. 2025 Mar 13;30(2):90–99. doi: 10.1177/17589983251322537

Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device

Elise Yee Yan Li 1,✉, Luke Steven Robinson 1, Claire Stockdale 1,2
PMCID: PMC11909645  PMID: 40092627

Abstract

Introduction

Evaluation of maximum grip strength (MGS) is commonly conducted to guide intervention planning and evaluate treatment outcomes for individuals with hand or upper limb impairments. While the hydraulic Jamar dynamometer is recognised as the gold standard measurement instrument, it has several limitations and there has not been a consensus on whether alternative instruments can replace or complement it. This study aimed to determine whether a new electronic rehabilitation device, the GripAble, is a reliable and valid instrument for measuring the MGS of healthy adults when compared to the gold standard hydraulic Jamar dynamometer. It also sought to compare the user’s experience when using the two instruments.

Methods

Fifty-five healthy adults completed two MGS assessments ten minutes apart using a standardised protocol. Test-retest and inter-instrument reliability were evaluated using intraclass correlation coefficient. Concurrent validity was established using a Bland-Altman plot. Relationship and level of agreement between measurements on the two devices were evaluated. Participants’ user experience was explored through a quantitative survey.

Results

The GripAble was found to have excellent test-retest reliability (ICC = 0.971-0.975) and good-to-excellent inter-instrument reliability (ICC = 0.898-0.922) with the Jamar dynamometer. Its measurement output is equivalent to approximately 63.3% of that of the Jamar dynamometer. The mean difference value between the measurements on the two devices was 9.44 kg (SD = 4.47). Users preferred the GripAble.

Discussion

While the GripAble is a reliable MGS assessment instrument, it should not be used interchangeably with the Jamar dynamometer for repeated measurements of the same individual owing to the differences in their readings.

Keywords: Hand grip strength, muscle strength dynamometer, reliability, validity

Introduction

Hand and upper limb injuries are common, accounting for 10%-30% of all emergency department presentations worldwide.1–3 These impairments commonly require extensive rehabilitation 4 which can significantly impact on one’s activity participation5,6 and overall well-being. 7 Accurate assessment and treatment of these conditions are crucial for guiding effective clinical interventions, optimising patient outcomes, and reducing the broader burden on the healthcare system and society.5,8

One of the most widely used assessments in hand and upper limb rehabilitation is the evaluation of maximum grip strength (MGS). It provides a simple yet reliable measure of hand and upper limb function, helping clinicians establish a patient’s baseline performance, develop appropriate treatment plans, and monitor recovery progress. 9 The hydraulic Jamar dynamometer (Patterson Medical, Warrenville, Illinois, USA) 10 is recognised as the gold standard MGS measurement instrument. 11 It has been proven reliable and valid and has comprehensive normative data available.12,13 Despite its widespread use in research and clinical practice, the device has several limitations, including the costs associated with regular recalibration, poor sensitivity to low MGS values, and that its analogue display requires manual recording of scores, which can pose risks of interpretation errors.14–16 Additionally, studies investigating participants’ user experience have reported that the Jamar dynamometer is commonly perceived as heavy, rigid, and uncomfortable17,18 and poses difficulties for older adults and individuals with arthritic hands.19,20

While alternative dynamometry devices, including hydraulic dynamometers, pneumatic devices, and work evaluation systems demonstrate a moderate-to-excellent agreement with the gold standard instrument17,21,22; they have limitations regarding user discomfort, costs, lack of portability, and use of a different measurement unit than the Jamar dynamometer.17,20,22 Despite offering additional advantages over these devices,14,23 electronic dynamometers were found to measure dissimilar values to the Jamar dynamometer15,23,24 and therefore cannot be used interchangeably.

Recently, an electronic rehabilitation device, the GripAbleTM (GripAble, London, England, UK), was made commercially available. 25 It incorporates force sensors and connects wirelessly via Bluetooth to its custom app on a Samsung tabletTM, which measures and digitally displays MGS to the nearest 0.1 kg. 25 It has the potential to be used as an alternative MGS measurement instrument that is lightweight and has increased precision and accuracy. 25 While recent studies demonstrate that the GripAble has excellent inter-rater reliability 26 and good-to-excellent inter-instrument reliability with the Jamar PLUS + digital dynamometer 27, there is still a lack of research on this new device. To the best of the authors’ knowledge, no published research has explored its test-retest reliability, inter-instrument reliability, concurrent validity, and user experience when compared to the gold standard hydraulic Jamar dynamometer.

This study aimed to:

  • 1. Establish the test-retest reliability of the GripAble for measuring MGS in healthy adults;

  • 2. Evaluate the inter-instrument reliability between the GripAble and Jamar dynamometer;

  • 3. Examine the concurrent validity between the two devices; and

  • 4. Compare the user’s experience when using the GripAble and Jamar dynamometer.

Methods

Definitions of psychometric properties

The Consensus-based Standards for Selection of Health Measurement Instrument (COSMIN) taxonomy is recognised as the international gold standard for defining the psychometric properties of health-related measurement instruments. 27 Test-retest reliability evaluates whether an instrument can measure consistent results when the test is repeated on the same subject at different time points. 27 Inter-instrument reliability, on the other hand, examines whether different measurement tools can provide consistent scores. 21 Criterion validity is defined by the COSMIN panel as “the degree to which the scores of a measurement instrument are an adequate reflection of a gold standard.” 27(p.743) Concurrent validity is a sub-type of criterion validity in which the target instrument and gold standard device are measured simultaneously. 28

Ethics

This study was reviewed and approved by the Monash University Human Research Ethics Committee on 13th December 2021 (approval reference number: 2022-30799-74585).

Study design

This study employed a psychometric, repeated-measures design, adhering to the Guidelines for Reporting Reliability and Agreement Studies 29 (Supplemental File 1). A quantitative survey was conducted to assess participants’ user experience with the two dynamometry devices through verbal self-report questions.

Participant recruitment

Using convenience sampling, students and academic staff members from Monash University and occupational therapists from Frankston Hospital were recruited. The explanatory statement and consent form were embedded into an online survey created using Qualtrics (a web-based survey development software). A QR code containing a direct link to the survey was disseminated to participants through invitation emails, learning management system announcements, social media posts and printed flyers. Participants were also verbally invited to participate by the first author, a final-year Bachelor of Occupational Therapy student, in various public spaces at the university campus.

Participants who were aged 18 years and over and had the capacity to complete surveys and MGS measurements were eligible to participate. Participants were excluded if they were diagnosed with an upper limb pathology or neurological disorder impacting MGS in either hand, or had experienced pain in their hand, wrist or forearm over the past week that interfered with their ability to complete daily tasks. Participants who met the inclusion criteria and provided their written consent would then schedule a suitable time using a calendar booking system for their MGS assessments. Completion of MGS testing was entirely voluntary, and no incentives were provided.

Data collection instruments

Demographic survey

A researcher-generated, online self-completion demographic survey (Supplemental File 2) was administered using Qualtrics to examine participants’ age, gender, and type of occupation.

Edinburgh Handedness Inventory-Short form (EHI-SF)

The EHI-SF (Supplemental File 3) is a four-item self-rating questionnaire which objectively measures handedness. 30 It has a high internal consistency (Cronbach alpha = 0.93) and good model fit (χ2 = 158.53) with the original 10-item questionnaire and is considered a reliable and valid screening tool for laterality. 31 Participants indicated their preferences for using their hands in four given activities (writing, throwing, using a toothbrush, using a spoon). Participants’ handedness is determined by obtaining the Laterality Quotient value, calculated by averaging the scores for the four items. 30

Dynamometry devices

The GripAble and its accompanying SamsungTM tablet, along with a hydraulic Jamar dynamometer, were used to record participants’ MGS. Both devices were calibrated prior to data collection commencing. Table 1 presents a comparison between the features of the GripAble and Jamar dynamometer.

Table 1.

Properties of the GripAble and hydraulic Jamar dynamometer.

Inline graphic
GripAble
Inline graphic
Hydraulic Jamar dynamometer
Material Plastic Metal and plastic
Mechanism for MGS measurement Through an electronic force measurement system with load cell sensors Through a sealed hydraulic system
Device weight (g) 240 1500
Front-to-back diameter (mm) 48 (isometric mode) 48 (second handle position)
Side-to-side diameter (mm) 40 25
Circumference (mm) 141 (isometric mode) 125 (second handle position)
Modes of measurement Isometric and isotonic Isometric only
Results display • Digital display (through its custom app on SamsungTM tablet)
• Displays MGS in both pounds (lb) and kilograms (kg)
• Analogue display
• Has a peak-hold needle which retains the highest reading until reset
• Displays MGS in both pounds (lb) and kilograms (kg)
Increments of measurement unit (kg) 0.1 (0-90) 2 (0-90)
Data recording Available through its custom app on SamsungTM tablet Not available, manual recording required
Sensitivity Able to detect 62.1±15.3 g changes in MGS Poor sensitivity to low MGS values, requires at least 1.5-2 kg of force to move its pin needle
Special features Has motion sensors that monitor the user’s adherence to the standardised body positioning required for MGS testing -

Note. MGS: maximum grip strength.

User experience survey

Five close-ended, self-report questions relating to the participant’s user experience of the devices (Supplemental File 4) were verbally asked following MGS measurements. The survey examined participant’s perceptions of the device’s weight, perceived comfort level, and preferred instrument for MGS testing through rating scales and multiple-choice questions. Participants were also asked to provide any additional comments regarding their user experience. The survey was pilot-tested with occupational therapy students (n = 5) and academic staff with experience in survey development (n = 3). Changes were made to the survey based on the feedback received to minimise ambiguity of response tasks.

The data collection procedures were pilot-tested with the second author, a senior hand therapist, to ensure the reliability and validity of the results.

Data collection procedures

Following the completion of the demographic survey, the starting order of the instrument and the first hand to be measured for each participant were randomised based on the sequential order in which they were recruited. Four orders of testing were used: (1) starting with the GripAble, using right hand first; (2) starting with the GripAble, using left hand first; (3) starting with the Jamar dynamometer, using right hand first; and (4) starting with the Jamar dynamometer, using left hand first. This followed previous research 32 and removed the potential effects that the starting order can have on MGS measurements. 23

Each participant took part in two MGS testing sessions conducted in a private consultation room ten minutes apart, aligning with similar previous studies.23,33 Participants were positioned according to the standardised protocol for MGS testing recommended by the American Society of Hand Therapists (ASHT) (Supplemental File 5). 11 They were seated in a chair without armrests, with feet fully resting on the floor, and hips as far back in the chair as possible. Hips and knees were positioned at approximately 90°, the testing arm adducted and neutrally rotated, the elbow flexed at 90°, the forearm in a neutral position, and the wrist positioned in 0° to 30° of extension and 0° to 15° of ulnar deviation.

The Jamar dynamometer was used in its second handle position, following the ASHT guidelines. 11 The GripAble was set in its isometric grip test mode (Supplemental File 6). 32 The Jamar dynamometer’s display and the SamsungTM tablet were positioned to face the assessor so that no visual or auditory feedback was provided to the participant. This was done to prevent provoking competitiveness in the participant during testing which could affect the sincerity and fidelity of measurements. 15 The assessor read the standard verbal instructions provided by the GripAble app (Supplemental File 7) to instruct participants to squeeze the device with maximal force for three seconds. Three MGS trials were performed on each hand using each device. The hand was alternated between trials (e.g., GripAble R-L-R-L-R-L, followed by Jamar R-L-R-L-R-L, therefore twelve trials in total). A 30-second rest, timed using a stopwatch, was provided between trials to allow for recovery from fatigue, in line with recommendations from similar previous studies.15,23 A timed two-minute rest was allowed between instruments. 34

After a timed 10-minute rest break, 23 the same assessor completed repeated measures, and the two devices were used in the same order as the first testing session. Following the second measurement of MGS, participants were asked to answer five verbal questions from the researcher and provide comments regarding their user experience of the two devices.

Data analysis

Survey data were extracted from Qualtrics and exported to IBM Statistical Package for the Social Sciences version 28.0 and Microsoft Excel for analysis. Scores on the GripAble (refer to Supplemental File 8) were recorded to the nearest 0.1 kg while measurements on the Jamar dynamometer were recorded to the nearest kg. In all analyses, a p value <.05 was considered statistically significant.

Participants’ demographic characteristics and user experience data were summarised using descriptive statistics. The mean MGS for each dynamometer was calculated by averaging the results of the three trials for each participant, as per the ASHT guidelines. 11 Participants’ comments from the user experience survey were analysed using content analysis to determine the frequency of commonly reported concepts and themes. 35

Reliability was measured by calculating the intraclass correlation coefficient (ICC), reflecting the degree of correlation and agreement between repeated measurements. 36 The two-way, mixed-effects model (i.e., ICC model 3), in which raters are regarded as fixed effects, was used as all participants were assessed by the same rater.37,38 Test-retest reliability was determined by calculating the absolute agreement of repeated measures [ICC (3, 3)], while inter-instrument reliability was determined by analysing the consistency of repeated measures [ICC (3, 3)]. 37 The 95% confidence interval (CI) was calculated for each ICC. An ICC value greater than 0.90 indicates excellent reliability, 0.75-0.90 being good, 0.50-0.75 being moderate, and less than 0.50 being poor. 38

As recommended by the COSMIN guidelines, when the target instrument and gold standard device both measure a continuous variable and are expressed in the same units, Bland-Altman plots should be used to establish their concurrent validity. 28 In a Bland-Altman plot, the differences in measurements between the two devices are plotted against their mean. 39 This was used to provide a graphical visualisation of the level of agreement between the GripAble and Jamar dynamometer. The 95% limits of agreement were calculated as mean difference ±1.96 standard deviation (SD). 40 The plot was visually assessed for the three characteristics indicating good agreement: (1) uniform dispersion of differences in scores between the two instruments; (2) mean difference value is close to zero; and (3) small limits of agreement. 39

A simple linear regression was performed to determine how well the Jamar dynamometer’s measurements could predict that of the GripAble. 38 A zero-intercept linear fit using total least squares (TLS) was used to create a line of best fit which minimises the degree of error in the regression line. 41 This allowed for a more accurate analysis of the proportionality between measurements on the two devices and reduced misinterpretation of data. 41 The Williamson-York Iterative Bivariate Fit Shell method, in which the perpendicular distances between the fitted line and the data points are minimised, 42 was used to complete the TLS fitting of data.

Results

Demographic characteristics

Sixty-three participants consented to participate in the study, with fifty-five participants completing all assessments during the data collection period (31st March 2022 – 1st June 2022). Seven of the participants failed to attend their scheduled assessment appointment despite follow-up email contact, and one participant reported mild pain during MGS assessments leading to the cessation of data collection. This equates to a participant retention rate of 87.3%.

Participants included in the data analysis (N = 55) were prominently female and students (78.2% respectively), with a mean age of 26.2 years (SD = 10.8, 95%CI [5.0-47.4]). As determined by the EHI-SF, most participants were right-handed (n = 49, 89.1%). The demographic characteristics of participants are summarised in Table 2.

Table 2.

Demographic characteristics of participants (N = 55).

n %
Gender Female 43 78.2
Male 12 21.8
Occupation Student 43 78.2
Professional 12 21.8
Handedness (as per EHI-SF) Right-handed 49 89.1
Left-handed 3 5.5
Mixed hander 3 5.5

Test-retest reliability of the GripAble and Jamar dynamometer

The test-retest reliability results of the GripAble and Jamar dynamometer are presented in Table 3. The GripAble demonstrates excellent test-retest reliability for both right (ICC = 0.975, 95% CI [0.957-0.986]) and left hands (ICC = 0.971, 95% CI [0.950-0.983]) across the two data collection time points. Similarly, the Jamar dynamometer demonstrates excellent test-retest reliability for both right (ICC = 0.982, 95% CI [0.970-0.990]) and left hands (ICC = 0.972, 95% CI [0.949-0.984]).

Table 3.

Test-retest reliability and inter-instrument reliability.

Test-retest reliability
Device Time points Hand Data points ICC ICC 95% CI Interpretation ICC Interpretation
ICC 95% CI
GripAble T1, T2 Right 55 0.975 [0.957-0.986] Excellent Excellent
T1, T2 Left 55 0.971 [0.950-0.983] Excellent Excellent
T1, T2 Both 110 0.973 [0.960-0.982] Excellent Excellent
Jamar T1, T2 Right 55 0.982 [0.970-0.990] Excellent Excellent
T1, T2 Left 55 0.972 [0.949-0.984] Excellent Excellent
T1, T2 Both 110 0.977 [0.965-0.985] Excellent Excellent
Inter-instrument reliability
Device Time point Hand Data points ICC ICC 95% CI Interpretation ICC Interpretation
ICC 95% CI
Jamar vs T1 Right 55 0.898 [0.826-0.941] Good Good to excellent
GripAble T1 Left 55 0.922 [0.866-0.954] Excellent Good to excellent
T1 Both 110 0.910 [0.869-0.939] Excellent Good to excellent
T2 Right 55 0.922 [0.867-0.955] Excellent Good to excellent
T2 Left 55 0.917 [0.858-0.952] Excellent Good to excellent
T2 Both 110 0.919 [0.882-0.945] Excellent Good to excellent
T1, T2 Right 110 0.910 [0.868-0.938] Excellent Good to excellent
T1, T2 Left 110 0.919 [0.883-0.945] Excellent Good to excellent
T1, T2 Both 220 0.915 [0.889-0.935] Excellent Good to excellent

Note. CI: confidence interval; ICC: intra-class correlation coefficient; T1: Time 1 (i.e., first MGS testing session); T2: Time 2 (i.e., repeated measurement session).

Inter-instrument reliability between the GripAble and Jamar dynamometer

The inter-instrument reliability results of the GripAble and Jamar dynamometer are presented in Table 3. The findings reveal good-to-excellent inter-instrument reliability between the two devices (ICCs range from 0.898, 95% CI [0.826-0.941] to 0.922, 95% CI [0.867-0.955]). When considering all measurements for both time points (i.e., 220 data points for Left/Right, Time 1/Time 2), the ICC was 0.915 (95% CI [0.889-0.935]) which demonstrates good-to-excellent inter-instrument reliability for all assessments completed.

Concurrent validity between the GripAble and Jamar dynamometer

A statistically significant regression equation was found (F (1, 218) = 1192.8, p < .001) with a high R2 value of 0.845. This suggests that the Jamar dynamometer’s measurements are a strong predictor of the GripAble’s measurements. As shown in Figure 1, participants’ predicted MGS as measured by the GripAble is equal to 1.4 + 0.6 times the value as measured by the Jamar dynamometer. The zero-intercept linear fit using TLS demonstrates that the GripAble’s measurement output is equivalent to approximately 63.3% of that of the Jamar dynamometer.

Figure 1.

Figure 1.

Relationship Between the MGS Measurements Obtained with the GripAble and Jamar Dynamometer. Note. The solid red line and red text indicate simple linear regression, while the dotted black line indicates the zero-intercept linear fit created using TLS. MGS: maximum grip strength; R2: coefficient of determination.

The Bland-Altman plot in Figure 2 demonstrates that all the differences fall above the zero line, indicating that the Jamar dynamometer consistently measured a higher MGS value than the GripAble. The mean difference value between the measurements on the two devices was 9.44 kg (SD = 4.47). The plot reveals a large span of the limits of agreement (0.68 kg-18.2 kg), meaning that measurements between the two instruments can vary up to 18.88 kg. Whilst 95% of the differences were within the limits of agreement, the dispersion of differences was not uniform. Several larger differences in scores (i.e., those approaching the upper limits of agreement) were found in participants with higher mean MGS values. Participants with lower mean MGS values did not show as wide a spread, and their differences in scores were mostly clustered around the mean difference line. Based on the three criteria suggested by Bland and Altman, 39 the findings demonstrate an insufficient level of agreement between the GripAble and Jamar dynamometer.

Figure 2.

Figure 2.

A Bland-Altman Plot Demonstrating the Level of Agreement Between Measurements on the GripAble and Jamar Dynamometer. Note. LoA: limits of agreement.

User experience of the GripAble and Jamar dynamometer

Responses to the user experience survey reveal that the GripAble’s handle was perceived to be more comfortable to hold (mean score = 8.3 [SD = 1.1]) compared to that of the Jamar dynamometer (mean score = 5.5 [SD = 2.0]). Most participants (n = 53, 96.4%) perceived the weight of the GripAble as ‘just right’, while 78.2% (n = 43) perceived the Jamar dynamometer as ‘somewhat heavy to hold’. Most participants (n = 46, 83.6%) preferred using the GripAble for MGS measurement. Thirty-five participants (63.6%) provided additional comments about their user experience with the two devices. Table 4 summarises the five most reported themes for the GripAble and Jamar dynamometer.

Table 4.

User experience themes generated through content analysis (Total n = 35).

Theme Frequency (%)
GripAble The slight retraction of finger plate provides feedback to the exertion of grip force which affirms users that they are able to generate a good grip 10 (28.6%)
The handle is comfortable and easy to grip due to its cylindrical shape 9 (25.7%)
The device is lighter 6 (17.1%)
Users are afraid of breaking the device as it is made up of plastic material 4 (11.4%)
The handle may be too small for someone with bigger hands and longer fingers 3 (8.6%)
Jamar Heavy to hold 11 (31.4%)
Dynamometer Rigidity of the handle causes discomfort in the palm 9 (25.7%)
The device provides no feedback as to whether anything is registering 9 (25.7%)
The rigid metal handle is incentive for users to grip more firmly 3 (8.6%)
The handle feels cold on the hand 2 (5.7%)

Discussion

The findings demonstrate comparable, excellent test-retest reliability (ICC all >0.97) for both the GripAble and Jamar dynamometer and both hands. This indicates that the GripAble is a reliable MGS assessment instrument that can measure consistent values when the test is repeated on the same subject under the same conditions at different time points. 36 The findings in this investigation compare favourably with previous research which reported excellent test-retest reliability (ICC = 0.967-0.996) for other electronic dynamometers, including the Biometrics, 17 DynEx, 26 Grippit, 32 and MyoGrip 18 dynamometers. As the sample size of the present study is comparable to these studies (N = 50-60), the findings are considered akin to previous research.

However, it should be noted that this study completed the test and retest sessions with a shorter interval compared to studies14,24 that completed the measurements several days to one week apart. While this approach helped reduce subject attrition and ensure the stability of the targeted variable (i.e., MGS), 38 it is acknowledged that repeated measurements of MGS are often taken at variable intervals in a clinical setting, such as after a rehabilitation program where weeks or months might have elapsed between test occasions. 43

The findings demonstrate good-to-excellent inter-instrument reliability (ICC = 0.898-0.922) between the GripAble and Jamar dynamometer. This is highly consistent with the values (ICC = 0.898-0.920) reported between GripAble and the Jamar PLUS + digital dynamometer. 32 This suggests that both the GripAble and Jamar dynamometer can record accurate, reliable, and consistent MGS scores across the two measures. Studies have also reported similar levels of inter-instrument reliability (ICC = 0.901-0.94) between the Jamar, GripTrack 34 and K-Force Grip dynamometers. 44

The zero-intercept linear fit using TLS reveals that the GripAble’s measurements are equivalent to approximately 63.3% of that of the Jamar dynamometer. Whilst this percentage is slightly lower than the 69% (95% CI [68-71]%) reported in a previous study, 32 it is acknowledged that their sample had a more equal representation of age groups and genders, compared to the present study. Additionally, Mutalib et al. 32 used the Jamar PLUS + digital dynamometer, while the current study used the hydraulic Jamar dynamometer, which is more prone to manual reading errors. 16 Another previous study also revealed that the hydraulic Jamar dynamometer measures higher values than the Jamar PLUS + by 10%. 45 These factors potentially account for such differences in the findings.

The Bland-Altman plot demonstrates that the Jamar dynamometer consistently measured a higher MGS value than the GripAble, with a mean difference of 9.44 kg and a large span of the limits of agreement between the two devices. This concurs with previous research findings which reported that the Jamar dynamometer consistently yielded higher mean MGS values than other electronic devices, such as the DynEx, 23 MyoGrip, 15 and Bodygrip 46 dynamometers. The mean difference value obtained in this investigation is higher than that previously reported in the comparison of the Jamar with the Grippit 24 and GripTrack 34 dynamometers, which revealed a mean difference of 2.2 and 2.75 kg respectively. Additionally, consistent with the observations by Mutalib et al., 32 differences between measurements on the GripAble and Jamar dynamometer were found to be proportional and increased with higher MGS values. These results suggest that MGS measurements from the two devices should not be interchanged for repeated measurements of the same individual, such as when comparing pre- and post-intervention.

Previous studies have suggested that variations in the measurements between the Jamar dynamometer and electronic devices may be attributed to the differences in their instrument design.14,34 The Jamar dynamometer’s handle, which has a smaller diameter, encourages flexion at the proximal interphalangeal joints. 47 This facilitates the recruitment of the extrinsic finger flexors and grip force generation, 47 which may account for the higher MGS readings obtained with the device. Conversely, the bigger, cylindrical handle of the GripAble encourages flexion at the metacarpophalangeal joints and recruitment of the smaller intrinsic muscles, 32 which have been shown to yield lower MGS. 48

The results concur with previous findings which suggest that the Jamar dynamometer is commonly perceived to be heavy and that its rigid handle can cause discomfort in the palm.17,18,33 This can have implications when used with older adults and patients with arthritic hands.19,20 Findings from the present study demonstrate that the GripAble is commonly perceived as lighter and more comfortable to hold and can potentially be the preferred device by users. It will be of interest for future studies to investigate the GripAble’s clinical utility with older adults and individuals with fragile hand structures and its potential to be used as an alternative measurement tool to minimise discomfort for these populations. However, it is important to note that the device is not interchangeable with the Jamar dynamometer.

Implications for hand and upper limb rehabilitation

Whilst the GripAble was found to be reliable and can potentially be the preferred device by users, the clinical use of this new device should be carefully considered owing to the proportional differences between its readings and measurements on the Jamar dynamometer. Adopting the GripAble as the MGS measurement device in clinical practice requires standardised reporting standards for documentation. As measurements obtained with the Jamar dynamometer are widely used in medico-legal documentation and as supporting evidence for funding applications, a consistent device should be used for repeated measurement of the same patient to reduce ambiguity or misinterpretation of results. As there is still limited research regarding the GripAble, further studies are required to build upon its psychometric evidence. In addition, it is important to consider the different levels of investment required for each of these devices. While the Jamar dynamometer costs less than the GripAble, it requires frequent recalibration to maintain reliability. 16 The GripAble, however, can be calibrated using its custom app at no extra cost. 32

Limitations

This study has several limitations. First, while the sample size (N = 55) is comparable to previous studies,14,34 the use of convenience sampling has resulted in a sample with an over-representation of female university students in their early 20s. This limits the representativeness and generalisability of the findings. 38 Second, despite the provision of rest breaks, the proximity of the two testing sessions could increase participants’ level of fatigue, which may have prevented them from exerting maximal effort during repeated trials. 16 Third, while participants were not provided with any visual or auditory feedback about their MGS results between trials, performance in their subsequent measurements may have been influenced by their motivation or competitiveness to improve. 15 Fourth, as the Jamar dynamometer’s analogue dial is differentiated by 2 kg increments, this can contribute to manual reading errors as the assessor must estimate the MGS score when the pin needle stops outside the marked numbers. 14

Future research directions

Future research using a similar design should be conducted with a larger and more representative sample with a greater diversity of age groups and genders. This will also enable the establishment of a normative dataset for the GripAble, which can be integrated into its software to allow for comparison of a patient’s MGS scores to the general population and assist with determining the presence of impairments. 32 A recent study has established the normative MGS values of the adult population in the United Kingdom using the GripAble. 26 Additionally, as the present study only recruited healthy participants, the findings of this investigation should be further validated with participants with hand injuries or upper limb impairments. This will help determine whether the GripAble can reliably measure minimal clinically significant changes in MGS when evaluating treatment effectiveness in rehabilitation. Future studies should also endeavour to build upon all the published psychometric evidence of the GripAble so that informed decisions can be made about the selection and use of this device in clinical practice.

Conclusion

While the GripAble was found to be a reliable instrument for evaluating the MGS of healthy adults and has the potential to be the preferred device by users, it should not be used interchangeably with the Jamar dynamometer for repeated assessments of the same individual, owing to the proportional differences in their readings.

Supplemental Material

Supplemental Material - Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device

Supplemental Material for Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device by Elise Yee Yan Li, Luke Steven Robinson, and Claire Stockdale in Hand Therapy

Acknowledgements

The authors sincerely thank all the participants for their time and participation in the grip strength assessments.

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

Supplemental Material: Supplemental material for this article is available online.

ORCID iDs

Elise Yee Yan Li https://orcid.org/0009-0008-0794-5759

Luke Steven Robinson https://orcid.org/0000-0003-4449-7357

References

  • 1.de Putter CE, Selles RW, Polinder S, et al. Economic impact of hand and wrist injuries: health-care costs and productivity costs in a population-based study. J Bone Joint Surg 2012; 94: 56. [DOI] [PubMed] [Google Scholar]
  • 2.Rosberg HE, Carlsson K, Cederlund RI, et al. Costs and outcome for serious hand and arm injuries during the first year after trauma – a prospective study. BMC Public Health 2013; 13: 501. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Şahin F, Akca H, Akkaya N, et al. Cost analysis and related factors in patients with traumatic hand injury. J Hand Surg 2012; 38: 673–679. [DOI] [PubMed] [Google Scholar]
  • 4.Robinson LS, O'Brien L. Description and cost-analysis of emergency department attendances for hand and wrist injuries. EMA 2019; 31: 772–779. [DOI] [PubMed] [Google Scholar]
  • 5.Amini D. Hand impairments. In: Dirette DP, Gutman S. (eds). Occupational Therapy and Physical Dysfunction. 8th ed. Philadelphia: Wolters Kluwer, 2021, pp. 895–921. [Google Scholar]
  • 6.Buchanan H, Niekerk L. Work transitions after serious hand injury: current occupational therapy practice in a middle-income country. Aust Occup Ther J 2022; 69: 151–164. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Gustafsson M, Hagberg L, Holmefur M. Ten years follow-up of health and disability in people with acute traumatic hand injury: pain and cold sensitivity are long-standing problems. J Hand Surg 2011; 36: 590–598. [DOI] [PubMed] [Google Scholar]
  • 8.Robinson LS, Sarkies M, Brown T, et al. Direct, indirect and intangible costs of acute hand and wrist injuries: a systematic review. Injury 2016; 47: 2614–2626. [DOI] [PubMed] [Google Scholar]
  • 9.Bobos P, Nazari G, Lu Z, et al. Measurement properties of the hand grip strength assessment: a systematic review with meta-analysis. Arch Phys Med Rehabil 2020; 101: 553–565. [DOI] [PubMed] [Google Scholar]
  • 10.Bechtol CO. Grip test: the use of a dynamometer with adjustable handle spacings. J Bone Joint Surg 1954; 36: 820–832. [PubMed] [Google Scholar]
  • 11.MacDermid J, Solomon G, Valdes K. Clinical Assessment Recommendations: Impairment-Based Conditions. 3rd ed. Mount Laurel. American Society of Hand Therapists, 2015. [Google Scholar]
  • 12.Mathiowetz V, Weber K, Volland G, et al. Reliability and validity of grip and pinch strength evaluations. J Hand Surg 1984; 9: 222–226. [DOI] [PubMed] [Google Scholar]
  • 13.Mathiowetz V, Kashman N, Volland G, et al. Grip and pinch strength: normative data for adults. Arch Phys Med Rehabil 1985; 66: 69–74. [PubMed] [Google Scholar]
  • 14.Allen D, Barnett F. Reliability and validity of an electronic dynamometer for measuring grip strength. Int J Ther Rehabil 2011; 18: 258–265. [Google Scholar]
  • 15.Hogrel JY. Grip strength measured by high precision dynamometry in healthy subjects from 5 to 80 years. BMC Musculoskelet Disord 2015; 16: 139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Roberts HC, Denison HJ, Martin HJ, et al. A review of the measurement of grip strength in clinical and epidemiological studies: towards a standardised approach. Age Ageing 2011; 40: 423–429. [DOI] [PubMed] [Google Scholar]
  • 17.Neumann S, Kwisda S, Krettek C, et al. Comparison of the grip strength using the Martin-Vigorimeter and the JAMAR-Dynamometer: establishment of normal values. In Vivo 2017; 31: 917–924. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Stephens JL, Pratt N, Michlovitz S. The reliability and validity of the Tekdyne hand dynamometer: part II. J Hand Ther 1996; 9: 18–26. [DOI] [PubMed] [Google Scholar]
  • 19.Kennedy D, Jerosch-Herold C, Hickson M. The reliability of one vs. three trials of pain-free grip strength in subjects with rheumatoid arthritis. J Hand Ther 2010; 23: 384–391. [DOI] [PubMed] [Google Scholar]
  • 20.Sipers WM, Verdijk LB, Sipers SJ, et al. The Martin Vigorimeter represents a reliable and more practical tool than the Jamar dynamometer to assess handgrip strength in the geriatric patient. J Am Med Dir Assoc 2016; 17: 466. [DOI] [PubMed] [Google Scholar]
  • 21.Mathiowetz V, Vizenor L, Melander D. Comparison of Baseline instruments to the Jamar dynamometer and the B and L Engineering pinch gauge. OTJR 2000; 20: 147–162. [Google Scholar]
  • 22.Beaton DE, O'Driscoll SW, Richards RR. Grip strength testing using the BTE work simulator and the Jamar dynamometer: a comparative study. J Hand Surg 1995; 20: 293–298. [DOI] [PubMed] [Google Scholar]
  • 23.Shechtman O, Gestewitz L, Kimble C. Reliability and validity of the DynEx dynamometer. J Hand Ther 2005; 18: 339–347. [DOI] [PubMed] [Google Scholar]
  • 24.Massy-Westropp N, Rankin W, Ahern M, et al. Measuring grip strength in normal adults: reference ranges and a comparison of electronic and hydraulic instruments. J Hand Surg 2004; 29: 514–519. [DOI] [PubMed] [Google Scholar]
  • 25.Mace M, Mutalib SA, Ogrinc M, et al. GripAble: an accurate, sensitive and robust digital device for measuring grip strength. J Rehabil Assist Technol Eng 2022; 9: 1–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Mutalib SA, Sharma D, Pike S, et al. GripAble: interrater reliability and normative grip strength of UK population. J Hand Ther 2024. [DOI] [PubMed] [Google Scholar]
  • 27.Mokkink LB, Terwee CB, Patrick DL, et al. The COSMIN study reached international consensus on taxonomy, terminology, and definitions of measurement properties for health-related patient-reported outcomes. J Clin Epidemiol 2010; 63: 737–745. [DOI] [PubMed] [Google Scholar]
  • 28.de Vet HCW, Terwee CB, Mokkink LB, et al. Validity. In: de Vet HCW, Terwee CB, Mokkink LB, et al. (eds). Measurement in Medicine - a Practical Guide. New York: Cambridge University Press, 2011, pp. 150–201. [Google Scholar]
  • 29.Kottner J, Audige L, Brorson S, et al. Guidelines for reporting reliability and agreement studies (GRRAS) were proposed. Int J Nurs Stud 2011; 48: 661–671. [DOI] [PubMed] [Google Scholar]
  • 30.Milenkovic S, Dragovic M. Modification of the Edinburgh handedness inventory: a replication study. Laterality 2013; 18: 340–348. [DOI] [PubMed] [Google Scholar]
  • 31.Veale JF. Edinburgh Handedness Inventory – Short Form: a revised version based on confirmatory factor analysis. Laterality 2014; 19: 164–177. [DOI] [PubMed] [Google Scholar]
  • 32.Mutalib SA, Mace M, Seager C, et al. Modernising grip dynamometry: inter-instrument reliability between GripAble and Jamar. BMC Musculoskelet Disord 2022; 23: 1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Conforto I, Samir C, Chausse F, et al. Comparison of psychometric properties between the Labin, a new electronic dynamometer, and the Jamar: preliminary results in healthy subjects. Hand Surg Rehabil 2019; 38: 293–297. [DOI] [PubMed] [Google Scholar]
  • 34.Svens B, Lee H. Intra- and inter-instrument reliability of grip-strength measurements: GRIPTRACK and JAMAR hand dynamometers. Hand Ther 2005; 10: 47–55. [Google Scholar]
  • 35.Grbich C. Qualitative Data Analysis: an Introduction. London: Sage, 2013. [Google Scholar]
  • 36.Pierson J. Data analysis in quantitative research. In: Liamputtong P. (ed). Research Methods in Health: Foundations for Evidence-Based Practice. 3rd ed. Melbourne: Oxford University Press, 2017, pp. 451–467. [Google Scholar]
  • 37.Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016; 15: 155–163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Portney LG. Foundations of Clinical Research: Applications to Evidence-based Practice. 4th ed. Philadelphia: F. A. Davis Company, 2020. [Google Scholar]
  • 39.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1. [PubMed] [Google Scholar]
  • 40.Giavarina D. Understanding Bland altman analysis. Biochem Med (Zagreb) 2015; 25: 141–151. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Field M. Discovering Statistics Using IBM SPSS Statistics. 4th ed. London: Sage, 2013. [Google Scholar]
  • 42.Cantrell CA. Technical note: review of methods for linear least-squares fitting of data and application to atmospheric chemistry problems. Atmos Chem Phys Discuss 2008; 8: 5477–5487. [Google Scholar]
  • 43.Hamilton A, Balnave R, Adams R. Grip strength testing reliability. J Hand Ther 1994; 7: 163–170. [DOI] [PubMed] [Google Scholar]
  • 44.Nikodelis T, Savvoulidis S, Athanasakis P, et al. Comparative study of validity and reliability of two handgrip dynamometers: K-Force Grip and Jamar. Biomechanics 2021; 1: 73–82. [Google Scholar]
  • 45.King T, II. Interinstrument reliability of the jamar electronic dynamometer and pinch gauge compared with the jamar hydraulic dynamometer and B&L engineering mechanical pinch gauge. Am J Occup Ther 2013; 67: 480–483. [DOI] [PubMed] [Google Scholar]
  • 46.Guerra RS, Amaral TF, Sousa AS, et al. Comparison of Jamar and Bodygrip dynamometers for handgrip strength measurement. J Strength Cond Res 2017; 31: 1931–1940. [DOI] [PubMed] [Google Scholar]
  • 47.Amaral JF, Mancini M, Novo JJM. Comparison of three hand dynamometers in relation to the accuracy and precision of the measurements. Braz J Phys Ther 2012; 16: 216–224. [DOI] [PubMed] [Google Scholar]
  • 48.Edgren CS, Radwin R, Irwin CB. Grip force vectors for varying handle diameters and hand sizes. Hum Factors 2004; 46: 244–251. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplemental Material - Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device

Supplemental Material for Establishing the reliability and validity for GripAble: A handheld sensor-based dynamometry device by Elise Yee Yan Li, Luke Steven Robinson, and Claire Stockdale in Hand Therapy


Articles from Hand Therapy are provided here courtesy of SAGE Publications