TABLE 4.

Comparison of Postoperative Manometric Characteristics by Device Size

Device size
Variables, median (IQR) 13
(N=58)
14
(N=90)
15
(N=66)
16
(N=28)
17
(N=13)
P
LES total length (cm) 2.9 (2–4) 3.0 (3–4) 2.9 (2–4) 3.2 (3–4) 2.3 (2–3) 0.150
LES intra-abdominal length (cm) 1.6 (1–3) 1.8 (1–2) 2.1 (1–3) 2.0 (2–3) 0.0 (0–2) 0.071
LES resting pressure (mm Hg) 30.6 (25–37) 29.1 (21–37) 26.2 (20–33) 24.8 (19–33) 19.8 (18–32) 0.064
Integrated relaxation pressure (mm Hg) 16.7 (13–20) 14.5 (11–18) 12.4 (10–16) 10.5 (8–14) 9.4 (6–13) <0.0001
Distal contractile integral (mm Hg-s-cm) 1853 (1.2–3.5k) 2644 (1.1–4.0k) 1925 (1.1–2.9k) 1649 (1.2–2.9k) 915.3 (0.7–1.6k) 0.038
Contractile amplitude (mm Hg) 88.7 (67–139) 95.6 (66–148) 83.5 (66–107) 82.9 (67–106) 60.3 (45–88) 0.053
Percent peristaltic 90.0 (80–100) 100.0 (80–100) 90.0 (80–100) 90.5 (98–100) 90.0 (90–100) 0.872
Incomplete bolus clearance (%) 20.0 (10–50) 10.0 (0–40) 10 (0–40) 9.5 (0–43) 10.0 (0–10) 0.100