TABLE 3.
Organ Motion Evaluation Due to Respiration Before and After Motion Correction for Different Tracer Studies Using Absolute Average Distance Between Liver Dome in Reference Frame and Liver Dome in All Other Frames
| PET/CT system | Tracer | Absolute average distance (mm) | P | |
|---|---|---|---|---|
| Before motion correction | After motion correction | |||
| uEXPLORER | 18F-FDG | 2.7 ± 1.9 (0.0–6.9) | 0 (0) | <0.01 |
| uEXPLORER | 18F-fluciclovine | 4.2 ± 3.9 (0.7–14) | 0.3 ± 0.4 (0.0–1.3) | <0.01 |
| uEXPLORER | 68Ga-DOTATATE | 3.3 ± 1.6 (1.0–7.0) | 0 (0) | <0.01 |
| Biograph Vision 600 | 18F-FDG | 5.3 ± 1.7 (1.6–8.4) | 1.2 ± 0.5 (0.4–2.0) | <0.01 |
| Biograph Vision 600 | 68Ga-PSMA | 4.8 ± 2.1 (0.9–8.3) | 0.6 ± 0.5 (0.0–1.1) | <0.01 |
| Biograph Vision 600 | 11C-PIB | 1.3 ± 1.0 (0.0–2.0) | 0 (0) | <0.1 |
Data in parentheses are range.