Abstract
Introduction
Patient-specific guides (PSGs) provide customized solutions and enhanced precision. However, the question remains: does clinical evidence support the added value of PSGs? This study critically appraises, summarizes, and compares the literature to assess the clinical value of PSGs in orthopedic surgery.
Materials and methods
PubMed and Embase were used to search for studies reporting on randomized controlled trials (RCTs) that compared the use of PSGs with a control group for an orthopedic intervention, excluding knee arthroplasty. The risk of bias was assessed using the Cochrane risk-of-bias tool (RoB 2). The clinical value was expressed as patient reported outcome measures (PROMs), complications, accuracy, surgery duration, blood loss, and radiation exposure. Relative and absolute differences were determined, and whether these were negative or positive for using PSGs.
Results
From 6310 studies, 27 RCTs were included, covering various interventions. The studies' heterogeneity prevented meta-analysis. Six (22.2%) of the included articles scored low risk of bias. Significant differences in the benefit of PSGs were reported across all included metrics: 32.2% in PROMs, 22.7% in complications, 69.8% in accuracy, 42.1% in surgery duration, 46.7% in blood loss, and 93.3% in radiation exposure. No significant negative differences were found in any of the studies.
Conclusion
PSGs generally show superior outcomes for accuracy and radiation exposure across multiple intervention types, while the reduction in complications was primarily significant in spinal fusion surgery. For PROMs, complications in other treatments, surgery duration, and blood loss, there may be clinical added value but future well-designed RCTs are needed to provide stronger evidence.
Supplementary Information
The online version contains supplementary material available at 10.1007/s00402-025-05775-2.
Keywords: Patient-specific, 3D, Three-dimensional, 3D printing, Orthopedic surgery, Surgical planning, Systematic review, RCT, Clinical added value, Clinical outcome
Introduction
In orthopedic surgery, patient specific guides (PSGs) could offer surgical guidance on pre-planned customized solutions, such as drilling, sawing, and (re-)positioning of bony structures [1]. The construction of a PSG begins with creating 3D bone models based on preoperative imaging, most commonly CT scans, to virtually plan the procedure. A digital 3D model is then designed to fit precisely onto a bony surface, incorporating the necessary functionality. These models are 3D printed and sterilized before surgery. During the operation, the bony surface is exposed, and the PSG is fitted onto the surface, enabling precision surgery.
3D printing is increasingly available for orthopedic surgeons. It is frequently used to create 3D printed anatomical models, offering surgeons supplementary insights into the nature of fractures or anatomy at relatively low cost. Studies have demonstrated that the use of 3D fracture models outperform conventional treatments across various outcomes [2–4]. PSGs can also potentially enhance the accuracy of specific treatments and provide customized solutions, but there is limited scientific evidence to support this. In maxillofacial and dental surgery, PSGs are more frequently utilized than in other specialties, and their use has been shown to result in superior outcomes [5, 6]. Just as 3D printed fracture models and PSGs in maxillofacial and dental surgery, PSGs could offer added value for various treatments in orthopedic surgery.
Despite the growing popularity in orthopedics, there is limited high level of evidence regarding the clinical added value. As described in the narrative review of Gauci et al., PSGs may offer several advantages including improved patient outcomes, reduced complication rates, increased accuracy, shorter surgery duration, decreased blood loss, and lower radiation exposure, compared to conventional interventions [1]. However, due to the narrative character, the clinical value of PSGs was not quantitatively assessed. One scoping review on 3D printing in orthopedic surgery, showed added value for 3D technology [7]. However, there remains a knowledge gap regarding PSGs specifically, as this review did not distinguish among types of 3D technology, e.g., PSGs, anatomical models, and virtual surgical planning. Furthermore, for total and unicompartimental knee arthroplasty, reviews have demonstrated minimal added value for PSGs [8–11]. This result could be explained by the fact that the standard knee arthroplasty operation kit is already somewhat patient-specific due to conventional sizing tools. However, for other types of orthopedic interventions where such kits are often lacking, PSGs may offer greater clinical value compared to conventional treatment.
This study aims to critically appraise, summarize, and compare the literature on randomized controlled trials (RCTs) that evaluate the clinical value of PSGs in orthopedic surgery compared to controls, excluding knee arthroplasty surgery.
Materials and methods
This systematic review was performed in accordance with the Preferred Reporting Items for Systematic review and Meta‐Analyses (PRISMA) [12].
Eligibility criteria
Studies were included that reported on RCTs in humans comparing the use of PSGs in orthopedic surgery with a control group. Studies investigating total or unicompartmental knee arthroplasty as well as phantom, animal, and cadaveric studies were excluded. This review focuses on recent literature, mainly because 3D technology has emerged and made substantial growth over the last decade. To ensure comprehensive coverage, studies of English origin and published in the past 15 years were considered.
Search strategy
A systematic search was performed in PubMed and Embase in September 2023. A combination of three main search strings was used, including terms related to orthopedic surgery, surgical guides and methods of how these guides can be made or for what purposes they can be used (e.g., 3D printed or drilling, respectively). A filter was applied to include publications since 2008. Additionally, a specified RCT filter was applied [13]. Articles that mentioned total- or unicompartmental knee arthroplasty in the title were excluded. The complete search strategy can be found in Supplementary data I.
Study selection
Two authors (N.K. and R.t.H.) independently screened the titles and abstracts according to the prespecified inclusion and exclusion criteria. After screening, the full text was read independently by two authors (N.K. and R.t.H.) to determine final inclusion. Discrepancies were discussed between reviewers and, if necessary, a third reviewer (F.S.) was involved to achieve consensus. Additionally, the references of included articles and some considered to be relevant reviews found during the screening process were hand‐searched to identify relevant studies that may have been overlooked by the search strategy.
Risk of bias assessment and data extraction
All studies were independently assessed for risk of bias by two reviewers (N.K. and R.t.H.) using the revised Cochrane Risk of Bias 2 (RoB2) tool [14]. This tool scores on five domains: randomization process, deviation from intended interventions, missing outcome data, measurement of the outcome, and selection of the reported results, with scores of low risk, some concerns, and high risk of bias. The overall bias is determined by the worst outcome in a domain. Any discrepancies were discussed by the two reviewers, optionally with the third reviewer (F.S.). Both reviewers (N.K. and R.t.H.) independently extracted half of the data and cross-reviewed each other's extractions. In case of missing, unclear data, or data provided in figures, the author or journal was contacted by email (Table 1).
Table 1.
Study Characteristics
| Intervention | References | Country/Region | N | N Intervention group | N Control group | Procedure | Function PSG |
|---|---|---|---|---|---|---|---|
| Spinal fusion | Cecchinato [50] | Italy | 29 | 14 | 15 | Surgery for spinal deformity | Screw positioning |
| Chen [51] | China | 43 | 20 | 23 | Lumbar pedicle screw fixation | Screw positioning | |
| Zhang [52] | China | 40 | 20 | 20 | Pedicle screw fixation for thoracolumbar fractures | Screw positioning | |
| Merc [53] | Slovenia | 24 | 11 | 13 | Pedicle screw positioning lumbar and sacral spine | Screw positioning | |
| Feng [54] | China | 12 | 6 | 6 | Cervical lateral mass screw fixation | Screw positioning | |
| Cui [55] | China | 84 | 42 | 42 | Sacral 2 alar iliac (S2AI) screws fixation | Screw positioning | |
| Merc [56] | Slovenia | 19 | 9 | 10 | Pedicle screw positioning lumbar and sacral spine | Screw positioning | |
| Total hip arthroplasty (THA) | Wang [57] | China | 104 | Divided into four subgroups based Crowe’s classification for hip dysplasia (Crow I, II, III, IV) | |||
| I | 21 | 21 | Total hip arthroplasty in patients with four stages of hip dysplasia | Positioning acetabular component | |||
| II | 13 | 14 | |||||
| III | 8 | 7 | |||||
| IV | 10 | 10 | |||||
| Jin [58] | China | 80 | 40 | 40 | Total hip arthroplasty | Femoral stem positioning | |
| Zhang [59] | China | 20 | 10 | 10 | Metal-on-metal hip resurfacing arthroplasty | Positioning acetabular and femoral components | |
| Small [60] | United States | 36 | 18 | 18 | Total hip arthroplasty | Positioning acetabular component | |
| Zhang [61] | China | 22 | 11 | 11 | Total hip arthroplasty in patients with hip dysplasia | Reaming acetabulum | |
| Zhang [62] | China | 53 | 23 | 30 | Hip arthroplasty (total and hemi) | Positioning femoral stem | |
| Anterior cruciate ligament (ACL) reconstruction | Zhu [63] | China | 78 | 40 | 38 | Anterior cruciate ligament reconstruction | Drilling tibial tunnel |
| Liu [64] | China | 41 | 22 | 19 | Anterior cruciate ligament reconstruction | Drilling femoral and tibial tunnel | |
| Lan [65] | China | 80 | 40 | 40 | Anterior cruciate ligament reconstruction | Drilling femoral tunnel | |
| High tibial osteotomy (HTO) | Gao [66] | China | 39 | 16 | 23 | High tibial osteotomy | Sawing and repositioning |
| Zhu [67] | China | 96 | 48 | 48 | High tibial osteotomy | Sawing and repositioning | |
| Total shoulder arthroplasty (TSA) | Boekel [68]† | Australia | 47 | 24 | 23 | Reversed total shoulder arthroplasty | Positioning glenoid component |
| Hendel [69] | United States | 31 | 15 | 16 | Total shoulder arthroplasty | Positioning glenoid component | |
| Percutaneous vertebroplasty | Hu [70] | China | 36 | 18 | 18 | Percutaneous vertebroplasty | Puncture assistance |
| Chen [71] | China | 97 | 47 | 50 | Percutaneous vertebroplasty | Puncture assistance | |
| Distal radius osteotomy | Buijze [72] | United States and Europe | 37 | 20 | 17 | Corrective Osteotomy Distal Radial Malunion | Sawing and repositioning |
| Distal humerus osteotomy | Hu [73] | China | 35 | 16 | 19 | Correction of cubitus varus deformity in children | Sawing assistance |
| Correction of lower limb deformities | Fan [74] | China | 55 | 21 | 24 | Correction of lower limb deformities in children | Drill assistance |
| Acetabular osteotomy | Ma [75] | China | 22 | 11 | 11 | Periacetabular osteotomy in acetabular dysplasia | Sawing assistance |
| Femoral neck fracture repair | Wang [76] | China | 60 | 30 | 30 | Femoral neck fracture repair | Screw positioning |
The studies are ordered per intervention type and within an intervention on risk of bias assessment. † In the control group virtual surgical planning was used
Outcome measures and analysis
The outcome measures of this systematic review included PROMs, complications, accuracy, surgery duration, blood loss, and radiation exposure. Accuracy metrics are included that compare the postoperative result quantitatively with the preoperative plan per individual, e.g., the positional (mm) or angular (°) deviation. Additionally, metrics were also included that describe the results of PSG use, such as the number of implants placed in safe zone or the number of insertions required to achieve the desired position. All six outcome measures are documented separately for the PSGs group and control group along with corresponding p-values (Table 2 and Supplementary data II). Due to the heterogeneity of orthopedic interventions and the variety of outcome measures (incl variety in follow-up), performing meta-analyses was deemed not appropriate. However, to summarize the outcomes and create a clear overview of the clinical value, the absolute changes (Δ) and relative changes (Δ%) of outcome measures have been displayed (Table 2 and Supplementary data II). These changes were determined by calculating the difference in mean or median values for continuous variables and by calculating the difference in percentages for discrete variables. If percentages were not explicitly stated, they were calculated based on the data provided by the study. For efficient interpretation of the results, the differences were described as positive or negative for PSGs in comparison to the controls. The notation used is: “–” for significant negative differences, “ + ” for significant positive differences, “(–)” for insignificant negative differences, “( +)” for insignificant positive differences, and “ = ” for no difference between the two groups. If it was unclear whether the difference was negative or positive for the use of PSG, it was marked as “?”. To assess the reliability of the studies, the overall risk of bias was calculated. Furthermore, to indicate if positive significant results originated more often from qualitative better studies with a lower risk of bias, the overall percentage of studies with low risk of bias was compared to the low risk of bias rate within all significant outcomes per outcome measure. In all tables and Fig. 2, the studies are ordered by intervention type, starting with those that have the most studies. Within each intervention type, studies are further ordered by their risk of bias score, as shown in Fig. 2.
Table 2.
Study outcomes
| Patient reported outcome measures (PROMs) | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Patient outcome description | Value in intervention group | Value in control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Total hip arthroplasty (THA) | Wang [57] | 104 | Divided into four subgroups based Crowe’s classification for hip dysplasia (Crow I, II, III, IV) | ||||||
| I | HHS pre-op | 56.5 ± 10.4 | 55.8 ± 13.3 | 0.847 | ~ | ~ | ~ | ||
| HHS 3 m post-op | 85.2 ± 3.6 | 84.5 ± 3.3 | 0.499 | 0.7 | 0.9 | ( +) | |||
| HHS 1y post-op | 88.4 ± 4.7 | 90.5 ± 3.8 | 0.088 | − 2.0 | − 2.2 | (-) | |||
| II | HHS pre-op | 52.9 ± 13.1 | 54.3 ± 9.0 | 0.757 | ~ | ~ | ~ | ||
| HHS 3 m post-op | 85.9 ± 4.4 | 84.3 ± 6.3 | 0.458 | 1.6 | 1.9 | ( +) | |||
| HHS 1y post-op | 89.8 ± 5.3 | 87.2 ± 3.3 | 0.170 | 2.5 | 2.9 | ( +) | |||
| III | HHS pre-op | 44.9 ± 17.1 | 43.0 ± 16.9 | 0.835 | ~ | ~ | ~ | ||
| HHS 3 m post-op | 83.1 ± 5.1 | 77.7 ± 4.3 | 0.043* | 5.4 | 7.0 | + | |||
| HHS 1y post-op | 88.9 ± 5.5 | 82.4 ± 2.9 | 0.015* | 6.4 | 7.8 | + | |||
| IV | HHS pre-op | 40.3 ± 13.8 | 41.0 ± 11.2 | 0.902 | ~ | ~ | ~ | ||
| HHS 3 m post-op | 71.9 ± 6.7 | 70.1 ± 7.1 | 0.566 | 1.8 | 2.6 | ( +) | |||
| HHS 1y post-op | 87.3 ± 3.5 | 83.2 ± 4.3 | 0.031* | 4.1 | 4.9 | + | |||
| Jin [58] | 80 | VAS pre-op | 6.4 ± 0.9 | 6.3 ± 1.1 | 0.657 | ~ | ~ | ~ | |
| VAS 7d post-op | 2.2 ± 0.6 | 2.3 ± 0.8 | 0.42 | − 0.1 | − 5.6 | ( +) | |||
| VAS 4w post-op | 1.6 ± 0.5 | 1.7 ± 0.8 | 0.41 | − 0.1 | − 7.5 | ( +) | |||
| VAS 3 m post-op | 1.4 ± 0.7 | 1.3 ± 0.6 | 0.857 | 0.1 | + 1.5 | (–) | |||
| VAS 6 m post-op | 0.8 ± 0.6 | 0.9 ± 0.6 | 0.713 | − 0.1 | − 10.8 | ( +) | |||
| VAS 12 m post-op | 0.7 ± 0.5 | 0.7 ± 0.6 | 0.845 | 0.0 | 0.0 | = | |||
| HHS pre-op | 48.9 ± 4.8 | 49.7 ± 5.6 | 0.536 | ~ | ~ | ~ | |||
| HHS 7d post-op | 82.8 ± 5.0 | 76.8 ± 6.4 | < 0.001* | 6.0 | 7.7 | + | |||
| HHS 4w post-op | 89.2 ± 2.8 | 85.9 ± 3.1 | < 0.001* | 3.3 | 3.8 | + | |||
| HHS 3 m post-op | 93.9 ± 1.7 | 92.7 ± 1.8 | 0.003* | 1.2 | 1.3 | + | |||
| HHS 6 m post-op | 95.0 ± 1.8 | 94.5 ± 2.2 | 0.156 | 0.5 | 0.5 | ( +) | |||
| High tibial osteotomy (HTO) | Gao [66] | 39 | Clinical AKSS pre-op | 57.5 ± 2.0 | 57.1 ± 2.4 | 0.616 | ~ | ~ | ~ |
| Clinical AKSS 3 m post-op | 75.6 ± 7.3 | 70.2 ± 8.4 | 0.042* | 5.5 | 7.8 | + | |||
| Clinical AKSS 6 m post-op | 85.0 ± 6.1 | 80.2 ± 7.4 | 0.040* | 4.8 | 6.0 | + | |||
| Clinical AKSS last follow-up | 90.3 ± 6.2 | 89.4 ± 6.3 | 0.654 | 0.9 | 1.0 | ( +) | |||
| Functional AKSS pre-op | 58.8 ± 6.2 | 59.1 ± 7.6 | 0.870 | ~ | ~ | ~ | |||
| Functional AKSS 3 m post-op | 70.9 ± 8.0 | 64.8 ± 9.0 | 0.034* | 6.2 | 9.5 | + | |||
| Functional AKSS 6 m post-op | 81.6 ± 4.7 | 80.4 ± 8.3 | 0.625 | 1.1 | 1.4 | ( +) | |||
| Functional AKSS last follow-up | 90.0 ± 6.3 | 89.6 ± 6.4 | 0.835 | 0.4 | 0.5 | ( +) | |||
| Total shoulder arthroplasty (TSA) | Boekel [68]† | 47 | CMS pre-op | 30.2 ± 17.6 | 26.6 ± 16.3 | 0.48 | ~ | ~ | ~ |
| CMS 6 m post-op | 68.6 ± 13.8 | 60.3 ± 16.3 | 0.06 | 8.3 | 13.8 | ( +) | |||
| ASES score pre-op | 33.7 ± 15.1 | 35.4 ± 10.5 | 0.64 | ~ | ~ | ~ | |||
| ASES score 6 m post-op | 78.1 ± 16.0 | 75.3 ± 18.8 | 0.58 | 2.8 | 3.7 | ( +) | |||
| OSS pre-op | 24.3 ± 7.5 | 23.1 ± 6.1 | 0.57 | ~ | ~ | ~ | |||
| OSS 6 m post-op | 40.7 ± 5.7 | 38.4 ± 6.6 | 0.20 | 2.3 | 6.1 | (-) | |||
| Complications | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Complication type | Number in intervention group | Number in control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Spinal fusion | Cecchinato [50] | 29 | Malpositioned implants, Grade B–C (n, [%]) | 29 (9.8) | 41 (16.9) | 0.014* | − 7.1 | − 42.0 | + |
| Malpositioned implants, Grade C (n, [%]) | 7 (2.4) | 22 (9.1) | < 0.001* | − 6.7 | − 73.6 | + | |||
| Dural lesion (n, [%]) | 1 (7.1) | 1 (6.7) | NR | 0.5 | 7.1 | (−) | |||
| Total hip arthroplasty (THA) | Wang [57] | 104 | Divided into four subgroups based Crowe’s classification for hip dysplasia (Crowe I, II, III, IV) | ||||||
| III | DVT | 1 (12.5) | 0 (0) | NR | 12.5 | NaN | (−) | ||
| IV | DVT | 0 (0) | 1 (10) | NR | − 10 | − 100 | (+) | ||
| Transient paralysis of peroneal nerve (n, [%]) | 0 (0) | 1 (10) | NR | − 10 | − 100 | (+) | |||
| Transient paralysis of femoral nerve (n, [%]) | 0 (0) | 1 (10) | NR | − 10 | − 100 | (+) | |||
| Jin [58] | 80 | No complications occurred in both groups | |||||||
| High tibial osteotomy (HTO) | Gao [66] | 39 | Lateral hinge fracture (n, [%]) | 0 (0) | 2 (8.7) | NR | − 8.7 | − 100 | (+) |
| Incisional exudation (n, [%]) | 1 (6.3) | 0 (0) | NR | 6.3 | NaN | (−) | |||
| Total shoulder arthroplasty (TSA) | Boekel [68]† | 47 | Anterior dislocation (n, [%]) | 1 (4.1) | 0 (0) | NR | 4.1 | NaN | (−) |
| Subacromial bursitis (n, [%]) | 0 (0) | 1 (4.3) | NR | − 4.3 | − 100 | (+) | |||
| Sirveaux grade I + II scapular notching (n, [%]) | 2 (8.3) | 3 (13.0) | 0.73 | − 4.7 | − 36.1 | (+) | |||
| Hendel [69] | 31 | Nonoptimal implant type used during surgery (n [%]) | 1 (7) | 10 (63) | < 0.001* | − 56 | − 88.9 | + | |
| Version or inclination malposition occurrences (n [%]) | 4 (13) | 14 (44) | < 0.001* | − 31 | − 221.4 | + | |||
| Malpositioned implants with > 10° deviation in version and/or inclination from the optimal preoperative plan (n [%]) | 4 (27) | 12 (75) | < 0.01* | − 48.0 | − 400 | + | |||
| Transient partial axillary nerve injury (n [%]) | 0 (0) | 1 (5.9) | – | − 5.9 | NaN | (+) | |||
| Accuracy | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Method of measuring accuracy | Outcomes intervention group | Outcomes control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Spinal fusion | Cecchinato [50] | 29 | Implants placed in safe zone, grade 0-A (n [%])‡ | 268 (90.2) | 202 (83.1) | NR | 7.1 | 8.5 | (+) |
| Total hip arthroplasty (THA) | Jin [58] | 80 | Absolute leg length discrepancy (mm) | 1.3 ± 1.0 | 4.7 ± 2.6 | < 0.001* | − 3.5 | − 73.5 | + |
| Leg length discrepancy ≤ 5 mm (n, [%]) | 40 (100) | 28 (70) | < 0.001* | 30 | 42.9 | + | |||
| Absolute femoral stem offset (mm) | 3.0 ± 1.6 | 6.0 ± 2.7 | < 0.001* | − 3.0 | − 49.8 | + | |||
| Absolute femoral stem offset ≤ 5 mm (n, [%]) | 36 (90) | 27 (67.5) | 0.014* | 22.5 | 33.3 | + | |||
| Absolute femoral stem anteversion (°) | 3.5 ± 1.1 | 6.9 ± 2.3 | < 0.001* | − 3.3 | − 48.5 | + | |||
| Absolute femoral stem varus/valgus (°) | 0.8 ± 0.4 | 2.4 ± 1.3 | < 0.001* | − 1.5 | − 64.4 | + | |||
| High tibial osteotomy (HTO) | Gao [66] | 39 | Absolute difference from designed target | ||||||
| Weight-bearing line ratio (AU) | 2.0 ± 1.8 | 5.4 ± 4.4 | 0.002* | − 3.5 | − 63.7 | + | |||
| Hip-knee-ankle angle (°) | 1.1 ± 0.9 | 2.3 ± 2.0 | 0.018* | − 1.2 | − 50.7 | + | |||
| Medial proximal tibial angle (°) | 1.0 ± 0.6 | 1.5 ± 1.0 | 0.068 | − 0.5 | − 31.8 | (+) | |||
| Correction angle (°) | 0.7 ± 0.6 | 0.6 ± 0.4 | 0.912 | 0.0 | 3.1 | (−) | |||
| Posterior tibial slope angle (°) | 1.3 ± 1.3 | 2.0 ± 2.0 | 0.244 | − 0.1 | − 34.0 | (+) | |||
| Total shoulder arthroplasty (TSA) | Boekel [68]† | 47 | Guidewires placed within 2 mm of the planned position in superior/inferior plane (n, [%]) | 22 (91.7) | 14 (14.0) | 0.01* | 30.8 | 50.6 | + |
| Guidewires placed within 2 mm of the planned position the AP plane (n, [%]) | 20 (83.3) | 18 (78.3) | 0.66 | 5.1 | 6.5 | (+) | |||
| Hendel [69] | 31 | Deviation in total offset (mm) | 2.4 ± 1.6 | 3.4 ± 1.8 | 0.11 | − 1.0 | − 29.4 | (+) | |
| Anteroposterior offset (mm) | 1.0 ± 0.9 | 1.9 ± 1.4 | 0.06 | − 0.9 | − 47.4 | (+) | |||
| Medial–lateral offset (mm) | 1.0 ± 0.9 | 1.9 ± 1.0 | 0.012* | − 0.9 | − 47.4 | + | |||
| Superior-inferior offset (mm) | 2.0 ± 1.5 | 2.3 ± 2.1 | 0.64 | − 0.3 | − 13.0 | (+) | |||
| Deviation in version (°) | 4.3 ± 4.5 | 6.9 ± 4.4 | 0.11 | − 2.6 | − 37.7 | (+) | |||
| Deviation in inclination (°) | 2.9 ± 3.4 | 11.6 ± 7.0 | < 0.0001* | − 8.7 | − 75.0 | + | |||
| Deviation in roll (°) | 6.5 ± 5.1 | 10.2 ± 9.7 | 0.13 | − 3.7 | − 36.3 | (+) | |||
| Deviation from plan least retroverted (°) | 7.0 ± 5.4 | 3.2 ± 2.1 | 0.14 | 3.8 | 118 | (−) | |||
| Deviation from plan most retroverted (°) | 1.2 ± 2.0 | 10.0 ± 4.4 | < 0.001* | − 8.8 | − 88.0 | + | |||
| Surgery duration | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Description duration | Value in intervention group | Value in control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Spinal fusion | Cecchinato [50] | 29 | Total surgery duration (min) in mean | 422 | 423 | > 0.05 | − 1 | − 0.2 | (+) |
| Total hip arthroplasty (THA) | Wang [57] | 104 | Divided into four subgroups based Crowe’s classification for hip dysplasia (Crow I, II, III, IV) | ||||||
| I | Total surgery duration (min) | 27.1 ± 3.4 | 24.6 ± 3.4 | 0.053 | 2.5 | 10.3 | (−) | ||
| II | Total surgery duration (min) | 33.0 ± 6.5 | 33.1 ± 4.4 | 0.994 | − 0.1 | − 0.1 | (+) | ||
| III | Total surgery duration (min) | 42.3 ± 4.2 | 50.0 ± 1.5 | 0.001* | − 7.7 | − 15.5 | + | ||
| IV | Total surgery duration (min) | 61.4 ± 14.4 | 70.5 ± 12.1 | 0.151 | − 18.8 | − 12.8 | (+) | ||
| Jin [58] | 80 | Total surgery duration (min) | 78.4 ± 16.4 | 74.2 ± 13.2 | 0.583 | 4.2 | 5.7 | (−) | |
| High tibial osteotomy (HTO) | Gao [66] | 39 | Total surgery duration (min) | 109.4 ± 20.8 | 131.7 ± 29.9 | 0.014* | − 22.3 | − 16.9 | + |
| Total shoulder arthroplasty (TSA) | Boekel [68]† | 47 | Total surgery duration (min) | 78.4 ± 16.3 | 74.8 ± 10.3 | 0.42 | 3.6 | 4.8 | (−) |
| Blood loss | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Description duration | Value in intervention group | Value in control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Total hip arthroplasty (THA) | Wang [57] | 104 | Divided into four subgroups based Crowe’s classification for hip dysplasia (Crow I, II, III, IV) | ||||||
| I | Blood loss (mL) | 333.3 ± 149.4 | 328.6 ± 133.8 | 0.914 | 4.7 | 1.4 | (−) | ||
| II | Blood loss (mL) | 361.5 ± 122.7 | 382.1 ± 156.4 | 0.706 | − 20.6 | − 5.4 | (+) | ||
| III | Blood loss (mL) | 412.5 ± 83.5 | 435.7 ± 102.9 | 0.643 | − 23.2 | − 5.3 | (+) | ||
| IV | Blood loss (mL) | 660.0 ± 206.6 | 875.0 ± 173.6 | 0.022* | − 215 | –2 4.6 | + | ||
| Jin [58] | 80 | Intraoperative blood loss (mL) | 435.8 ± 73.3 | 427.5 ± 69.8 | 0.395 | 8.3 | 1.9 | (−) | |
| Total blood loss (mL) | 711.3 ± 159.5 | 680.6 ± 148.2 | 0.528 | 30.7 | 4.5 | (−) | |||
| High tibial osteotomy (HTO) | Gao [66] | 39 | Intraoperative blood loss (mL) | 50.6 ± 14.4 | 97.8 ± 67.4 | 0.003* | − 47.2 | − 48.2 | + |
| Radiation exposure | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Application | References | N | Description duration | Value in intervention group | Value in control group | P-value | Absolute change (Δ) | Relative change (Δ%) | Positive or negative significant outcome for PSG |
| Spinal fusion | Cecchinato [50] | 29 | C-arm fluoroscopy shots (n) | 11 ± 9.87 | 47.5 ± 15.3 | 0.001* | − 36.5 | − 77 | + |
| DAP (cGycm2) | 133.5 ± 59.6 | 473.3 ± 448.3 | 0.001* | − 339.8 | − 71.8 | + | |||
| Fluoroscopy images (n) | 13.2 ± 8.7 | 32 ± 20.9 | 0.001* | − 18.8 | − 58.7 | + | |||
| Exposure time (s) | 9.4 ± 2.9 | 28.3 ± 27.7 | NR | − 19.0 | − 66.8 | + | |||
| High tibial osteotomy (HTO) | Gao [66] | 39 | Radiation exposure (n) | 18.5 ± 4.8 | 28.22 ± 4.28 | 0.003* | − 9.7 | − 34.4 | + |
Values are given in means with standard deviation unless indicated otherwise. Absolute change (Δ) and relative change (Δ%) are based on mean or median values in the case of continuous variables. In the case of discrete variables, these are based on percentages. If authors used more than one decimal place in their study, values are rounded to one decimal place unless rounding results in zero. To calculate the absolute and relative change the unrounded values are used. “–” for significant negative differences, “ + ” for significant positive differences, “(–)” for insignificant negative differences, “( +)” for insignificant positive differences, and “ = ” for no difference between the two groups. “?” was used when it was unclear whether a result was positive or negative for the use of PSG. ‡This article performed an intention-to-treat and per protocol analysis but in this table we show the results of the intention-to-treat analysis. † In the control group virtual surgical planning was used. ~ Should not be determined because this is a preoperative measurement. * Indicates a statistical significant difference. Abbreviations: American Knee Society Score (AKSS), American Shoulder and Elbow Surgeons (ASES), Anterior Cruciate Ligament (ACL), Arbitraty Unit (AU), Constant Murley Score (CMS), Deep Venous Trombosis (DVT), Dose Area Product (DAP), Harris Hip Score (HHS), Not Applicable (NA), Not Reported (NR), Not a Number (NaN), Oxford Shoulder Score (OSS), Visual Analogue Score (VAS)
Fig. 2.
Risk of Bias assessment using Rob 2 tool. The studies are ordered per intervention type and within an intervention on risk of bias assessment
Results
Search and RCT characteristics
In the initial search, 7574 articles were identified. After removing duplicates, 6310 remained for title and abstract screening (Fig. 1). After screening, 6250 articles were excluded, and two potentially relevant articles were identified from reviews, resulting in 62 eligible studies for full-text screening. Thirty-five of these articles were excluded for the following reasons: twenty-one were not randomized [15–35], six did not use a PSG [36–41], three were non-original articles [42–44], two were not conducted on humans [45, 46], two were not prospective [47, 48], and one did not have a control group [49] (Fig. 1.). So, twenty-seven studies met the inclusion criteria. The references of the included studies revealed no relevant new publications. All included studies and characteristics are presented in Table 1 [50–76]. Seven studies reported on screw positioning for spine surgery [50–56]. Six studies reported on aiding implant positioning in total hip arthroplasty (THA) [57–62]. Three studies investigated guided drilling for anterior cruciate ligament (ACL) reconstruction [63–65]. Two studies focused on the sawing and repositioning in high tibia osteotomy (HTO) [66, 67]. Two studies focused on the positioning of the glenoid component in total shoulder arthroplasty (TSA) [68, 69]. Two studies investigated a puncture assisting device for percutaneous vertebroplasty [70, 71]. One study reported on sawing and repositioning of a distal radius malunion [72]. One study aided in sawing a distal humerus with a cubitus varus deformity [73]. One study assisted in drilling for the correction of lower limb deformities [74]. One study assisted in sawing during a periacetabular osteotomy [75]. Lastly, one study guided screw positioning for femoral neck fracture repair [76].
Fig. 1.
Flowchart of the screening and selection process
Risk of bias assessment
An overview of the risk of bias assessment is shown in Fig. 2. Six (22.2%) of the included articles scored a low risk of bias [50, 57, 58, 66, 68, 69]. Twenty-one (77.8%) of the included studies scored unclear or high risk of bias for at least one domain of the RoB2 tool. Eleven (40.7%) scored unclear [51–54, 59, 63, 64, 67, 70, 72, 73], and ten (37.0%) scored high risk of bias [55, 56, 60–62, 65, 71, 74–76].
Outcome measures
The outcome measures of the studies with low risk of bias are shown in Table 2. A complete overview of the outcome measures of all studies can be found in Supplementary data II.
Patient reported outcome measures (PROMs)
PROMs were assessed in 18 (66.7%) studies, with some studies using multiple outcomes at multiple time points. Therefore, 87 measures were assessed: 27 (31.0%) were significantly positive for the PSG group [52, 57, 58, 66, 67, 71], 41 (47.1%) were insignificantly positive [52–55, 57, 58, 63–68, 72, 73, 75, 76], three (3.4%) showed no difference between groups [58, 71, 72], and 16 (18.4%) were insignificantly negative [57, 58, 65, 68, 70, 72] (Table 2).
Complications
Twenty-four (88.9%) studies reported the occurrence or absence of complications. Six of these reported no complications in either group [54, 58, 63, 65, 73, 75]. The remaining 18 studies reported complications 44 times, of which ten times (22.7%) significant reductions was reported [50, 52, 53, 55, 56, 69, 76], with absolute reductions in complications ranging from 6.7% to 56%. Remarkably, 6 out of these 10 significant reductions were observed in spinal fusion [50, 52, 53, 55]. Twenty-four complications (54.5%) were reported as insignificant reductions [51, 53, 55–57, 60, 66, 68–72, 74], three (6.8%) had the same number of complications in both groups [55, 67] and seven (15.9%) were insignificantly negative [50, 53, 57, 64, 66, 68, 71] (Table 2).
Accuracy
Out of the 27 articles included, twenty (74.1%) reported one or more objective accuracy outcomes, totaling 63 outcomes: 44 (69.8%) outcomes indicated significantly positive results for accuracy when using PSG compared to the control procedure[52, 54–56, 58–64, 66–69, 72, 73, 76], 11 (17.5%) were insignificantly positive[50, 56, 66, 68, 69], one (1.6%) showed no difference [65], four (6.3%) were insignificantly negative[60, 66, 69, 72], and three (4.8%) were unclear[65] (Table 2). In one article, some accuracy outcomes were shown in figures [68]. The authors were contacted but no response was obtained. Therefore, the data was denoted as not reported.
Surgery duration
The surgery duration was reported in 16 (59.3%) studies. One study had four different subgroups [57], resulting in 19 assessments of duration. In eight (42.1%) times, the duration was significantly decreased [52, 57, 59, 66, 70, 71, 74, 75], with relative time savings ranging from 15.4% to 30.2%. In five (26.3%) assessments the duration was insignificantly decreased [50, 54, 56, 57], and in six (31.6%) assessments the duration was insignificantly increased [57, 58, 60, 61, 68, 72] (Table 2).
Blood loss
Nine articles (33.3%) reported on blood loss with 15 different outcomes. Seven (46.7%) showed a significant decrease in blood loss related outcome measures [52, 57, 59, 66, 75] with relative decreases in blood loss ranging from 14.5% to 48.2%. Three measures (20.0%) showed an insignificant decrease [57, 61], and five (33.3%) showed an insignificant increase [54, 57, 58, 60] (Table 2). Eighteen (66.7%) articles did not report blood loss [50, 51, 53, 55, 56, 62–65, 67–74, 76].
Radiation exposure
Radiation exposure was reported in nine (33.3%) studies. In eight (29.6%) studies, it was not applicable [57–62, 68, 69], and in ten (37.0%) studies [53–56, 63–65, 67, 73, 76], it was not reported. In total, 15 outcomes were assessed: 14 (93.3%) measures showed significant decrease [50–52, 66, 70–72, 75] and one (6.7%) showed an insignificant decrease in radiation exposure [74] (Table 2).
Positive significance and risk of bias
The percentage of positive significant outcomes originating from studies with a low risk of bias was 33.3%, 50.0%, 23.8%, 25.0%, 28.6%, and 33.3% for PROMs, complications, accuracy, surgery duration, blood loss, and radiation exposure, respectively.
Discussion
This systematic review aimed to critically appraise, summarize, and compare the literature on RCTs evaluating the clinical added value of PSGs in orthopedic surgery compared to conventional treatments. This review provides evidence that PSGs generally show superior outcomes for accuracy and radiation exposure across various intervention types, and specifically reductions of complications were primarily observed in spinal fusion (Table 2 and Supplementary data II). Other outcome measures, including PROMS, complications in other interventions than spinal fusion, surgery duration and blood loss, may also be superior, though the evidence for these measures is less strong. Furthermore, no significant negative effects for the use of PSGs were reported in any of the RCTs.
Thirty-one percent of the PROMS outcomes (27 out of 87) showed that PSGs were significantly better than conventional treatment [52, 57, 58, 66, 67, 71]. Remarkably, the rate of positive but insignificant results was relatively high at 47.1% (41 out of 87). This may suggest that studies lacked statistical power or that PSGs offer limited added value for certain interventions. Note, that the time intervals of the measurements varied considerably between studies; some studies had solely two intervals [54, 63, 68] (e.g., preoperative and 6 months postoperative [68]), while others included multiple intervals [52, 58, 66] (e.g. preoperative, 1 day, 7 days, 1 month, 3 months [52]). Some studies found a significant difference shortly after the surgery but no significant difference at later follow up measurements [52, 52, 58, 58, 66, 67, 67, 71]. It is debatable whether intervals shortly after surgery (e.g. 1 day, 7 days, 1 month) are clinically relevant, as the ultimate goal is to achieve superior long-term clinical outcomes. Therefore, to optimize the assessments of PROMs in future studies, it is recommended to standardize the intervals for each intervention type to preoperative, 3 months postoperative, 6 months, and 1 year, similar to the included studies on distal radius osteotomies [68] and HTOs [67]. To properly assess long term clinical outcome, 2 and 5 year follow up should be included in follow up studies. Furthermore, there is difference between specific types of PROMs that were used. This is of course depending on the type of surgery but even within the same intervention type different scores were used, such as American Knee Society Score (AKSS), the International Knee Documentation Committee (IKDC) score, and Lysholm Knee Score. This variability makes it difficult to compare studies and highlights the need for standardized PROMs per intervention type. Good practice was seen in studies on hip-related surgeries, as the Harris Hip Score (HHS) was consistently used in all studies reporting PROMs for such surgeries [57, 58, 75, 76].
There is evidence that PSGs can reduce the risk of complications. Specifically, there was a significant difference between the PSG and control groups in 22.7% (10 out of 44) of the complication measures [50, 52, 53, 55, 56, 69, 76], with absolute reductions in complications ranging from 6.7% to 56%. Six out of 10 significant differences were found in spinal fusion surgery [50, 52, 53, 55]. This indicates that for this type of surgery, where the complication rate is relatively high, PSGs can help reduce complications. Similar to the PROMs, the amount of positive but insignificant complications outcomes was substantially higher compared to the negative but insignificant outcomes. Again, we can argue that studies lack the statistical power to detect significant differences, as 63% of the included studies mention sample size as a study limitation [47, 49–51, 54, 55, 57, 60, 61, 63, 65, 66, 71, 73–76].
Accuracy was frequently assessed (20 studies) with results indicating that PSGs can significantly improve accuracy in 69.8% (44 out of 63) of the outcomes across all interventions [52, 54–56, 58–64, 66–69, 72, 73, 76]. Similarly to PROMs, accuracy metrics should be standardized for each intervention type. Ideally, these metrics should be based on absolute 3D positional and angular deviations between preoperative plan and outcomes, typically determined on postoperative CT imaging [58, 66, 69, 77, 78]. Five included studies determined the relative difference instead of absolute, which may overestimate the accuracy of a PSG [55, 56, 60–62]. Therefore, these studies scored high risk of bias in the fourth domain “Measurement of the outcome”. One article compared mean postoperative implant position metrics, such as the femoral anteversion angle, between the PSG group and control group [57]. Because this approach does not overestimate the accuracy of the PSG, this did not negatively affect the risk of bias assessment. However, this irrelevant metric was not included in the review.
The surgery duration was significantly decreased in 42.1% (8 out of 19) of the outcomes [52, 57, 59, 66, 70, 71, 74, 75], with the use of PSGs resulting in relative time savings ranging from 15.4% to 30.2% compared to the control group. Although it is not the most clinically relevant outcome measure, it might be of added value because shorter surgery duration can reduce the risk of infection [79] and may allow for more surgeries to be performed in a day. A decrease in surgery duration may be underestimated due to the learning curve associated with implementing PSGs. On the other hand, for certain applications, surgery duration might increase due to the need for sufficient bone exposure for proper guide placement.
Blood loss was the least reported outcome measure, yet 46.7% (seven out of 15) outcomes showed a significant decrease in blood loss [52, 57, 59, 66, 75]. This indicates that PSGs can reduce blood loss, possibly in close relation with the reduced surgery duration, as some studies who showed significant decrease in blood loss also reported significant decrease in surgery duration [52, 66, 75].
PSGs can clearly reduce the radiation exposure. In the results (Table 2) is shown that nearly all radiation exposure outcomes showed a significant radiation decrease, with only one [74] of the 15 [50–52, 66, 70–72, 75] outcomes showing an insignificant positive effect (Table 2). This reduction is likely due to the decreased need for additional fluoroscopic guidance, as the PSGs provide the necessary guidance, resulting in lower radiation exposure not only for the patient but also for the surgical team. Depending on the type of surgery, a preoperative scan may already be part of standard care. However, if an additional CT scan is required due to the use of a PSG, this could result in higher overall radiation exposure.
Radiation exposure shows the highest positive significance rate of 93.9%, though it was assessed less frequently compared to accuracy. Therefore, we assume that the evidence for added value of PSG is the strongest in terms of accuracy.
Two other reviews regarding 3D technology also found that 3D printed devices (not exclusively PSGs) are clinically effective [7, 80]. However, these reviews did not solely focus on PSGs [7] or orthopedics [80]. In oral and maxillofacial surgery studies have reported on PSGs clinical value in particular, which align with the findings of the present review. In a meta-analysis for jaw reconstructions, PSGs were associated with significant better aesthetic outcomes and reduced surgery duration of 21.2% (95% CI: 10–33%) [6]. These findings are consistent with some of the PROMs and surgery duration outcomes (significant findings ranging from 15.4% to 30.2%) in this review. Furthermore, an RCT that investigated the accuracy for PSGs in dental implantations, found a significantly better accuracy for the PSG group compared to the freehand control group, 3.04° ± 1.5 and 7.03° ± 3.44, respectively [5]. Other outcome measures, such as complications and blood loss are neither frequently investigated, nor show significant differences in oral and maxillofacial surgery [5, 6, 81]. These measures have been better assessed in orthopedic applications as has become apparent in this review. Though, it seems that, when larger studies are conducted with outcome measures standardized per intervention type, superior outcomes will be more consistently detected in orthopedic surgery similar to the studies that were performed in oral and maxillofacial surgery [5, 6, 81].
In Table 2, which contains the study characteristics, it is shown that solely 25.9% of the studies were conducted in the Western world. This is remarkably low considering the high standard of healthcare and research in these regions. This difference could be attributed to the Medical Device Regulation (MDR) legislation in the European Union [82] and the significant amount of time, money, and effort involved in conducting RCTs.
It should be noted that the overall low risk of bias rate was relatively low (22%). However, the percentage of positive significant outcomes originating from studies with low risk of bias were 33.3%, 50.0%, 23.8%, 25.0%, 28.6%, and 33.3% for PROMs, complications, accuracy, surgery duration, blood loss, and radiation exposure, respectively. This means that the low risk of bias studies, i.e., the higher quality studies, reported positive significant outcomes more frequently compared to studies with unclear and high risk of bias. In other words, the studies with unclear or high risk of bias do not seem to report significant findings more frequently.
The strength of this systematic review lies in its comprehensive overview of RCTs investigating the use of PSG in orthopedic surgery for multiple outcome measures. By focusing exclusively on RCTs, it ensures reliable comparisons with control groups. However, the restriction to RCTs can also be seen as a limitation, as it excludes case–control studies, such as for tumor resections [83] or HTOs [84]. In some cases, it might be more ethically appropriate to conduct a case–control study, especially when the new method is already considered superior. Nonetheless, we chose to include only RCTs because randomization, controlled and prospectively gathered data provides higher level of evidence. There may be publication bias, as studies that did not demonstrate added value for PSG are less likely to be published. Observer bias could also affect intraoperative outcomes, such as surgery duration or blood loss, since it is impossible to blind the surgeon. Another limitation is the heterogeneity of the included studies, limiting the feasibility of conducting a meta-analysis. This heterogeneity also affects the outcome measures, making it difficult to compare different guides for the same intervention. Furthermore, the limited number of trials within each subspecialty makes it difficult to draw conclusions within specific subspecialties, however this review gives a good general sense of patient specific guides in orthopedic surgery. Finally, it is important to note that significant differences in outcome measures do not necessarily result in clinically relevant differences. For example, Jin et al., observed a significant decrease in absolute femoral stem varus/valgus deviation from 2.4 ± 1.3° in the control group to 0.8 ± 0.4° (p < 0.001) in the PSG group, representing a mean reduction of 1.6° in varus/valgus deviation (Table 2) [58]. However, this difference may not be clinically relevant for this particular intervention.
The clinical implications based on this review suggest that the use of PSGs should be more widely adopted in orthopedic practice. Furthermore, we would plea for research to standardize outcome measures and follow up intervals to objectively compare conventional treatment with various PSG designs for specific procedures. In the included articles, the costs regarding the use of PSGs were barely mentioned. However, the costs that were mentioned for materials and total expenses (it is unclear what exactly these costs consist of) varied between 30 dollars and 840 euros [53, 58, 61, 63, 64], which may be dependent on local factors such as availability, requirements due to regulations, personnel costs, etc. The cost prevention associated with the use of PSGs, for example through a reduction in complications, was never mentioned. However, we suggest that future studies should also investigate the cost-effectiveness, as the broader implementation of PSGs is likely influenced by their financial viability. Additionally, while this review focused on comparing PSGs with conventional treatments, other technologies with similar goals, such as stereotactic navigation or augmented reality (AR), could also be included in comparisons to determine the most (cost) effective treatment method. For example, one of the included articles compared a PSG, a navigated technique, and the conventional treatment in HTO [67].
Although surgeons can accurately recreate a preoperative 3D plan using PSGs, navigation systems or AR, the optimal physiological, anatomical or mechanical 3D plan often remain unclear. For example the ideal patient-specific position of an implant can be unknown, while this could have effect on patient outcome. Therefore, future research should focus not only on the implementation and evaluation of PSGs, but also on determining the ideal implant positions or bone corrections to achieve best patient outcomes in orthopedic surgery.
Conclusion
This systematic review summarizes the current literature on RCTs, evaluating the clinical added value of PSGs in orthopedic surgery, with exclusion of PSGs used in knee arthroplasty surgery. The findings demonstrate that PSGs often outperform conventional treatment in terms of accuracy and radiation exposure across multiple intervention types. Additionally, PSGs seem to reduce complications, particularly in spinal fusion surgery. Therefore, PSGs offer clinical value and should be more widely adopted in orthopedic practice. For the other outcome measures, including PROMs, complications in other interventions than spinal fusion surgery, surgery duration, and blood loss, there is moderate to limited evidence that PSGs are superior. To draw more definitive conclusions on the added value of PSGs, future research with well-designed RCTs are needed to provide stronger evidence.
Supplementary Information
Below is the link to the electronic supplementary material.
Abbreviations
- AKSS
American knee society score
- ASES
American shoulder and elbow surgeons
- ACL
Anterior cruciate ligament
- AU
Arbitraty unit
- CMS
Constant murley score
- DVT
Deep venous trombosis
- DAP
Dose area product
- HHS
Harris hip score
- HTO
High tibial osteotomy
- MDR
Medical device regulation
- NA
Not applicable
- NR
Not reported
- NaN
Not a number
- OSS
Oxford shoulder score
- PSG
Patient specific guide
- PROMS
Patient reported outcome measures
- RCT
Randomized controlled trial
- VAS
Visual analogue score
- THA
Total hip arthroplasty
- TSA
Total shoulder arthroplasty
Author contribution
All authors contributed to the writing, editing, and proofreading of this manuscript and have approved the final submitted version.
Funding
This research received no external funding, and all authors have no financial disclosures.
Data availability
No datasets were generated or analysed during the current study.
Declarations
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Gauci MO. Patient-specific guides in orthopedic surgery. Orthopaedics and traumatology: surgery and research. Elsevier Masson. 2022. [DOI] [PubMed]
- 2.He Y, Zhou P, He C (2022) Clinical efficacy and safety of surgery combined with 3D printing for tibial plateau fractures: systematic review and meta-analysis. Ann Transl Med 10:403–403 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Yammine K, Karbala J, Maalouf A, Daher J, Assi C. 2022. Clinical outcomes of the use of 3D printing models in fracture management a meta-analysis of randomized studies European Journal of Trauma and Emergency Surgery. Springer. Cham. [DOI] [PubMed]
- 4.Bai J, Wang Y, Zhang P, Liu M, Wang P, Wang J et al (2018) Efficacy and safety of 3D print-assisted surgery for the treatment of pilon fractures: a meta-analysis of randomized controlled trials. J Orthop Surg Res BioMed Central Ltd. 10.1186/s13018-018-0976-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Varga E, Antal M, Major L, Kiscsatári R, Braunitzer G, Piffkó J (2020) Guidance means accuracy: a randomized clinical trial on freehand versus guided dental implantation. Clin Oral Implants Res 31:417–430 [DOI] [PubMed] [Google Scholar]
- 6.Serrano C, van den Brink H, Pineau J, Prognon P, Martelli N (2019) Benefits of 3D printing applications in jaw reconstruction: a systematic review and meta-analysis. J Cranio-Maxillo Surg. 10.1016/j.jcms.2019.06.008 [DOI] [PubMed] [Google Scholar]
- 7.Wong RMY, Wong PY, Liu C, Chung YL, Wong KC, Tso CY et al (2021) 3D printing in orthopaedic surgery: a scoping review of randomized controlled trials Aims. Bone Joint Res 10:807–819 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Hinloopen JH, Puijk R, Nolte PA, Schoones JW, de Ridder R, Pijls BG (2023) The efficacy and safety of patient-specific instrumentation in primary total knee replacement: a systematic review and meta-analysis. Expert Rev Med Devices 20:245–252 [DOI] [PubMed] [Google Scholar]
- 9.Sassoon A, Nam D, Nunley R, Barrack R (2015) Systematic review of patient-specific instrumentation in total knee arthroplasty: new but not improved. Clin Orthop Relat Res 473:151–158 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Rudran B, Magill H, Ponugoti N, Williams A, Ball S (2022) Functional outcomes in patient specific instrumentation vs conventional instrumentation for total knee arthroplasty a systematic review and meta analysis of prospective studies. BMC Musculoskelet Disord. 10.1186/s12891-022-05620- [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Li M, Zeng Y, Wu Y, Liu Y, Wei W, Wu L et al (2020) Patient-specific instrument for unicompartmental knee arthroplasty does not reduce the outliers in alignment or improve postoperative function: a meta-analysis and systematic review. Arch Orthop Trauma Surg 140:1097–1107 [DOI] [PubMed] [Google Scholar]
- 12.Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD et al (2021) The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ Publishing Group, Londan [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ket JCF MMPTRI and VR de. Search blocks—Randomized controlled trials (RCTs). 2013. https://blocks.bmi-online.nl/catalog/296
- 14.Sterne JAC, Savović J, Page MJ, Elbers RG, Blencowe NS, Boutron I et al (2019) RoB 2: a revised tool for assessing risk of bias in randomised trials. The BMJ 10:366 [DOI] [PubMed] [Google Scholar]
- 15.Hananouchi T, Saito M, Koyama T, Sugano N, Yoshikawa H (2010) Tailor-made surgical guide reduces incidence of outliers of cup placement. Clin Orthop Relat Res 468:1088–1095 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Chen B, Zhang Y, Xiao S, Gu P, Lin X (2012) Personalized image-based templates for iliosacral screw insertions: a pilot study. Int J Med Robot Computer Assisted Surg 8:476–482 [DOI] [PubMed] [Google Scholar]
- 17.Hirao M, Oka K, Ikemoto S, Nakao R, Tsuboi H, Nampei A et al (2014) Use of a custom-made surgical guide in total ankle arthroplasty in rheumatoid arthritis cases. Techniques Orthopaed. 10.1097/BTO.0b013e318294cdbb [Google Scholar]
- 18.Schweizer A, Mauler F, Vlachopoulos L, Nagy L, Fürnstahl P (2016) Computer-assisted 3-dimensional reconstructions of scaphoid fractures and nonunions with and without the use of patient-specific guides: early clinical outcomes and postoperative assessments of reconstruction accuracy. J Hand Surg 41:59–69 [DOI] [PubMed] [Google Scholar]
- 19.Spencer-Gardner L, Pierrepont J, Topham M, Baré J, Mcmahon S, Shimmin AJ, et al. Patient-specific instrumentation improves the accuracy of acetabular component placement in total hip arthroplasty. [DOI] [PubMed]
- 20.Guo F, Dai J, Zhang J, Ma Y, Zhu G, Shen J et al (2017) Individualized 3D printing navigation template for pedicle screw fixation in upper cervical spine. PLoS One. 10.1371/journal.pone.0171509 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Hamid KS, Matson AP, Nwachukwu BU, Scott DJ, Mather RC, Deorio JK (2017) Determining the cost-savings threshold and alignment accuracy of patient-specific instrumentation in total ankle replacements. Foot Ankle Int 38:49–57 [DOI] [PubMed] [Google Scholar]
- 22.Aqil A, Patel S, Wiik A, Jones G, Bridle A, Cobb JP (2018) Patient-specific guides improve hip arthroplasty surgical accuracy. Comput Methods Biomech Biomed Engin 21:579–584 [DOI] [PubMed] [Google Scholar]
- 23.Duan X, He P, Fan H, Zhang C, Wang F, Yang L (2018) Application of 3D-printed personalized guide in arthroscopic ankle arthrodesis. Biomed Res Int 2018:1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Sugawara T, Kaneyama S, Higashiyama N, Tamura S, Endo T, Takabatake M et al (2018) Prospective multicenter study of a multistep screw insertion technique using patient-specific screw guide templates for the cervical and thoracic spine. Spine 43:1685–94 [DOI] [PubMed] [Google Scholar]
- 25.Cai X, Xu Y, Yu K, He X, Luo H, Duan J et al (2020) Clinical application of 3-dimensional printed navigation templates in treating femoral head osteonecrosis with pedicled iliac bone graft. Ann Plast Surg 84:S230–S234 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Di Laura A, Henckel J, Hothi H, Hart A (2020) Can 3D surgical planning and patient specific instrumentation reduce hip implant inventory? A prospective study. Print Med. 10.1186/s41205-020-00077-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Mao Y, Xiong Y, Li Q, Chen G, Fu W, Tang X et al (2020) 3D-printed patient-specific instrumentation technique vs conventional technique in medial open wedge high tibial osteotomy: a prospective comparative study. Biomed Res Int. 10:12 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Xing Q et al (2020) A comparative study of patients’ subjective feelings toward total hip arthroplasty with patient-specific instruments and traditional total hip arthroplasty. Orthop Surg. 12:269–76 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Du H, Li S, Tu S, Wang K, Fang Z (2022) The comparison of clinical efficacy of minimally invasive tarsal sinus approach and L-type incision approach combined with 3D printing technology in calcaneal fracture. Comput Math Methods Med. 10.1155/2022/5706341 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Liu F, Tang K, Zheng P (2022) Performance of Tönnis triple osteotomy in older children with developmental dysplasia of the hip (DDH) assisted by a 3D printing navigation template. BMC Musculoskelet Disord. 10.1186/s12891-022-05669-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Yang J, Ni P, Zhang L, Lu Z, Liu D, Mo F et al (2021) Clinical application of a 3D-printed positioning module and navigation template for percutaneous vertebroplasty. Surg Innov. 29:760–8. 10.1177/15533506211062404 [DOI] [PubMed] [Google Scholar]
- 32.Moralidou M, Henckel J, Di Laura A, Hart A (2023) Guiding prosthetic femoral version using 3D-printed patient-specific instrumentation (PSI): a pilot study. Print Med. 10.1186/s41205-023-00168-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Xiong X, Chen YL, Zhao L, Li H, Xu M, Shuang F (2023) Individualized 3D-printed navigation template-assisted tension band wiring for olecranon fractures. J Orthop Surg Res. 10.1186/s13018-023-03892-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Heylen S, Van Haver A, Vuylsteke K, Declercq G, Verborgt O (2016) Patient-specific instrument guidance of glenoid component implantation reduces inclination variability in total and reverse shoulder arthroplasty. J Shoulder Elbow Surg 25:186–192 [DOI] [PubMed] [Google Scholar]
- 35.Jiang L, Dong L, Tan M, Qi Y, Yang F, Yi P et al (2017) A modified personalized image-based drill guide template for atlantoaxial pedicle screw placement: A clinical study. Med Sci Monit 23:1325–1333 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Andresen AK, Wickstrøm LA, Holm RB, Carreon LY, Andersen MØ (2023) Instrumented versus uninstrumented posterolateral fusion for lumbar spondylolisthesis: a randomized controlled trial. JBJS. 10.2106/JBJS.22.00941 [DOI] [PubMed] [Google Scholar]
- 37.Warme WJ, Papadonikolakis A. The Use of a PCL Drill Guide to Increase Safety in the Reconstruction of Unstable Sternoclavicular Joints. 2012. www.shoulderelbowsurgery.com
- 38.Yang L, Zheng Z, Chen W, Wang J, Zhang Y (2015) Femoral neck osteotomy guide for total hip arthroplasty. BMC Surg. 10.1186/s12893-015-0015-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Xu H, Yin W, Hsu P, Qin H, An Z, Zhang C et al (2016) Improve the efficiency of surgery for femoral shaft fractures with a novel instrument: a randomized controlled trial. PLoS One. 10.1371/journal.pone.0154332 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Yoshii Y, Kusakabe T, Akita K, Tung WL, Ishii T (2017) Reproducibility of three dimensional digital preoperative planning for the osteosynthesis of distal radius fractures. J Orthop Res 35:2646–2651 [DOI] [PubMed] [Google Scholar]
- 41.Maini L, Verma T, Sharma A, Sharma A, Mishra A, Jha S (2018) Evaluation of accuracy of virtual surgical planning for patient-specific pre-contoured plate in acetabular fracture fixation. Arch Orthop Trauma Surg 138:495–504 [DOI] [PubMed] [Google Scholar]
- 42.Leong NL, Buijze GA, Fu EC, Stockmans F, Jupiter JB (2010) Computer-assisted versus non-computer-assisted preoperative planning of corrective osteotomy for extra-articular distal radius malunions: a randomized controlled trial. BMC Musculoskelet Disord. 11:110 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Hamilton WG, Parks NL (2014) Patient-specific instrumentation does not shorten surgical time: a prospective, randomized trial. J Arthropl. 10.1016/j.arth.2014.01.029 [DOI] [PubMed] [Google Scholar]
- 44.Iannotti JP, Walker K, Rodriguez E, Patterson TE, Jun BJ, Ricchetti ET (2019) Accuracy of 3-Dimensional planning, implant templating, and patient-specific instrumentation in anatomic total shoulder arthroplasty. J Bone Joint Surg Am Vol 101:446–457 [DOI] [PubMed] [Google Scholar]
- 45.Buller L, Smith T, Bryan J, Klika A, Barsoum W, Iannotti JP (2013) The use of patient-specific instrumentation improves the accuracy of acetabular component placement. J Arthroplasty 28:631–636 [DOI] [PubMed] [Google Scholar]
- 46.Kitada M, Sakai T, Murase T, Hanada T, Nakamura N, Sugano N (2013) Validation of the femoral component placement during hip resurfacing: a comparison between the conventional jig, patient-specific template, and CT-based navigation. Int J Med Robot Computer Assisted Surg 9:223–229 [DOI] [PubMed] [Google Scholar]
- 47.Zhang YW, You MR, Zhang XX, Yu XL, Zhang L, Deng L et al (2021) The novel application of three-dimensional printing assisted patient-specific instrument osteotomy guide in the precise osteotomy of adult talipes equinovarus. Biomed Res Int. 10.1155/2021/1004849 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.He Y, Shang Q, Feng L, Li H, Han L, Zhou D et al (2022) Comparison of the safety and efficacy of three-dimensional guiding templates and free hand technique for cervical pedicle screw fixation: a retrospective study. Surg Innov 29:652–61. 10.1177/15533506221127381 [DOI] [PubMed] [Google Scholar]
- 49.Zhang M, Li J, Fang T, Yan J, Wu L, Zhou Q (2022) Application of 3-dimensional printing guide template and pointed lotus-style regulator in percutaneous pedicle screw fixation for thoracolumbar fractures. Sci Rep. 10.1038/s41598-022-06256-x [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Cecchinato R, Berjano P, Zerbi A, Damilano M, Redaelli A, Lamartina C (2019) Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial. Eur Spine J 28:1712–1723 [DOI] [PubMed] [Google Scholar]
- 51.Chen H, Wu D, Yang H, Guo K (2015) Clinical use of 3D printing guide plate in posterior lumbar pedicle screw fixation. Med Sci Monit 21:3948–3954 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Zhang M, Li J, Fang T, Zhao J, Pan W, Wang X et al (2020) Evaluation of a three-dimensional printed guide and a polyoxymethylene thermoplastic regulator for percutaneous pedicle screw fixation in patients with thoracolumbar fracture. Med Sci Monitor. 10.12659/MSM.920578 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Merc M, Recnik G, Krajnc Z (2017) Lumbar and sacral pedicle screw placement using a template does not improve the midterm pain and disability outcome in comparison with free-hand method. Eur J Orthop Surg Traumatol 27:583–589 [DOI] [PubMed] [Google Scholar]
- 54.Feng S, Lin J (2020) 3-Dimensional printing templates guiding versus free hand technique for cervical lateral mass screw fixation: a prospective study. J Clin Neurosci 78:252–8 [DOI] [PubMed] [Google Scholar]
- 55.Cui L, Gong S, Xie S, Zhang L, Peng W (2021) Efficacy of three-dimensional guide plate technique guided sacral 2 alar iliac screws fixation in patients with degenerative kyphoscoliosis. Am J Transl Res 10:12 [PMC free article] [PubMed] [Google Scholar]
- 56.Merc M, Drstvensek I, Vogrin M, Brajlih T, Recnik G (2013) A multi-level rapid prototyping drill guide template reduces the perforation risk of pedicle screw placement in the lumbar and sacral spine. Arch Orthop Trauma Surg 133:893–899 [DOI] [PubMed] [Google Scholar]
- 57.Wang C, Li Y, Hu Y, Liu H, Wang L, Xie J et al (2021) Patient-specific total hip arthroplasty is superior to conventional methods for Crowe III and IV adult developmental hip dysplasia: a randomized controlled trial. Ann Transl Med 9:212–212 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Jin X, Chen M, Cheema AN, Liu X, Yang S, Xu W (2022) Effectiveness of a patient-specific guide for femoral stem implantation in primary total hip arthroplasty: a randomized control trial. Int Orthop 46:805–814 [DOI] [PubMed] [Google Scholar]
- 59.Zhang YZ, Lu S, Yang Y, Xu YQ, Li YB, Pei GX (2011) Design and primary application of computer-assisted, patient-specific navigational templates in metal-on-metal hip resurfacing arthroplasty. J Arthroplasty 26:1083–1087 [DOI] [PubMed] [Google Scholar]
- 60.Small T, Krebs V, Molloy R, Bryan J, Klika AK, Barsoum WK (2014) Comparison of acetabular shell position using patient specific instruments vs. standard surgical instruments: a randomized clinical trial. J Arthropl 29:1030–7 [DOI] [PubMed] [Google Scholar]
- 61.Zhang YZ, Chen B, Lu S, Yang Y, Zhao JM, Liu R et al (2011) Preliminary application of computer-assisted patient-specific acetabular navigational template for total hip arthroplasty in adult single development dysplasia of the hip. Int J Med Robot Computer Assisted Surg 7:469–474 [DOI] [PubMed] [Google Scholar]
- 62.Zhang Y, Rao Z, Zhang J, Li S, Chang S, Sun Y (2021) 3D printed guides and preoperative planning for uncemented stem anteversion reconstruction during hip arthroplasty: a pilot study. Biomed Res Int. 10.1155/2021/6621882 [DOI] [PMC free article] [PubMed] [Google Scholar] [Retracted]
- 63.Zhu M, Li S, Su Z, Zhou X, Peng P, Li J et al (2018) Tibial tunnel placement in anatomic anterior cruciate ligament reconstruction: a comparison study of outcomes between patient-specific drill template versus conventional arthroscopic techniques. Arch Orthop Trauma Surg 138:515–525 [DOI] [PubMed] [Google Scholar]
- 64.Liu D, Li Y, Li T, Yu Y, Cai G, Yang G et al (2020) The use of a 3D-printed individualized navigation template to assist in the anatomical reconstruction surgery of the anterior cruciate ligament. Ann Transl Med 8:1656–1656 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Lan X, Huang J, Tian F, Wang S, Liu C, Wang J. 3D-Printed Guide for Lateral Femoral Positioning in ACL Reconstruction Application of 3D-Printed Personalized Guide for Lateral Femoral Positioning in Anterior Cruciate Ligament Reconstruction of the Knee ORIGINAL RESEARCH. 360 ALTERNATIVE THERAPIES. 2023. [PubMed]
- 66.Gao F, Yang X, Wang C, Su S, Qi J, Li Z et al (2022) Comparison of clinical and radiological outcomes between calibratable patient-specific instrumentation and conventional operation for medial open-wedge high tibial osteotomy: a randomized controlled trial. Biomed Res Int. 10.1155/2022/1378042 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Zhu X, Qian Y, Liu A, Xu P, Guo JJ (2022) Comparative outcomes of patient-specific instrumentation, the conventional method and navigation assistance in open-wedge high tibial osteotomy: a prospective comparative study with a two-year follow up. Knee 39:18–28 [DOI] [PubMed] [Google Scholar]
- 68.Boekel P, Rikard-Bell M, Grant A, Brandon B, Doma K, O’Callaghan WB et al (2023) Image-derived instrumentation vs. conventional instrumentation with 3D planning for glenoid component placement in reverse total shoulder replacements: a randomized controlled trial. JSES Int. 7:614–22 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Hendel MD, Bryan JA, Barsoum WK, Rodriguez EJ, Brems JJ, Evans PJ et al (2012) Comparison of patient-specific instruments with standard surgical instruments in determining glenoid component position: a randomized prospective clinical trial. J Bone Joint Surg 94:2167–2175 [DOI] [PubMed] [Google Scholar]
- 70.Hu PL, Lin JS, Meng H, Su N, Yang Y, Fei Q (2021) A novel “three-dimensional-printed individual guide template-assisted percutaneous vertebroplasty” for osteoporotic vertebral compression fracture: a prospective, controlled study. J Orthop Surg Res. 10.1186/s13018-021-02471-w [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Chen J, Lin X, Lv Z, Chen M, Huang T (2022) Use of a 3D-printed body surface percutaneous puncture guide plate in vertebroplasty for osteoporotic vertebral compression fractures. PLoS One. 10.1371/journal.pone.0276930 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Buijze GA, Leong NL, Stockmans F, Axelsson P, Moreno R, Sörensen AI et al (2018) Three-dimensional compared with two-dimensional preoperative planning of corrective osteotomy for extra-articular distal radial malunion: a multicenter randomized controlled trial. J Bone Joint Surg Am Vol 100:1191–1202 [DOI] [PubMed] [Google Scholar]
- 73.Hu X, Zhong M, Lou Y, Xu P, Jiang B, Mao F et al (2020) Clinical application of individualized 3D-printed navigation template to children with cubitus varus deformity. J Orthop Surg Res. 10.1186/s13018-020-01615-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Fan M, Wang Y, Pang H, Wang Y, Xu P, Lou Y et al (2022) Application of three-dimensional printed navigation templates to correct lower limb deformities in children by the guided growth technique. World J Pediatr Surg 5:e000349 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Ma S, Xiao L, Guo D, Shi Q, Shen R, Li X (2022) Application of 3D-printed osteotomy guides in periacetabular osteotomy: a short-term clinical study. Int J Artif Organs 45:945–951 [DOI] [PubMed] [Google Scholar]
- 76.Wang S, Wang B, Lan X, Xu Z, Huang H, Wang X et al (2021) Application of a personalized finite element analysis and 3D-printed navigation template in the treatment of femoral neck fracture with cannulated screw. J Med Biol Eng 41:934–941 [Google Scholar]
- 77.Kampkuiper N, Nellensteijn J, Hekman E, Tuijthof G, Lankheet S, Koenrades M et al (2023) Patient-specific 3D virtual surgical planning using simulated fluoroscopic images to improve sacroiliac joint fusion. Biomechanics 3:511–522 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 78.Pijpker PAJ, Kraeima J, Witjes MJH, Oterdoom DLM, Coppes MH, Groen RJM et al (2019) Accuracy assessment of pedicle and lateral mass screw insertion assisted by customized 3D-printed drill guides: a human cadaver study. Operative Neurosurgery 16:94–102 [DOI] [PubMed] [Google Scholar]
- 79.Cheng H, Chen BPH, Soleas IM, Ferko NC, Cameron CG, Hinoul P (2017) Prolonged operative duration increases risk of surgical site infections: a systematic review. Surg Infect. 10.1089/sur.2017.089 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.Diment LE, Thompson MS, Bergmann JHM (2017) Clinical efficacy and effectiveness of 3D printing: a systematic review. BMJ Publishing Group, BMJ Open [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Joda T, Derksen W, Wittneben JG, Kuehl S (2018) Static computer-aided implant surgery (s-CAIS) analysing patient-reported outcome measures (PROMs), economics and surgical complications: a systematic review. Clin Oral Implants Res. 10.1111/clr.13136 [DOI] [PubMed] [Google Scholar]
- 82.Vasiljeva K, van Duren BH, Pandit H (2020) Changing device regulations in the european union: impact on research, innovation and clinical practice. Indian J Orthop 54:123–129 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Evrard R, Schubert T, Paul L, Docquier PL (2019) Resection margins obtained with patient-specific instruments for resecting primary pelvic bone sarcomas: a case-control study. Orthop Traumatol Surg Res 105:781–787 [DOI] [PubMed] [Google Scholar]
- 84.Stimolo D, Leggieri F, Matassi F, Barra A, Civinini R, Innocenti M (2024) Learning curves for high tibial osteotomy using patient-specific instrumentation: a case control study. Innov Surg Sci. 10.1515/iss-2024-0007/html [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
No datasets were generated or analysed during the current study.