Skip to main content
Eye logoLink to Eye
. 2025 Jan 13;39(3):544–547. doi: 10.1038/s41433-025-03598-4

Anatomical and functional results of patients with late-stage age-related macular degeneration 6 months after smaller-incision new-generation implantable miniature telescope (SING IMT™) implantation

Ginevra Giovanna Adamo 1,2,✉, Marco Pellegrini 1,2,3, Francesco Nasini 2, Pietro Maria Talli 2, Laura Sarti 1,2, Piera Perri 2, Francesco Parmeggiani 1,2, Marco Mura 1,2,4
PMCID: PMC11794482  PMID: 39805963

Abstract

Objectives

The aim of the study was to evaluate anatomical and functional outcomes of the Smaller-Incision New Generation Implantable Miniature Telescope (SING IMT™) in patients with bilateral advanced age-related macular degeneration (AMD).

Methods

This non-comparative retrospective single-surgeon interventional case series included patients with bilateral late-stage AMD who underwent cataract surgery and SING IMT™ implantation at the Sant’Anna University Hospital, University of Ferrara, Italy. The main outcome measures included corrected distance (CDVA) and near visual acuity (CNVA), endothelial cell loss (ECL), and incidence of complications.

Results

11 eyes of 11 patients were included. The mean follow-up duration was 6.5 ± 2.4 months. After surgery, CDVA significantly improved (from 17.00 ± 9.74 to 26.00 ± 8.53) letters (P = 0.008). Significant improvement of CNVA was also observed (from 12.27 ± 4.36 to 8 ± 2.61 Jaeger levels; P = 0.004). Mean ECL was 4.8 ± 5.5% at 3 months. No intraoperative complications were observed, while postoperative complications included iris incarceration (9.1%), pigment deposition on the device (9.1%), and transient corneal oedema (27.3%). Nevertheless, 10 of 11 patients (90.9%) began to complain of blurred or hazy vision within 3 months of surgery. The device was ultimately explanted in 3 patients (27.3%) because of this symptom.

Conclusions

Although SING IMT™ implantation is associated with promising objective results, unexplained blurred or hazy vision represents common postoperative complaints which may lead to patient dissatisfaction. Further studies including patient-reported outcomes are warranted to evaluate the effect of the intervention on patients’ visual function and quality of life.

Subject terms: Retinal diseases, Eye manifestations

Introduction

Age-related macular degeneration (AMD) is an age-associated disease and represents the third-leading cause of blindness worldwide after cataracts and glaucoma [1]. The disease includes non-exudative and exudative forms, accounting for 85–90% and 10–15% of cases, respectively [2]. The progressive central visual function deterioration due to the macular damage has a significant impact on patients’ quality of life, affecting their independence in daily activities involving the use of central vision. Over the last decades, many different therapies for exudative AMD have been proposed, ranging from laser procedures to anti-neovascular agents [3]. Conversely, there are currently no commercially available treatment for geographic atrophy and significant research efforts are underway to develop therapies that may slow the progression of non-exudative AMD [4, 5] As a result, visual rehabilitation in patients with bilateral central visual field loss related to late AMD include a variety of magnifying devices, such as low vision spectacles, external telescopes or magnifiers [6–9]. On the other hand, surgical options are limited and include macular translocation surgery and retinal-pigmented epithelium transplant [10]. Therefore, innovative alternatives to non-surgical devices such as intraocular lens implants or implantable telescopic visual-enhancing devices have been recently developed [11].

The Smaller-Incision New Generation Implantable Miniature Telescope (SING IMT™, Samsara Vision Ltd) has been approved in 2020 as surgical treatment for bilateral central vision loss due to end-stage AMD within the European Union. It is a miniaturised Galilean telescope providing a 2.7× magnification of the central visual field and projecting it onto healthy retinal areas which surround the degenerated macula, at the cost of losing peripheral vision [12]. Compared to its predecessor IMT™, the design of the second generation of the implant provides several improvements, including a smaller overall diameter combined with a minimally invasive implantation system requiring an 8.0 mm corneal incision size, associated with a greater postoperative corneal clearance and lower rates of endothelial cell density (ECD) loss [6, 13–17].

In this study, we retrospectively evaluated the outcomes and safety of the second generation of the IMT implant, in patients with central visual impairment due to bilateral advanced AMD.

Materials and methods

Study design

This retrospective single-surgeon interventional case series included 11 eyes of 11 patients with bilateral late-stage AMD who underwent cataract surgery and SING IMT implantation at the Sant’Anna University Hospital, University of Ferrara, Italy, from June 2022 to May 2023. The study was performed in accordance with the tenets of the Declaration of Helsinki and approved by the Ethics Committee of the University of Ferrara. Written informed consent for the processing of personal data was obtained from all patients.

Patient screening, inclusion and exclusion criteria

Inclusion criteria for SING IMT implantation were: age of 55 years or older, central visual acuity loss caused by untreatable bilateral end-stage AMD (geographic atrophy, disciform scar, or both), a diagnosis of cataract in the study eye, good peripheral vision in the fellow eye. Moreover, patients were required to have an anterior chamber depth (ACD) greater than 2.5 mm, ECD greater than 1600 cells/mm2, intraocular pressure (IOP) less than 22 mmHg, and bilateral CDVA comprised between 20/80 and 20/800 (0.6–1.6 logMAR) on the ETDRS visual acuity chart. The eye selected for implantation had to be able to demonstrate an improvement of at least five letters on the Early Treatment Diabetic Retinopathy Study Charts (ETDRS) scale through the use of an external hand-held telescope simulator (ETS 3.0X-FF; VisionCare Ophthalmic Technologies Ltd, Petah Tikva, Israel) preoperatively. If both eyes were eligible for implantation, the SING IMT™ was implanted in the eye with the potential for greater gain in ETDRS letters, as determined with the ETS. Patients were also selected based on their willingness to participate in all follow-up visits and postoperative vision training.

Exclusion criteria were: active choroidal neovascularization (CNV) or CNV treatment in the preceding 6 months, previous intraocular surgery, corneal stromal or endothelial dystrophies, myopia greater than 6.0 D, hyperopia greater than 4.0 D, history of steroid-induced increases in IOP, uncontrolled glaucoma in the study eye, other ophthalmic pathologies compromising the peripheral vision, marked cognitive impairment potentially interfering with understanding the instructions or appropriate visual training and rehabilitation, unwillingness to participate to the visual training.

Importantly, patients were made aware that the functional visual result obtained after the implant would constitute a new type of functional vision, rather than a return to the vision before advanced AMD, and visual rehabilitation would consequently require considerable effort and time.

Examination methods and rehabilitation training program

Patient preoperative and postoperative data were collected from the electronic medical records. All patients underwent a complete ophthalmic examination at baseline and postoperative day 1, day 15, month 1, month 3, month 6, month 9. Examinations included slit lamp biomicroscopy, gonioscopy, fundus examination, IOP measurement, and specular microscopy. IOP was measured through Goldmann applanation tonometry. Corrected distance visual acuity (CDVA) and corrected near visual acuity (CDVA) were measured with the ETDRS chart and Jaeger levels (J), respectively.

Axial length and ACD were evaluated using optical biometry (ZEISS IOLMaster 500), whereas corneal ECD was measured with specular microscopy (TOMEY Specular Microscope EM 3000). Macular optical coherence tomography was performed using the Spectralis (Heidelberg Engineering, Heidelberg, Germany).

Since the functioning of the SING IMT assumes that the patient learns to use alternating binocular vision, excellent collaboration between the two eyes is essential to be able to suppress one of the two images, depending on the daily activities carried out. Therefore, patients were entered into the rehabilitation programme starting from the fourth postoperative week onwards, rehabilitation sessions were organised at month 1, month 2, month 3, month 5, and month 9.

Surgical procedure and postoperative care

All surgical procedures were performed by the same surgeon (M.M.) at the S. Anna Hospital, University of Ferrara, Italy, using a previously described technique (Savastano et al. [14]). Anaesthesia was obtained using sub-tenon block 10 min before surgery and a sclerocorneal tunnel was carried out at the 12 o’clock position, then continuous manual circular anterior capsulorhexis with a diameter of at least 6.0 mm was created. After standard phacoemulsification, the superior sclerocorneal incision was enlarged up to 8.0 mm and the SING IMT (Samsara Vision Inc., Far Hills, NJ) was implanted in the capsular bag, two haptics positioned inferiorly and one superiorly, using the preloaded delivery system. Finally, the sclerocorneal incision was sutured with the use of 3 single 10-0 monofilament nylon sutures and an iridectomy was performed at 12 o’clock position. Following surgery, a standardised postoperative regimen of topical ophthalmic medications was used, with antibiotic and anti-inflammatory drops for 1 month and cycloplegic drops for 2 weeks.

Statistical analysis

All continuous variables were reported as mean ± standard deviation. Demographic information included gender, race, diagnosis, implanted eye, and age in years. Efficacy parameters included CDVA, CNVA, ACD, ECD, and IOP. The Wilcoxon test was used to compare continuous variables before and after surgery. Statistical significance was set at p < 0.05.

Results

A total of 11 patients were enrolled in the study. Baseline demographics and pertinent clinical characteristics are outlined in Table 1. The study population consisted of 6 (54.5%) females and 5 (45.5%) males. The mean age of the study population was 78.6 ± 7.0 years (range: 68–92 years), all patients were Caucasian. Ten patients (90.9%) had with geographic atrophy in the study eye while 1 patient (9.1%) had disciform a scar associated secondary to CNV. Mean follow up was 6.5 ± 2.4 months.

Table 1.

Baseline demographics and clinical characteristics of included patients.

Characteristic Value
Age (years)
 Mean (±SD) 78.6 ± 7.0
 Range 68–92
Gender, n (%)
 Female 6 (54.5)
 Male 5 (45.5)
Implanted eye, n (%)
 Right 2
 Left 9
Race, n (%)
 Caucasian 11 (100)
 CDVA (ETDRS letters) ± SD 17.00 ± 9.74
 CNVA (Jaeger levels) ± SD 12.27 ± 4.36
Macular lesion (implanted eye), n (%)
 Geographic atrophy 10 (90.9)
 Disciform scar associated with CNV 1 (9.1)
 ACD (mm) ± SD 3.246 ± 0.726
 ECD (cells/mm2) ± SD 2150.2 ± 297.5
 IOP (mmHg) ± SD 13.7 ± 1.4

SD standard deviation, CDVA corrected distance visual acuity, ETDRS Early Treatment Diabetic Retinopathy Study, CNVA corrected near visual acuity, ACD anterior chamber depth, ECD endothelial cell density, IOP intraocular pressure.

Mean CDVA was 17.00 ± 9.74 letters before surgery and significantly improved to 26.00 ± 8.53 at the last follow-up (P = 0.008) (Fig. 1). Mean CNVA in the study eyes significantly increased from 12.27 ± 4.36 to 8 ± 2.61 J-levels (P = 0.004), while mean ECD loss was 4.8 ± 5.5% at 3 months (P = 0.003) (Fig. 2). The mean ACD of the study eyes by postoperative month 5 was not significantly different from baseline (3.092 ± 0.617 mm and 3.246 ± 0.726 mm respectively). No intraoperative complications were observed.

Fig. 1. Mean corrected distance visual acuity (CDVA) in study eyes.

Fig. 1

CDVA was measured preoperatively (baseline) and at the last follow-up. Data are expressed as mean ± standard deviation.

Fig. 2. Mean endothelial cell density (ECD) in study eyes over 3 months.

Fig. 2

ECD was measured preoperatively (baseline) and at 1 month and 3 months. Data are expressed mean ± standard deviation.

Postoperative complications included iris incarceration in 1 patient (9.1%) which required surgical repositioning, pigment deposition on the device in 1 patient (9.1%), transient corneal oedema in 3 patients (27.3%) that was successfully resolved through a regimen of topical steroids and hypertonic eyedrops. One month after surgery, 2 patients (18.2%) had high corneal astigmatism (>4.5 D), while no cases of high astigmatism were recorded at 3 months following complete suture removal.

Despite the good results in terms of CDVA maintained over time, 10 of 11 patients (90.9%) started to complain of blurred or hazy vision within 3 months of surgery, reporting that this symptom had a significant impact on their quality of life. In 3 of these patients (27.3%), the SING IMT was ultimately explanted and exchanged with a conventional monofocal IOL.

Discussion

Over the years, several approaches to improve the visual function in patients with late-stage AMD have been attempted. The IMT is an intraocular magnifying device based on the principle of fixed-focus Galilean telescope. The device is implanted only in one eye and magnifies the objects in the centre of the field of view, with two wide-angle magnifications available (×2.2 and ×2.7) [6, 18, 19] Compared to its predecessor, the SING IMT implant has solely a ×2.7 magnification and is provided with a pre-loaded delivery system which requires a less invasive procedure with a smaller corneal wound (6.5–8 mm vs. 10–12 mm), reducing the postoperative astigmatism and endothelial cell loss [12, 20–22]. Indeed, Toro et al. reported good short-term (3 months) outcomes with the SING IMT device, highlighting a significant lower ECD loss (10.4% versus 20.0% at 3 months) while maintaining the visual acuity outcomes achieved with the first generation IMT device. Indeed, short-term visual outcomes were in line with those observed at 12 months in the IMT pivotal study. Moreover, the cumulative incidence of increased IOP in their study (4.2%) resulted considerably lower compared to the rate obtained in the IMT-002 pivotal trial (28.0%) [20, 21].

During the course of our study, significant improvements in mean CDVA and CNVA were observed, with no major intraoperative nor postoperative complication developed. There were no reports of corneal decompensation, glaucoma, iris atrophy, iritis, reactivation of CNV, vitreous haemorrhage, retinal detachment, endophthalmitis, or posterior dislocation of the telescope.

With regards to ECD, a mean loss of 4.9% was observed in our series. This slightly lower compared to those reported by Toro et al. and Savastano et al. (10.4% and 8.3%, respectively) [14, 21].

However, within 3 months from surgery, 10 of 11 patients (90.9%) started to complain of blurred or hazy vision, while maintaining the achieved visual outcomes. 3 of 11 patients (27.3%) finally required the explanation of the device due to their inability to tolerate this visual discomfort. After SING IMT explantation, all three patients reported a subjective improvement in their visual function, with resolution of the discomfort caused by the blurred vision.

Despite a comprehensive eye examination including retinal imaging to rule out cystoid macular oedema or CNV reactivation, no cause for these visual symptoms could be identified. Interestingly, Savastano et al. also documented SING IMT explantation in three unsatisfied patients who were unable to adapt to the device and complained about an unbearable fog in the visual field [23]. Although the cause of cloudy vision after SING IMT implantation remains unclear, we have identified some critical issues: Firstly, the external telescope simulator offers the unique opportunity of preoperatively assess the potential benefits for each individual patient that meets the inclusion criteria. Patient’s satisfaction with the simulator is the result of several aspects, including the indication for surgery (stage and type of AMD) and patient’s psychosocial and cognitive profile. Indeed, one of the main difficulties in the use of this technology are related to the criteria employed for patient selection. Nevertheless, based on our results, patient’s visual expectations are often disregarded after surgery. For this reason, the simulator should be revised to allow a more faithful reproduction of what the visual result will be after the device implantation. Secondly, SING IMT often does not allow the patient to recover the ability to read a text, but rather that of reading individual characters or words, even in very cooperative subjects participating in training of saccadic movements. Moreover, all patients inevitably experience a narrowing of the visual field due to magnification, and consequently the difficulty in alternating the use of the two eyes for central and peripheral vision in daily activities.

Considering the high costs of SING IMT, patients would benefit from a better counselling regarding the expected visual results, the need to adapt to the new visual situation and the long-term expectations of independence. Furthermore, it remains unclear whether objective parameters such as CDVA and CNVA represent the best metrics to measure the success of SING IMT implantation. Further studies should include patient-reported outcomes to evaluate the effect of the intervention on patients’ visual function and quality of life.

In conclusion, SING IMT implantation is associated with good functional results, with a significant improvement in distance and near vision, and acceptable rates of complications. Nevertheless, unexplained hazy or blurry vision is a common postoperative complaint, which may lead to patient dissatisfaction and ultimately require the explantation of the device.

Summary

What was known before

  • Visual rehabilitation in patients with bilateral central visual field loss related to late AMD includes a variety of magnifying devices, such as low-vision spectacles, external telescopes, or magnifiers.

  • On the other hand, surgical options are limited and include macular translocation surgery and retinal-pigmented epithelium transplant.

  • Therefore, innovative alternatives to non-surgical devices such as intraocular lens implants or implantable telescopic visual-enhancing devices have been recently developed.

What this study adds

  • This study represents a single-surgeon interventional case series of patients with bilateral late-stage AMD undergoing cataract surgery and SING IMT™ implantation.

  • Although promising objective results were observed, unexplained blurred or hazy vision represented a common postoperative complaint.

Author contributions

GGA: Conceptualisation; Methodology; Analysis; Investigation; Writing—original draft, review, editing. MP: Conceptualisation; Methodology; Resources; Investigation; Writing—review, editing. FN: Investigation; Writing—review, editing. PMT: Investigation; Writing—review, editing. LS: Investigation; Writing—review, editing. PP: Investigation. FP: Investigation; Writing—review, editing. MM: Conceptualisation; Methodology; Resources; Investigation; Writing—review, editing.

Data availability

The datasets generated and analysed during the current study are available from the corresponding author upon reasonable request.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Wong WL, Su X, Li X, Cheung CM, Klein R, Cheng CY, et al. Global prevalence of age-related macular degeneration and disease burden projection for 2020 and 2040: a systematic review and meta-analysis. Lancet Glob Health. 2014;2:e106–16. [DOI] [PubMed] [Google Scholar]
  • 2.Schultz NM, Bhardwaj S, Barclay C, Gaspar L, Schwartz J. Global burden of dry age-related macular degeneration: a targeted literature review. Clin Ther. 2021;43:1792–818. [DOI] [PubMed] [Google Scholar]
  • 3.Galindo-Camacho RM, Blanco-Llamero C, da Ana R, Fuertes MA, Señoráns FJ, Silva AM, et al. Therapeutic approaches for age-related macular degeneration. Int J Mol Sci. 2022;23:11769. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Gehrs KM, Anderson DH, Johnson LV, Hageman GS. Age-related macular degeneration-emerging pathogenetic and therapeutic concepts. Ann Med. 2006;38:450–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Rubner R, Canto-Soler MV, Li KV. Progress of clinical therapies for dry age-related macular degeneration. Int J Ophthalmol. 2022;15:157–66. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Grzybowski A, Wasinska-Borowiec W, Alio JL, Amat-Peral P, Tabernero J. Intraocular lenses in age-related macular degeneration. Graefes Arch Clin Exp Ophthalmol. 2017;255:1687–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Singer MA, Amir N, Herro A, Porbandarwalla SS, Pollard J. Improving quality of life in patients with end-stage age-related macular degeneration: focus on miniature ocular implants. Clin Ophthalmol. 2012;6:33–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Teh BL, Megaw R, Borooah S, Dhillon B. Optimizing cataract surgery in patients with age-related macular degeneration. Surv Ophthalmol. 2017;62:346–56. [DOI] [PubMed] [Google Scholar]
  • 9.Tabernero J, Qureshi MA, Robbie SJ, Artal P. An aspheric intraocular telescope for agerelated macular degeneration patients. Biomed Opt Express. 2015;6:1010–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Raimondi R, Zollet P, De Rosa FP, Tsoutsanis P, Stravalaci M, Paulis M. Where are we with RPE replacement therapy? A translational review from the ophthalmologist perspective. Int J Mol Sci. 2022;23:682. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Borkenstein AF, Borkenstein EM, Augustin AJ. Implantable vision-enhancing devices and postoperative rehabilitation in advanced age-related macular degeneration. Eye. 2023;37:597–606. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Grzybowski A, Wang J, Mao F, Wang D, Wang N. Intraocular vision-improving devices in age-related macular degeneration. Ann Transl Med. 2020;8:1549. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Colby KA, Chang DF, Stulting RD, Lane SS. Surgical placement of an optical prosthetic device for end-stage macular degeneration: the implantable miniature telescope. Arch Ophthalmol. 2007;125:1118–21. [DOI] [PubMed] [Google Scholar]
  • 14.Savastano A, Caporossi T, Sasso P, De Vico U, Rizzo S. A new intraocular telescopic device for age-related macular degeneration. Ophthalmol Retin. 2022;6:971–2. [DOI] [PubMed] [Google Scholar]
  • 15.Lane SS, Kuppermann BD, Fine IH, Hamill MB, Gordon JF, Chuck RS, et al. A prospective multicenter clinical trial to evaluate the safety and effectiveness of the implantable miniature telescope. Am J Ophthalmol. 2004;137:993–1001. [DOI] [PubMed] [Google Scholar]
  • 16.Hudson HL, Stulting RD, Heier JS, Lane SS, Chang DF, Singerman LJ, et al. Implantable telescope for end-stage age-related macular degeneration: long-term visual acuity and safety outcomes. Am J Ophthalmol. 2008;146:664–73. [DOI] [PubMed] [Google Scholar]
  • 17.Virgili G, Acosta R, Bentley SA, Giacomelli G, Allcock C, Evans JR. Reading aids for adults with low vision. Cochrane Database Syst Rev. 2018;4:CD003303. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Dunbar HMP, Dhawahir-Scala FE. A discussion of commercially available intraocular telescopic implants for patients with age-related macular degeneration. Ophthalmol Ther. 2018;7:33–48. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Boyer D, Freund KB, Regillo C, Levy MH, Garg S. Long-term (60-month) results for the implantable miniature telescope: efficacy and safety outcomes stratified by age in patients with end-stage age-related macular degeneration. Clin Ophthalmol. 2015;9:1099–107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Hudson HL, Lane SS, Heier JS, Stulting RD, Singerman L, Lichter PR, et al. IMT-002 Study Group. Implantable miniature telescope for the treatment of visual acuity loss resulting from end-stage age-related macular degeneration: 1-year results. Ophthalmology. 2006;113:1987–2001. [DOI] [PubMed] [Google Scholar]
  • 21.Toro MD, Vidal-Aroca F, Montemagni M, Xompero C, Fioretto G, Costagliola C. Three-month safety and efficacy outcomes for the smaller-incision new-generation implantable miniature telescope (SING IMT™). J Clin Med. 2023;12:518. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Avisol Capital Partners. Samsara vision: developing implantable miniature telescopes for late-stage AMD patients. Seeking Alpha 2021. Updated Nov 9, 2021. https://seekingalpha.com/article/4467579-samsara-vision-smsa-implantabletelescopes-for-late-stage-amd-patients.
  • 23.Savastano A, D’Onofrio NC, Francione G, Sasso P, Hu L, Rizzo S. SING IMT removal for unsatisfied patients: step-by-step surgery for a safe explant. Retina. 2024. 10.1097/IAE.0000000000004209. [DOI] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets generated and analysed during the current study are available from the corresponding author upon reasonable request.


Articles from Eye are provided here courtesy of Nature Publishing Group

RESOURCES