Skip to main content
. 2025 Mar 3;15:7473. doi: 10.1038/s41598-025-91751-0

Table 3.

The protocol parameters for T2 FLAIR, MUSE DWI, and PROPELLER DWI sequences.

T2 FLAIR (Sag.) MUSE DWI (Sag.) PROPELLER DWI (Sag.) MUSE DWI (Cor.) PROPELLER DWI (Cor.)
TR 10,000 ms 5000 ms 5000 ms 5000 ms 5000 ms
TE 157 ms 95 ms 81 ms 95 ms 81 ms
TI 2358 ms NA NA NA NA
b value NA 0, 500, 800 0,1000 0, 500, 800 0, 1000
number of shots NA 4 NA 4 NA
NEX 1

2 (b = 0)

3 (b = 500)

6 (b = 800)

1.5 (b = 0)

6 (b = 1000)

2 (b = 0)

3 (b = 500)

6 (b = 800)

1.5 (b = 0)

6 (b = 1000)

FOV 190 × 190 mm 200 × 200 mm 200 × 200 mm 190 × 190 mm 190 × 190 mm
Matrix Size 224 × 224 320 × 320 256 × 256 288 × 288 256 × 256
Slice Thickness 1 mm 3 mm 3 mm 3 mm 3 mm
Scan Time 9:44 10:20 9:50 10:20 9:50

MUSE multiplexed sensitivity encoding, PROPELLER parallel lines with enhanced reconstruction, DWI diffusion weighted image, Sag sagittal, Cor coronal.