Table 3. System Suitability Parameters of the Studied HPLC Method.

Item HPLC
Reference values
  GAT BEN
 
    C12 C14  
Tailing factor 0.85 1.1 1.2 T ≤ 1.5
Injection precision 0.24 0.27 0.25 RSD ≤ 2%
Number of theoretical plates (N) 3875 4250 4356 N> 2000
Resolution 3.1 4.5 - Rs > 1.5
Retention time (Rt) 0.2 0.3 0.2 RSD ≤ 2%