Skip to main content
. 2025 Feb 1;58(Pt 1):138–153. doi: 10.1107/S1600576724011038

Table 3. Indexing rate obtained with xgandalf (Gevorkov et al., 2019 ▸) from peak positions extracted with different picking algorithms available from CrystFEL (White et al., 2012 ▸) on a subset of 1000 frames of microcrystals of lysozyme (HEWL + Ga) collected with an Eiger 4M detector at ESRF-ID30A3.

  xgandalf (default) xgandalf (fast)
Peak-picking method Index rate (%) Runtime (s) Index rate (%) Runtime (s)
zaef † 10.0 2178 10.0 430
peakfinder8 ‡ 49.5 10397 48.5 1757
peakfinder9 § 44.2 8328 43.5 1436
RobustPF 22.4 6314 21.2 1628
pyFAI †† 49.7 9325 49.2 1595
†

Zaefferer (2000 ▸).

‡

Barty et al. (2014 ▸).

§

Gevorkov et al. (2024 ▸).

Hadian-Jazi et al. (2021 ▸).

††

This contribution.