting in improved SNR efficiency while achieving much lower SAR deposition than for diffusion‐prepared TSE.
Our spinal cord experiments were performed using thin slabs (32 mm) where 2D phase correction is adequate. 80 For thicker slabs, the 2D navigator phase correction could be less accurate due to through‐slab phase variations, 81 and in these cases it may be necessary to implement an accelerated 3D navigator for robust phase correction. 65 , 82 , 83 , 84 , 85 Furthermore, alternative self‐navigated phase correction approaches for 3D multi‐shot diffusion imaging have been proposed 25 and could be used to replace the 2D phase navigators.
A limitation of the proposed technique is that it may be sensitive to imperfections in the tip‐down and tip‐up pulses of the DP. B1 inhomogeneities, which are more pronounced at 3 T compared to 1.5T, 86 affect both pulses and may lead to a decrease in the available diffusion‐prepared magnetization within the reduced FOV. We would like, however, to note that this limitation is not specific to this sequence and applies to diffusion‐prepared sequences using standard nonselective tip‐down and tip‐up pulses. A further effect affecting the nonselective tip‐up pulse is that B1 inhomogeneities outside of the reduced FOV would also affect the amount of residual unprepared longitudinal magnetization from the outer volume at the end of the DP, which may increase sensitivity to imperfections in the spoiling using magnetization stabilizer rewinders. The performance of the FOV reduction may also depend on the slice profile of the tip‐down pulse, and we have mitigated this in the proposed sequence by applying a small amount of phase oversampling and using a tip‐down pulse with a high time‐bandwidth product.
A further limitation of the technique is that the required phase FOV may be subject‐dependent, especially for imaging the spinal cord, for which adaptation of the phase FOV may be needed depending on spine curvature. This would lead to variability in the achieved shot durations for the proposed sequence and hence variability in the benefits of the technique such as reducing T1 decay across the shot. In addition, the proposed sequence is inherently limited in its use to organs in which a reduced FOV approach is suitable and therefore has less applicability than other multi‐shot DWI techniques with reduced distortion such as readout‐segmented EPI. 13
Pulsatile cord and CSF motion are known to introduce artifacts when imaging the spinal cord, 87 leading to signal voids in diffusion‐weighted images due to incoherent intravoxel motion, 88 even for single‐shot diffusion EPI sequences. 89 These signal voids pose a challenge for accurate and reproducible ADC mapping along the spinal cord, 87 , 89 which may give rise to some of the variability in ADC values observed in vivo in the current study (Figure 5). Similarly to a previous study, 90 we have therefore focused on the upper cervical spine for our ADC mapping study because it is less susceptible to cardiac and respiratory motion. Whereas cardiac or pulse triggering have also been used to mitigate against the pulsatility of the spinal cord, 91 this is not desirable in the sequence presented because it would introduce a variable magnetization recovery delay. An alternative solution is to use optimized diffusion gradient waveforms, such as motion‐compensated diffusion gradients to minimize the effect of bulk motion 50 , 89 ; however, these may lead to prolonged DP times and more pronounced signal loss due to T2 relaxation.
This is a proof‐of‐concept study, and thus we limited our exploration of this technique to the cervical spinal cord in a small number of healthy volunteers. Other applications with restricted imaging regions could also benefit from this RFOV technique, such as the prostate, cervix, lumbosacral/sciatic, and optic nerves. Further testing of this sequence is required in regions with increased off‐resonance and for radiotherapy applications in which geometric fidelity requirements are stringent. In our sequence evaluation, we did not include a comparison with reduced FOV SS‐EPI diffusion sequences, which are expected to exhibit reduced distortions compared to standard SS‐EPI sequences, as shown in previous spinal cord imaging studies. 38 , 40 , 90 The phase FOV of the DW‐SS‐EPI acquisition was not optimized for acquisition in the cervical spine; however, DW‐SS‐EPI is not the gold standard clinical sequence for spinal cord DWI, and pronounced distortions would be expected to persist. 90 The sequence was acquired with a relatively large slice thickness (4 mm) to achieve acceptable SNR; therefore, more advanced reconstruction techniques such as deep learning–based reconstruction for diffusion imaging 92 , 93 , 94 could be considered to enable an isotropic resolution 3D scan with diagnostic image quality without a significant scan time penalty.
5. CONCLUSION
In this work, we proposed a novel 3D reduced FOV, diffusion‐prepared GRE sequence and showed initial results for feasibility of spinal cord diffusion imaging in healthy volunteers. A clinical study in patients is warranted to further evaluate the potential of this technique.
CONFLICT OF INTEREST STATEMENT
Dr Sarah McElroy and Dr Raphael Tomi‐Tricot are employed by Siemens Healthineers. The rest of the authors do not have any conflicts of interest related to this work.
Supporting information
Figure S1. Example in‐vivo images of the T2‐weighted TSE (top row), b0 (left, bottom 3 rows) and b500 trace‐weighted images (right, bottom 3 rows) for the proposed sequence, DW‐SS‐EPI and RESOLVE for all 5 volunteers with contour plots of CSF and spinal cord overlaid on the images to better highlight the distortion properties of each sequence.
McElroy S, Tomi‐Tricot R, Cleary J, et al. 3D distortion‐free, reduced FOV diffusion‐prepared gradient echo at 3 T . Magn Reson Med. 2025;93:1471‐1483. doi: 10.1002/mrm.30357
REFERENCES
- 1. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval‐Jeantet M. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology. 1986;161:401‐407. [DOI] [PubMed] [Google Scholar]
- 2. Le Bihan D. Looking into the functional architecture of the brain with diffusion MRI. Nat Rev Neurosci. 2003;4:469‐480. [DOI] [PubMed] [Google Scholar]
- 3. Stejskal EO, Tanner JE. Spin diffusion measurements: spin echoes in the presence of a time‐dependent field gradient. J Chem Phys. 1965;42:288‐292. [Google Scholar]
- 4. Turner R, Le Bihan D, Chesnick AS. Echo‐planar imaging of diffusion and perfusion. Magn Reson Med. 1991;19:247‐253. [DOI] [PubMed] [Google Scholar]
- 5. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952‐962. [PubMed] [Google Scholar]
- 6. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
- 7. Bammer R, Keeling SL, Augustin M, et al. Improved diffusion‐weighted single‐shot echo‐planar imaging (EPI) in stroke using sensitivity encoding (SENSE). Magn Reson Med. 2001;46:548‐554. [DOI] [PubMed] [Google Scholar]
- 8. Bammer R, Auer M, Keeling SL, et al. Diffusion tensor imaging using single‐shot SENSE‐EPI. Magn Reson Med. 2002;48:128‐136. [DOI] [PubMed] [Google Scholar]
- 9. Skare S, Newbould RD, Clayton DB, Albers GW, Nagle S, Bammer R. Clinical multishot DW‐EPI through parallel imaging with considerations of susceptibility, motion, and noise. Magn Reson Med. 2007;57:881‐890. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med. 1995;34:65‐73. [DOI] [PubMed] [Google Scholar]
- 11. Butts K, Riederer SJ, Ehman RL, Thompson RM, Jack CR. Interleaved echo planar imaging on a standard MRI system. Magn Reson Med. 1994;31:67‐72. [DOI] [PubMed] [Google Scholar]
- 12. Butts K, Pauly J, De Crespigny A, Moseley M. Isotropic diffusion‐weighted and spiral‐navigated interleaved EPI for routine imaging of acute stroke. Magn Reson Med. 1997;38:741‐749. [DOI] [PubMed] [Google Scholar]
- 13. Porter DA, Heidemann RM. High resolution diffusion‐weighted imaging using readout‐segmented echo‐planar imaging, parallel imaging and a two‐dimensional navigator‐based reacquisition. Magn Reson Med. 2009;62:468‐475. [DOI] [PubMed] [Google Scholar]
- 14. Robson MD, Anderson AW, Gore JC. Diffusion‐weighted multiple shot echo planar imaging of humans without navigation. Magn Reson Med. 1997;38:82‐88. [DOI] [PubMed] [Google Scholar]
- 15. Ordidge RJ, Helpern JA, Qing Z, Knight RA, Nagesh V. Correction of motional artifacts in diffusion‐weighted MR images using navigator echoes. Magn Reson Imaging. 1994;12:455‐460. [DOI] [PubMed] [Google Scholar]
- 16. Anderson AW, Gore JC. Analysis and correction of motion artifacts in diffusion weighted imaging. Magn Reson Med. 1994;32:379‐387. [DOI] [PubMed] [Google Scholar]
- 17. Pipe JG, Farthing VG, Forbes KP. Multishot diffusion‐weighted FSE using PROPELLER MRI. Magn Reson Med. 2002;47:42‐52. [DOI] [PubMed] [Google Scholar]
- 18. Miller KL, Pauly JM. Nonlinear phase correction for navigated diffusion imaging. Magn Reson Med. 2003;50:343‐353. [DOI] [PubMed] [Google Scholar]
- 19. Chen N‐K, Guidon A, Chang HC, Song AW. A robust multi‐shot scan strategy for high‐resolution diffusion weighted MRI enabled by multiplexed sensitivity‐encoding (MUSE). Neuroimage. 2013;72:41‐47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Chu M, Chang H, Chung H, Truong T, Bashir MR, Chen N. POCS‐based reconstruction of multiplexed sensitivity encoded MRI (POCSMUSE): a general algorithm for reducing motion‐related artifacts. Magn Reson Med. 2015;74:1336‐1348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Guo H, Ma X, Zhang Z, Zhang B, Yuan C, Huang F. POCS‐enhanced inherent correction of motion‐induced phase errors (POCS‐ICE) for high‐resolution multishot diffusion MRI. Magn Reson Med. 2016;75:169‐180. [DOI] [PubMed] [Google Scholar]
- 22. Hu Z, Ma X, Truong T, Song AW, Guo H. Phase‐updated regularized SENSE for navigator‐free multishot diffusion imaging. Magn Reson Med. 2017;78:172‐181. [DOI] [PubMed] [Google Scholar]
- 23. Truong T, Guidon A. High‐resolution multishot spiral diffusion tensor imaging with inherent correction of motion‐induced phase errors. Magn Reson Med. 2014;71:790‐796. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Zhang Z, Huang F, Ma X, Xie S, Guo H. Self‐feeding MUSE: a robust method for high resolution diffusion imaging using interleaved EPI. Neuroimage. 2015;105:552‐560. [DOI] [PubMed] [Google Scholar]
- 25. Moeller S, Ramanna S, Lenglet C, et al. Self‐navigation for 3D multishot EPI with data‐reference. Magn Reson Med. 2020;84:1747‐1762. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Steinhoff M, Nehrke K, Mertins A, Börnert P. Segmented diffusion imaging with iterative motion‐corrected reconstruction (sediment) for brain echo‐planar imaging. NMR Biomed. 2020;33:e4185. [DOI] [PubMed] [Google Scholar]
- 27. Mani M, Jacob M, Kelley D, Magnotta V. Multi‐shot sensitivity‐encoded diffusion data recovery using structured low‐rank matrix completion (MUSSELS). Magn Reson Med. 2017;78:494‐507. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Mani M, Aggarwal HK, Magnotta V, Jacob M. Improved MUSSELS reconstruction for high‐resolution multi‐shot diffusion weighted imaging. Magn Reson Med. 2020;83:2253‐2263. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Hu Y, Wang X, Tian Q, et al. Multi‐shot diffusion‐weighted MRI reconstruction with magnitude‐based spatial‐angular locally low‐rank regularization (SPA‐LLR). Magn Reson Med. 2020;83:1596‐1607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Hu Y, Levine EG, Tian Q, et al. Motion‐robust reconstruction of multishot diffusion‐weighted images without phase estimation through locally low‐rank regularization. Magn Reson Med. 2019;81:1181‐1190. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Chen X, Wu W, Chiew M. Motion compensated structured low‐rank reconstruction for 3D multi‐shot EPI. Magn Reson Med. 2024;91:2443‐2458. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Li Z, Miller KL, Chen X, Chiew M, Wu W. Self‐navigated 3D diffusion MRI using an optimized CAIPI sampling and structured low‐rank reconstruction. IEEE Trans Med Imaging. 2024. Sep 6; PP:1. doi: 10.1109/TMI.2024.3454994. [DOI] [PMC free article] [PubMed]
- 33. Dong Y, Koolstra K, Li Z, Riedel M, van Osch MJ, Börnert P. Structured low‐rank reconstruction for navigator‐free water/fat separated multi‐shot diffusion‐weighted EPI. Magn Reson Med. 2024;91:205‐220. [DOI] [PubMed] [Google Scholar]
- 34. Hu Y, Xu Y, Tian Q, et al. RUN‐UP: accelerated multishot diffusion‐weighted MRI reconstruction using an unrolled network with U‐Net as priors. Magn Reson Med. 2021;85:709‐720. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Aggarwal HK, Mani MP, Jacob M. MoDL‐MUSSELS: model‐based deep learning for multishot sensitivity‐encoded diffusion MRI. IEEE Trans Med Imaging. 2019;39:1268‐1277. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Bilgic B, Chatnuntawech I, Manhard MK, et al. Highly accelerated multishot echo planar imaging through synergistic machine learning and joint reconstruction. Magn Reson Med. 2019;82:1343‐1358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Dai E, Mani M, McNab JA. Multi‐band multi‐shot diffusion MRI reconstruction with joint usage of structured low‐rank constraints and explicit phase mapping. Magn Reson Med. 2023;89:95‐111. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Wilm BJ, Svensson J, Henning A, Pruessmann KP, Boesiger P, Kollias SS. Reduced field‐of‐view MRI using outer volume suppression for spinal cord diffusion imaging. Magn Reson Med. 2007;57:625‐630. [DOI] [PubMed] [Google Scholar]
- 39. Wheeler‐Kingshott CA, Hickman SJ, Parker GJ, et al. Investigating cervical spinal cord structure using axial diffusion tensor imaging. Neuroimage. 2002;16:93‐102. [DOI] [PubMed] [Google Scholar]
- 40. Saritas EU, Cunningham CH, Lee JH, Han ET, Nishimura DG. DWI of the spinal cord with reduced FOV single‐shot EPI. Magn Reson Med. 2008;60:468‐473. [DOI] [PubMed] [Google Scholar]
- 41. In MH, Posnansky O, Speck O. High‐resolution distortion‐free diffusion imaging using hybrid spin‐warp and echo‐planar PSF‐encoding approach. Neuroimage. 2017;148:20‐30. [DOI] [PubMed] [Google Scholar]
- 42. Dong Z, Wang F, Reese TG, et al. Tilted‐CAIPI for highly accelerated distortion‐free EPI with point spread function (PSF) encoding. Magn Reson Med. 2019;81:377‐392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Dong Z, Wang F, Wald L, Setsompop K. SNR‐efficient distortion‐free diffusion relaxometry imaging using accelerated echo‐train shifted echo‐planar time‐resolving imaging (ACE‐EPTI). Magn Reson Med. 2022;88:164‐179. [DOI] [PubMed] [Google Scholar]
- 44. Robson MD, Gore JC, Constable RT. Measurement of the point spread function in MRI using constant time imaging. Magn Reson Med. 1997;38:733‐740. [DOI] [PubMed] [Google Scholar]
- 45. Zeng H, Constable RT. Image distortion correction in EPI: comparison of field mapping with point spread function mapping. Magn Reson Med. 2002;48:137‐146. [DOI] [PubMed] [Google Scholar]
- 46. Wang F, Dong Z, Reese TG, et al. Echo planar time‐resolved imaging (EPTI). Magn Reson Med. 2019;81:3599‐3615. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Lee H, Price RR. Diffusion imaging with the MP‐RAGE sequence. J Magn Reson Imaging. 1994;4:837‐842. [DOI] [PubMed] [Google Scholar]
- 48. Thomas DL, Pell GS, Lythgoe MF, Gadian DG, Ordidge RJ. A quantitative method for fast diffusion imaging using magnetization‐prepared turboFLASH. Magn Reson Med. 1998;39:950‐960. [DOI] [PubMed] [Google Scholar]
- 49. Kirsch J. IVIM turbo FLASH with insensitivity to gradient field errors. In Proceedings of the 10th Annual Scientific Meeting & Exhibition of the SMRM, San Francisco, CA 1991. p. Click here to enter text.853.
- 50. Nguyen C, Fan Z, Sharif B, et al. In vivo three‐dimensional high resolution cardiac diffusion‐weighted MRI: a motion compensated diffusion‐prepared balanced steady‐state free precession approach. Magn Reson Med. 2014;72:1257‐1267. [DOI] [PubMed] [Google Scholar]
- 51. Roccia E, Neji R, Benkert T, Kiefer B, Goh V, Dregely I. Distortion‐free 3D diffusion imaging of the prostate using a multishot diffusion‐prepared phase‐cycled acquisition and dictionary matching. Magn Reson Med. 2021;85:1441‐1454. [DOI] [PubMed] [Google Scholar]
- 52. Alsop DC. Phase insensitive preparation of single‐shot RARE: application to diffusion imaging in humans. Magn Reson Med. 1997;38:527‐533. [DOI] [PubMed] [Google Scholar]
- 53. Sinha U, Sinha S. High speed diffusion imaging in the presence of eddy currents. J Magn Reson Imaging. 1996;6:657‐666. [DOI] [PubMed] [Google Scholar]
- 54. Van AT, Cervantes B, Kooijman H, Karampinos DC. Analysis of phase error effects in multishot diffusion‐prepared turbo spin echo imaging. Quant Imag Med Surg. 2017;7:238‐250. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55. Merboldt K, Hänicke W, Bruhn H, Gyngell ML, Frahm J. Diffusion imaging of the human brain in vivo using high‐speed STEAM MRI. Magn Reson Med. 1992;23:179‐192. [DOI] [PubMed] [Google Scholar]
- 56. Nolte UG, Finsterbusch J, Frahm J. Rapid isotropic diffusion mapping without susceptibility artifacts: whole brain studies using diffusion‐weighted single‐shot STEAM MR imaging. Magn Reson Med. 2000;44:731‐736. [DOI] [PubMed] [Google Scholar]
- 57. Merrem A, Hofer S, Voit D, et al. Rapid diffusion‐weighted magnetic resonance imaging of the brain without susceptibility artifacts: single‐shot STEAM with radial undersampling and iterative reconstruction. Invest Radiol. 2017;52:428‐433. [DOI] [PubMed] [Google Scholar]
- 58. Merrem A, Hofer S, Seif Amir Hosseini A, et al. Diffusion‐weighted MRI of the prostate without susceptibility artifacts: undersampled multi‐shot turbo‐STEAM with rotated radial trajectories. NMR Biomed. 2019;32:e4074. [DOI] [PubMed] [Google Scholar]
- 59. Voit D, Kalentev O, Frahm J. Diffusion‐weighted magnetic resonance imaging (MRI) without susceptibility artifacts: single‐shot stimulated echo acquisition mode (STEAM) MRI with iterative reconstruction and spatial regularization. Quant Imag Med Surg. 2021;11:831‐837. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60. Zhang Z, Zhang B, Li M, et al. Multishot cartesian turbo spin‐echo diffusion imaging using iterative POCSMUSE reconstruction: multishot TSE DWI. J Magn Reson Imaging. 2017;46:167‐174. [DOI] [PubMed] [Google Scholar]
- 61. Zhang Q, Coolen BF, Versluis MJ, Strijkers GJ, Nederveen AJ. Diffusion‐prepared stimulated‐echo turbo spin echo (DPsti‐TSE): an eddy current‐insensitive sequence for three‐dimensional high‐resolution and undistorted diffusion‐weighted imaging. NMR Biomed. 2017;30:e3719. [DOI] [PubMed] [Google Scholar]
- 62. Cervantes B, Van AT, Weidlich D, et al. Isotropic resolution diffusion tensor imaging of lumbosacral and sciatic nerves using a phase‐corrected diffusion‐prepared 3D turbo spin echo. Magn Reson Med. 2018;80:609‐618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Lee PK, Hargreaves BA. A joint linear reconstruction for multishot diffusion weighted non‐Carr‐Purcell‐Meiboom‐Gill fast spin echo with full signal. Magn Reson Med. 2022;88:2139‐2156. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64. Jeong E, Kim S, Parker DL. High‐resolution diffusion‐weighted 3D MRI, using diffusion‐weighted driven‐equilibrium (DW‐DE) and multishot segmented 3D‐SSFP without navigator echoes. Magn Reson Med. 2003;50:821‐829. [DOI] [PubMed] [Google Scholar]
- 65. O'Halloran RL, Aksoy M, Van AT, Bammer R. 3D isotropic high‐resolution diffusion‐weighted MRI of the whole brain with a motion‐corrected steady‐state free precession sequence. Magn Reson Med. 2013;70:466‐478. [DOI] [PubMed] [Google Scholar]
- 66. Martinez JA, Moulin K, Yoo B, et al. Evaluation of a workflow to define low specific absorption rate MRI protocols for patients with active implantable medical devices. J Magn Reson Imaging. 2020;52:91‐102. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67. Ma S, Nguyen CT, Han F, et al. Three‐dimensional simultaneous brain T1, T2, and ADC mapping with MR multitasking. Magn Reson Med. 2020;84:72‐88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68. McElroy S, Tomi‐Tricot R, Jeljeli S, Kinsella S, Goh V, Neji R. 3D high resolution, distortion‐free, reduced field of view diffusion‐prepared MRI at 3T. In Proceedings of the Joint ISMRM & ISMRT Annual Meeting, Singapore 2024. p. 4427.
- 69. Lee PK, Hess JJ, Gomella AA, Loening AM, Hargreaves BA. A diffusion‐prepared reduced FOV sequence for prostate MRI near metallic implants. Magn Reson Med. 2025;93:261‐275. doi:10.1002/mrm.30280 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70. Cohen‐Adad J. High‐resolution DWI in brain and spinal cord with syngo RESOLVE. Siemens Magnetom‐Clin Neurol. 2012;510:16‐23. [Google Scholar]
- 71. Hayes L, Jones R, Porter D, et al. Diffusion‐weighted imaging of the pediatric spine using readout‐segmented Echo planar imaging: a pictorial review of clinical applications. Neurographics. 2015;5:197‐208. [Google Scholar]
- 72. Le Bihan D, Poupon C, Amadon A, Lethimonnier F. Artifacts and pitfalls in diffusion MRI. J Magn Reson Imaging. 2006;24:478‐488. [DOI] [PubMed] [Google Scholar]
- 73. Dong Y, Koolstra K, Riedel M, van Osch MJ, Börnert P. Regularized joint water–fat separation with B0 map estimation in image space for 2D‐navigated interleaved EPI based diffusion MRI. Magn Reson Med. 2021;86:3034‐3051. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74. Dong Y, Riedel M, Koolstra K, van Osch MJ, Börnert P. Water/fat separation for self‐navigated diffusion‐weighted multishot echo‐planar imaging. NMR Biomed. 2023;36:e4822. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75. Hiepe P, Herrmann KH, Ros C, Reichenbach JR. Diffusion weighted inner volume imaging of lumbar disks based on turbo‐STEAM acquisition. Z Med Phys. 2011;21:216‐227. [DOI] [PubMed] [Google Scholar]
- 76. Prost JEH, Wehrli FW, Drayer B, Froelich J, Hearshen D, Plewes D. SAR reduced pulse sequences. Magn Reson Imaging. 1988;6:125‐130. [DOI] [PubMed] [Google Scholar]
- 77. Nezafat R, Stuber M, Ouwerkerk R, Gharib AM, Desai MY, Pettigrew RI. B1‐insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T. Magn Reson Med. 2006;55:858‐864. [DOI] [PubMed] [Google Scholar]
- 78. Finsterbusch J, Frahm J. Half‐Fourier single‐shot STEAM MRI. Magn Reson Med. 2002;47:611‐615. [DOI] [PubMed] [Google Scholar]
- 79. Frahm J, Merboldt KD, Hänicke W, Haase A. Stimulated echo imaging. J Magn Reson Imaging. 1985;64:81‐93. [Google Scholar]
- 80. Engström M, Skare S. Diffusion‐weighted 3D multislab echo planar imaging for high signal‐to‐noise ratio efficiency and isotropic image resolution. Magn Reson Med. 2013;70:1507‐1514. [DOI] [PubMed] [Google Scholar]
- 81. Frost R, Miller KL, Tijssen RH, Porter DA, Jezzard P. 3D multi‐slab diffusion‐weighted readout‐segmented EPI with real‐time cardiac‐reordered k‐space acquisition. Magn Reson Med. 2014;72:1565‐1579. [DOI] [PubMed] [Google Scholar]
- 82. Zhang Q, Coolen BF, Nederveen AJ, Strijkers GJ. Three‐dimensional diffusion imaging with spiral encoded navigators from stimulated echoes (3D‐DISPENSE). Magn Reson Med. 2019;81:1052‐1065. [DOI] [PubMed] [Google Scholar]
- 83. McNab JA, Gallichan D, Miller KL. 3D steady‐state diffusion‐weighted imaging with trajectory using radially batched internal navigator echoes (TURBINE). Magn Reson Med. 2010;63:235‐242. [DOI] [PubMed] [Google Scholar]
- 84. Chang H, Hui ES, Chiu P, Liu X, Chen N. Phase correction for three‐dimensional (3D) diffusion‐weighted interleaved EPI using 3D multiplexed sensitivity encoding and reconstruction (3D‐MUSER). Magn Reson Med. 2018;79:2702‐2712. [DOI] [PubMed] [Google Scholar]
- 85. Van AT, Hernando D, Sutton BP. Motion‐induced phase error estimation and correction in 3D diffusion tensor imaging. IEEE Trans Med Imaging. 2011;30:1933‐1940. [DOI] [PubMed] [Google Scholar]
- 86. Bernstein MA, Huston J III, Ward HA. Imaging artifacts at 3.0 T. J Magn Reson Imaging. 2006;24:735‐746. [DOI] [PubMed] [Google Scholar]
- 87. Andre JB, Bammer R. Advanced diffusion‐weighted magnetic resonance imaging techniques of the human spinal cord. Top Magn Reson Imaging. 2010;21:367‐378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88. Morozov D, Lopez Rios N, Duval T, Foias A, Cohen‐Adad J. Effect of cardiac‐related translational motion in diffusion MRI of the spinal cord. Magn Reson Imaging. 2018;50:119‐124. [DOI] [PubMed] [Google Scholar]
- 89. Zhang Z, Ma X, Yuan C, Guo H. Effect of velocity‐compensated diffusion preparation for spinal cord diffusion imaging. Proceedings of the 25th Annual Meeting of ISMRM, Honolulu, HI; 2017:1859. [Google Scholar]
- 90. Zaharchuk G, Saritas E, Andre J, et al. Reduced field‐of‐view diffusion imaging of the human spinal cord: comparison with conventional single‐shot echo‐planar imaging. Am J Neuroradiol. 2011;32:813‐820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91. Holder CA, Muthupillai R, Mukundan S Jr, Eastwood JD, Hudgins PA. Diffusion‐weighted MR imaging of the normal human spinal cord in vivo. Am J Neuroradiol. 2000;21:1799‐1806. [PMC free article] [PubMed] [Google Scholar]
- 92. Ueda T, Ohno Y, Yamamoto K, et al. Deep learning reconstruction of diffusion‐weighted MRI improves image quality for prostatic imaging. Radiology. 2022;303:373‐381. [DOI] [PubMed] [Google Scholar]
- 93. Bae SH, Hwang J, Hong SS, et al. Clinical feasibility of accelerated diffusion weighted imaging of the abdomen with deep learning reconstruction: comparison with conventional diffusion weighted imaging. Eur J Radiol. 2022;154:110428. [DOI] [PubMed] [Google Scholar]
- 94. Afat S, Herrmann J, Almansour H, et al. Acquisition time reduction of diffusion‐weighted liver imaging using deep learning image reconstruction. Diagn Interv Imaging. 2023;104:178‐184. [DOI] [PubMed] [Google Scholar]
- 95. Haase A, Frahm J, Hanicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30:341‐344. [DOI] [PubMed] [Google Scholar]
- 96. Bydder GM, Young IR. MR imaging: clinical use of the inversion recovery sequence. J Comput Assist Tomogr. 1985;9:659‐675. [PubMed] [Google Scholar]
- 97. Nagy Z, Weiskopf N. Efficient fat suppression by slice‐selection gradient reversal in twice‐refocused diffusion encoding. Magn Reson Med. 2008;60:1256‐1260. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Figure S1. Example in‐vivo images of the T2‐weighted TSE (top row), b0 (left, bottom 3 rows) and b500 trace‐weighted images (right, bottom 3 rows) for the proposed sequence, DW‐SS‐EPI and RESOLVE for all 5 volunteers with contour plots of CSF and spinal cord overlaid on the images to better highlight the distortion properties of each sequence.