How to compare algorithms for automated insulin delivery using different sensors?

Klavs Würgler Hansen 1,
PMCID: PMC11965003  PMID: 39910766

Recent caution has been advised against comparing automatic insulin delivery (AID) systems that use different sensors. 1 , 2 While the same two AID systems (Tandem Control IQ (CoIQ) and Minimed 780 G (MM780G)) have shown clear differences in time in range (TIR) outcomes, studies report no clinically relevant difference in HbA1c 3 , 4 or only minimal changes in HbA1c and TIR. 5 , 6 A simplified overview of these four studies is presented in Table 1. Of note, all TIR values are numerically higher in persons using MM780G (from 2.4% to 7.0% points), but all HbA1c values are minimally lower in persons using CoIQ (about 0.3% points). Tree studies 3 , 4 , 6 provide information on sensor mean glucose or glucose monitoring indicator (GMI) 7 over a 2‐week period and HbA1c, which can be translated into estimated p‐glucose from the ADAG equation (A1c‐derived average glucose). 8 Although this indirect approach contains several pitfalls, the result may indicate differences in sensor performances. Specifically, the Dexcom sensor connected to CoIQ appears to measure glucose levels higher than the Guardian sensor connected to the MM780G. This observation aligns with a recent small‐scale study in which persons used both sensors simultaneously, reporting a slight positive bias for the Dexcom and a negative bias for the Guardian sensor when compared to self‐monitored blood glucose. 9

TABLE 1.

An overview of four studies comparing HbA1c and CGM data follow‐up values for two different AID systems.

CoIQ MM780G p‐value

Henry, 3 Retrospective observational study

Follow‐up time: 90 days

Number of participants 64 167
HbA1c % 6.96 7.05 NS
TIR % (CGM sampling minimum 2 weeks) 67.0 73.77 <0.001
Glucose monitoring indicator (GMI) (%) 7.14 6.89 <0.01
Mean sensor glucose from GMI (mmol/L) 8.9 a 8.3 a <0.01
Mean P‐glucose estimated from HbA1c (mmol/L) 8.5 b 8.6 b NS
Piona, 4 Cross‐sectional study
Number of participants 284 329
HbA1c % 6.7 6.9 <0.001
TIR % (CGM sampling 2 weeks) 68.2 75.2 <0.001
Mean sensor glucose (mmol/L) 8.9 8.2 <0.001
Mean P‐glucose estimated from HbA1c (mmol/L) 8.1 b 8.4 b <0.001

Singh, 5 Retrospective observational study

Follow‐up time: median 1.7 years

Number of participants 92 50
HbA1c % 6.7 c 6.8 c NA
TIR % (CGM sampling up to 4 weeks, minimum 5 days) 69.2 c 74.5 c NA

Beato‐Vibora, 6 Prospective observational study

Follow‐up time: 3 months

Number of participants 75 75
HbA1c % 6.56 6.88 NS
TIR % (CGM sampling 2 weeks) 74.99 77.38 NS
Mean sensor glucose (mmol/L) 8.2 7.9 NS
P‐glucose estimated from HbA1c (mmol/L) 7.8 8.3 NS
a

Calculated from: GMI (%) = 0.02392 × mean sensor glucose (mg/dL) + 3.3.

b

Calculated from Glucose (mmol/L) = 1.59 × HbA1c (%) – 2.59.

c

Calculated from original data of baseline and change. Changes from baseline were compared but not the follow‐up values.

[Correction added on 28 February 2025, after first online publication: The study by Beato‐Vibora6 has been corrected to ‘Prospective observational study’ in this version.]

This discrepancy between sensor‐driven and sensor‐independent glycaemic metrics indicates the influence of varying study populations and sensor performance across brands. However, does this necessarily imply that the insulin delivery algorithm and continuous glucose monitoring (CGM) data hold no significance?

Consider a scenario in which an insulin pump using algorithm 1 with sensor A is compared with an insulin pump using algorithm 2 with sensor B. Suppose users of algorithm 1 achieve a TIR of 68% and a sensor‐based mean glucose of 9.0 mmol/L, while a comparable group of users of algorithm 2 achieves a TIR of 75% and a mean glucose of 8.5 mmol/L, yet with a similar HbA1c. In such cases, TIR and sensor mean glucose metrics may be misleading from a clinical perspective, as sensor A likely measures glucose levels higher than sensor B. If algorithm 1 receives glucose input from sensor B, this is a new product with unknown results requiring new validation.

The goal of developing algorithms for insulin delivery is to enhance safety, minimize the risk of hypoglycaemia, ensure patient adherence and increase TIR, guided by input from the specific sensor. In this context, algorithm 2 is more effective than algorithm 1. A 7 percentage point difference in TIR could translate into a meaningful difference in the risk of diabetic complications. 10 The apparent ‘rescue’ of algorithm 1 due to differences in sensor performance fails to capture the fundamental intellectual and clinical objectives underlying algorithm design.

Sensor performance has traditionally been reported as the mean absolute relative difference (MARD) between a sensor and a reference glucose concentration from a blood sample. There is a growing understanding that a similar and low MARD value does not guarantee the absence of any clinically relevant differences between sensors. 2 , 11 One of the drawbacks of MARD is that this metric does not provide information about the direction of any deviation from the reference value. 12

To reliably evaluate the clinical performance of different AID systems based solely on CGM data, an internationally accepted standardization of glucose sensors is essential, with adherence mandated for manufacturers. 13 , 14 , 15 However, this standardization process is complicated by the lack of a definitive reference value for interstitial glucose and the inherent difficulties in estimating plasma glucose from interstitial glucose under varying physiological conditions. 13

While we fully agree that sensor standardization is essential, and HbA1c is still needed for clinical assessment, we believe that insulin delivery algorithms per se can still be evaluated effectively with sensor‐based data despite differences in sensor performance. In other words, the insulin delivery algorithm, as a data engineering product, can be evaluated in its own right from the sensor data that served as input during its development and optimisation process.

CONFLICT OF INTEREST STATEMENT

KWH has received grants for an investigator‐initiated study from Abbott Diabetes Care and Novo Nordisk.

PEER REVIEW

The peer review history for this article is available at https://www.webofscience.com/api/gateway/wos/peer‐review/10.1111/dom.16234.

Hansen KW. How to compare algorithms for automated insulin delivery using different sensors? Diabetes Obes Metab. 2025;27(5):2319‐2321. doi: 10.1111/dom.16234

DATA AVAILABILITY STATEMENT

Data sharing is not applicable to this article as no new data were created or analyzed in this study.

REFERENCES

  • 1. Messer LH, Welsh JB, Habif S, Pinsker JE, Walker TC. Regarding Singh et al, “effects, safety, and treatment experience of advanced hybrid closed‐loop systems in clinical practice among adults living with type 1 diabetes”. J Diabetes Sci Technol. 2024;18(5):1265‐1266. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Oliver N, Reddy M, Leelarathna L. Continuous glucose sensor accuracy: beyond the headline metric. Lancet Diabetes Endocrinol. 2024;12:934‐938. [DOI] [PubMed] [Google Scholar]
  • 3. Henry Z, Villar Fimbel S, Bendelac N, Perge K, Thivolet C. Beneficial effects of automated insulin delivery over one‐year follow‐up in real life for youths and adults with type 1 diabetes irrespective of patient characteristics. Diabetes Obes Metab. 2024;26(2):557‐566. [DOI] [PubMed] [Google Scholar]
  • 4. Piona C, Passanisi S, Bombaci B, et al. Time in tight range in automated insulin delivery system users: real‐world data from children and adolescents with type 1 diabetes. Diabetes Obes Metab. 2024;26(10):4767‐4771. [DOI] [PubMed] [Google Scholar]
  • 5. Singh R, Imberg H, Ahmadi SS, et al. Effects, safety, and treatment experience of advanced hybrid closed‐loop systems in clinical practice among adults living with type 1 diabetes. J Diabetes Sci Technol. 2024;19322968241242386. 10.1177/19322968241242386 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Beato‐Vibora PI, Chico A, Moreno‐Fernandez J, et al. A multicenter prospective evaluation of the benefits of two advanced hybrid closed‐loop systems in glucose control and patient‐reported outcomes in a real‐world setting. Diabetes Care. 2024;47(2):216‐224. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Bergenstal RM, Beck RW, Close KL, et al. Glucose management indicator (GMI): a new term for estimating A1C from continuous glucose monitoring. Diabetes Care. 2018;41(11):2275‐2280. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Nathan DM, Kuenen J, Borg R, et al. Translating the A1C assay into estimated average glucose values. Diabetes Care. 2008;31(8):1473‐1478. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Lundemose SB, Laugesen C, Ranjan AG, Norgaard K. Factory‐calibrated continuous glucose monitoring systems in type 1 diabetes: accuracy during in‐clinic exercise and home use. Sensors (Basel). 2023;23(22):9256. 10.3390/s23229256 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Shah VN, Kanapka LG, Akturk HK, et al. Time in range is associated with incident diabetic retinopathy in adults with type 1 diabetes: a longitudinal study. Diabetes Technol Ther. 2024;26(4):246‐251. [DOI] [PubMed] [Google Scholar]
  • 11. Vigersky RA, Shin J. The myth of MARD (mean absolute relative difference): limitations of MARD in the clinical assessment of continuous glucose monitoring data. Diabetes Technol Ther. 2024;26(S3):38‐44. [DOI] [PubMed] [Google Scholar]
  • 12. Heinemann L, Schoemaker M, Schmelzeisen‐Redecker G, et al. Benefits and limitations of MARD as a performance parameter for continuous glucose monitoring in the interstitial space. J Diabetes Sci Technol. 2020;14(1):135‐150. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Freckmann G, Nichols JH, Hinzmann R, et al. Standardization process of continuous glucose monitoring: traceability and performance. Clin Chim Acta. 2021;515:5‐12. [DOI] [PubMed] [Google Scholar]
  • 14. Freckmann G, Eichenlaub M, Waldenmaier D, et al. Clinical performance evaluation of continuous glucose monitoring systems: a scoping review and recommendations for reporting. J Diabetes Sci Technol. 2023;17(6):1506‐1526. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Eichenlaub M, Pleus S, Rothenbuhler M, et al. Comparator data characteristics and testing procedures for the clinical performance evaluation of continuous glucose monitoring systems. Diabetes Technol Ther. 2024;26(4):263‐275. [DOI] [PMC free article] [PubMed] [Google Scholar]