Skip to main content
. 2025 Mar 19;26(4):e70051. doi: 10.1002/acm2.70051

TABLE 2.

Dosimeter pros and cons.

Detector Uses a Pros Cons Resource(s)
Gafchromic EBT‐XD Film TAD BC

Widely utilized

DR independent

Better dynamic range and lower uncertainty than EBT3

Calibration in CONV

Can be cut and shaped to desired size

Large uncertainty (up to 4%)

Passive and requires read out time. Usually, 24 hrs. to a week

Many calibration measurements required with known dose delivered

Orientation effects

Oh et al. 30 Cetnar et al. 37

No et al. 34

Petusseau et al. 38

Rahman et al. 31

Sloop et al. 17

Jorge et al. 28

Konradsson et al. 26

Garty et al. 39

Adrian et al. 40

Deut et al. 41

Konradsson et al. 42

Kyle et al. 35

Lempart et al. 25

Levy et al. 43

Szpala et al. 44

Xie et al. 45

Gafchromic EBT3 Film TAD BC

Widely utilized

DR independent

Calibration in CONV

Can be cut and shaped to desired size

Large uncertainty (up to 4%) Passive and requires read out time. Usually, 24 hrs. to a week Many calibration measurements required with known dose delivered Worse dynamic range and higher uncertainty than EBT‐XD Orientation effects
OC‐1 OrthoChromic Film TAD BC

Higher saturation level to dose than EBT3

Better high dose rate performance than EBT3 and EBT‐XD

Utilize for high dose measurements 

Requires reflection mode scanning Garty et al 39
microDiamond PTW TAD BC

Linear response up to 1.5 Gy/pulse

Could be used as primary detector for beam characterization (SN 122325)

PDDs and profiles agreement with EBT3 Gafchromic film

Could be used to measure output factors and output vs gantry angle 

Above 1.5 Gy/pulse is uncertain Required cross calibration There is another more optimized model from PTW flashDiamond  Dal Bello et al 23
Edge Diode SNC 1118 (N‐Type)

TAD

BC

DPP

Linear response

DR, energy, and DPP independent

Good temporal resolution (0.1 ms)

Comparable to IC in terms of precision

No recombination correction

Agreement with EBT‐XD Gafchromic film to within 3%

Energy independent within 4% of film

Good SR

Dose conversion given

Agreement with W1 scintillator within ‐7 to 5%

Tested during ramp up as well (various DPP) and had 4.5% error bars (only due to threshold setting with device)

Independent of repetition rate 

Film has a higher SR

Lost 0.4% per kGy of sensitivity

Needs further characterization

Electrometer requires customization (simple and provided by manufacturer though)

Not tested above 0.78 Gy/pulse and 180 Gy/s

USB not fast enough to transfer data in real time

Rahman et al 24

Sloop et al 17

EDD 2–3G Diode IBA PC

Agreement with an external monitor chamber (monitor chamber was able to measure DPP with large uncertainty after corrections applied)

Was able to control delivery by # pulses and was accurate every time (> 1000 trials)

Required construction of a MCU to actually count

No information on DPP

Lempart et al. 25

Konradsson et al 26

EDP 20–3G Diode IBA TAD

Can be used to tune a LINAC to FLASH

Could be calibrated to measure output in FLASH

Linear with dose

Linear with DPP compared with film

Study did not mention the calibration procedure Lempart et al. 25
Ion Chambers TAD BC

Can provide accurate data for profiles in UHDR

Can be used to approximate DR

Logistic model can accurately track saturation

Saturation and recombination effects at UHDR outside of clinical tolerance

Requires many verification tests alongside other detectors first

Requires models to correct for recombination

Saturation at “low” DPP

Profilers need to be used at a low voltage and far away from the source to compensate

Kyle et al. 35

Oh et al. 36

Poirier et al. 21

No et al 34

W1 Scintillator Exradin

TAD

DPP

Agreement with film measurements

Independent of DR up to 360 Gy/s, DPP up to 1.1 Gy/pulse, and FS

Microsecond temporal resolution

Major damage issues of 16.2% loss / kGy (declines after accumulation of more dose)

Recalibration required almost every use

Must account for Cherenkov production

Tedious calibration (10 measurements, 2 factors)

Ashraf et al 29
W2 Scintillator Exradin with MAX SD TAD

Agreement with 2% of film

MAX SD corrects for Cherenkov itself

Smaller model linear with dose up to the max settings of the LINAC

No pulse repetition frequency non‐linearity

MAX SD required modifications by vendor (otherwise saturates)

Large volume model saturates at “low” DPP

Optical fiber damages 2% per kGy

Oh et al. 30
TLDs TAD

Close agreement with EBT3 film

Calibration can be traced to primary standard

Reproducible response with different total dose at high DPP

DR independent (up to 3 × 10^5 Gy/s)

Requires correction factors for electrons

Farther off from dose than film and alanine in UHDR

For short SSD, may require energy correction

Some sensitization effects 

Motta et al 27

Jorge et al 28

OSLDs TAD

Reproducible response with different total dose at high DPP

DR independent (up to 3 × 10^5 Gy/s)

Al2O3:C was most precise—agreement with film within 3%

There was a BeO OSLD with a higher sensitivity that was affecting the DR plots (package to package variability)

Conflicting results with Al2O3:C across studies

Some sensitization effects 

Motta et al 27
Alanine Pellets  TAD

Close to the expected dose (more than TLD and film in the study) in UHDR

Agreement to film within 2%

Reproducible 

Farther off in dose from film and TLD

Study had no absolute reference, results may need further testing

Passive, requires EPR

Jorge et al 28
Quad Channel CXP‐12 GigaSens (Concurrent EDA) CMOS Camera

BC

RPP

Can gate to LINAC signal and match repetition rate with capture

Excellent SR

Temporal resolution to capture each pulse

Used in vivo

Can maybe obtain estimates of dose, DR, and DPP

Picks up respiratory motion effects on DPP when utilized in vivo

Worked up to 360 Hz (FPS was matched to 360)

Can provide spatial dose distribution info on a pulse‐by‐pulse basis, not just total dose

Low SNR without intensifier

Intensifier reduces SR

Optical scattering causes measured profile to extend out of the treatment field

Requires dark and flat field corrections

Requires median filtering

Did not convert pixel density to dose

No information on DPP

Rahman et al 33  
DoseOptics iCMOS Camera

TAD

BC

Linear response

Dose rate independent (tested up to 400 Gy/s)

DPP independent (tested up to 0.91 Gy/pulse)

Scintillation imaging was better than Cherenkov, especially with bandpass filtering

Good SR

Can reconstruct or directly measure profiles/PDDs per pulse

Can achieve high passing rates for certain profiles against film

High spatial resolution (1 mm^3)

Cherenkov did not work for PDDs/ profiling due to lateral scatter

Technique needs further improvement based on optical alignment, better viewing, speed

Must filter Cherenkov signal from Scintillation signal due to energy dependance

cintillation requires a quinine sulfate solution

Disagreement with films on PDD buildup and some profiles

Applicator leakage imaging to remove

Lots of processing steps

Ashraf et al. 32

Rahman et al 31

HC‐120 PMT with OF read with 3000 series Picoscope PC

DR independent

Linear with dose (up to 200 Gy tested) and DR (up to 300 Gy/s tested) (360 Hz)

Could be used in place of target signal that tunes the beam

No information on the absolute dose in a pulse

Was placed at the edge of the field to just monitor the output pulse by pulse

Read out system limits temporal resolution (PMT alone has nan‐second resolution)

Rahman et al 31  
Hyperscint RP100 scintillator system TAD

Comes with integrated software system that tells exactly how to calibrate and upload measurements for calibration, and auto‐removes Cherenkov/background

Software has a built‐in pulse counter

Linear with dose up to several thousand Gy

 < 0.4% signal loss per kGy

Agreement to films within 4%

Not all photodetectors are read out simultaneously causes under recording especially if using for dose per pulse (counting)

Dose calibration must be performed everyday

Temporal resolution is close to that of a 180 Hz beam (2.5 ms)

Poirier et al. 20
Remote Trigger Unit DoseOptics PC Pulses measured matches expected for low frequencies 

Positional dependance with respect to the field

PC has some errors in higher repetition rates (maybe due to lag in gating—see control system section)

Decay time causes issues

Ashraf et al. 29

Rahman et al. 16

a

Uses of detectors are one of the following options: Total absorbed dose (TAD), beam characteristics (BC), dose per pulse (DPP), reading per pulse (RPP, same as DPP but not calibrated), or pulse counting (PC).