Skip to main content
. 2025 Feb 13;12:16. doi: 10.1186/s40658-025-00717-8

Table 4.

ASICs in PET: Performance overview in terms of coincidence time resolution, single-photon time resolution and energy resolution of different ASICs designed for HEP and PET applications. Only results from single detector pixel measurements are displayed

ASIC SiPM Crystal CTR (FWHM) / ps SPTR (FWHM) / ps dE/E (FWHM) / % References
TOFPET2c BRCM AFBR-S4N33C013 2×2×3 mm3 LYSO:Ce,Ca 127 n.a. 10.4 [52]
TOFPET2c BRCM AFBR-S4N44P014M 2×2×3 mm3 LYSO:Ce,Ca 128 n.a. n.a. [71]
TOFPET2c BRCM AFBR-S4N44P014M 3×3×19 mm3 LYSO:Ce,Ca 157 n.a. n.a. [52]
TOFPET2c BRCM AFBR-S4N44P014M 2×2×3 mm3 BGO 480 n.a. n.a. [71]
TOFHiR 2A
Liroc FBK NUV-HD-LF M3 n.a. n.a. 90 n.a. [25]
Radioroc BRCM AFBR-S4N44P014M 2×2×3 mm3 LYSO:Ce,Ca 83 n.a. n.a. [25]
Radioroc BRCM AFBR-S4N44P014M 3×3×20 mm3 LYSO:Ce,Ca 127 n.a. 9.2 [25]
Radioroc FBK NUV-HD-LF M3 n.a. n.a. 73 n.a. [25]
NINO HPK S13360-3050CS 2×2×5 mm3 LSO:Ce,0.4%Ca 93 151 n.a. [87]
NINO FBK NUV-HD 2×2×3 mm3 LSO:Ce,0.4%Ca 73 n.a. n.a. [41]
NINO HPK S13360-3050CS n.a. n.a. 160 n.a. [24]
NINO FBK NUV-HD n.a. n.a. 135 n.a. [24]
FlexToT v2 HPK S13360-3050CS 2×2×5 mm3 LSO:Ce,0.4%Ca 123 214 9.6 [87]
HRFlexToT HPK S13360-3050CS 2×2×5 mm3 LSO:Ce,0.4%Ca 117 167 12.0 [24]
HRFlexToT FBK NUV-HD 2×2×5 mm3 LSO:Ce,0.4%Ca 119 142 n.a. [24]
FastIC HPK S13360-3050CS n.a. n.a. 140 n.a. [108]
FastIC FBK NUV-HD LF2 M0 2×2×3 mm3 LSO:Ce,0.2%Ca 76 n.a. n.a. [127]
FastIC FBK NUV-HD LF2 M0 3.13×3.13×20 mm3 LSO:Ce,0.2%Ca 127 n.a. n.a. [127]
FastIC FBK NUV-MT 2×2×3 mm3 BGO 330 n.a. n.a. [128]
FastIC FBK NUV-MT 2×2×20 mm3 BGO 490 n.a. n.a. [128]
PETA4 FBK NUV-HD 4×4×25 mm3 LYSO:Ce 460 n.a. 12.8 [23]
EXYT SSPM 2×2×14 mm3 LYSO:Ce 363 n.a. 10.0 [102]
DIET SensL 30020 3×3×16 mm3 LYSO:Ce 244 n.a. 11.0 [122]