Skip to main content
. 2002 Jan;40(1):44–51. doi: 10.1128/JCM.40.1.44-51.2002

TABLE 1.

Results of HHV-7-specific peptide-based ELISA with six reference HHV-7-seronegative sera

Tested serum Antibody titer
Intra-assay variabilitya
Interassay variabilitya
HHV-6 HHV-7 No.b OD
No. OD
Mean SD CV (%) Mean SD CV (%)
None (blank reaction) NAc NA 8 0.089 0.063 7.1 14 0.086 0.015 17.1
N1 >1,280 <10 4 0.199 0.019 9.5 5 0.207 0.018 8.8
N2 >1,280 <10 4 0.174 0.014 8.3 3 0.160 0.020 13.8
N3 >1,280 <10 3 0.182 0.006 3.3 2 0.177 0.006 3.6
N4 >1,280 <10 3 0.173 0.019 11.0 2 0.173 0.019 11.0
N5 >1,280 <10 4 0.183 0.024 13.2 2 0.177 0.008 4.8
N6 >1,280 <10 3 0.217 0.097 44.7 2 0.162 0.077 47.0
a

The variability of OD signal was studied by comparing the values obtained from different wells of the same microplate within the same assay (intra-assay variability) and the values obtained in independent experiments (interassay variability).

b

Number of independent wells (intra-assay variability) or independent assays (interassay variability).

c

NA, not applicable.