Table 1.

Typical Imaging parameters of the T2-weighted and DWI sequences of both protocols.

mpMRI ultra-fast bpMRI
Acquisition time (full protocol) 20 min 33 s 3 min 28 s (approx. - 80 %)
Sequence T2 TSE T2 TSE
Acquisition time 3 min 12 s 1 min 56 s (approx. - 40 %)
Orientation axial axial
Number of slices 27 27
Slice thickness (mm) 3 3
Gap 0 % 0 %
Field of view (mm2) 180 × 180 179 × 179
Matrix size 704 × 704 640 × 640
Pixel spacing (mm) 0.26 × 0.26 0.28 × 0.28
PAT factor 3 4
Number of excitations 3 2
TR (ms) 4100 4030
TE (ms) 102 101
Flip Angle (º) 140 140
Echo train length 25 25
Sequence DWI DWI
Sequence type EPI EPI
Acquisition time 4 min 50 s 1 min 32 s (approx. - 70 %)
Orientation axial axial
Number of slices 27 27
Slice thickness (mm) 3 3
Field of view (mm2) 101 × 199 101 × 199
Matrix size 116 × 228 116 × 224
Pixel spacing (mm) 0.88 × 0.88 0.88 × 0.88
Number of excitations 2/3/10 2/3
B-values (s/mm2) 100, 600, 1000, calculated 2000 50, 800, calculated 1400
TR (ms) 4800 4500
TE (ms) 70 68
Flip Angle (º) 90 90
Echo train length 61 61