Assay detection limit established by titration of pool 5 reference antiserum

Expected value from dilution factor (WR U/ml) Value observed in assay (WR U/ml)a
Accurate quanti- tation
Expt 1 Expt 2 Expt 3
180.0 177.1 (−1.6) 179.8 (0.1) 182.9 (1.6) Yes
90.0 74.2 (−17.6) 83.4 (−7.3) 82.8 (−8.0) Yes
45.0 36.6 (−18.7) 38.4 (−14.7) 39.5 (−12.2) Yes
22.5 17.5 (−22.2) 20.0 (−11.1) 19.1 (−15.1) Yes
11.2 9.9 (−11.6) 10.2 (−8.9) 9.0 (−19.6) Yes
5.6 4.9 (−12.5) 4.8 (−14.3) 3.6 (−35.7) Yes
2.8 0.3 (−89.3) 1.0 (−64.3) 0.6 (−78.6) No
a