TABLE 2.
Dosimeter pros and cons.
| Detector | Uses a | Pros | Cons | Resource(s) |
|---|---|---|---|---|
| Gafchromic EBT‐XD Film | TAD BC |
Widely utilized DR independent Better dynamic range and lower uncertainty than EBT3 Calibration in CONV Can be cut and shaped to desired size |
Large uncertainty (up to 4%) Passive and requires read out time. Usually, 24 hrs. to a week Many calibration measurements required with known dose delivered Orientation effects |
No et al. 34 Petusseau et al. 38 Rahman et al. 31 Sloop et al. 17 Jorge et al. 28 Konradsson et al. 26 Garty et al. 39 Adrian et al. 40 Deut et al. 41 Konradsson et al. 42 Kyle et al. 35 Lempart et al. 25 Levy et al. 43 Szpala et al. 44 Xie et al. 45 |
| Gafchromic EBT3 Film | TAD BC |
Widely utilized DR independent Calibration in CONV Can be cut and shaped to desired size |
Large uncertainty (up to 4%) Passive and requires read out time. Usually, 24 hrs. to a week Many calibration measurements required with known dose delivered Worse dynamic range and higher uncertainty than EBT‐XD Orientation effects | |
| OC‐1 OrthoChromic Film | TAD BC |
Higher saturation level to dose than EBT3 Better high dose rate performance than EBT3 and EBT‐XD Utilize for high dose measurements |
Requires reflection mode scanning | Garty et al 39 |
| microDiamond PTW | TAD BC |
Linear response up to 1.5 Gy/pulse Could be used as primary detector for beam characterization (SN 122325) PDDs and profiles agreement with EBT3 Gafchromic film Could be used to measure output factors and output vs gantry angle |
Above 1.5 Gy/pulse is uncertain Required cross calibration There is another more optimized model from PTW flashDiamond | Dal Bello et al 23 |
| Edge Diode SNC 1118 (N‐Type) |
TAD BC DPP |
Linear response DR, energy, and DPP independent Good temporal resolution (0.1 ms) Comparable to IC in terms of precision No recombination correction Agreement with EBT‐XD Gafchromic film to within 3% Energy independent within 4% of film Good SR Dose conversion given Agreement with W1 scintillator within ‐7 to 5% Tested during ramp up as well (various DPP) and had 4.5% error bars (only due to threshold setting with device) Independent of repetition rate |
Film has a higher SR Lost 0.4% per kGy of sensitivity Needs further characterization Electrometer requires customization (simple and provided by manufacturer though) Not tested above 0.78 Gy/pulse and 180 Gy/s USB not fast enough to transfer data in real time |
Rahman et al 24 Sloop et al 17 |
| EDD 2–3G Diode IBA | PC |
Agreement with an external monitor chamber (monitor chamber was able to measure DPP with large uncertainty after corrections applied) Was able to control delivery by # pulses and was accurate every time (> 1000 trials) |
Required construction of a MCU to actually count No information on DPP |
Lempart et al. 25 Konradsson et al 26 |
| EDP 20–3G Diode IBA | TAD |
Can be used to tune a LINAC to FLASH Could be calibrated to measure output in FLASH Linear with dose Linear with DPP compared with film |
Study did not mention the calibration procedure | Lempart et al. 25 |
| Ion Chambers | TAD BC |
Can provide accurate data for profiles in UHDR Can be used to approximate DR Logistic model can accurately track saturation |
Saturation and recombination effects at UHDR outside of clinical tolerance Requires many verification tests alongside other detectors first Requires models to correct for recombination Saturation at “low” DPP Profilers need to be used at a low voltage and far away from the source to compensate |
Kyle et al. 35 Oh et al. 36 Poirier et al. 21 No et al 34 |
| W1 Scintillator Exradin |
TAD DPP |
Agreement with film measurements Independent of DR up to 360 Gy/s, DPP up to 1.1 Gy/pulse, and FS Microsecond temporal resolution |
Major damage issues of 16.2% loss / kGy (declines after accumulation of more dose) Recalibration required almost every use Must account for Cherenkov production Tedious calibration (10 measurements, 2 factors) |
Ashraf et al 29 |
| W2 Scintillator Exradin with MAX SD | TAD |
Agreement with 2% of film MAX SD corrects for Cherenkov itself Smaller model linear with dose up to the max settings of the LINAC No pulse repetition frequency non‐linearity |
MAX SD required modifications by vendor (otherwise saturates) Large volume model saturates at “low” DPP Optical fiber damages 2% per kGy |
Oh et al. 30 |
| TLDs | TAD |
Close agreement with EBT3 film Calibration can be traced to primary standard Reproducible response with different total dose at high DPP DR independent (up to 3 × 10^5 Gy/s) |
Requires correction factors for electrons Farther off from dose than film and alanine in UHDR For short SSD, may require energy correction Some sensitization effects |
Motta et al 27 Jorge et al 28 |
| OSLDs | TAD |
Reproducible response with different total dose at high DPP DR independent (up to 3 × 10^5 Gy/s) Al2O3:C was most precise—agreement with film within 3% |
There was a BeO OSLD with a higher sensitivity that was affecting the DR plots (package to package variability) Conflicting results with Al2O3:C across studies Some sensitization effects |
Motta et al 27 |
| Alanine Pellets | TAD |
Close to the expected dose (more than TLD and film in the study) in UHDR Agreement to film within 2% Reproducible |
Farther off in dose from film and TLD Study had no absolute reference, results may need further testing Passive, requires EPR |
Jorge et al 28 |
| Quad Channel CXP‐12 GigaSens (Concurrent EDA) CMOS Camera |
BC RPP |
Can gate to LINAC signal and match repetition rate with capture Excellent SR Temporal resolution to capture each pulse Used in vivo Can maybe obtain estimates of dose, DR, and DPP Picks up respiratory motion effects on DPP when utilized in vivo Worked up to 360 Hz (FPS was matched to 360) Can provide spatial dose distribution info on a pulse‐by‐pulse basis, not just total dose |
Low SNR without intensifier Intensifier reduces SR Optical scattering causes measured profile to extend out of the treatment field Requires dark and flat field corrections Requires median filtering Did not convert pixel density to dose No information on DPP |
Rahman et al 33 |
| DoseOptics iCMOS Camera |
TAD BC |
Linear response Dose rate independent (tested up to 400 Gy/s) DPP independent (tested up to 0.91 Gy/pulse) Scintillation imaging was better than Cherenkov, especially with bandpass filtering Good SR Can reconstruct or directly measure profiles/PDDs per pulse Can achieve high passing rates for certain profiles against film High spatial resolution (1 mm^3) |
Cherenkov did not work for PDDs/ profiling due to lateral scatter Technique needs further improvement based on optical alignment, better viewing, speed Must filter Cherenkov signal from Scintillation signal due to energy dependance cintillation requires a quinine sulfate solution Disagreement with films on PDD buildup and some profiles Applicator leakage imaging to remove Lots of processing steps |
Ashraf et al. 32 Rahman et al 31 |
| HC‐120 PMT with OF read with 3000 series Picoscope | PC |
DR independent Linear with dose (up to 200 Gy tested) and DR (up to 300 Gy/s tested) (360 Hz) Could be used in place of target signal that tunes the beam |
No information on the absolute dose in a pulse Was placed at the edge of the field to just monitor the output pulse by pulse Read out system limits temporal resolution (PMT alone has nan‐second resolution) |
Rahman et al 31 |
| Hyperscint RP100 scintillator system | TAD |
Comes with integrated software system that tells exactly how to calibrate and upload measurements for calibration, and auto‐removes Cherenkov/background Software has a built‐in pulse counter Linear with dose up to several thousand Gy < 0.4% signal loss per kGy Agreement to films within 4% |
Not all photodetectors are read out simultaneously causes under recording especially if using for dose per pulse (counting) Dose calibration must be performed everyday Temporal resolution is close to that of a 180 Hz beam (2.5 ms) |
Poirier et al. 20 |
| Remote Trigger Unit DoseOptics | PC | Pulses measured matches expected for low frequencies |
Positional dependance with respect to the field PC has some errors in higher repetition rates (maybe due to lag in gating—see control system section) Decay time causes issues |
Ashraf et al. 29 Rahman et al. 16 |
Uses of detectors are one of the following options: Total absorbed dose (TAD), beam characteristics (BC), dose per pulse (DPP), reading per pulse (RPP, same as DPP but not calibrated), or pulse counting (PC).