Published in final edited form as: Ultrasound Med Biol. 2025 Mar 11;51(6):941–950. doi: 10.1016/j.ultrasmedbio.2025.01.017

Table 1.

Sensitivity and specificity for complicated cyst detection in B-mode and short-lag spatial coherence (SLSC) images, aggregated among the three readers, compared with that achieved with a generalized contrast-to-noise ratio (gCNR) = 0.73 and with the gCNR threshold closest to (0,1) on the real-time SLSC receiver operating characteristic curve (i.e., gCNR = 0.62).

Real-time Offline
Sensitivity Specificity Sensitivity Specificity
B-mode
Readers 1–3 0.67 0.70 0.67 0.55
SLSC
Readers 1–3 1.00 0.58 0.83 0.48
gCNR = 0.73 0.67 0.95 0.67 1.00
gCNR = 0.62 1.00 0.95 1.00 0.95

gCNR, generalized contrast-to-noise ratio; SLSC, short-lag spatial coherence.