Abstract
Purpose
Dephasing gradients can be introduced within a variety of gradient‐echo pulse sequences to delineate local susceptibility changes (“White‐Marker” phenomenon), e.g., for the visualization of metallic interventional devices which are otherwise difficult to display. We investigated dephased balanced steady‐state free precession (d‐bSSFP) and compared it with similar contrast techniques: dephased RF‐spoiled fast low‐angle shot (d‐FLASH) and dephased steady‐state free precession (d‐SSFP).
Methods
A signal model was formulated to describe the positive contrast in d‐bSSFP. For the example of an MR‐compatible aspiration needle, the positive contrast artifact appearance was theoretically derived, and the model was verified in a water phantom at B0 = 0.55 T. Model accuracy was evaluated by comparing the measured artifact size (for TEs between 3.4 ms and 50 ms) and the signal magnitude to the model prediction.
Results
While positive contrast artifacts for d‐FLASH and d‐SSFP are axisymmetric with respect to the generating object, for d‐bSSFP, a point‐symmetric susceptibility artifact arises for a cylindrical needle due to the characteristic signal formation. The observed d‐bSSFP artifact size was in accordance with the model (error < 1 mm). Measured (predicted) cumulated artifact signal was 1.13 ± 0.07 (1.27) times higher and 5.9 ± 0.4 times higher than the d‐SSFP and d‐FLASH cumulated artifact signal, respectively. In contrast to d‐SSFP, the d‐bSSFP artifact was robust against banding artifacts.
Conclusion
d‐bSSFP contrast is well described by the introduced model. Positive contrast artifacts show higher cumulated signal magnitude, symmetry, and homogeneity compared with d‐FLASH and d‐SSFP and can therefore improve device visualization and potentially device localization.
Keywords: bSSFP, dephased MRI, interventional MRI, positive contrast, white‐marker imaging
1. INTRODUCTION
In MR‐guided interventions, metallic interventional devices can locally disturb the magnetic field, creating a characteristic device artifact in the MR image which is often used for passive device localization. 1 , 2 , 3 , 4 , 5 Due to the difference in magnetic susceptibilty of metallic interventional devices and the surrounding tissue, a local magnetic field gradient is induced. In conventional gradient‐echo (GRE)–based imaging, this leads to dephasing of local magnetization, and therefore signal loss. 6 , 7 , 8 As signal voids in MR images are ambiguous and large signal voids near devices can impede localization, several methods for the generation of susceptibilty‐based, positive contrast have been proposed in the past. 9 , 10 , 11 , 12 , 13 , 14 , 15 These methods make use of the distortion in the magnetic field induced by the metallic device to generate high signal intensities in proximity of the interventional device.
A prominent example of these positive susceptibility‐based contrast techniques is so‐called White‐Marker (WM) imaging. 9 By applying an additional external gradient moment before image readout, the local susceptibilty‐induced gradient moment in proximity of the device can be counteracted, reestablishing phase coherence of the local magnetization, and therefore restoring signal. Simultaneously, the external gradient dephases the signal further away from the device where the magnetic field is not disturbed by the device susceptibilty, hence suppressing anatomy in the image and exclusively highlighting the device. WM imaging has been used for the localization of paramagnetic markers and interventional devices. 9 , 16 A comparable gradient‐compensating approach has also been used to create positive contrast of superparamagnetic iron oxide particle (SPIO)–labeled cells. 17 In the original work by Seppenwoolde et al., WM contrast was established by shortening the slice‐selection rephaser gradient and incidentally creating a residual excess WM gradient moment in the slice‐selection direction to counteract the local susceptibilty‐induced gradient moment. 9 Bakker et al. later generalized the concept of gradient dephasing to arbitrary gradient directions (corresponding to discrete shifts of the acquisition k‐space) and established the term “dephased MRI”. 10
In general, the generation of dephased contrast is primarily possible with GRE‐type sequences, as susceptibility‐induced dephasing is rephased with spin‐echo sequences. GRE‐type sequences typically comprise radiofrequency (RF)‐spoiled GRE or also called spoiled fast low‐angle shot (FLASH), steady‐state free precession (SSFP), and balanced SSFP (bSSFP). 18 Whereas the original paper by Seppenwoolde et al. used dephased spoiled FLASH contrast, 9 which we termed d‐FLASH in the context of this paper, Patil et al. introduced “echo‐dephased SSFP”, 15 which we will call dephased SSFP (d‐SSFP) for simplicity. Dephased bSSFP (d‐bSSFP) has also been used in the past by some authors for delineation of interventional catheters and needles. 19 , 20 Some preliminary experimental work has been carried out to characterize the d‐bSSFP contrast and found it favorable in terms of achievable signal intensity compared with d‐FLASH. 21 However, d‐bSSFP has never been formally introduced or characterized, nor has the signal mechanism been compared with d‐FLASH or d‐SSFP.
In this work, we provide a comprehensive investigation of d‐bSSFP contrast. We introduce a signal model for d‐bSSFP and compare the mechanism of positive signal realization with the related contrast mechanisms of d‐FLASH and d‐SSFP. Using an aspiration biopsy needle as an example, we derive an analytical description of the device artifact that we validate in phantom measurements.
2. THEORY
Let G(t) be a magnetic field gradient that generates a linear field variation in ‐direction (without loss of generality), acting on local magnetization at time t during the repetition time (TR). At the echo time (TE), the gradient will have introduced a gradient moment . At the end of TR, effects of an acting gradient moment on the local magnetization can either be spoiled (disruption of transverse coherences) or balanced (zero net gradient moment at the end of TR), or remain unspoiled/unbalanced. For spoiled or balanced gradient moments, local phase at TE will depend linearly on . For unspoiled/unbalanced gradient moments, phase at TE relates to the absolute induced gradient moment across TR in a complex fashion that is dependent on T1, T2, and the flip angle. 22 For gradients acting uniformly across TR (in the sense that the absolute gradient moment at the end of TR equals ), Lebel et al. proposed to model the signal phase at TE with a simple square wave (SW) function (“shell model”). 23 The phase introduced at TE to the local magnetization at position for unspoiled/unbalanced as well as for spoiled or balanced gradients is therefore given by
| (1) |
where is the gyromagnetic ratio, , and τ is a discrete phase offset. Figure 1 shows how unspoiled/unbalanced and spoiled or balanced gradients, acting uniformly during TR, influence the phase at TE according to Eq. (1).
FIGURE 1.

Phase of local magnetization at TE for an unspoiled/unbalanced or a spoiled or balanced gradient moment acting along ‐direction with a constant phase offset . For spoiled or balanced gradients, the phase of the local magnetization at TE depends linearly on . For an unspoiled/unbalanced gradient, the local magnetization rephases at TE, either at 0 or π (“shell model” 23 , requiring an absolute induced gradient moment of at the end of TR).
The signal magnitude of a voxel with edge length and unit length in ‐ and ‐direction (neglecting relaxation effects) for an intravoxel phase distribution (in ‐direction) is generically described as follows 10 :
| (2) |
where rect is the rectangular function (rect(x) = 1 for ; rect(x) = 0.5 for x = ±0.5; rect(x) = 0 elsewhere), ϕ is the phase of the local magnetization and describes the (location‐independent) steady‐state signal magnitude per voxel. The value of depends on the type of steady state that is established for the sequence (coherent steady‐state for sequences without spoiling or incoherent steady‐state for sequences with spoiling 24 ). While both and depend on the tissue (spin density, T1, T2) and the chosen sequence parameters (TR, flip angle), also varies with the locally induced phase advance during TR. 22 In the context of the model and following Lebel et al., 23 we assume a constant coherent steady‐state signal magnitude , independent of the local phase advance during TR.
Seppenwoolde et al. investigated the emergence of bright contrast near metallic devices after the application of an external gradient . 9 They explained this effect, which they termed the WM phenomenon, as the consequence of local gradient moment compensation. A varying, susceptibility‐induced, magnetic field distortion near a metallic device results in a local susceptibility‐induced magnetic field gradient . For small voxels, we can linearize the induced field gradient (first‐order Taylor approximation of the magnetic field distortion). Considering an external WM gradient moment introducing a phase at TE, we now investigate an underlying susceptibility‐induced gradient moment (also assumed to act in ‐direction) and the additional external WM gradient acting simultaneously. We distinguish three cases: (i) and both being spoiled at the end of TR (d‐FLASH); (ii) and both being unspoiled/unbalanced at the end of TR (d‐SSFP); (iii) being unspoiled/unbalanced and being balanced at the end of TR (d‐bSSFP). The type of the acting gradients for the different contrast methods is summarized in Table 1. Introducing the phase relations from Eq. (1) into Eq. (2), depending on the type of the acting gradients, we can calculate the average voxel signal magnitudes , , and :
| (3) |
TABLE 1.
Type of acting gradient moments in dephased MRI sequences (d‐FLASH, d‐SSFP, and d‐bSSFP). Two gradient moments (susceptibility‐induced and WM gradient moment) act simultaneously on the local magnetization. While spoiled or balanced gradient moments will introduce a linear phase at TE, local magnetization will rephase with a phase of 0 or π for unspoiled/unbalanced gradient moments (“shell model” 23 , see Eq. [1]).
| Susceptibilty‐induced gradient moment | White‐Marker gradient moment | |
|---|---|---|
| d‐FLASH | Spoiled | Spoiled |
| d‐SSFP | Unspoiled/unbalanced | Unspoiled/unbalanced |
| d‐bSSFP | Unspoiled/unbalanced | Balanced |
While for d‐FLASH, a discrete phase offset is not considered when integrating the signal across the voxel (as it only adds a constant factor that does not affect integrated signal magnitude), we have to average across a discrete phase offset τ for d‐SSFP and d‐bSSFP to calculate the average signal magnitude. A closed solution for the signal equations (up to first signal maximum at for ) is given in the Appendix by Eq. (A1). Figure 2 shows the signal magnitude curves for the three investigated contrasts ( numerically continued for ). The result for is the well‐known sinc function. 10 The curve is shifted such that maximum signal appears at full gradient compensation (). For , the signal maximum equally appears at full gradient compensation. Unlike for d‐FLASH and d‐SSFP, two instead of one signal maxima emerge for d‐bSSFP. Signal is maximal for , but also for . While full and signal magnitude can be reached for d‐FLASH and d‐SSFP at maximum gradient compensation, only about 64% of the full signal can be achieved with d‐bSSFP according to the model (see Eq. [A1]). For better illustration of the signal formation at the signal maxima, Figure 3 shows the intravoxel phase for , as well as for d‐FLASH, d‐SSFP, and d‐bSSFP. While for d‐FLASH and d‐SSFP, phase coherence is only reestablished for and having opposite polarity, limited phase coherence is reestablished for d‐bSSFP independent of the respective gradient polarities.
FIGURE 2.

Average voxel signal magnitude for d‐FLASH, d‐SSFP, and d‐bSSFP contrast () for a varying susceptibility‐induced gradient moment . The signal magnitude is calculated in relation to the normalized maximum signal magnitude of the underlying steady state, i.e., for d‐FLASH and for d‐SSFP and d‐bSSFP.
FIGURE 3.

Intravoxel phase distribution of local magnetization at TE for d‐FLASH, d‐SSFP, and d‐bSSFP across one pixel (pixel edge length ) for the two cases: (expected signal maximum for d‐FLASH, d‐SSFP, and d‐bSSFP) and (expected additional signal maximum for d‐bSSFP). The phase is plotted for and acting independently (contributing phases following Eq. [1] with phase offset ) as well as for both gradients acting simultaneously (summed resulting phase). Phase contributions from spoiled or balanced gradient moments are drawn in (light‐) blue and from unspoiled/unbalanced gradient moments in (light‐) green (overlapping lines are drawn with a slight offset for better visibility). For d‐FLASH and d‐SSFP, adding WM–induced and susceptibility‐induced phases leads to complete restoration of phase coherence for . For d‐bSSFP, partial rephasing of the magnetization is achieved for .
3. METHODS
To evaluate d‐bSSFP contrast and compare it with d‐FLASH and d‐SSFP, we implemented three different research MR pulse sequences. Using a cylindrical aspiration needle as an example, we derived a description of the artifact (points of maximum signal intensity) with the introduced d‐bSSFP signal model. We verified it using an MR‐compatible needle placed in a water phantom.
3.1. MR pulse sequences
To generate d‐bSSFP, d‐FLASH, and d‐SSFP contrast, we implemented three research MR pulse sequences for 3D imaging (Figure 4B–D). A radial k‐space trajectory (Figure 4A) was chosen to mitigate potential distortion in the image caused by spatial frequency mismatch due to changed Larmor frequency near metallic devices (due to the changing frequency encoding direction of the acquired k‐space spokes, spatial frequency mismatch will lead to blurring rather than to cause discrete spatial shifts). 25 , 26 All three sequence types include a WM gradient, shifting the acquired k‐space in ‐direction with . To generate d‐FLASH contrast, the readout gradient is reverted at the end of each TR and a spoiler gradient inducing a dephasing of 2π per voxel is added in ‐direction with implemented RF spoiling. 27 , 28 To generate d‐SSFP contrast, an additional gradient moment equal to the WM gradient moment is played out after TE, leading to a net external gradient moment of at the end of TR. 15 Additionally, TE is fixed at TE = TR/2 to ensure that a net susceptibility‐induced gradient moment of acts across TR and Eq. (1) applies. To generate d‐bSSFP contrast, all played‐out gradient moments are balanced at the end of TR, including the WM gradient, and TE is again fixed at TE = TR/2. Image reconstruction using a NUFFT was implemented in Python with SigPy. 29 , 30
FIGURE 4.

Example implementation of three MR pulse sequences to acquire images with d‐FLASH (B), d‐SSFP (C), and d‐bSSFP (D) contrast following a three‐dimensional radial k‐space trajectory (A). In all three sequence types, a WM gradient (red) shifts the acquired k‐space in ‐direction with . For d‐FLASH contrast, all gradient moments are reverted, and a 2π‐spoiler gradient is played out in ‐direction at the end of TR (+ additional RF spoiling). For d‐SSFP contrast, an equal gradient moment to the WM gradient moment is played out at the end of TR. For d‐bSSFP, all gradient moments are balanced at the end of TR.
3.2. Contrast characterization for aspiration needle
Metallic needles introduce a local inhomogeneity in the magnetic field. Following Ladd et al., we modeled the needle shaft as an infinite cylinder, 7 placed along the ‐axis in a right‐handed coordinate system (Figure 5A). For a magnetic field that encloses an angle θ with the ‐axis, the local change in magnetic field magnitude ΔB susc is given by
| (4) |
where χ i is the magnetic susceptibility of the needle material itself, whereas χ e is the susceptibility of the surrounding material, and R is the radius of the biopsy needle. Calculating the derivative of Eq. (4), we can derive the local susceptibility‐induced magnetic field gradient moment as follows:
| (5) |
FIGURE 5.

(A) Schematic of the infinite cylinder model, representing a needle, oriented in ‐direction in a right‐handed coordinate system. The field encloses an angle with the needle. The WM gradient moment is oriented parallel to the field. (B) The needle induces a local magnetic field gradient in the respective ‐plane. The figure shows the needle piercing the image plane (black dot), and the orientation of the susceptibility‐induced field gradient and the WM field gradient (gray and red arrows, respectively). At radius from the needle center as defined in Eq. (8) (solid black line), the induced gradient moment magnitude from the susceptibility‐induced field distortion matches the ‐component of the WM gradient moment magnitude. Along the dashed black lines, the ‐component of the susceptibility‐induced gradient moment vanishes. Using Eq. (9), we find three points (yellow triangles) with (points of maximum signal for d‐FLASH, d‐SSFP, and d‐bSSFP) and three points (yellow crosses) with (additional points of maximum signal for d‐bSSFP).
We additionally introduce a WM gradient moment acting on the local magnetization at TE as follows:
| (6) |
As described in Section 2, maximum WM signal for d‐FLASH and d‐SSFP contrast will be reached for points where susceptibility‐induced and WM gradient moments have the same magnitude but opposite polarity. Hence, maximum signal magnitude will be reached when , as any susceptibility‐induced gradient moment in ‐direction cannot be compensated by a WM gradient as defined in Eq. (6), and the needle‐induced gradient moment magnitude is fully compensated by the WM gradient moment: (the ‐component of induces a global dephasing and will always be non‐zero for needle orientations with a component parallel to ). For d‐bSSFP contrast, maximum WM signal is equivalently reached for points where susceptibility‐induced and WM gradient have the same magnitude and polarity: . The criterion for maximum WM signal can therefore be expressed by the following system of equations:
| (7) |
We solve Eq. (7) for a WM gradient induced phase of . While the condition is fulfilled for and , we find to be true on a concentric circle in the ‐plane with radius as follows:
| (8) |
We therefore find the location of the WM maxima in the ‐plane as follows:
| (9) |
Figure 5B shows the induced local gradient field for a cylindrical needle in the ‐plane and the WM gradient field as an overlay. The ‐component of the susceptibility‐induced gradient moment vanishes along the dashed black lines. Along the solid black line with the radius , the magnitude of the susceptibility‐induced gradient moment is equal to the magnitude of the z‐component of the WM gradient moment. Three intersection points () can be identified where the susceptibility‐induced gradient moment has the opposite polarity of the WM gradient moment (yellow triangles). At these points, highest WM signal can be achieved for d‐FLASH, d‐SSFP and d‐bSSFP. Three additional points (, , ) can be identified, where gradients have the same polarity (yellow crosses). At these points, additional WM signal maxima in the ‐plane are expected for d‐bSSFP. In three dimensions, the WM needle artifact therefore consists of either three or six bright stripes flanking the needle shaft. For the maximum signal magnitude, a ratio of about 0.64 between d‐bSSFP and d‐SSFP contrast is expected (see Eq. [A1]). Due to the higher symmetry of the d‐bSSFP needle artifact compared with the d‐SSFP artifact, the d‐bSSFP artifact contains twice as many pixels with maximum signal; thus, we expect a cumulative signal magnitude ratio twice as large (≈1.27). For a comparison of d‐FLASH with d‐SSFP and d‐bSSFP, a general prediction of the maximum signal magnitude ratio or the cumulative signal magnitude ratio is not feasible, as two different steady states are reached (coherent for d‐SSFP and d‐bSSFP, incoherent for d‐FLASH). While the maximum signal magnitude for the incoherent steady‐state signal depends on T1 and TR (maximized for Ernst angle), maximum coherent steady‐state signal magnitude is T1‐dependent and T2‐dependent, 31 making the signal intensity ratio between d‐FLASH and d‐SSFP, as well as between d‐FLASH and d‐bSSFP, tissue‐dependent.
3.3. Measurements
To verify the introduced model and compare the different investigated contrast techniques, a 20‐gauge (R = 0.04059 cm) MR‐compatible aspiration needle (Cook Medical, Bloomington, IN, USA), made from Inconel (Nickel‐Chromium alloy with magnetic susceptibility χ i = 568 × 10−6), 32 was placed in a flask phantom (at an angle θ = 90° to as defined in Figure 5A) filled with deionized water (magnetic susceptibility χ e = −9.05 × 10−6). 33 Images with d‐FLASH, d‐SSFP, and d‐bSSFP contrast were acquired using the implemented sequences (see Section 3.1). The phantom was imaged at B 0 = 0.55 T (MAGNETOM Free.Max; Siemens Healthineers, Erlangen, Germany) using a 12‐channel head coil and the following sequence parameters: field of view = (128 mm)3; res = (2 mm)3; TR = 40 ms; TE = 20 ms, bandwidth = 900 Hz/px, flip angle = 70° (d‐SSFP, d‐bSSFP) / 10° (d‐FLASH, Ernst angle as determined from a parameter sweep with step size of 2°). An additional reference image of the needle phantom, using a spin‐echo sequence, was acquired with the following sequence parameters: field of view = (128 mm)3; res = (1 mm)3; TE = 152 ms; TR = 3000 ms; bandwidth = 501 Hz/px. Maximum artifact signal magnitude (averaged across 3 or 6 points of maximum artifact signal magnitude in the ‐plane for a slab of 10 slices) was compared between the different contrast mechanisms and, for the ratio between d‐bSSFP and d‐SSFP, compared with the theoretical prediction (see Section 3.2). Additionally, the cumulated artifact signal magnitude was evaluated (summed signal of 3 or 6 points of maximum artifact signal magnitude in the ‐plane, averaged across a slab of 10 slices) and, for d‐SSFP and d‐bSSFP, again compared with the theoretically expected ratio. To further evaluate the accuracy of the model, d‐bSSFP contrast was acquired with varying TE (TE1 = 3.4 ms, TE2 = 10 ms, TE3 = 20 ms, TE4 = 30 ms, TE5 = 40 ms, TE6 = 50 ms), and the radius of the needle artifact was determined from the images by averaging over the distance from the true needle position as extracted from the spin‐echo reference and the position of the pixels with maximum artifact magnitude (assumed measurement uncertainty of one pixel, corresponding to half of the resolution). The measured artifact radius was then compared with the artifact radius predicted from Eq. (8) to evaluate the model's accuracy.
4. RESULTS
Figure 6 shows a reference spin‐echo image of the flask phantom, along with the d‐FLASH, d‐SSFP, and d‐bSSFP images. While for d‐FLASH and d‐SSFP only three signal maxima emerge in the ‐plane (three stripes flanking the needle shaft, axisymmetric artifact with respect to ‐axis), six maxima (point‐symmetric artifact) are visible for d‐bSSFP. While the artifact appears spatially homogeneous for d‐FLASH and d‐bSSFP, multiple dark bands appear in d‐SSFP that break up the artifact (banding).
FIGURE 6.

A 20‐gauge needle was placed in a phantom bottle filled with deionized water (orthogonal to ). A reference spin‐echo image (‐plane, ‐plane, and ‐plane as defined in Figure 5A; needle axis indicated by dotted line) as well as images acquired with d‐FLASH, d‐SSFP, and d‐bSSFP are shown (‐plane view is slightly offset from the needle axis for the dephased contrast compared with the reference to depict the artifact). An additional zoomed view on the artifact is displayed for the ‐plane view. For a slab of 10 slices (between white solid lines drawn in the ‐plane and ‐plane images), we calculated the mean signal averaged across the three, respectively six, pixels with maximum artifact signal in each slice (see Table 2).
The maximum artifact magnitude for d‐bSSFP and d‐FLASH was found to be (0.56 ± 0.09) and (0.19 ± 0.04) relative to d‐SSFP (Figure 6). The theoretical signal ratio for d‐bSSFP and d‐SSFP (≈0.64) is therefore found to be within one standard deviation of the experimental value. The measured cumulated signal magnitude for d‐bSSFP and d‐FLASH was (1.13 ± 0.07) and (0.19 ± 0.01) relative to d‐SSFP. The theoretical cumulative signal ratio for d‐bSSFP and d‐SSFP (≈1.27) is therefore found to be within two standard deviations of the experimental value. Directly comparing d‐bSSFP and d‐FLASH, d‐bSSFP exhibits a (2.9 ± 0.5) and a (5.9 ± 0.4) times larger maximum artifact signal and cumulated signal magnitude than d‐FLASH.
Figure 7 compares the needle artifact for different TEs. d‐bSSFP images for four example TEs are shown in Figure 7A. Some stripe‐like artifacts can be seen in the water phantom that appear unrelated to the needle (see discussion in Section 5). Needle WM artifact size increased with increasing TE. The true location of the needle shaft was extracted from the reference spin‐echo image and added to the d‐bSSFP images (white dot). Dashed and solid lines represent the presumed location of vanishing and , respectively, according to the introduced model and corresponding to Figure 5B. Pixels with maximum intensity can be seen to lie on the dashed lines with a maximum deviation of one pixel. The measured artifact radius was plotted against TE in Figure 7B. The artifact radius as predicted by the model is also plotted. The measured artifact radius was in accordance with the model (within measurement uncertainty) for all investigated TEs.
FIGURE 7.

(A) The needle phantom (compare with Figure 6) was imaged with the d‐bSSFP sequence at six different echo times: TE1 = 3.4 ms, TE2 = 10 ms, TE3 = 20 ms, TE4 = 30 ms, TE5 = 40 ms, and TE6 = 50 ms. Slices corresponding to the ‐plane and ‐plane of the needle coordinate system, as defined in Figure 5A, are displayed for TE1,2,4,6. An additional zoomed view on the artifact is displayed for the ‐plane view. The estimated location where (compare with Figure 5B) is indicated with a white circle with the radius r, calculated as the average distance between the reference needle position and pixels of maximum signal. Some stripe‐like artifacts (orange arrow) can be seen in the water phantom (see discussion in Section 5). (B) The measured radius is plotted against TE (with half‐pixel measurement uncertainty). Additionally, the modeled radius from Eq. (8) is plotted using the literature values for the magnetic susceptibility of Inconel and water.
Results, comparing the three investigated contrast types (exerted artifact symmetry, sensitivity to banding artifacts, theoretically expected and measured signal magnitude), are summarized in Table 2.
TABLE 2.
Comparison of the investigated contrast mechanisms (d‐FLASH, d‐SSFP, and d‐bSSFP) comprising the exhibited symmetry, the sensitivity to banding artifacts, and the maximum as well as the cumulated artifact signal magnitude (normalized to d‐SSFP magnitude for theoretical model prediction and measurements, given with standard deviation).
| d‐FLASH | d‐SSFP | d‐bSSFP | |
|---|---|---|---|
| Symmetry | Axial | Axial | Point |
| Sensitive to banding artifacts | No | Yes | No |
| Maximum artifact magnitude (normalized to d‐SSFP) | Model: n.a. | Model: 1 | Model: 0.64 |
| Exp.: 0.19 ± 0.04 | Exp.: 1.0 ± 0.1 | Exp.: 0.56 ± 0.09 | |
| Cumulated artifact magnitude (normalized to d‐SSFP) | Model: n.a. | Model: 1 | Model: 1.27 |
| Exp.: 0.19 ± 0.01 | Exp.: 1.00 ± 0.04 | Exp.: 1.13 ± 0.07 |
5. DISCUSSION
In this work, d‐bSSFP was investigated as a positive, susceptibility‐based contrast method and compared with the related techniques d‐FLASH and d‐SSFP. A signal model was established and used to describe the artifact of a cylindrical metallic needle. Artifact shape, size, and signal magnitude as predicted by the model were compared with phantom measurements.
The d‐bSSFP signal model predicts, in contrast to d‐FLASH and d‐SSFP, a point‐symmetric artifact. Matching the theoretical result, the higher artifact symmetry was confirmed in phantom measurements and can be explained by the unique combination of a discrete and a continuous phase evolution, as demonstrated in Figure 3. In this paper, the artifact shape was evaluated for the example geometry of a metallic cylinder, but a point‐symmetric artifact is equally expected to emerge for spherical devices, such as fiducial markers, as they also induce a symmetric field distortion. 6 A point‐symmetric artifact might provide some benefits for applications of a positive contrast technique, e.g., for localization tasks. The artifact appears more homogeneously around the device due to the additional symmetry (six instead of three signal maxima as shown in Figure 5), which might make it easier to detect the artifact center when viewing the images. The artifact center could, for example, be identified by simply connecting maximum signal points that are diagonally opposite of each other and taking the intersection. Also, the additional signal maxima in the d‐bSSFP artifact increase redundancy, which could, in particular, pose an advantage if the artifact is investigated in heterogeneous tissue instead of a homogeneous water phantom. As the focus of this paper remains the analysis of the d‐bSSFP contrast mechanism itself, further analysis of potential benefits in applications, such as localization tasks, shall be addressed in future work.
As described in Section 3.1, the three investigated research pulse sequences (d‐FLASH, d‐SSFP, and d‐bSSFP) were implemented to follow a radial acquisition scheme, as it is inherently immune to spatial signal shifts that can occur in the proximity to magnetic disturbers. For a potential Cartesian implementation, any possible distortions of the artifacts could be counteracted by deliberately choosing a high receiver bandwidth. The WM gradient moment in the investigated pulse sequences was implemented such that it acted in the direction. In general, a different direction for the WM gradient moment could also be chosen, and the position of the artifact signal maxima could be similarly determined from an adapted Eq. (7). Although orientation of the artifact can change with the direction of the WM gradient moment, no fundamental change in the artifact's general appearance and symmetry is expected, as the investigated signal model is independent of the WM orientation.
While in Figure 6 the artifact appears homogeneous for d‐FLASH and d‐bSSFP, dark banding artifacts interrupt the artifact for d‐SSFP. This sensitivity of the d‐SSFP contrast to banding can be explained from the mechanism of contrast formation. For d‐SSFP, the unbalanced susceptibility‐induced gradient is counterbalanced by an equally unbalanced external WM gradient as described in Section 2. At points of maximum gradient compensation, no net gradient remains, and the sequence therefore mimics the classic balanced SSFP sequence type. The signal can be said to locally “morph” itself into a bSSFP echo. 15 , 34 This automatically makes the artifact prone to banding artifacts that appear in bSSFP. In d‐bSSFP, banding artifacts are not an issue, as the local magnetization at points of maximum signal compensation experiences an unbalanced susceptibility gradient purposely introducing a phase flip in all voxels. A much smaller background gradient due to imperfect shimming will only lead to a negligible change in the unbalanced gradient. The sensitivity to banding puts d‐SSFP at a disadvantage compared with d‐FLASH and d‐bSSFP, as artifact homogeneity is crucial for the visualization and delineation of a device.
For d‐SSFP and d‐bSSFP, the ratio of the maximum artifact signal and the cumulated artifact signal was theoretically determined using the introduced models. Due to the different types of established steady‐state magnetization ( and ), the maximum signal magnitude ratio between d‐FLASH and d‐SSFP, as well as d‐FLASH and d‐bSSFP, is tissue dependent, and a general prediction is not feasible (see Section 3.2). For all three investigated contrast methods, maximum and cumulated signal magnitude were compared experimentally. Generally, a higher signal level can be reached with SSFP imaging techniques than with RF‐spoiled GRE (FLASH) imaging techniques. 31 , 35 This is reflected accordingly by the measured maximum artifact signal magnitude being lowest for d‐FLASH (Table 2). For d‐bSSFP, a slightly lower maximum (within one standard deviation) and cumulative signal ratio (within two standard deviations) were measured experimentally than were predicted by the model. This observation can potentially be attributed to the model's assumptions of a spatially constant gradient and a constant coherent steady‐state signal independent of the locally induced phase advance during TR, which are not perfectly matched by reality. Equation (2) furthermore assumes a uniform voxel sensitivity function. Due to the nature of the discrete Fourier transform used in MRI, imaged magnitude is however not sampled equally from the original distribution of transverse magnetization throughout the voxel. 36 This might cause an additional deviation of the measurement from the model in terms of a reduced measured signal magnitude, as the measured signal might not fully represent a phase coherence that is established by the WM gradient across the complete voxel. However, the cumulated artifact signal being highest for d‐bSSFP, as theoretically predicted, was experimentally confirmed and might, similar to the higher artifact symmetry, prove useful in potential applications of the positive contrast, as it could improve visibility of devices or SPIO‐labeled cells.
For further validation of the signal model, d‐bSSFP needle artifact size was investigated for six different TEs. For the investigated TEs, the measured artifact radius was well in accordance with the curve in Figure 7B representing the relationship in Eq. (8), as all measured values matched the predicted radius within the margin of error. Stripe‐like artifacts in the water phantom are visible in some of the d‐bSSFP images in Figure 7A. The origin of these artifacts, which appear to be unrelated to the needle due to the spatial distance, is not clear, but they could be eddy current–related and emerge due to the applied radial readout trajectory. 37 In this case, a reduction in artifact magnitude would be expected for a Cartesian readout trajectory.
To evaluate the different contrast methods, signal models for d‐FLASH, d‐SSFP, and d‐bSSFP were investigated. While the signal magnitude for d‐FLASH is described by the well‐known relation, 10 the investigated signal models for d‐SSFP and d‐bSSFP have, to the best of our knowledge, not yet been introduced in the literature. As described in Section 2, the models predict the signal for a pixel that experiences a certain dephasing due to spoiled or balanced and unspoiled/unbalanced gradients acting on the local magnetization either solely or jointly. Due to the discrete nature of the established phase for unspoiled/unbalanced gradients, the d‐SSFP and d‐bSSFP signal also depend on the local discrete phase offset, and we calculated an average voxel signal magnitude. For a spatially‐constant unspoiled/unbalanced gradient, a modulation of signal magnitude across multiple voxels, depending on the local phase offset, would be expected (appearance of bright bands for d‐bSSFP that appear closer to each other for more effective gradient compensation). Effects of this discrete behavior could, however, not be seen in the conducted experiments. As, for the investigated needle geometry, the magnitude of the susceptibility‐induced gradient decreases quickly (with as derived from Eq. [4]) with increasing distance from the device, no bands are formed.
In this work, we studied d‐bSSFP contrast and compared it with the related contrast techniques d‐FLASH and d‐SSFP. A signal model for d‐bSSFP was established and used for artifact prediction of a cylindrical aspiration needle. Compared with the related techniques d‐FLASH and d‐SSFP, we found the d‐bSSFP needle artifact to present with higher artifact symmetry and higher cumulated artifact signal. Compared with the d‐SSFP artifact, the d‐bSSFP artifact showed robustness against banding artifacts. The properties of the d‐bSSFP contrast encourage further studies to investigate the application of the contrast technique in interventional MRI and potentially in imaging of SPIO‐labeled cells or fiducial marker localization, as they could render d‐bSSFP beneficial compared with d‐FLASH and d‐SSFP for delineation and localization tasks.
CONFLICT OF INTEREST
Jonas Frederik Faust, Peter Speier, Axel Joachim Krafft, Sunil Patil, Ravi Teja Seethamraju, and Florian Maier are employees of Siemens Healthineers.
ACKNOWLEDGMENTS
The authors thank Dr. Jesse Roll of Cook Advanced Technologies for providing the needle.. Open Access funding enabled and organized by Projekt DEAL.
APPENDIX A.
A.1.
For the signal equations set up in Eq. (3), we can find closed solutions. For the d‐FLASH signal equation, the solution is the well‐known sinc function. 10 For d‐SSFP and d‐bSSFP, we can determine the solution to the signal equations using geometrical arguments (here given up to first signal maximum at for ):
| (A1) |
Here, is the rectangular function ( for ; for ; elsewhere); is the triangular function (); and describes the modulo operation. We therefore find for the case of maximum signal magnitude.
Faust JF, Speier P, Krafft AJ, et al. Positive susceptibility‐based contrast imaging with dephased balanced steady‐state free precession. Magn Reson Med. 2025;94:59‐72. doi: 10.1002/mrm.30421
REFERENCES
- 1. Rothgang E, Gilson WD, Wacker F, Hornegger J, Lorenz CH, Weiss CR. Rapid freehand MR‐guided percutaneous needle interventions: an image‐based approach to improve workflow and feasibility. J Magn Reson Imaging. 2013;37:1202‐1212. [DOI] [PubMed] [Google Scholar]
- 2. Bock M, Wacker FK. MR‐guided intravascular interventions: techniques and applications. J Magn Reson Imaging. 2008;27:326‐338. [DOI] [PubMed] [Google Scholar]
- 3. Weiss CR, Nour SG, Lewin JS. MR‐guided biopsy: a review of current techniques and applications. J Magn Reson Imaging. 2008;27:311‐325. [DOI] [PubMed] [Google Scholar]
- 4. Veltri A, Bargellini I, Giorgi L, Almeida PAMS, Akhan O. CIRSE guidelines on percutaneous needle biopsy (PNB). Cardiovasc Intervent Radiol. 2017;40:1501‐1513. [DOI] [PubMed] [Google Scholar]
- 5. Lewin JS, Duerk JL, Jain VR, Petersilge CA, Chao CP, Haaga JR. Needle localization in MR‐guided biopsy and aspiration: effects of field strength, sequence design, and magnetic field orientation. Am J Roentgenol. 1996;166:1337‐1345. [DOI] [PubMed] [Google Scholar]
- 6. Lüdeke KM, Röschmann P, Tischler R. Susceptibility artefacts in NMR imaging. Magn Reson Imaging. 1985;3:329‐343. [DOI] [PubMed] [Google Scholar]
- 7. Ladd ME, Erhart P, Debatin JF, Romanowski BJ, Boesiger P, McKinnon GC. Biopsy needle susceptibility artifacts. Magn Reson Med. 1996;36:646‐651. [DOI] [PubMed] [Google Scholar]
- 8. Posse S, Aue WP. Susceptibility artifacts in spin‐echo and gradient‐echo imaging. J Magn Reson Imaging. 1990;88:473‐492. [Google Scholar]
- 9. Seppenwoolde JH, Viergever MA, Bakker CJG. Passive tracking exploiting local signal conservation: the white marker phenomenon. Magn Reson Med. 2003;50:784‐790. [DOI] [PubMed] [Google Scholar]
- 10. Bakker CJG, Seppenwoolde J, Vincken KL. Dephased MRI. Magn Reson Med. 2006;55:92‐97. [DOI] [PubMed] [Google Scholar]
- 11. Stuber M, Gilson WD, Schär M, et al. Positive contrast visualization of iron oxide‐labeled stem cells using inversion‐recovery with ON‐resonant water suppression (IRON). Magn Reson Med. 2007;58:1072‐1077. [DOI] [PubMed] [Google Scholar]
- 12. Dharmakumar R, Koktzoglou I, Li D. Generating positive contrast from off‐resonant spins with steady‐state free precession magnetic resonance imaging: theory and proof‐of‐principle experiments. Phys Med Biol. 2006;51:4201‐4215. [DOI] [PubMed] [Google Scholar]
- 13. Dahnke H, Liu W, Herzka D, Frank JA, Schaeffter T. Susceptibility gradient mapping (SGM): a new postprocessing method for positive contrast generation applied to superparamagnetic iron oxide particle (SPIO)‐labeled cells. Magn Reson Med. 2008;60:595‐603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Vonken EPA, Schär M, Yu J, Bakker CJG, Stuber M. Direct in vitro comparison of six three‐dimensional positive contrast methods for susceptibility marker imaging: comparison of positive contrast methods. J Magn Reson Imaging. 2013;38:344‐357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Patil S, Bieri O, Scheffler K. Echo‐dephased steady state free precession. Magn Reson Mater Phys Biol Med. 2009;22:277‐285. [DOI] [PubMed] [Google Scholar]
- 16. Zijlstra F, Viergever MA, Seevinck PR. SMART tracking: simultaneous anatomical imaging and real‐time passive device tracking for MR‐guided interventions. Phys Med. 2019;64:252‐260. [DOI] [PubMed] [Google Scholar]
- 17. Mani V, Briley‐Saebo KC, Itskovich VV, Samber DD, Fayad ZA. Gradient echo acquisition for superparamagnetic particles with positive contrast (GRASP): sequence characterization in membrane and glass superparamagnetic iron oxide phantoms at 1.5T and 3T. Magn Reson Med. 2006;55:126‐135. [DOI] [PubMed] [Google Scholar]
- 18. Hargreaves B. Rapid gradient‐echo imaging. J Magn Reson Imaging. 2012;36:1300‐1313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Campbell‐Washburn AE, Rogers T, Xue H, Hansen MS, Lederman RJ, Faranesh AZ. Dual echo positive contrast bSSFP for real‐time visualization of passive devices duringmagnetic resonance guided cardiovascular catheterization. J Cardiovasc Magn Reson. 2014;16:88. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Weine J, Schneider R, Kägebein U, Hensen B, Wacker F, Maier F. Interleaved white marker contrast with bSSFP real‐time imaging for deep learning based needle localization in MR‐guided percutaneous interventions. Proceedings of the 27th Annual Meeting of ISMRM, 11‐16 May 2019, International Society for Magnetic Resonance in Medicine, Montréal, QC, Canada; 2019:3832. [Google Scholar]
- 21. Reiß S, Krafft AJ, Düring K, von zur Mühlen C, Bock M. To spoil or to balance? A comparison of the white marker phenomenon in gradient Echo pulse sequences. In: Proceedings of the 23rd Annual Meeting of ISMRM, 30 May ‐ 5 June 2015, Toronto, ON, Canada, 2015:4160. [Google Scholar]
- 22. Scheffler K, Hennig J. Is TrueFISP a gradient‐echo or a spin‐echo sequence? Magn Reson Med. 2003;49:395‐397. [DOI] [PubMed] [Google Scholar]
- 23. Lebel RM, Menon RS, Bowen CV. Relaxometry model of strong dipolar perturbers for balanced‐SSFP: application to quantification of SPIO loaded cells. Magn Reson Med. 2006;55:583‐591. [DOI] [PubMed] [Google Scholar]
- 24. Norman Cheng Y, Haacke EM. Short‐TR, coherent, gradient echo imaging. Curr Protoc Magn Reson Imaging. 2001;2:B5.2.1–B5.2.12. [Google Scholar]
- 25. Chan RW, Ramsay EA, Cunningham CH, Plewes DB. Temporal stability of adaptive 3D radial MRI using multidimensional golden means. Magn Reson Med. 2009;61:354‐363. [DOI] [PubMed] [Google Scholar]
- 26. Brown RW, Cheng YN, Haacke EM, Thompson MR, Venkatesan R. Magnetic Resonance Imaging: Physical Principles and Sequence Design. 1st Edition. Hoboken, New Jersey, USA: Wiley; 2014. [Google Scholar]
- 27. Leupold J, Hennig J, Scheffler K. Moment and direction of the spoiler gradient for effective artifact suppression in RF‐spoiled gradient echo imaging. Magn Reson Med. 2008;60:119‐127. [DOI] [PubMed] [Google Scholar]
- 28. Zur Y, Wood ML, Neuringer LJ. Spoiling of transverse magnetization in steady‐state sequences. Magn Reson Med. 1991;21:251‐263. [DOI] [PubMed] [Google Scholar]
- 29. Pipe JG, Menon P. Sampling density compensation in MRI: rationale and an iterative numerical solution. Magn Reson Med. 1999;41:179‐186. [DOI] [PubMed] [Google Scholar]
- 30. Ong F, Lustig M. SigPy: a python package for high performance iterative reconstruction. Proceedings of the 27th Annual Meeting of ISMRM, 11–16 May 2019, Montréal, QC, Canada; 2019:4819. [Google Scholar]
- 31. Bieri O, Scheffler K. Fundamentals of balanced steady state free precession MRI: fundamentals of balanced SSFP MRI. J Magn Reson Imaging. 2013;38:2‐11. [DOI] [PubMed] [Google Scholar]
- 32. Griebel AJ, Anttila E, Baker G, Schaffer JE, Gross DC. Magnetic susceptibility of common metals and alloys used in medical devices. J Test Eval. 2023;51: 4554–4566. [Google Scholar]
- 33. Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys. 1996;23:815‐850. [DOI] [PubMed] [Google Scholar]
- 34. Bieri O, Patil S, Quick HH, Scheffler K. Morphing steady‐state free precession. Magn Reson Med. 2007;58:1242‐1248. [DOI] [PubMed] [Google Scholar]
- 35. Scheffler K, Lehnhardt S. Principles and applications of balanced SSFP techniques. Eur Radiol. 2003;13:2409‐2418. [DOI] [PubMed] [Google Scholar]
- 36. Parker DL, Du YP, Davis WL. The voxel sensitivity function in Fourier transform imaging: applications to magnetic resonance angiography. Magn Reson Med. 1995;33:156‐162. [DOI] [PubMed] [Google Scholar]
- 37. Scheffler K, Hennig J. Eddy current optimized phase encoding schemes to reduce artifacts in balanced SSFP imaging. In: Proceedings of the 11th Annual Meeting of ISMRM, 10–16 July 2003, Toronto, ON, Canada, 2003:294. [Google Scholar]