TABLE 5.

Reproducibility testing results across three brands of commercial Mueller Hinton agar—Hardy Diagnostics, Becton–Dickinson, and Thermo Fisher Scientific (Remel)

Isolate Susceptibility phenotype Strip cross MIC (μg/mL)
Hardy BD Remel
E. cloacae AR38 ATM resistant, CZA resistant
ATM–CZA combination susceptible
1 1 1
K. pneumoniae AR40 1 1 1
E. coli AR48 4 4 4
M. morganii AR57 0.25 0.25 0.25
K. pneumoniae AR68 2 2 2
K. aerogenes AR161 0.25 0.25 0.25
K. pneumoniae AR453 2 2 2
S. maltophilia 160 4 4 4
E. coli AR1055 ATM resistant, CZA resistant
ATM–CZA combination resistant
≥256 ≥256 ≥256
P. aeruginosa AR235 ≥256 ≥256 ≥256
E. coli AR348 ≥256 ≥256 ≥256
P. aeruginosa AR1112 128 64 64
E. coli AR434 ≥256 ≥256 ≥256
S. maltophilia 159 ATM resistant, CZA susceptible
ATM–CZA combination susceptible
0.5 0.5 0.5
E. cloacae AR02 1 1 1
K. pneumoniae AR03 2 2 1