. 2025 Mar 19;87(4):2307–2323. doi: 10.1097/MS9.0000000000003155

Comparison of diagnostic accuracy of electrocardiogram-based versus photoplethysmography-based smartwatches for atrial fibrillation detection

A Systematic Review and Meta-Analysis

Dan Yang Luo a, Zhi Wei Zhang b, Olivier Sibomana c,*, Salomon Izere c
PMCID: PMC11981249  PMID: 40212135

Abstract

Background:

Atrial fibrillation (AF), the most prevalent cardiac arrhythmia, significantly affects morbidity and mortality, making early detection crucial for preventing stroke and heart failure. Recent advancements in wearable technology have introduced smartwatches as potential tools for continuous non-invasive AF detection.

Objective:

This systematic review and meta-analysis aimed to evaluate and compare the diagnostic accuracy of electrocardiography (ECG)-and photoplethysmography (PPG)-based smartwatches in detecting AF.

Methodology:

A comprehensive search was conducted on PubMed, Google Scholar, and other databases from 18 August to 23 September 2024, to fetch original studies that evaluated performance metrics of ECG and PPG smartwatches in AF detection. The obtained literature was screened according to preset inclusion and exclusion criteria. For included studies, the random-effects model was used to calculate their pooled sensitivity and specificity in AF detection using Jamovi 2.3.28 software. A significance threshold of P <0.05 was applied to all statistical analyses.

Results:

Out of the 2564 studies screened, 25 met the inclusion criteria: 11 on PPG and 14 on ECG smartwatches. PPG smartwatches exhibited higher diagnostic performance with a pooled sensitivity of 97.4% (95% CI: 96.5–98.3) and specificity of 96.6% (95% CI: 94.9–98.3). Conversely, ECG smartwatches showed a pooled sensitivity of 83% (95% CI: 78–88) and specificity of 88.4% (95% CI: 84.5–92.2), lower than PPG smartwatches.

Conclusion:

PPG-based smartwatches outperformed ECG-based devices in AF detection, offering higher sensitivity and specificity. Even though both modalities are effective in AF detection, the considerable variability in ECG smartwatch performance highlights the need for further research and standardization.

Keywords: atrial fibrillation, cardiac arrythmia, ECG smartwatches, PPG smartwatches, wearable technology

Introduction

Atrial fibrillation (AF) is the most common form of cardiac arrhythmia[1]. It is characterized by disorganized atrial activity recognized in the electrocardiogram (ECG) as an irregularly irregular rhythm lasting more than 30 seconds, with no discernible P-waves preceding the QRS complex[2,3]. Estimates suggest that the number of individuals affected by AF could double or even triple by 2050[2,4]. While AF affects about 1% of the general population, this figure escalates to approximately 9% for those over 75, and the lifetime risk by age 80 reaches 22%[2]. Its prevalence and incidence are on the rise, varying across different genders and racial/ethnic groups, influenced by a combination of genetic, environmental, and socioeconomic factors. Men have a higher incidence compared to women. Even though women experience longer-lasting and more frequent symptomatic episodes, have a lower prevalence of coronary artery disease, and are less likely to receive appropriate anticoagulation therapy, leading to a higher adjusted risk of death[5].

HIGHLIGHTS

  • Atrial fibrillation (AF) is the most common cardiac arrhythmia.

  • This meta-analysis compared ECG and PPG smartwatches accuracy for AF detection.

  • PPG smartwatches showed 97.4% sensitivity and 96.6% specificity in detecting AF.

  • ECG smartwatches showed 83% sensitivity and 88.4% specificity in in detecting AF.

  • Both ECG and PPG technologies can aid in early AF detection for early intervention.

It also varies among different racial and ethnic groups. In the United States, studies have shown that Black patients have a significantly lower compared to White patients. For instance, in a cohort of heart failure patients, Black patients had markedly less AF than White patients (20.8% vs. 44.8%, P < 0.001[6]. Similarly, research has found that Black patients had a significantly lower risk of AF compared to White patients, with an incidence rate of 21.4% in Black versus 25.5% in White patients[7]. Additionally, it is noted that AF is more commonly diagnosed in men and is relatively more prevalent in Caucasian populations compared to Black individuals[8,9].

When considering both gender and ethnic factors, disparities in AF Prevalence become more pronounced. A study reported that the prevalence of subclinical AF was 3.3% in White men, 2.5% in White women, 2.1% in Black men, and 1.6% in Black women[7]. Several risk factors contribute to its development including aging, hypertension, obesity, diabetes mellitus, coronary artery disease, etc[10].

Traditional diagnostic methods, particularly 12-lead ECG, are widely used in clinical settings; however, they provide a snapshot at a specific time instead of allowing for continuous monitoring required for capturing the transient nature of paroxysmal AF episodes[11,12]. Though these conventional techniques are reliable, there is an evident need for advanced screening devices to offer real-time monitoring and enhance clinical detection rates.

In recent years, wearable technology, especially smartwatches with integrated cardiac monitoring features, has emerged as a promising avenue for noninvasive AF detection[13]. They can identify arrhythmias independently, allowing patients to take control of their therapy. ECG and photoplethysmography (PPG) smartwatches are among the leading technologies developed for this purpose despite the rise of other innovations, such as bright rings, chest patches, smartphone-based sensors, and earring sensors.

ECG-based smartwatches, including popular models such as the Apple Watch and Samsung Galaxy Watch, monitor the heart’s electrical activity to detect AF[14]. They are equipped with single-lead ECG capabilities, allowing users to record their heart’s electrical signal directly from the wrist. In this setup, the watch’s back serves as a positive electrode, and the user places a fingertip on the opposite hand to act as a negative electrode creating a lead similar to the standard ECG Lead I. This configuration enables the detection of AF by identifying irregularities in the heart’s electrical patterns[15]. In contrast, PPG-based smartwatches are based on the optical technique that measures blood volume changes in the microvascular bed of tissue[16]. Sensors emit light into the skin and detect the amount of light that is either absorbed or reflected, which varies with blood flow and is used to calculate heart rate and rhythm[17].

Despite these advancements, challenges remain in AF detection. PPG signals are highly susceptible to motion artifacts, Even minimal arm movements can degrade signal quality, leading to false positives or missed detections. Additionally, PPG signals can be poor even without arm movement, due to various factors such as skin tone, ambient light, and sensor placement[17]. The intermittent nature of PPG monitoring can result in missed detections during short AF episodes[18]. On the other hand, ECG-based detection requires user compliance and may be less convenient for continuous monitoring[19].

While their effectiveness in detecting AF is less well-established than standard 12-lead ECG, various studies have assessed their diagnostic performance, with some elaborating on their potential impact on the diagnosis of those patients at higher risk, such as stroke patients[20]. Additionally, Smartwatches are emerging as valuable tools in healthcare, offering potential benefits for various populations. They can contribute to increased physical activity and weight loss in chronic patients, especially when integrated into broader lifestyle interventions[21]. These devices show promise in screening for AF and monitoring conditions such as Parkinson’s disease, epilepsy, and diabetes[21,22]. However, most studies have been conducted in hospital settings, emphasizing more studies to evaluate their effectiveness in long-term ambulatory monitoring among diverse populations[23,24]. These studies demonstrate the potential of wearable devices with integrated AF detection algorithms for improving AF screening, diagnosis, and monitoring, demonstrating non-inferiority to standard care in arrhythmia detection and improving patient care while, on the other hand, reducing healthcare costs[25]. However, no thorough analysis has examined these devices’ diagnostic accuracy in a way that considers how well their underlying algorithms work.

This knowledge gap is substantial, given consumers’ and doctors’ growing usage of smartwatches for ongoing AF monitoring and early intervention. Comprehending the relative efficacy of different devices is crucial to inform clinical judgments and assist users in choosing the most dependable alternative for AF detection.

This review aims to address this gap by evaluating and comparing the diagnostic performance of ECG and PPG smartwatches for detecting AF. We will clarify these technologies’ sensitivity, specificity, and overall diagnostic effectiveness differences by analyzing available studies. Our findings aim to inform clinical practice and assist consumers in making informed choices regarding the most reliable option for continuous AF monitoring and early intervention.

Methodology

Registration

This systematic review and meta-analysis protocol was registered in Open Science Framework (OSF) (Registries: 10.17605/OSF.IO/D3QT8).

Reporting

The current systematic review and meta-analysis were conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines[26], and reported according to the Assessing the Methodological Quality of Systematic Reviews (AMSTAR) guidelines[27].

Databases literature searches

Between 18 August and 23 September 2024, an extensive and systematic search was carried out across several databases, including PubMed/MEDLINE, Google Scholar, DOAJ, AJOL, and the Cochrane Library, to locate original studies evaluating the accuracy of ECG-based smartwatches or PPG-based smartwatches in detecting atrial fibrillation. Alongside these database searches, additional manual searches were performed using Google to gather relevant grey literature. The search strategy utilized vital terms such as “ECG smartwatch,” “PPG smartwatch,” “smartwatch,” “scan watch,” “Apple watch,” “Samsung watch,” “wrist-wearables,” “wearable ECG,” “wearable PPG,” “cardiac monitoring,” “atrial fibrillation,” and “cardiac arrhythmia.” As outlined in Table 1, Boolean operators are used for PubMed search. To ensure the inclusion of recent studies, the search was limited to literature published from 2014 onward, covering the most recent decade. The references collected, including those from gray literature, were imported into Rayyan software for deduplication and subsequently screened based on predefined inclusion and exclusion criteria.

Table 1.

Boolean operators based search parameters in PubMed

Search Search string Number of results
#1 ((((((((ECG smartwatch) OR (PPG smartwatch)) OR (smartwatch)) OR (scan watch)) OR (Apple watch)) OR (Samsung watch)) OR (wrist wearable)) OR (wearable ECG)) OR (wearable PPG) Filters: from 2014 to 2024 5547
#2 ((atrial fibrillation) OR (cardiac arrhythmia)) OR (cardiac monitoring) Filters: from 2014 to 2025 159 523
#3 #1 AND #2 1810

Eligibility criteria

The inclusion criteria established for the studies analyzed in this systematic review and meta-analysis were meticulously crafted to ensure the selection of high-quality research studies focusing on the diagnostic efficacy of ECG or PPG-based smartwatches in detecting AF. The review encompassed only primary research articles, including clinical trials, cross-sectional studies, case-control studies, and prospective cohort studies that reported diagnostic metrics such as sensitivity, specificity, accuracy, positive predictive value (PPV), and negative predictive value (NPV).

The studies had to evaluate adult populations and concentrate specifically on smartwatches fitted with ECG or PPG sensors for AF detection, with a comparative analysis against reference standards such as 12-lead ECG, Holter monitors, or telemetry ECG systems. Only publications in English from 2014 to 2024 were considered to capture recent advancements in wearable technology. To uphold methodological rigor, studies were mandated to meet a minimum quality threshold and evaluated using the Joanna Briggs Institute (JBI) Critical Appraisal Tool, with only those rated as medium or high quality (JBI scores above 50%) being included.

Studies categorized as reviews, commentaries, case reports, or those unrelated to AF detection, along with non-English or inaccessible studies, were excluded to ensure consistency and relevance. Table 2 details all inclusion and exclusion criteria of this systematic review and meta-analysis.

Table 2.

Inclusion and exclusion criteria

Criterion Included Excluded
Study design Diagnostic validation studies, clinical trials, cross-sectional, case-control, and prospective cohort studies Commentaries, perspectives, case reports, conference proceedings, reports, reviews, opinions, and letters to the editors
Year of publication 2014 to 2024 Before 2014
Outcome of interest Reporting performance of ECG or PPG smartwatches in the detection of atrial fibrillation (AF)
  • Not reporting both sensitivity and specificity

  • Wrist-worn devices other than smartwatches

  • ECG or PPG smart watches for other cardiac monitoring purposes, not AF detection (ex: Heart rate or blood pleasure monitoring)

Accessibility Abstract and full text assessable
  • Abstract and full text inaccessible

  • Abstract accessible, full text inaccessible

Language English Other remaining languages besides English
Quality Medium and high-quality studies Low-quality studies

Data extraction

Two authors independently extracted data from the selected studies using a standardized template in Microsoft Excel. The extracted data included details such as author identification, study design, study location, total and actual number of participants, mean age, male percentage, type of sensor used (ECG or PPG smartwatch), reference standard ECG measurements, and crucial performance metrics such as accuracy, sensitivity, specificity, PPV, and NPV. Any disagreements that arose during this process were resolved through discussion. If consensus could not be reached, a third reviewer was consulted to ensure the accuracy and consistency of the extracted data.

Quality assessment

The quality of the included studies was assessed using the Joanna Briggs Institute (JBI) critical appraisal tool[28]. Studies were categorized into three distinct quality tiers: high-quality (JBI scores exceeding 70%), medium-quality (JBI scores ranging from 50% to 70%), and low-quality (JBI scores below 50%). Only studies rated as medium or high quality were deemed suitable for inclusion in this systematic review and meta-analysis, ensuring the analysis was based on sufficient rigor and reliability research.

Assessment of risk of bias

The risk of bias and applicability of the included studies was evaluated using the Quality Assessment of Diagnostic Accuracy Studies–2 (QUADAS-2) tool, as recommended by the Cochrane Collaboration and the U.K. National Institute for Health and Care Excellence (NICE). This assessment encompassed four domains: (1) participant selection, (2) index test, (3) reference standard, and (4) flow and timing. The results were visualized utilizing the Robvis tool, which facilitates the graphical representation of risk of bias assessments.

Statistical analysis

The meta-analysis employed a Restricted Maximum Likelihood (REML) random-effects model to separately calculate ECG and PPG smartwatches’ aggregated sensitivity and specificity in detecting atrial fibrillation to address the anticipated variability among studies. The degree of heterogeneity was evaluated using the I2 statistics. Meta-analyses were performed using Jamovi 2.3.28 software. A significance threshold of P <0.05 was applied to all statistical analyses.

Ethical approval and consent form

Ethical approval and consent forms were unnecessary since this study utilized publicly available secondary data and did not involve direct interaction with human or animal subjects.

Results

Study selection

A total of 2564 studies were initially identified through searches of electronic databases and grey literature sources. After removing 631 duplicate records, 1933 studies were available for the preliminary evaluation. Of these, 1796 were excluded during title and abstract screening due to irrelevance, leaving 137 studies for in-depth full-text assessment. Among these, 112 studies were discarded for various reasons: 47 did not address the outcomes of interest, 18 involved smart wearables other than smartwatches, 23 focused on cardiac monitoring unrelated to atrial fibrillation detection, 17 were not primary research articles, and seven had full texts that were inaccessible. As a result, 25 studies (11 on PPG smartwatches and 14 on ECG smartwatches) met all the specified criteria and were included in the systematic review and meta-analysis, as depicted in Fig. 1.

Figure 1.

Figure 1.

PIRSMA flowchart demonstrating selection of eligible studies.

Characteristics of included studies

As detailed in Table 3, among the 25 entries analyzed, eight studies[29-36] (32%) were conducted in the USA, six studies[37-42] (24%) in France, and two studies (8%) each in Switzerland[43,44], Taiwan[45,46], and the Netherlands[47,48]. One study (4%) was conducted in the UK[49], Norway[50], Japan[51], and Australia[52]. One study[53] (4%) was international and involved Germany and Switzerland. Most of these were diagnostic clinical validation studies and clinical trials, encompassing 16 440 participants, of whom 14 622 actively participated. The mean age of the participants was 62.44 years, with a sex distribution of 65.66% males and 34.34% females. These studies evaluated two types of smartwatches: 11 ECG smartwatches (44%) and 14 PPG smartwatches (56%). To assess the accuracy of these sensors, the most commonly used gold standards were the 12-lead ECG, Holter ECG, and telemetry ECG.

Table 3.

Summary and characteristics of included studies

Author ID Study design Country Number of subjects Total number screened Mean age Male (%) Type of sensor Gold standard Accuracy Sensitivity Specificity PPV NPV
Dörr et al, 201955 Prospective case-control trial Germany and Switzerland 672 508 76.4 55.7 PPG Smartwatch 12-Lead ECG 96.1 93.7 98.2 97.8 94.7
Chang et al, 202247 ProspectiveDiagnostic accuracy study Taiwan 200 112 66.1 63.5 PPG Smartwatch 24-h Holter ECG 93.5 97.3 88.6 91.6 96.3
Selder et al, 202349 ProspectiveDiagnostic accuracy study Netherlands 78 78 66 53.5 PPG Smartwatch 12-Lead ECG 97 98 96 96 99
Nonoguchi et al, 202253 Prospective single-centre study Japan 1500 166 66.5 68 PPG Smartwatch Single-lead Telemetry ECG 98 90.6 69.4 99.5
Bashar et al, 201935 Prospective single-centre study USA 46 46 - - PPG Smartwatch 7-Lead Holter ECG 97.54 98.18 97.43 - -
Bashar et al, 201934 ProspectiveDiagnostic accuracy study USA 20 20 - - PPG Smartwatch 7-Lead Holter ECG 97.11 96.15 97.37 - -
Bonomi et al, 201850 Prospective cohort clinical trial Netherlands 20 18 59.6 55.5 PPG Smartwatch Single-lead ECG apparatus (Actiwave Cardio) 98 96 100 100 98
40 34 67.4 61.1 PPG Smartwatch 24-h Holter ECG 97 93 100 100 95
Liao et al, 202248 Prospective single-centre study Taiwan 116 116 59.6 67 PPG Smartwatch 24-h Holter ECG 95.8 97 94 96 95.4
Han et al, 202032 Prospective Diagnostic accuracy study USA 37 37 70.49 78.3 PPG Smartwatch 7-Lead Holter ECG 97.95 98.18 97.9 91.53 99.57
Tison et al, 201836 Prospective Diagnostic accuracy study USA 9750 9750 42 63 PPG Smartwatch 12-Lead ECG 97 98 90.2 90.9 97.8
Ding et al, 201933 Prospective Diagnostic accuracy study USA 40 40 71 80 PPG Smartwatch 7-Lead Holter ECG 98.1 98.2 98.1 91.5 99.6
Mannhart et al, 202346 Clinical diagnostic validation study Switzerland 201 201 66.7 69 Single-lead ECG Apple Watch 6 12-Lead ECG 85 75 58 93
Single-lead ECG Samsung Galaxy Watch 3 12-Lead ECG 85 75 57 93
Single-lead ECG Withings Scanwatch 12-Lead ECG 58 75 48 82
Single-lead ECG Fitbit Sense 12-Lead ECG 66 79 55 85
Fiorina et al, 202443 Clinical diagnostic validation study France 400 400 64.8 70.8 ECG smartwatch with DNN 12-Lead ECG 91 95
ECG smartwatch with Apple app 12-Lead ECG 61 97
Müller et al, 202452 Prospective observational study Norway 105 93 68 78.5 Single-lead ECG Apple watch 5 2–4 days telemetry ECG 95 91 96 83 81
Single-lead ECG Apple watch 5 12-Lead ECG 71 92 72 96
Gala et al, 202351 Clinical diagnostic validation study UK 400 393 63 73 Single-lead ECG Apple Watch Series 6 12-Lead ECG 67.4 74.9
Fiorina et al, 202242 Clinical diagnostic validation study France 101 101 ECG Apple smartwatch with ECG 2.0 App 12-Lead ECG 89 81 97
ECG Apple smartwatch with DNN 12-Lead ECG 91 92 90
Campo et al, 202241 Clinical diagnostic validation study France 262 262 67.7 61.1 Single-lead ECG ScanWatch 12-Lead ECG 96.3 100
Rajakariar et al, 202054 Clinical diagnostic validation study Australia 200 200 67 56.5 ECG smartwatch (AliveCor KardiaBand) 12-Lead ECG 84.3 94.4 81.9 54.8 98.4
Wasserlauf et al, 201938 Clinical diagnostic validation study USA 26 24 72.1 65.4 ECG smartwatch (Apple Watch with KardiaBand) Insertable cardiac monitor (ICM; Reveal LINQ) 97.7 98.9 76.8 99.9
Abu-Alrub et al, 202244 Clinical diagnostic validation study France 200 200 62 56 ECG smartwatch (Apple Watch) 12-Lead ECG 87 86 86 87
ECG smartwatch (Samsung) 12-Lead ECG 88 81 82 87
ECG smartwatch (Withings) 12-Lead ECG 78 80 80 78
Racine et al, 202239 Clinical diagnostic validation study France 734 734 66 58 ECG smartwatch (Apple Watch series 5) 12-Lead ECG 69 81
Badertscher et al, 202245 Clinical diagnostic validation study Switzerland 319 319 67 52 ECG Withings Scanwatch 12-Lead ECG 76 99
Velraeds et al, 202340 Clinical diagnostic validation study France 723 723 ECG smartwatch (Apple Watch Series 5) 12-Lead ECG 91.6 89.6 92.1 74.2 97.2
Seshadri et al, 202026 Clinical diagnostic validation study USA 50 47 ECG smartwatch (Apple Watch Series 4) Telemetry ECG 96 100
Bumgarner et al, 201837 Clinical diagnostic validation study USA 100 100 68.2 83 ECG smartwatch (Apple Watch) 12-Lead ECG 93 84

Assessment of risk of bias

Of the included studies, five demonstrated a high risk of bias in patient selection, three in the index test, one in the reference standard, and three in flow and timing. The remaining studies were assessed as having either low risk or unclear risk of bias. Overall, nine studies were categorized as having a high risk of bias, three as having some concerns, and twelve as having a low risk of bias (Fig. 2).

Figure 2.

Figure 2.

Risk of bias graph’: review authors’ judgments about each risk of bias item presented as percentages across all included studies, and “risk of bias summary”: review authors’ judgments about each risk of bias item for each included study.

Performance of ECG and PPG smartwatches in atrial fibrillation detection

The current meta-analysis evaluated the efficacy of ECG and PPG smartwatches in detecting atrial fibrillation, highlighting differences in their pooled sensitivity and specificity. ECG smartwatches exhibited a pooled sensitivity of 83% (95% CI: 78–88), accompanied by significant heterogeneity (I2 = 93.75%) (Fig. 3). The pooled specificity for ECG smartwatches was 88.4% (95% CI: 84.5–92.2), also with substantial heterogeneity (I2 = 96.15%) (Fig. 4). In contrast, PPG smartwatches showed a higher pooled sensitivity of 97.4% (95% CI: 96.5–98.3) with minimal heterogeneity (I2 = 3.16%) (Fig. 5) and a pooled specificity of 96.6% (95% CI: 94.9–98.3) with moderate heterogeneity (I2 = 75.94%) (Fig. 6).

Figure 3.

Figure 3.

Pooled sensitivity of ECG smartwatches in detecting atrial fibrillation.

Figure 4.

Figure 4.

Pooled specificity of ECG smartwatches in detecting atrial fibrillation.

Figure 5.

Figure 5.

Pooled sensitivity of PPG smartwatches in detecting atrial fibrillation.

Figure 6.

Figure 6.

Pooled specificity of PPG smartwatches in detecting atrial fibrillation.

When comparing the two technologies, PPG-based smartwatches were 17.3% more sensitive than ECG-based smartwatches in detecting atrial fibrillation, with sensitivity rates of 97.4% and 83%, respectively. Moreover, PPG smartwatches demonstrated 9.2% higher specificity than ECG smartwatches, achieving 96.6% specificity compared with 88.4%. This significant difference suggests that PPG-based smartwatches may offer superior accuracy in detecting atrial fibrillation compared to their ECG counterparts.

The Summary ROC (SROC) curves for both ECG and PPG smartwatches in detecting atrial fibrillation revealed distinct performance characteristics. The ECG smartwatch curve demonstrated high sensitivity at lower false positive rates (FPR), gradually plateauing as FPR increases, indicating strong initial diagnostic ability but diminishing returns with higher false positive rates. Meanwhile, the curve stayed above the chance line, showing a reliable ability to detect atrial fibrillation (Fig. 7). In contrast, the PPG smartwatch SROC curve was almost flat, remaining near the upper left corner, suggesting consistently high sensitivity with few false positives. However, the PPG curve showed little improvement with increasing FPR, indicating that while it performs well at low FPRs, its sensitivity does not significantly improve with higher thresholds (Fig. 8). Both devices demonstrated effectiveness, though the ECG smartwatch exhibits a more pronounced trade-off between sensitivity and false positive rates, while the PPG smartwatch maintains stable sensitivity across varying FPRs.

Figure 7.

Figure 7.

Summary ROC (SROC) curve for performance of ECG smartwatches in detecting atrial fibrillation.

Figure 8.

Figure 8.

Summary ROC (SROC) curve for performance of PPG smartwatches in detecting atrial fibrillation.

Publication bias

The publication bias assessment for sensitivity and specificity in studies evaluating ECG smartwatch performance for detecting atrial fibrillation revealed consistent findings. For sensitivity, the fail-safe N was 137 743 (P < .001), indicating that a large number of unpublished “null” studies would be required to change the observed effect size, suggesting robust results. However, both the Rank Correlation Test (Kendall’s Tau = −1.000, P < .001) and Regression Test (Egger’s Test: Z = −13.362, P < .001) showed funnel plot asymmetry (Fig. 9), pointing to potential publication bias or selective reporting. For specificity, the fail-safe N was notably higher at 371 468 (P < .001), further supporting the robustness of the findings. For specificity, similar asymmetry was observed in both the Rank Correlation Test (Kendall’s Tau = −1.000, P < .001) and Regression Test (Egger’s Test: Z = −14.472, P < .001), showing funnel plot asymmetry (Fig. 10).

Figure 9.

Figure 9.

Funnel plot for the sensitivity of ECG smartwatches in detecting atrial fibrillation.

Figure 10.

Figure 10.

Funnel plot for specificity of ECG smartwatches in detecting atrial fibrillation.

The publication bias assessment for sensitivity and specificity in studies evaluating PPG-based smartwatches for detecting atrial fibrillation indicates robust findings, though potential bias may influence the results. For sensitivity, the fail-safe N was 192 463 (P < .001), suggesting that a large number of unpublished “null” studies would be needed to alter the observed effect, supporting the stability of the findings. However, significant funnel plot asymmetry was observed (Fig. 11), with Kendall’s Tau of −1.000 (P < .001) and Egger’s Test showing a Z value of −2.789 (P = 0.005), pointing to possible publication bias. For specificity, the fail-safe N was 236 072 (P < .001), further reinforcing the robustness of the findings. Yet, similar asymmetry was observed in both the Rank Correlation Test (Kendall’s Tau = −1.000, P < .001) and Regression Test (Egger’s Test: Z = −5.688, P < .001), suggesting publication bias (Fig. 12).

Figure 11.

Figure 11.

Funnel plot for sensitivity of PPG smartwatches in detecting atrial fibrillation.

Figure 12.

Figure 12.

Funnel plot for specificity of PPG smartwatches in detecting atrial fibrillation.

Subgroup analysis

Subgroup analysis was conducted to explore the significant heterogeneity identified in the main studies. Stratification was performed based on the type of ECG used as the gold standard, the country of study, and the mean age of participants. For ECG-based smartwatches, heterogeneity remained high across all subgroups for the type of ECG and mean age of participants (P < 0.001), suggesting these variables did not explain the observed variability. However, stratification by the country of study revealed that studies conducted in the USA demonstrated lower heterogeneity for sensitivity (I2 = 18.6%, P = 0.277). Despite this reduction for studies conducted in the US, heterogeneity remained overall high, indicating that other unaccounted factors might contribute to the observed variation (Table 4).

Table 4.

Subgroup analysis for studies on the performance of ECG smartwatches in detecting AF.

Subgroup Category Sensitivity (95% CI) I2 P value Specificity (95% CI) I2 P value
Gold standard* 12 Lead ECG 81.9 (76.6, 87.2) 93.4 <0.001 87.4 (83.3, 91.4) 94.8 <0.001
The mean age of the participants Below 65 81.2 (73.5, 88.8) 89.5 <0.001 88.3 (82.5, 94.1) 89.8 <0.001
Above 65 82.4 (74.9, 89.9) 95.2 <0.001 87 (81.3, 92.8) 96.7 <0.001
Country of study France 83.8 (77.2, 90.5) 91 <0.001 90.6 (86.1, 95.1) 91.5 <0.001
USA 96.2 (93.8, 98.6) 18.6 <0.277 94.7 (85.4, 100) 97.7 <0.001

*Other types of ECG used as the gold standard (telemetry ECG and ICM) had insufficient observations for meta-analysis.

For PPG-based smartwatches, heterogeneity was successfully explained across specific subgroups (Table 5). For sensitivity, studies utilizing 7-lead ECG, conducted in the USA, or including participants with a mean age >65 years exhibited no heterogeneity at all. Conversely, minimal heterogeneity was observed in studies employing 24-hour Holter ECG and those with a mean age <65 years, with values below 1%. The highest heterogeneity occurred in studies conducted in the Netherlands (I2 = 34.35%, P = 0.214), although it was not statistically significant. For specificity, heterogeneity remained non-significant except for studies utilizing 24-hour Holter ECG (I2 = 86.3%, P < 0.001), 12-lead ECG (I2 = 71.0%, P = 0.046), and those with a mean age <65 years (I2 = 88.2%, P < 0.001).

Table 5.

Subgroup analysis for studies on the performance of PPG smartwatches in detecting AF

Subgroup Category Sensitivity (95% CI) I[2] P value Specificity (95% CI) I[2] P value
Gold standard 12 Lead ECG 97.4 (95.6, 99.2) 1.95 0.261 95.4 (91.2, 99.6) 71 0.046
7 Lead ECG 97.9 (96.5, 99.3) 0 0.809 97.7 (96.3,99.2) 0 0.982
24 hrs Holter ECG 96.4 (94.3, 98.5) 0.06 0.332 94.6 (88.4, 100) 86.3 <0.001
The mean age of participants Below 65 96.9 (956, 98.2) 0.78 0.472 96 (93, 98.9) 88.2 <0.001
Above 65 98.1 (96.6, 99.7) 0 0.994 96.2 (92.3, 100) 74.8 0.055
Country of study Netherlands 96.3 (93.6, 98.9) 34.35 0.214 99.3 (98.3, 100) 3 0.222
USA 97.9 (96.7, 99.2) 0 0.914 97.3 (95.9, 98.7) 0.19 0.183

Discussion

The current systematic review and meta-analysis analyzed 25 high-quality studies to compare the accuracy of ECG smartwatches versus PPG smartwatches in detecting atrial fibrillation. The findings from this systematic review and meta-analysis are pivotal in addressing the increasing prevalence of AF and its associated complications, such as stroke, which contributes to rising healthcare costs and mortality rates worldwide. With advances in wearable technology and brilliant watches, there is a growing interest in their potential to serve as convenient tools for early AF detection and continuous heart rhythm monitoring, particularly outside clinical settings.

Diagnostic accuracy of PPG-based smartwatches

Among the 11 PPG-based smartwatch studies, pooled sensitivity and specificity estimates were remarkably high, with a pooled sensitivity of 0.97 (0.96, 0.98) and a pooled specificity of 0.97 (0.95, 0.98). This suggests that PPG-based smartwatches perform well at detecting AF episodes. For example, studies by Dörr et al (2019)[53] and Tison et al (2018)[34] demonstrated high diagnostic accuracy, with a sensitivity of 93.7% and 98% and specificity of 98.2% and 90.2%, respectively.

Interestingly, Bashar et al (2019)[32] evaluated two separate PPG-based devices, showing consistent results in detecting AF, with both studies reporting sensitivities above 96% and specificities above 97%. Similarly, Bonomi et al (2018)[48] reported a specificity of 100%, which underscores the reliability of PPG-based devices for detecting abnormal heart rhythms, particularly AF.

Recent studies have demonstrated the potential of PPG smartwatches to detect AF (Ding et al, 2019)[54]. Using a smartwatch algorithm, we have reported high accuracy (98.1%) in identifying irregular pulses among older adults. Similarly, Bashar et al (2019) achieved 97.54% accuracy in AF detection from wrist PPG signals, which aligns with our findings.

PPG-based smartwatches are more accessible to the general population because of their affordability, ease of use, and non-invasive nature. However, some limitations, such as the potential for false positives from motion artifacts, remain a concern, as indicated by Nonoguchi et al (2022)[51], who reported a specificity of 69.4%, significantly lower than that reported in other studies with different results. This highlights the variability that can occur depending on the device’s algorithm and the user’s activity levels.

Diagnostic accuracy of ECG-based smartwatches

Fourteen studies investigating the accuracy of ECG-based smartwatches showed a slightly lower overall diagnostic accuracy than PPG-based smartwatches. The pooled sensitivity was 0.83 (0.78, 0.88), and the specificity was 0.88 (0.85, 0.92). Some studies have reported higher and lower accuracy levels of ECG smartwatches in AF detection compared to our findings. Avram et al (2021)[55] validated a novel algorithm combining PPG and ECG, showing excellent performance with 96.9% sensitivity and 99.3% specificity when using both modalities. Additionally, Selder et al (2023)[47] reported a high sensitivity of 98% and specificity of 96% for ECG smartwatches, aligning with another study that revealed that deep neural networks applied to smartwatch ECGs achieved a sensitivity of 91% and a specificity of 95%, outperforming traditional methods[41]. Neri et al (2024)[56] validated an algorithm for ECG signal quality, achieving a sensitivity of 94.7% and specificity of 95.3%, demonstrating the reliability of wearable ECG devices, further contrasting the findings of our review. Similarly, Perera et al (2016)[57] reported a sensitivity of 80% and a specificity of 97% for ECG smartwatches, reaffirming their reliability in clinical settings.

However, other studies have reported lower accuracy for ECG-based devices (Mannhart et al, 2023)[44]. The Apple Watch 6 had a sensitivity of 75% and a specificity of 58% in detecting AF. Similarly, Abu-Alrub et al (2022)[42] reported mixed results across different ECG smartwatch models, with sensitivities ranging from 81% to 92% and specificities ranging from 86% to 97%. Differences likely influence these variations in device hardware, algorithm sensitivity, and the gold standard used for comparison, such as telemetry or 12-lead ECG.

A notable finding from studies such as Rajakariar et al (2020)[52] and Wasserlauf et al (2019)[36] was that while ECG-based smartwatches generally provided accurate results, they are more commonly used in clinical or higher-risk populations where detailed heart rhythm readings are crucial. These devices, which generate a direct ECG trace, are more suitable for individuals with known cardiovascular conditions or those requiring closer monitoring by healthcare professionals.

The role of wearable technology in promoting patient engagement and self-monitoring has been well-documented. Smartwatches facilitate continuous monitoring of heart rhythm, which can lead to increased patient awareness of their health status[58]. Studies indicate that patients using smartwatches are more likely to report irregularities, thereby seeking medical advice sooner, which is vital for timely interventions[59]. According to Del-Valle-Soto et al (2024)[60], these devices facilitate real-time health monitoring and enhance user engagement through personalized feedback and gamification, encouraging widespread adoption among patients.

High perceived ease of use and favorable attitudes toward wearable technology have been reported, particularly among younger demographics[61], with over 80% of global consumers expressing willingness to utilize these wearable devices for health monitoring[60]. Despite these advantages, challenges, such as data privacy and demographic disparities in usage, persist, necessitating ongoing research and tailored strategies to enhance accessibility.

Overall, the findings from this systematic review and meta-analysis reinforce the existing literature and highlight the significant potential of ECG and PPG smartwatches in the early detection of atrial fibrillation. The pooled sensitivity and specificity rates and comparisons to more than 25 studies demonstrated a consistent trend toward high diagnostic accuracy. They underscored the importance of technological advancements in improving patient care.

Clinical implications of the findings

The high diagnostic accuracy of PPG-based smartwatches makes them particularly well suited for general population screening and long-term AF monitoring, where continuous heart rate variability tracking is essential. Their non-invasive and easy-to-use nature can enable constant cardiac monitoring. In addition, PPG-based devices are relatively affordable, increasing their accessibility to larger populations, particularly in community settings.

By contrast, ECG-based smartwatches, which are slightly less sensitive and specific, provide more precise diagnostic information by directly measuring the heart’s electrical activity. This makes them more appropriate for clinical use, especially in patients with a higher risk of AF or pre-existing cardiovascular conditions. The ability to generate an ECG waveform that healthcare professionals can review adds value to ECG-based devices, particularly in high-risk populations, where more detailed heart rhythm information is necessary.

The role of smart wearables, such as smartwatches, is increasingly significant in managing high-risk patients. Early detection and prediction of atrial fibrillation (AF) risks are crucial for identifying individuals who may benefit from these devices[20,62]. This early diagnosis is essential as it facilitates timely interventions, potentially reducing the disease’s progression and associated complications[63].

Study strengths and limitations

The primary strength of this review is the inclusion of high-quality studies with rigorous methodologies. The consistency of the results across studies further reinforces the validity of the pooled estimates for sensitivity and specificity. However, this study has several limitations that need to be acknowledged.

First, the studies used different gold standards for AF detection, including 12-lead ECG, Holter monitors, and telemetry ECG, which could have introduced variability in diagnostic accuracy estimates. Second, the demographic characteristics of the study participants varied significantly, with some studies focusing on older populations while others included younger cohorts. These differences could affect the generalizability of the results, mainly when applied to diverse populations.

In addition, while most studies have reported high diagnostic accuracy for PPG and ECG smartwatches, issues such as motion artifacts, environmental conditions, and user adherence may affect real-world performance, particularly for PPG-based devices. These factors should have been consistently addressed across studies, warranting further investigation in future research.

Conclusion and future direction

In conclusion, PPG- and ECG-based smartwatches demonstrate significant potential for detecting AF, with PPG-based devices showing superior sensitivity and specificity. Although ECG-based smartwatches remain valuable for clinical applications, PPG-based devices offer a reliable and accessible solution for continuous monitoring in the general population.

Future research should focus on improving the accuracy of these devices, particularly by refining algorithms to reduce false positives and account for factors such as motion artifacts and skin tone variability. Direct comparisons between PPG- and ECG-based smartwatches in real-world settings would further clarify their role in AF detection. Moreover, Concerns about equity and bias in wearable device adoption persist, with lower usage rates among underserved populations[64]. A study of smartwatch ECG users in the US found that younger individuals and those in states with higher innovation indices were likelier to use these devices[65]. The observed AF prevalence among users was lower than expected, suggesting a potential demographic gap in access to this technology. As smartwatches become more prevalent in healthcare, addressing these disparities and ensuring equitable distribution of innovation is crucial for maximizing their clinical impact across diverse populations[64]. Moreover, studies with larger sample sizes and more varied populations would help generalize the findings and ensure broader applicability across different patient demographics.

Footnotes

Sponsorships or competing interests that may be relevant to content are disclosed at the end of this article.

Published online 19 March 2025

Contributor Information

Dan Yang Luo, Email: 1540116648@qq.com.

Zhi Wei Zhang, Email: 324002995@qq.com.

Olivier Sibomana, Email: oliviersibomana917@gmail.com.

Salomon Izere, Email: izesajw73@gmail.com.

Ethical approval

No ethical approval was needed as the study was secondary, using publicly available data.

Consent

No consent was needed as study did not directly involve human subjects.

Sources of funding

No funding was received.

Author’s contribution

D.Y.L.: Conceptualization, methodology, supervision, resources, manuscript writing; Z.W.Z.: Conceptualization, methodology, resources, manuscript writing; O.S.: Conceptualization, methodology, resources, data analysis, visualizations, manuscript writing

S.I.: Conceptualization, methodology, resources, data analysis, visualizations, manuscript writing. All authors reviewed and approved the final draft.

Conflicts of interest disclourse

Authors have no conflict of interest to declare.

Guarantor

Olivier Sibomana.

Research Registration Unique Identifying Number (UIN)

Registry name: Open Science Framework (OSF) registries.

Registration DOI: 10.17605/OSF.IO/D3QT8.

Provenance and peer review

Not invited.

Data availability statement

Essential data are contained in the manuscript. Additional data are available from the corresponding author upon reasonable request.

References

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Essential data are contained in the manuscript. Additional data are available from the corresponding author upon reasonable request.


Articles from Annals of Medicine and Surgery are provided here courtesy of Wolters Kluwer Health