Key Points
Question
Is automated insulin delivery (AID) safe and effective for use in adults with type 2 diabetes who are being treated with insulin with or without other glucose-lowering medications?
Findings
In this nonrandomized clinical trial including 305 adults with type 2 diabetes, there was a significant decrease in hemoglobin A1c (HbA1c) levels from 8.2% to 7.4% after 13 weeks of AID use, representing a 0.8–percentage point reduction.
Meaning
In this study, a reduction in HbA1c levels was observed in a socioeconomically, racially, and ethnically diverse cohort of adults with type 2 diabetes who were being treated with insulin with and without other glucose-lowering treatments following AID use.
This nonrandomized clinical trial evaluates whether use of an automated insulin delivery system is associated with lower hemoglobin A1c levels among adults with type 2 diabetes.
Abstract
Importance
There is a need for additional treatment options for people with type 2 diabetes treated with insulin. Given the limited data on the use of automated insulin delivery (AID) systems in type 2 diabetes, studies evaluating their safety and efficacy are important.
Objective
To evaluate the association of AID with hemoglobin A1c (HbA1c) levels in a diverse cohort of adults with type 2 diabetes.
Design, Setting, and Participants
This single-arm prospective trial was conducted at 21 clinical centers in the United States among individuals aged 18 to 75 years with type 2 diabetes who had been using insulin for at least 3 months prior to screening. Participants with AID system use were excluded. The study started with a 14-day standard therapy phase, followed by 13 weeks of treatment with the investigational device. The first participant was enrolled April 11, 2023, and the last participant follow-up visit was February 29, 2024.
Intervention
Participants used the Omnipod 5 AID System for 13 weeks following the 14-day standard therapy phase.
Main Outcomes and Measures
Primary outcome was change in HbA1c level at 13 weeks, tested sequentially for noninferiority (0.3% margin) and superiority, compared with baseline.
Results
Among 305 participants (mean [SD] age, 57 [11] years; 175 [57%] female; 72 [24%] Black, 66 [22%] Hispanic or Latino, and 153 [50%] White), 289 (95%) completed the trial. At baseline, 223 (73%) were using multiple daily injections, 63 (21%) were using basal insulin without bolus, 17 (6%) were using an insulin pump, 188 (62%) were using continuous glucose monitoring, 168 (55%) were using glucagon-like peptide-1 receptor agonists (GLP-1RAs), and 134 (44%) were using sodium-glucose transport protein 2 inhibitors (SGLT-2is). Following AID use, HbA1c levels decreased from a mean (SD) of 8.2% (1.3) at baseline to 7.4% (0.9) at 13 weeks (mean difference, −0.8 [95% CI, −1.0 to −0.7] percentage points; P < .001 for noninferiority and superiority). Improvement was seen across various subgroups (age, sex, race and ethnicity, insurance), and notably with or without use of GLP-1RAs or SGLT-2is and regardless of pretrial mealtime insulin regimen. Time in target glucose range (70-180 mg/dL) increased from a mean (SD) of 45% (25) to 66% (17) (mean difference, 20 [95% CI, 18 to 22] percentage points; P < .001). Percentage of time in hypoglycemic ranges of less than 54 mg/dL and less than 70 mg/dL was noninferior compared with standard therapy. There was 1 episode of severe hypoglycemia and none of diabetic ketoacidosis or hyperosmolar hyperglycemic syndrome.
Conclusions and Relevance
In this nonrandomized clinical trial, HbA1c levels were lower in a diverse cohort of adults with type 2 diabetes following AID initiation, suggesting that AID may be a beneficial and safe option for people with type 2 diabetes using insulin.
Trial Registration
ClinicalTrials.gov Identifier: NCT05815342
Introduction
Diabetes is a major public health concern, affecting more than 38 million Americans,1 with high rates of morbidity and mortality.2 Despite the growing use of glucagon-like peptide-1 receptor agonists (GLP-1RAs) and sodium-glucose transport protein 2 inhibitors (SGLT-2is), many individuals with type 2 diabetes have hemoglobin A1c (HbA1c) levels greater than the target recommendations of the American Diabetes Association.3 Automated insulin delivery (AID) systems have been demonstrated to improve glycemia in type 1 diabetes.4,5,6,7,8 In type 2 diabetes, prior studies of AID have been limited to small (30 participants or less) feasibility trials.9,10,11,12,13
At the time of study start, the Omnipod 5 AID System (Insulet Corporation) was cleared by the US Food and Drug Administration (FDA) and Conformité Européenne–marked for individuals with type 1 diabetes aged 2 years and older. This wearable, tubeless pump communicates with an interoperable continuous glucose monitor (CGM) to adjust insulin delivery based on sensor glucose readings.14 To determine the outcomes of AID in individuals with type 2 diabetes, this study evaluated the Omnipod 5 AID System in a multicenter, single-arm, pivotal trial enrolling a large and diverse population of adults with insulin-treated type 2 diabetes. Data from this study were used to obtain FDA clearance for individuals with type 2 diabetes aged 18 years and older.
Methods
Trial Conduct and Oversight
The trial protocol was approved by central and local institutional review boards with oversight from an independent medical monitor and an investigational device exemption from the FDA. Written informed consent was obtained from each participant. The protocol and statistical analysis plan are available in Supplement 1.
Trial Design and Participants
This 13-week single-arm, prospective trial evaluated the safety and efficacy of AID in adults diagnosed with type 2 diabetes in 21 centers in the United States. The single-arm noninferiority, followed by superiority, trial design was selected because it was appropriate for FDA clearance of a medical device and maximized intervention exposure across a large and heterogeneous group of people with type 2 diabetes. Participants were aged 18 to 75 years and treated with a stable insulin regimen for at least 3 months prior to screening. They could be additionally treated with other antihyperglycemic and weight loss medications without dose change for at least 4 weeks prior to the trial. All participants were required to have a baseline HbA1c of less than 12% (to convert to proportion of total hemoglobin, multiply by 0.01), and for basal insulin–only users, there was a lower HbA1c limit of 7%. Complete eligibility criteria in eTable 1 in Supplement 2. The study defined enrollment goals to recruit a diverse population (eTable 2 in Supplement 2).
CGM data were collected in a 14-day standard therapy phase using the Dexcom G6 CGM (Dexcom), masked for those not already using this CGM. During this period participants continued using their prestudy treatment.
Prior to initiating AID, investigators assessed insulin dosing at mealtimes and provided carbohydrate counting instructions or advised participants to use a simplified meal bolus technique (ie, small, medium, and large, fixed-dose, or correction-only). Participants then transitioned to the 13-week treatment phase using AID, during which those receiving stable doses of noninsulin antihyperglycemic medications continued use of these medications into the treatment phase. Initial pump settings were at the discretion of the investigator. The visit schedule and detailed methods are available in eTables 3 and 4 in Supplement 2.
Investigational Device
The investigational device includes a tubeless insulin pump (Pod) with embedded AID algorithm and an application on a provided locked-down Android phone or a personal Android smartphone with English or Spanish language options. The model predictive control algorithm used in the trial is the same AID algorithm cleared by FDA for use among patients with type 1 diabetes. When used in Automated Mode with an interoperable CGM (study CGM: Dexcom G6), the AID algorithm delivers dynamic microboluses of insulin every 5 minutes or pauses insulin based on glucose values to approach the user configurable target glucose value of 110 mg/dL to 150 mg/dL in 10mg/dL increments, adjustable by time of day (to convert glucose to micromoles per liter, multiply by 0.0555).14
Outcomes
The primary outcome was the change in HbA1c at 13 weeks from baseline, tested sequentially for noninferiority (limit 0.3%) and superiority. Secondary glycemic outcomes, tested in a hierarchal fashion to preserve type I error (eTable 5 in Supplement 2), included change from baseline in mean sensor glucose; percentage of time with sensor glucose in ranges of interest (70-180 mg/dL [time in range, TIR], 70-140 mg/dL, ≥300 mg/dL, >250 mg/dL, >180 mg/dL, <70 mg/dL [noninferiority, 2.0% limit], and <54mg/dL [noninferiority, 0.5% limit]); Type 2 Diabetes Distress Assessment System (T2-DDAS),15 Pittsburgh Sleep Quality Index (PSQI),16 and Hypoglycemia Confidence Scale (HCS) total scores17; proportion with high distress (T2-DDAS ≥2.0), poor sleep quality (PSQI >5.0), and low hypoglycemia confidence (HCS <3.0); percentage of time with glucose less than 70 mg/dL (superiority); percentage of time with glucose less than 54 mg/dL (superiority); and coefficient of variation of sensor glucose. Reportable adverse events are listed in eTable 4 in Supplement 2, and additional outcomes not included in the hierarchy are listed in eTable 6 in Supplement 2.
Sample Size and Power Calculation
A sample of 300 participants initiating the treatment phase with the goal of 275 completers was planned to ensure sufficient exposure of system use by participants using multiple daily injections (MDI), basal without bolus insulin (referred to as basal insulin only), and other noninsulin glucose-lowering medications along with insulin. Based on prior studies,13,18,19 an SD of 0.8 for the change in HbA1c from baseline to 13 weeks was estimated. Assuming 275 completers, statistical power was calculated at greater than 99% for the primary outcome of change in HbA1c levels from baseline to 13 weeks, tested for noninferiority with a limit of 0.3% (1-sided type I error rate of 2.5%; true mean difference of zero is assumed) and for superiority assuming a true mean difference of 0.4% (2-sided type I error rate of 5%).
Statistical Analysis
The primary analysis was conducted with complete cases. Sensitivity analyses were performed to handle missing HbA1c values for both the noninferiority and superiority primary outcomes (statistical significance of P < .025 for 1-sided tests and P < .05 for 2-sided tests). A per-protocol analysis was performed including participants with sufficient system use as defined within eTable 4 in Supplement 2. The primary analysis and most secondary and exploratory analyses were analyzed using paired t tests. If not normally distributed, robust regression using M estimation with the Huber weight function or a nonparametric test was used.
Preplanned superiority analyses were conducted, stratified by insulin regimen at baseline (MDI [not including pump users] and basal without bolus insulin) using the method described for the primary analysis. For this and other planned secondary analyses outside the statistical hierarchy, the false discovery rate was controlled using the adaptive Benjamini-Hochberg procedure.20 For preplanned assessments of differences across subgroups, linear or robust regression models were used to assess interaction between outcome and subgroup of interest adjusted for baseline outcome.
Additional details of the statistical testing methods are included in eTable 4 in Supplement 2. Data analyses were performed using SAS version 9.4 (SAS Institute).
Results
Participants
Between April 11, 2023, and November 14, 2023, 343 individuals were enrolled and 305 initiated AID, with 289 completing the 13-week follow-up period (Figure 1). Of the 305 initiating AID (175 [57%] female; mean [SD] age, 57 [11] years), 223 (73%) were using MDI at baseline; 17 (6%), an insulin pump without AID; and 63 (21%), basal insulin only; 188 (62%) were using CGM. One hundred sixty-eight (55%) were using a GLP-1RA, 135 (44%) an SGLT-2i, and 82 (27%) both. Other notable population characteristics included race (72 [24%] Black, not Hispanic or Latino; 153 [50%] White, not Hispanic or Latino), ethnicity (66 [22%] Hispanic or Latino), insurance (124 [41%] publicly funded or no health insurance). A full list of baseline characteristics is provided in Table 1. Baseline characteristics for the 16 noncompleters are summarized in eTable 7 in Supplement 2.
Figure 1. Trial Flow Diagram.
aHemoglobin A1c (HbA1c) analyses included all participants who had values at both baseline and 13 weeks or early withdrawal if done 42 to 119 days from the initiation of automated insulin delivery system.
bContinuous glucose monitoring (CGM) analyses included all participants with at least 168 hours of CGM data during each of the standard therapy period and the 13-week treatment period.
Table 1. Baseline Characteristics of Trial Population.
| Characteristic | Participants, No. (%) (N = 305) |
|---|---|
| Age, mean (SD) [range], y | 57 (11) [20-75] |
| Duration of type 2 diabetes, median (IQR), y | 17 (11-24) |
| Sex | |
| Female | 175 (57) |
| Male | 130 (43) |
| Race and ethnicitya | |
| American Indian or Alaskan Native, not Hispanic or Latino | 2 (<1) |
| Asian or Pacific Islander, not Hispanic or Latino | 7 (2) |
| Black, not Hispanic or Latino | 72 (24) |
| Hispanic or Latino | 66 (22) |
| White, not Hispanic or Latino | 153 (50) |
| ≥1 Race, not Hispanic or Latino | 3 (1) |
| HbA1c level, mean (SD) [range], % | 8.2 (1.3) [4.9-12.0] |
| HbA1c category, % | |
| <7 | 45 (15) |
| 7 to <8 | 107 (35) |
| 8 to <9 | 82 (27) |
| 9 to <10 | 32 (10) |
| ≥10 | 39 (13) |
| Body mass index, mean (SD) [range]b | 35 (8) [22-70] |
| Standard period of insulin therapy | |
| Multiple daily injections | 223 (73) |
| Basal insulin only injections | 63 (21) |
| Insulin pump | 17 (5.6) |
| Premix insulin injections | 2 (<1) |
| Previous continuous glucose monitor use | |
| Never | 75 (25) |
| In past, but not current | 42 (14) |
| Current | 188 (62) |
| Total daily dose of insulin, mean (SD) [range], U/kg | 0.8 (0.5) [0.1-2.9] |
| Use of select noninsulin glucose-lowering medications | |
| GLP-1RA | 168 (55) |
| SGLT-2i | 135 (44) |
| DPP4i | 8 (3) |
| GLP-1RA and/or SGLT-2i | 221 (72) |
| GLP-1RA and SGLT-2i | 82 (27) |
| Neither GLP-1RA nor SGLT-2i | 84 (28) |
| Health insurancec | |
| Private | 161 (53) |
| Medicaid | 27 (9) |
| Medicare | 52 (17) |
| Other government | 27 (9) |
| No coverage | 18 (6) |
| Educationd | |
| High school diploma or less | 99 (32) |
| Technical, vocational, or associate’s degree | 73 (24) |
| Bachelor degree | 70 (23) |
| Advanced degree | 40 (13) |
| Income, $e | |
| <50 000 | 84 (28) |
| 50 000 to <100 000 | 75 (25) |
| ≥100 000 | 67 (22) |
| Other key medical conditions | |
| Hypertension | 223 (73) |
| Myocardial infarction | 12 (4) |
| Stroke | 5 (2) |
| Lipid abnormality | 265 (87) |
| GAD antibody level, IU/mLf | |
| <5 | 284 (93) |
| 5 to <250 | 7 (2) |
| ≥250 | 13 (4) |
Abbreviations: DPP4i, dipeptidyl peptidase 4 inhibitor; GAD, antiglutamic acid decarboxylase; GLP-1RA, glucagon-like peptide 1 receptor agonist; HbA1c, hemoglobin A1c; SGLT-2i, sodium-glucose cotransporter 2 inhibitor.
SI conversion factors: To convert HbA1c to proportion of total hemoglobin, multiply by 0.01.
Race and ethnicity were reported by the participants and are displayed exactly as reported. Race and ethnicity were unknown or not reported for 2 participants (<1%).
Body mass index is calculated as weight in kilograms divided by height in meters squared.
Health insurance coverage was unknown or not reported for 20 participants (7%).
Educational background was unknown or not reported by 23 participants (8%).
Income was unknown or not reported for 79 participants (26%).
GAD antibody level was unknown or not reported for 1 participant (<1%).
Primary Outcome
HbA1c decreased by mean (SD) of 0.8 (1.0) percentage points (95% CI, −1.0 to −0.7 percentage points) from 8.2% (1.3) at baseline to 7.4% (0.9) at treatment phase end (P < .001 for noninferiority and superiority) (Table 2). Results were similar for the 247 participants included in the per-protocol analysis (eTable 8 in Supplement 2) and similar in analyses that excluded participants who were positive for glutamic acid decarboxylase (GAD) antibodies (eTable 9 in Supplement 2). HbA1c reduction was observed in both prior MDI (−0.8 [95% CI, −0.9 to −0.7] percentage points; P < .001) and prior basal insulin only (−1.2 [95% CI, −1.5 to −0.9]; P < .001) users.
Table 2. Primary and Secondary Outcomes.
| Outcome | Phase | P value | ||
|---|---|---|---|---|
| Mean (SD) | Mean difference (95% CI)b | |||
| Baseline or ST (2 weeks)a,b | End of treatment or treatment phase (13 weeks)a,b | |||
| Primary outcome | ||||
| HbA1c, % | ||||
| Overall | 8.2 (1.3) | 7.4 (0.9) | −0.8 (−1.0 to −0.7)c | <.001c,d |
| Prior MDI userse | 8.2 (1.4) | 7.4 (0.9) | −0.8 (−0.9 to −0.7) | <.001f |
| Prior basal insulin only users | 8.6 (1.2) | 7.5 (0.8) | −1.2 (−1.5 to −0.9) | <.001f |
| Secondary outcomes in prespecified hierarchical orderg | ||||
| Mean sensor glucose, mg/dL | 202 (50) | 170 (24) | −32 (−37 to −28) | <.001 |
| Time in glucose range, % | ||||
| 70-180 mg/dL | 45 (25) | 66 (17) | 20 (18 to 22) | <.001 |
| 70-140 mg/dL | 21 (18) | 33 (17) | 12 (10 to 13) | <.001 |
| ≥300 mg/dLh | 8 (10) | 2 (2) | −5 (−6 to −4) | <.001 |
| >250 mg/dLh | 20 (22) | 7 (8) | −12 (−14 to −11) | <.001 |
| >180 mg/dL | 54 (25) | 34 (17) | −20 (−22 to −18) | <.001 |
| <70 mg/dLi | 0.2 (0.3) | 0.2 (0.2) | 0.0 (−0.1 to 0.0) | <.001 |
| <54 mg/dLi | 0.01 (0.02) | 0.04 (0.05) | 0.01 (0.00 to 0.01) | <.001 |
| T2-DDASj | ||||
| Total intensity score | 2.5 (1.0) | 2.2 (0.9) | −0.3 (−0.4 to −0.2) | <.001 |
| Score ≥2.0, No. (%) | 201 (66) | 167 (55) | NA | <.001 |
| PSQIj | ||||
| Total score | 7.3 (4.0) | 7.0 (4.1) | −0.4 (−0.7 to 0.0) | .04 |
| Score >5.0, No. (%) | 190 (63) | 174 (59) | NA | .10 |
| HCS | ||||
| Total score | 3.2 (0.6) | 3.3 (0.6) | 0.1 (0.0 to 0.1) | NAk |
| Score <3.0, No. (%) | 98 (32) | 75 (25) | NA | NAk |
| Coefficient of variation of sensor glucose, % | 27.8 (6.3) | 27.1 (5.1) | −0.7 (−1.4 to −0.1) | NAk |
Abbreviations: HbA1c, hemoglobin A1c; HCS, Hypoglycemia Confidence Scale; MDI, multiple daily injections; NA, not applicable; PSQI, Pittsburgh Sleep Quality Index; ST, standard therapy; T2-DDAS, Type 2 Diabetes Distress Assessment System.
SI conversion factors: To convert HbA1c to proportion of total hemoglobin, multiply by 0.01; glucose to millimoles per liter, multiply by 0.0555.
Baseline and follow-up (end of trial) data were used for change in HbA1c, T2-DDAS, PSQI, and HCS total score. The remaining outcomes compared the ST phase with the treatment phase.
The number of participants with available data ranged from 294 to 305 across outcomes.
A sensitivity analysis was done using Rubin multiple imputation for the 9 participants who dropped out without a 13-week HbA1c value. This gave the same point estimate, 95% CI, and P value as listed.
P value is testing for superiority. Testing for noninferiority with a limit of 0.3% also gives P < .001.
Does not include prior insulin pump users.
This analysis was completed outside of the statistical hierarchy. The false discovery rate was controlled using the Benjamini-Hochberg procedure for HbA1c level.
A hierarchical approach was used to control the type I error. Hypothesis testing for secondary outcomes was performed sequentially in the order listed in the table. When a P value of .05 or higher was observed, the outcomes below that finding on the list were not formally tested.
Robust mean.
Noninferiority margin of 2.0% for glucose less than 70 mg/dL or 0.5% for glucose less than 54 mg/dL.
Lower scores indicate better outcomes (eg, less distress related to diabetes, better sleep quality).
As part of the hierarchy testing and because of the P value for PSQI score greater than 5 was .10, the formal testing stopped, and P values were not given for any outcomes further down on the list, including superiority testing for percentage of time with glucose levels less than 70 mg/dL and less than 54 mg/dL, which would have been tested before coefficient of variation of sensor glucose.
HbA1c reduction was observed across a wide range of participant characteristics (Figure 2; eTable 10 in Supplement 2). A greater decrease in HbA1c was associated with a higher baseline HbA1c level (eFigure 1A in Supplement 2): those with baseline HbA1c of 9.0% or greater had a decrease of −2.1 (95% CI, −2.3 to −1.9) percentage points, while those with baseline HbA1c level of less than 7.0% saw no change (0.0 [95% CI, −0.3 to 0.2] percentage points) (P < .001). After adjustment for baseline HbA1c, a similar change in HbA1c level was observed among subgroups based on sex, age, socioeconomic status, pretrial mealtime insulin regimen, noninsulin glucose-lowering medication use, C-peptide level, and previous CGM use. There was an apparent interaction between change in HbA1c and race and ethnicity, with non-Hispanic Black participants having a lower observed reduction following adjustment for baseline HbA1c, although all racial and ethnic groups saw benefit (Figure 2; eTable 10 in Supplement 2). Further exploration into these data showed a difference in adoption of the 110 mg/dL glucose target setting between race and ethnicity groups (White [not Hispanic or Latino], Black [not Hispanic or Latino], and Hispanic or Latino used the 110 mg/dL target 56%, 37%, and 55% of cumulative total study time, respectively). A reduction in HbA1c level was observed in both GLP-1RA users and nonusers: HbA1c decreased from a mean (SD) of 8.1% (1.2) at baseline to 7.3% (0.8) at end of treatment in GLP-1RA users (adjusted difference, −0.9 [95% CI, −1.0 to −0.8] percentage points) and from a mean (SD) 8.4% (1.4) to 7.5% (1.0) for GLP-1RA nonusers (adjusted difference, −0.8 [95% CI, −0.9 to −0.6] percentage points). A similar pattern was observed for SGLT-2i users and nonusers. A similar decrease in HbA1c was observed across varying baseline HbA1c ranges in those who used or did not use GLP-1RA (eFigure 2A in Supplement 2). The proportion of participants achieving HbA1c of less than 7% was 14% at baseline (42 of 296 participants) and 37% at treatment phase end (110 participants) (eTable 11 in Supplement 2).
Figure 2. Change in Hemoglobin A1c (HbA1c) Levels in Demographic and Clinical Subgroups.
To convert C-peptide to nanograms per mililiter, divide by 0.331; HbA1c to proportion of total hemoglobin, multiply by 0.01. AID indicates automated insulin delivery; CGM, continuous glucose monitoring; GED, General Educational Development; GLP-1RA, glucagon-like peptide-1 receptor agonist; MDI, multiple daily injections; SGLT-2i, sodium-glucose transport protein 2 inhibitor; ST, standard therapy.
aA total of 9 participants are missing a final HbA1c result and are excluded from this analysis.
bInteraction P values were adjusted for false discovery rate and compare the change in outcome between the characteristic levels after adjusting for baseline HbA1c, except for the stratification by baseline HbA1c group.
cPrior insulin therapy and prior CGM use subgroups were added to the interaction testing post hoc.
dPost hoc interaction P values were adjusted for false discovery rate for prior insulin therapy and prior CGM use subgroups within their own category.
Secondary Outcomes
Glycemic Outcomes
TIR increased from a mean (SD) of 45% (25) at baseline to 66% (17) at 13 weeks (mean difference, 20 [95% CI, 18-22] percentage points; P < .001) or an additional 4.8 h/d in target range (Table 2; eFigure 2B in Supplement 2). The percentage of participants achieving TIR of greater than 70% of the time was 19% during standard therapy (57 of 299 participants) and 42% at end of treatment phase (127 participants) (eTable 11 in Supplement 2). A significant decrease occurred for percentage of time in hyperglycemia ranges, and the noninferiority outcomes were met for percentage of time with glucose less than 70 mg/dL and less than 54 mg/dL. Glycemic improvement was evident during both daytime and nighttime (eFigure 2C and eTable 12 in Supplement 2).
Patient-Reported Outcomes
The T2-DDAS total intensity score decreased by −0.3 points (95% CI, −0.4 to −0.2 points; P < .001). The percentage of participants with high distress (T2-DDAS ≥2.0) decreased from 66% at baseline (164 of 247 participants) to 55% at the end of treatment phase (133 participants) (P < .001). While the PSQI total score decreased by −0.4 points (95% CI, −0.7 to 0.0 points, P = .04), the percentage of participants with poor sleep quality (PSQI >5.0) did not change significantly (Table 2).
Additional Outcomes
Body mass index (BMI; calculated as weight in kilograms divided by height in meters squared) changed from a mean (SD) of 34.9 (7.5) at baseline to 35.1 (7.6) at 13 weeks (change, 0.3 [95% CI, 0.1 to 0.4]; P < .001), corresponding to a body weight increase of 0.8 kg (P < .001). Total daily insulin requirements decreased from mean (SD) of 0.80 (0.46) U/kg/d at baseline to 0.57 (0.29) U/kg/d with AID (change, −0.23 U/kg/d [95% CI, −0.27 to −0.20 U/kg/d]; P < .001) (eTable 13 in Supplement 2). At baseline, 94 participants (31%) used 100 U/d of insulin or more, which decreased to 31 (10%) at 13 weeks.
System Use
In addition to the 2 participants who were lost to follow-up, 14 participants discontinued the AID system prior to 13 weeks. Participants spent a median (IQR) of 94% (87%-97%) of time in Automated Mode during trial participation (n = 305). The 110 mg/dL, 120 mg/dL, 130 mg/dL, 140 mg/dL, and 150 mg/dL targets were each used by 183, 133, 122, 80, and 89 participants, respectively, and were in use for 52%, 27%, 15%, 2%, and 4% of cumulative total study time.
Adverse Events and Device Issues
During the treatment period, 43 participants (14%) reported 53 nonserious adverse events (Table 3). There were no serious adverse device effects, and 19 nonserious adverse device effects reported in 17 participants (6%). There was 1 severe hypoglycemia event, deemed unrelated to malfunction of the trial device, and no cases of diabetic ketoacidosis or hyperosmolar hyperglycemic syndrome. There were 13 nonglycemic serious adverse events, all unrelated to the trial device (eTable 14 in Supplement 2).
Table 3. Adverse Events.
| Event type | Participants, No. (%) (n = 305) |
|---|---|
| Nonserious adverse event | |
| Participants with ≥1 event | 43 (14) |
| Events, No. | 53 |
| Events per 100 person-years, No. | 72 |
| Adverse device effects, No. | 19 |
| Hyperglycemiaa | 12 (11) |
| Skin irritation or bleeding at insertion site | 5 (4) |
| Hypoglycemiab | 1 (1) |
| Otherc | 1 (1) |
| Serious adverse event | |
| Participants with ≥1 event | 13 (4) |
| Events, No.d | 13 |
| Events per 100 person-years, No. | 18 |
| Adverse device effects, No. | 0 |
| Severe hypoglycemiae | 1 (1) |
| Diabetic ketoacidosisf | 0 |
| Hyperosmolar hyperglycemic syndromeg | 0 |
Hyperglycemia requiring evaluation or treatment from a health care facility for an acute event involving hyperglycemia or the site was contacted to receive guidance on how to manage the hyperglycemia.
Participant was experiencing hypoglycemia that required evaluation and treatment at the emergency department but did not meet the criteria for a severe hypoglycemic event.
One participant administered an excessive bolus and was brought to the emergency department for monitoring but did not experience hypoglycemia.
Detailed listing of the 13 serious adverse events are included in eTable 13 in Supplement 2.
Severe hypoglycemia requiring the assistance of another person due to altered consciousness and requiring another person to actively administer carbohydrate, glucagon, or other resuscitative actions.
Hyperglycemia with the presence of polyuria, polydipsia, nausea or vomiting, serum ketones greater than 1.5 mmol/L or large to moderate urine ketones, either arterial blood pH less than 7.30, venous blood pH less than 7.24, or serum bicarbonate less than 15 mmol/L, and treatment provided in a health care facility.
Hyperglycemia with the presence of polyuria, polydipsia, nausea, or vomiting, plasma glucose greater than 600 mg/dL, plasma effective osmolarity >320 mOsm/L, absence of significant ketones, and treatment provided in a health care facility.
Overall, 73 device issues were reported during the treatment phase. A total of 58% of these were related to the Pod, 21% related to the controller or personal smartphone, 12% related to the CGM sensor, and 10% related to the CGM transmitter (eTable 15 in Supplement 3).
Acceptance of Treatment Technology Survey
At study end, participants completed a system opinion survey to assess their acceptance of the new treatment technology. The survey found that most participants strongly agreed or agreed that it was easy to use the bolus calculator (91%), they would recommend the AID system to family or friends (90%), the AID system was easy to wear (82%), they would continue using the AID system after completion of the trial (78%), and they barely noticed wearing the AID System (72%) (eTable 16 in Supplement 2).
Discussion
In this multisite nonrandomized clinical trial of AID use in a heterogenous adult cohort with type 2 diabetes on insulin therapy, there was a substantial reduction in HbA1c, with a particularly notable decrease of 2.1 percentage points in those with a high baseline HbA1c (≥9.0%). Improvement in HbA1c was observed in participants using MDI and in those using basal insulin only at baseline and across diverse racial, ethnic, and socioeconomic backgrounds; pretrial mealtime insulin regimen for those receiving meal time insulin; range of C-peptide levels; and noninsulin antihyperglycemic medication use, including GLP-1RAs and SGLT-2is.
Improvements also were observed in TIR, with an increase of 20 percentage points, amounting to an additional 4.8 h/d in target range. These glycemic improvements were achieved with no increase in hypoglycemia. Notably, a similar improvement in TIR was observed for both daytime and nighttime hours, which differs from what has been frequently observed in AID studies of individuals with type 1 diabetes, for whom a larger improvement in TIR is observed during nighttime, when meals are not being consumed, compared with the daytime.4,19,21,22,23 The majority of participants (83%) did not experience a meaningful change in weight (within 5% of baseline); however, weight gain was observed primarily in those with highest baseline HbA1c levels, indicative of a catabolic state in need of recovery.
A significant reduction in distress related to diabetes and improved quality of sleep also were observed. Of note, 94% of participants were pump naive at baseline and were able to initiate the study AID system without a run-in period of pump training prior to automation of insulin: whether this approach could be implemented across all available AID systems is unknown.
Although GLP-1RA and SGLT-2i therapy can produce substantial benefits in many people with type 2 diabetes,24 most participants receiving stable doses in this trial were not achieving glycemic targets3 at baseline; only 16% of GLP-1RA users and 15% of SGLT-2i users had baseline HbA1c levels less than 7%, similar to the 13% frequency in participants not using GLP-1RAs or SGLT-2is. In this trial, similar improvements in glycemic outcomes with AID were observed in users and nonusers of GLP-1RA and SGLT-2i therapy, suggesting that adults with type 2 diabetes who are taking stable doses of GLP1-RA and SGLT-2i and require insulin therapy can benefit from AID.
There have been a small number of clinical trials assessing the feasibility of the more recently available hybrid and fully closed-loop AID systems in adults with type 2 diabetes.9,10,11,12,13 The feasibility trial of the Omnipod 5 System used in this study showed similar glycemic improvements in 24 adults with type 2 diabetes.13 Two trials of the t:slim X2 insulin pump with Control-IQ technology, a hybrid closed-loop AID system, each demonstrated improvements in glycemic outcomes in groups of 30 adults, with 1 being a 6-week single-arm trial9 and the other being a 12-week randomized, parallel group trial.10 Two randomized crossover trials evaluating the CamAPS HX fully closed-loop AID system over 20 days to 8 weeks of use also demonstrated improved glycemic outcomes in groups of 26 adults.11,12 These small prior studies, while limited in their direct comparison to this present study due to differences in study type and hybrid vs fully closed-loop systems, further support the benefits of AID in adults with insulin-treated type 2 diabetes.
Limitations and Strengths
This study has limitations, including its single-arm design that did not include a concurrent control group. The amount of improvement that might be expected with CGM alone or attributed to a study effect can be estimated from a previous study evaluating CGM in type 2 diabetes. In a randomized clinical trial of CGM intervention in 158 people with type 2 diabetes using MDI,25 the subgroups with baseline HbA1c level of 8.5% or greater had a 1.1 percentage point improvement in HbA1c with CGM vs a 0.7 percentage point improvement in the control group. The comparable sample of those with baseline HbA1c level of 8.5% or greater in our study (113 participants) had an improvement of 1.7 percentage points for those not using CGM at baseline and 1.5 percentage points for those already using CGM. Looking within the groups of the previous CGM randomized clinical trial provides some perspective on how much improvement observed in this study might have been due to the addition of CGM or study effect alone.
Notable strengths of this trial include the large and diverse population recruited with varying racial and ethnic backgrounds, prior insulin regimen and medication use, representing the diverse general population with type 2 diabetes. This trial intentionally enrolled a higher proportion of Black and Hispanic participants to encompass communities disproportionately affected by adverse social determinants of health, who also have higher prevalence rates of diabetes26 and have often been underrepresented in clinical trials.27 Compared with the general US population,28 this study sample was similar in terms of education level, insurance, and skewed slightly more to the lower household income range, likely due to higher mean age.
Conclusions
During a single-arm nonrandomized trial in adults with type 2 diabetes, glycemia improved over 13 weeks of AID use with no increase in hypoglycemia. Improvement in HbA1c levels was observed in participants using MDI and in those using basal insulin only at baseline, across diverse racial, ethnic, and socioeconomic backgrounds, and among individuals using noninsulin glucose-lowering medications, including GLP-1RAs and SGLT-2is.
Trial Protocol and Statistical Analysis Plan
eTable 1. Eligibility and Exclusion Criteria
eTable 2. Study Enrollment Goals
eTable 3. Schedule of Participant Visits and Procedures
eTable 4. Additional Research Design and Methods
eTable 5. Hierarchical Listing of Primary and Secondary Efficacy Outcomes
eTable 6. Listing of Additional Outcomes Included
eTable 7. Baseline Characteristics of the 16 Participants Who Did Not Complete the Trial
eTable 8. Per-Protocol Analyses
eTable 9. Analyses Excluding Participants Positive for GAD Antibodies
eTable 10. Subgroup Analysis for Primary Outcome
eTable 11. Percentage of Participants Achieving Targets
eTable 12. CGM Metrics by Time of Day Overall and by Prior Insulin Therapy Method
eTable 13. BMI, Weight, and Insulin Use
eTable 14. Summary of Serious Adverse Events
eTable 15. Summary of Device Issues
eTable 16. System Opinion Questionnaire
eFigure 1. Scatterplot of Change in HbA1c Levels and Time With Glucose Between 70 and 180 mg/dL
eFigure 2. Glycemic Outcomes Stratified by Baseline HbA1c and by Time of Day
Nonauthor Collaborators
Data Sharing Statement
References
- 1.US Centers for Disease Control and Prevention. National diabetes statistics report. May 15, 2024. Accessed January 6, 2025. https://www.cdc.gov/diabetes/php/data-research/index.html
- 2.Ekanayake P, Edelman S. Identifying patients with type 2 diabetes who might benefit from insulin pump therapy: literature review, clinical opportunities, potential benefits and challenges. Diabetes Obes Metab. 2023;25(S2)(suppl 2):3-20. doi: 10.1111/dom.15059 [DOI] [PubMed] [Google Scholar]
- 3.American Diabetes Association Professional Practice Committee . 6. Glycemic goals and hypoglycemia: standards of care in diabetes—2024. Diabetes Care. 2024;47(suppl 1):S111-S125. doi: 10.2337/dc24-S006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sherr JL, Bode BW, Forlenza GP, et al. ; Omnipod 5 in Preschoolers Study Group . Safety and glycemic outcomes with a tubeless automated insulin delivery system in very young children with type 1 diabetes: a single-arm multicenter clinical trial. Diabetes Care. 2022;45(8):1907-1910. doi: 10.2337/dc21-2359 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Forlenza GP, Ekhlaspour L, DiMeglio LA, et al. Glycemic outcomes of children 2-6 years of age with type 1 diabetes during the pediatric MiniMed 670G system trial. Pediatr Diabetes. 2022;23(3):324-329. doi: 10.1111/pedi.13312 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Choudhary P, Kolassa R, Keuthage W, et al. ; ADAPT study Group . Advanced hybrid closed loop therapy versus conventional treatment in adults with type 1 diabetes (ADAPT): a randomised controlled study. Lancet Diabetes Endocrinol. 2022;10(10):720-731. doi: 10.1016/S2213-8587(22)00212-1 [DOI] [PubMed] [Google Scholar]
- 7.Kanapka LG, Wadwa RP, Breton MD, et al. ; iDCL Trial Research Group . Extended use of the Control-IQ closed-loop control system in children with type 1 diabetes. Diabetes Care. 2021;44(2):473-478. doi: 10.2337/dc20-1729 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Ware J, Boughton CK, Allen JM, et al. ; DAN05 Consortium . Cambridge hybrid closed-loop algorithm in children and adolescents with type 1 diabetes: a multicentre 6-month randomised controlled trial. Lancet Digit Health. 2022;4(4):e245-e255. doi: 10.1016/S2589-7500(22)00020-6 [DOI] [PubMed] [Google Scholar]
- 9.Levy CJ, Raghinaru D, Kudva YC, et al. Beneficial effects of control-IQ automated insulin delivery in basal-bolus and basal-only insulin users with type 2 diabetes. Clin Diabetes. 2024;42(1):116-124. doi: 10.2337/cd23-0025 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Reznik Y, Carvalho M, Fendri S, et al. Should people with type 2 diabetes treated by multiple daily insulin injections with home health care support be switched to hybrid closed-loop? the CLOSE AP+ randomized controlled trial. Diabetes Obes Metab. 2024;26(2):622-630. doi: 10.1111/dom.15351 [DOI] [PubMed] [Google Scholar]
- 11.Daly AB, Boughton CK, Nwokolo M, et al. Fully automated closed-loop insulin delivery in adults with type 2 diabetes: an open-label, single-center, randomized crossover trial. Nat Med. 2023;29(1):203-208. doi: 10.1038/s41591-022-02144-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Boughton CK, Tripyla A, Hartnell S, et al. Fully automated closed-loop glucose control compared with standard insulin therapy in adults with type 2 diabetes requiring dialysis: an open-label, randomized crossover trial. Nat Med. 2021;27(8):1471-1476. doi: 10.1038/s41591-021-01453-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Davis GM, Peters AL, Bode BW, et al. Safety and efficacy of the Omnipod 5 automated insulin delivery system in adults with type 2 diabetes: from injections to hybrid closed-loop therapy. Diabetes Care. 2023;46(4):742-750. doi: 10.2337/dc22-1915 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Forlenza GP, Buckingham BA, Brown SA, et al. First outpatient evaluation of a tubeless automated insulin delivery system with customizable glucose targets in children and adults with type 1 diabetes. Diabetes Technol Ther. 2021;23(6):410-424. doi: 10.1089/dia.2020.0546 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Fisher L, Polonsky WH, Perez-Nieves M, Desai U, Strycker L, Hessler D. A new perspective on diabetes distress using the type 2 diabetes distress assessment system (T2-DDAS): prevalence and change over time. J Diabetes Complications. 2022;36(8):108256. doi: 10.1016/j.jdiacomp.2022.108256 [DOI] [PubMed] [Google Scholar]
- 16.Buysse DJ, Reynolds CF III, Monk TH, Berman SR, Kupfer DJ. The Pittsburgh Sleep Quality Index: a new instrument for psychiatric practice and research. Psychiatry Res. 1989;28(2):193-213. doi: 10.1016/0165-1781(89)90047-4 [DOI] [PubMed] [Google Scholar]
- 17.Polonsky WH, Fisher L, Hessler D, Edelman SV. Investigating hypoglycemic confidence in type 1 and type 2 diabetes. Diabetes Technol Ther. 2017;19(2):131-136. doi: 10.1089/dia.2016.0366 [DOI] [PubMed] [Google Scholar]
- 18.Brown SA, Forlenza GP, Bode BW, et al. ; Omnipod 5 Research Group . Multicenter trial of a tubeless, on-body automated insulin delivery system with customizable glycemic targets in pediatric and adult participants with type 1 diabetes. Diabetes Care. 2021;44(7):1630-1640. doi: 10.2337/dc21-0172 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Brown SA, Kovatchev BP, Raghinaru D, et al. ; iDCL Trial Research Group . Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med. 2019;381(18):1707-1717. doi: 10.1056/NEJMoa1907863 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Benjamini Y, Hochberg Y. On the Adaptive Control of the False Discovery Rate in Multiple Testing With Independent Statistics. J Educ Behav Stat. 2000;25(1):60-83. doi: 10.3102/10769986025001060 [DOI] [Google Scholar]
- 21.Thabit H, Lubina-Solomon A, Stadler M, et al. Home use of closed-loop insulin delivery for overnight glucose control in adults with type 1 diabetes: a 4-week, multicentre, randomised crossover study. Lancet Diabetes Endocrinol. 2014;2(9):701-709. doi: 10.1016/S2213-8587(14)70114-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Collyns OJ, Meier RA, Betts ZL, et al. Improved glycemic outcomes with Medtronic MiniMed advanced hybrid closed-loop delivery: results from a randomized crossover trial comparing automated insulin delivery with predictive low glucose suspend in people with type 1 diabetes. Diabetes Care. 2021;44(4):969-975. doi: 10.2337/dc20-2250 [DOI] [PubMed] [Google Scholar]
- 23.Carlson AL, Sherr JL, Shulman DI, et al. Safety and glycemic outcomes during the MiniMed advanced hybrid closed-loop system pivotal trial in adolescents and adults with type 1 diabetes. Diabetes Technol Ther. 2022;24(3):178-189. doi: 10.1089/dia.2021.0319 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Young KG, McInnes EH, Massey RJ, et al. ; ADA/EASD PDMI . Treatment effect heterogeneity following type 2 diabetes treatment with GLP1-receptor agonists and SGLT2-inhibitors: a systematic review. Commun Med (Lond). 2023;3(1):131. doi: 10.1038/s43856-023-00359-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Billings LK, Parkin CG, Price D. Baseline glycated hemoglobin values predict the magnitude of glycemic improvement in patients with type 1 and type 2 diabetes: subgroup analyses from the DIAMOND study program. Diabetes Technol Ther. 2018;20(8):561-565. doi: 10.1089/dia.2018.0163 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.US Centers for Disease Control and Prevention. National Diabetes Statistics Report Appendix A: Detailed Tables. May 15, 2024. Accessed January 6, 2025. https://www.cdc.gov/diabetes/php/data-research/appendix.html#cdc_report_pub_study_section_2-table-2
- 27.Akturk HK, Agarwal S, Hoffecker L, Shah VN. Inequity in racial-ethnic representation in randomized controlled trials of diabetes technologies in type 1 diabetes: critical need for new standards. Diabetes Care. 2021;44(6):e121-e123. doi: 10.2337/dc20-3063 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.US Census Bureau. Census Bureau releases new educational attainment data. February 24, 2022. Accessed January 6, 2025. https://www.census.gov/newsroom/press-releases/2022/educational-attainment.html
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Trial Protocol and Statistical Analysis Plan
eTable 1. Eligibility and Exclusion Criteria
eTable 2. Study Enrollment Goals
eTable 3. Schedule of Participant Visits and Procedures
eTable 4. Additional Research Design and Methods
eTable 5. Hierarchical Listing of Primary and Secondary Efficacy Outcomes
eTable 6. Listing of Additional Outcomes Included
eTable 7. Baseline Characteristics of the 16 Participants Who Did Not Complete the Trial
eTable 8. Per-Protocol Analyses
eTable 9. Analyses Excluding Participants Positive for GAD Antibodies
eTable 10. Subgroup Analysis for Primary Outcome
eTable 11. Percentage of Participants Achieving Targets
eTable 12. CGM Metrics by Time of Day Overall and by Prior Insulin Therapy Method
eTable 13. BMI, Weight, and Insulin Use
eTable 14. Summary of Serious Adverse Events
eTable 15. Summary of Device Issues
eTable 16. System Opinion Questionnaire
eFigure 1. Scatterplot of Change in HbA1c Levels and Time With Glucose Between 70 and 180 mg/dL
eFigure 2. Glycemic Outcomes Stratified by Baseline HbA1c and by Time of Day
Nonauthor Collaborators
Data Sharing Statement