Skip to main content
. 2002 Mar;40(3):881–885. doi: 10.1128/JCM.40.3.881-885.2002

TABLE 2.

Performance characteristics of EIAs in different patient groups

Characteristica Valueb
Ortho
Organon
Calypte
Group A Group B Group A Group B Group A Group B
Sensitivity (%) 100 100 NA 100 100 100
Specificity (%) 100 100 NA 97.3 95.1 98.7
NPV 1.0 1.0 NA 1.0 1.0 1.0
PPV 1.0 1.0 NA 0.25 0.96 0.4
a

Sensitivity, true positives/(true positives + false negatives); specificity, true negatives/(true negatives + false positives); NPV, negative predictive value [true negatives/(true negatives + false negatives)]; PPV, positive predictive value [true positives/(true positives + false positives)].

b

NA, not applicable.