Skip to main content
. 2024 Dec 1;26(4):e14577. doi: 10.1002/acm2.14577

TABLE 1.

Percentage of fractions with 3D repositioning shifts less than or equal to clinically relevant thresholds.

Type‐S, % of 3D shifts BPL1, % of 3D shifts
Region (# Fx) ≤1.5 mm ≤3.0 mm ≤5.0 mm Region (# Fx) ≤1.5 mm ≤3.0 mm ≤5.0 mm
All (N = 60) 50.0 91.7 100.0 All (N = 112) 46.4 74.1 97.3
Cervical (N = 16) 68.8 100.0 100.0 Cervical (N = 0)
Thoracic, T1‐T4 (N = 44) 43.2 88.6 100.0 Thoracic, T5‐T12 (N = 69) 46.5 74.4 95.3
Lumbar (N = 0) Lumbar (N = 43) 46.4 73.9 98.6

Abbreviation: BPL1, Body Pro‐Lok ONE setup.