Skip to main content
Proceedings (Baylor University. Medical Center) logoLink to Proceedings (Baylor University. Medical Center)
. 2025 Mar 17;38(3):278–284. doi: 10.1080/08998280.2025.2474907

Clinical utilization and culture concordance of categorical and semiquantitative concentration values on the BioFire® Pneumonia Panel at a major academic quaternary referral center

Tiana R Endicott-Yazdani a,b,c,✉, Chris Boseovski d, Neelam Dhiman e, Yun X Ying e, Susan K Mathai a,b,c, Todd J Grazia a,b,c, Raul Benavides d
PMCID: PMC12026127  PMID: 40291090

Abstract

Background

Pneumonia mortality can be decreased by early antibiotic administration. Pathogen identification aims to minimize inappropriate, nontargeted antibiotic exposure. Molecular assays expedite organism identification through nucleic acid detection and antimicrobial resistance by screening for genetic markers.

Methods

We evaluated concordance of organism identification, resistance markers, and semiquantitative results between a Food and Drug Administration–approved molecular diagnostic test and traditional culture methods. We performed a retrospective analysis of BioFire® Pneumonia Panel (PN Panel) orders during a 2-month period.

Results

Organism identification was 97% concordant between paired culture and polymerase chain reaction (PCR) detection. Probability of growth in culture varied proportionally with the “semiquantitative” PN Panel result, with only 4% of organisms with 104 copies/mL growing in culture, versus 53% of organisms with 107 copies/mL growing in culture. Additionally, in 2.5% of cases, the PN Panel identified an organism that did not grow from culture. In comparison, 0.1% of paired organisms were detected by culture but were not seen by BioFire PCR. Concordance of resistance detection with various culture-based methods was 99%.

Conclusion

Combining PN Panel and culture results can maximize early, targeted resistance detection and organism treatment, and the semiquantitative result is a proxy for the probability of growth in culture and the clinical burden of each organism.

Keywords: BioFire, colonization, concordance, pneumonia, resistance


Pneumonia is a leading cause of death in the United States, with significant associated morbidity, particularly in the elderly and immunocompromised populations.1,2 It accounted for 1.4 million emergency room visits and over 41,108 deaths in the US in 2022.1 Pneumonia is typically the result of bacterial or viral infection, and the two are often indistinguishable based on clinical presentation alone. Low-risk patients are often treated empirically with antibiotics, while those with additional risk factors are treated with broad-spectrum antibiotics that are subsequently narrowed as additional data become available.3 In many cases, specific organisms causing the pneumonia are not identified.

The American Thoracic Society and the Infectious Diseases Society of America recommend that patients who require hospitalization should have a respiratory sample collected and sent for bacterial culture.3 Culture techniques are slow, often unreliable, and can lead to prolonged antibiotic exposure in the absence of organism identification; a specific pathogen is identified in <40% of cases.4–6 Rapid identification of causative organisms theoretically allows for narrowing and possibly removal of isolation precautions, and molecular identification can expedite organism identification.7,8 The BioFire® FilmArray® Pneumonia Panel (hereafter, “PN Panel”) is a rapid method to identify causative pneumonia pathogens through the identification of bacterial or viral DNA and RNA.9–11 The panel includes eight viruses and three atypical bacteria and provides semiquantitative results for 15 bacteria in under 1 hour with <5 minutes of hands-on time. Additionally, it screens for seven resistance genes that impact antibiotic choices: methicillin resistance (MRSA) genes (mecA/C. MREJ), carbapenemase genes KPC (blaKPC), NDM (blaNDM), Oxa-48-like (blaOxa-48 family), VIM (blaVIM), IMP (blalIMP), and extended spectrum beta-lactamase gene CTX-M (blaCTX-M) (Table 1).

Table 1.

Bacteria and resistance genes detected by BioFire® FilmArray® Pneumonia Panel*

Bacteria Resistance genes
Acinetobacter calcoaceticus-baumanni complex Pseudomonas aeruginosa Methicillin:
Enterobacter cloacae complex Serratia marcescens mecA/C and MREJ
Escherichia coli Staphylococcus aureus Carbapenemases:
Haemophilus influenzae Streptococcus agalactiae KPC VIM
Klebsiella aerogenes Streptococcus pneumoniae NDM IMP
Klebsiella oxytoca Streptococcus pyogenes Oxa-48-like
Klebsiella pneumoniae group Chlamydia pneumoniae ESBL:
Moraxella catarrhalis Legionella pneumophila CTX-M
Proteus spp Mycoplasma pneumoniae  
*

All bacteria listed are reported with semi-quantitative results except for Chlamydia pneumoniae, Legionella pneumophila, and Mycoplasma pneumoniae.

Several studies have evaluated how the PN Panel performs in the detection of organisms and concordance with traditional culture methods.9–12 There is a reported 60% to 70% increase in organism detection via the PN Panel compared to traditional culture.10,11 However, due to the high sensitivity of the PN Panel, there is risk of overdetection of nonviable or “colonizing” bacteria.

We performed a retrospective chart review of inpatient and outpatient PN Panel bacterial results over a 2-month period at Baylor University Medical Center (BUMC) to evaluate real-world usage of the PN Panel in a mixed-specialty setting. We evaluated concordance of the PN Panel and bacterial culture results and resistance detection and assessed if the results of the PN Panel testing were used in clinical care to modify antimicrobial choice. Semiquantitative bin data results were also compared to the levels of eventual growth on culture of the specific organism.

METHODS

This study was approved by the BUMC institutional review board (Protocol #020-346). We performed a retrospective chart review of BioFire® FilmArray Pneumonia Panel (PN Panel) orders performed at BUMC between December 27, 2019 and February 29, 2020. All PN Panel orders were placed prior to the start of the retrospective chart review. The reviewed period predated the detection of SARS-CoV-2 in the North Texas area (March 2020); therefore, no patient in this study was suspected of having SARS-CoV-2 infection. This study focused on bacterial pathogen reporting utilization; therefore, viral results of the PN Panel were not evaluated. Samples collected were categorized as expectorated (including sputum and tracheal aspirate) or bronchoalveolar lavage (BAL) and processed according to the methods listed below.

To evaluate concordance of bacterial organism detection and antimicrobial resistance markers, results were tabulated by unique organisms or resistance genes identified and not by number of panels ordered. For example, a PN Panel report consisting of 18 discrete bacterial detection results that returned with two organisms detected would be counted as two positive results and 16 negative results, as each organism or antibiotic resistance marker identified in a sample has the potential for concordance. Secondly, a subanalysis of concordance between the genetic antibiotic resistance markers detected in BioFire PN Panel compared to culture-based methods was performed.

PN Panel orders that did not have a corresponding culture sent were excluded from concordance analysis; however, for utilization in antibiotic escalation or deescalation, all PN Panel orders were chart reviewed for actions by ordering physicians.

BioFire® filmarray Pneumonia Panel Protocol

The PN Panel was performed on the BioFire® Torch Array system (bioMérieux). The patient sample (BAL, tracheal aspirate, sputum) was directly inoculated along with buffer into a loading pouch, hydrated, and amplified via multiplex real-time polymerase chain reaction (PCR). Identification was then performed via high-resolution melt-curve analysis. For those organisms that were quantitative, the instrument calculated an approximate concentration for each target based on real-time PCR amplification data relative to an internal reference standard.

Bacterial culture protocol

For bacterial culture, a sample was inoculated onto appropriate plates for routine bacteria including blood agar, chocolate agar, and MacConkey agar plates as per standard laboratory policy and read at 24 and 48 hours. For respiratory pathogen growth, Gram stain was performed, and identification was obtained using the MALDI-TOF Vitek MS instrument. Susceptibility and minimum inhibitory concentration (MIC) of each pathogen were determined using Vitek 2 and/or Microscan. Questionable resistance patterns were confirmed with ETEST MIC and/or disk diffusion.

RESULTS

For concordance evaluation, PN Panel orders that did not have a corresponding bacterial culture order at the same approximate time (N = 37) were excluded from analysis, leaving 283 PN Panel orders that accounted for 5094 organism identification results compared to growth on culture (Figure 1). A total of 166 BAL results and 117 sputum (expectorated and tracheal aspirate) results were evaluated. Results demonstrated concordance of 97.4%, consisting of 56 detected results (1.1%) and 4908 nondetected results (96.3%). Discordant results consisted of 130 (2.5%) PN Panel–detected organisms negative on culture, and 4 (0.1%) culture-based detections negative on PN Panel. Of the 186 detected bacteria results on PN Panel, only 56 (30%) also grew out on culture.

Figure 1.

Figure 1.

(a) Organisms identified in either PCR only or culture only are listed to the side with the number of occurrences in parentheses. Due to the possibility of multiple organisms in a single specimen, the number of occurrences per organism on the left and right sides do not necessarily add up to the corresponding number of specimens in the main chart. (b) The number of unique organisms detected (0 to 4) in the BioFire® FilmArray® Pneumonia Panel and if the sample was collected by either bronchoalveolar lavage or sputum.

An additional “per specimen/organism” analysis of concordance rates was performed post hoc on the same data to better capture the experience of utility on a per submitted specimen basis. To account for organisms detected only in culture, these occurrences were considered an additional sample with discordant “negative PN Panel and positive culture” status. For example, a single physical sample could potentially be counted more than once if both (a) an organism was detected by the PN Panel but not culture and also (b) a separate organism grew in culture but was not detected on the PN Panel. This led to a total sample increase from 281 to 303 (Figure 1a). The concordance was 62.1% (either both positive or both negative), with 31.4% PN Panel-only detections and 6.6% culture-only detections.

The most detected organisms via PN Panel included three fastidious growers (Haemophilus influenzae, N = 24; Streptococcus pneumoniae, N = 21; and Moraxella catarrhalis, N = 12) and four nonfastidious growers (Pseudomonas aeruginosa, N = 19; Staphylococcus aureus, N = 16; Escherichia coli, N = 9; and Klebsiella pneumoniae, N = 8). Rarely, bacteria grew exclusively in culture: S. pneumoniae (N = 2), H. influenzae (N = 1), and S. aureus (N = 1) (Figure 1a). BAL results were more likely to have no organisms detected via PN Panel than sputum results, and detection of multiple targets was more likely in expectorated sputum samples (Figure 1b).

The PN Panel utilizes real-time PCR and reports semiquantitative bins for estimates of nucleic acid present (104, 105, 106, and 107 copies/mL) for 15 organisms (Figure 2). Concordance was evaluated by the categorization into semiquantitative bins and corresponding rate of positive cultures in each semiquantitative bin. Analysis revealed minimal concordance at bin 104, with only 2 of 48 (4.2%) PN Panel detections also seen in culture (Table 2). However, in bin 107, growth was seen in 32 of 60 PN Panel detections, for a concordance of 53.3% (Table 2). Two frequent airway respiratory pathogens, P. aeruginosa and S. aureus, did not grow in culture when detected by PN Panel at bin 104 (0/9 and 0/4, respectively), but were seen in culture in 86% (13/15) and 63% (7/11) of cultures, respectively, when in the PN Panel bin 107 (Table 2). Overall, of the 184 detected pathogens that could be semiquantified by the BioFire PN Panel (which excludes two atypical bacteria that were detected), only 56 of those organisms also grew in culture, for an overall concordance rate of 30%.

Figure 2.

Figure 2.

Panel schematic indicating 104 copies/mL as the lowest bin of detection and bin 107 copies/mL as the highest. The bins are representative of the nucleic acid for a given organism present in a sample during real-time PCR.

Table 2.

Semiquantitative Pneumonia Panel bin results for organisms (copies/mL) detected in the 320 samples

Organism Bin 104
Bin 105
Bin 106
Bin 107
PCR CX PCR CX PCR CX PCR CX
Acinetobacter baumanni 0 0 0 0 0 0 0 0
Enterobacter cloacae complex 2 0 2 1 1 0 0 0
Escherichia coli 6 0 1 0 2 1 4 2
Haemophilus influenzae 4 0 6 0 5 0 11 2
Klebsiella aerogenes 1 0 0 0 0 0 1 1
Klebsiella oxytoca 1 0 0 0 0 0 0 0
Klebsiella pneumoniae group 6 1 3 1 0 0 5 2
Moraxella catarrhalis 4 0 3 0 4 2 2 0
Proteus mirabilis 0 0 2 0 0 0 0 0
Pseudomonas aeruginosa 9 0 6 2 8 4 15 13
Serratia marcescens 2 0 2 0 0 0 2 1
Staphylococcus aureus 4 0 10 3 7 5 11 7
Streptococcus agalactiae 3 0 1 0 1 1 0 0
Streptococcus pneumoniae 6 1 8 0 2 0 8 2
Streptococcus pyogenes 0 0 0 0 2 0 1 1
Total 48 2 44 7 32 13 60 31
Concordance 4.2% 15.9% 40.6% 51.7%

CX indicates culture; PCR, BioFire® FilmArray® Pneumonia Panel polymerase chain reaction.

Fifty-four sputum samples were retrospectively analyzed for possible oral contamination by reviewing the Gram stain results. Of the 54 samples, 6 (11.1%) samples resulted with >25 squamous epithelial cells per low power field, which is generally accepted as consistent with possible oral contamination. Four of these samples (67%) had multiple specific organisms identified by PN Panel, but the routine culture results were only reported as “normal respiratory flora.” In the remaining two samples, three organisms total were detected on PN Panel but none grew on culture, and only one had a semiquantitative bin >105 (P. aeruginosa, bin = 106). Of note, 6 of the 8 (75%) organisms with discrepancies were in the lower semiquantitative bin category (104–105).

PN panel antibiotic resistance screening includes seven distinct genomic markers, which in this study produced 1981 distinct antibiotic resistance genomic marker results within the 283 PN Panel orders (Table 3). Detection of genetic markers of antibiotic resistance by PCR versus demonstration of similar resistance pattern on culture-based method was concordant 99.2% of the time (1966/1981). PCR detection of resistance genes not demonstrated on culture-based methods accounted for the remaining 0.8% (15/1981) of results (Table 4).

Table 3.

Concordance of BioFire® FilmArray® Pneumonia Panel and culture results when any resistance was detected

Resistance +PCR, -CX +PCR, +CX -PCR, +CX Concordance
CTX 3 3 0 50%
KPC 0 1 0 100%
mecA/B and MREJ 11 8 0 42%
NDM 1 0 0 0%
Other 0 0 21 n/a
Total 15 12 21  
Concordance 31.2% 25% 43.7%  

CX indicates culture; PCR, BioFire® FilmArray® Pneumonia Panel polymerase chain reaction.

Table 4.

Concordance of BioFire® FilmArray® Pneumonia Panel and culture semiquantitative bin results for organisms (copies/mL) when resistance was detected via PCR

Resistance Bin 104
Bin 105
Bin 106
Bin 107
PCR CX PCR CX PCR CX PCR CX
CTX 1 0 0 0 2 1 3 2
KPC 0 0 1 1 0 0 0 0
mecA/B and MREJ 3 0 4 1 4 3 8 4
NDM 1 0 0 0 0 0 0 0
Total 5 0 5 2 6 4 11 6
Concordance 0% 40% 66.7% 54.5%

Finally, we reviewed the clinical effects of the data from this diagnostic testing by examining antibiotic escalation or deescalation after receipt of testing results. Evaluation was on all 320 PN Panel orders, regardless of whether a corresponding culture was ordered. Antibiotic escalation was defined as adding an antibiotic, broadening coverage, or covering more resistant bacteria. Antibiotic deescalation was defined as removing an antibiotic or narrowing coverage. Based on review of the clinicians’ responses to the results, the PN Panel result led to overall change (deescalation and escalation) in antibiotic regimens 40% of the time. Deescalation of antibiotics was observed 25.6% (82/320) of the time and escalation of antibiotics was observed in 14.4% (46/320) of subjects.

DISCUSSION

This study demonstrates that combining PN Panel and culture results can maximize early pathogen and targeted resistance detection that can aid change in patient management. The semiquantitative result may be used to estimate the probability of growth in culture and deduce the clinical burden of the organism.

Lower respiratory tract infection is the fourth leading cause of death worldwide and a common cause of sepsis.13 Early initiation of antibiotics for pneumonia has become standard of care, as studies have shown increased mortality when antibiotics are delayed more than 4 hours.14,15 PCR detection of microorganisms is a sensitive method for detection for select bacteria and viruses. Previous studies have evaluated the performance of the PN Panel in the setting of hospitalized patients and those in the intensive care unit (ICU). The current study evaluated all PN Panel results from inpatient and outpatient settings with respiratory samples (sputum, tracheal aspirate, and BAL) over a 2-month period at BUMC prior to the SARS-CoV2 pandemic, with concordance analysis on 283 samples that also had a simultaneously ordered bacterial culture.

The total percentage of PN Panel orders that demonstrated any bacterial detections reported in our study, 44% (125/283), is similar to other published reports.10,11 The PN Panel exclusively identified an organism in approximately 30% of specimens (Figure 1a). This number is lower than other published reports, which could be due to differences in study populations: previous studies evaluated hospitalized or ICU patients, while we evaluated samples from a diverse group of hospitalized patients, ICU patients, outpatients, and transplant surveillance encounters.10,11 Thus, a lower rate of organism detection is not unexpected in this more diverse population.

A potential benefit of the semiquantitative bins in PCR detection is the potential to differentiate noninfectious or inactive bacterial ‘detections’ from those causing active infection, either because of low nonpathologic concentrations or detection of residual DNA/RNA from treated/dead organisms. Colonization in this article is defined as bacterial presence without causing disease,16 whereas infection is defined as tissue invasion by an organism.17 Indeed, our study suggests that organisms with a low semiquantitative bin result (104) may correspond to a lower pathogen concentration, which could reflect colonization and not an acute infectious process. Similarly, in our study, organisms with a high semiquantitative bin result (107) were more likely to grow in culture. However, this comparison may be more complicated than concordance of simple semiquantification results to culture results, as fastidious organism are inherently difficult to grow, even at higher colony counts.4,9,18 In addition, for some organisms such as Legionella pneumoniae, special media is needed, and that may not be ordered for all applicable patients. For fastidious organisms identified by the PN Panel, validating the semiquantitative results for all organisms as a tool to differentiate colonization versus active infection remains difficult. Finally, in the setting of respiratory infectious symptoms, a positive but low quantitative bin result, especially in the absence of any other identified pathogen, is probably still highly likely to represent the causative pathogen.

It is true that DNA copy numbers detected by the assay do not directly correlate with culture colony-forming units (CFU/mL). However, several previous reports have noted reasonable concordance of BioFire PN Panel and culture results based on semiquantitative thresholds, and the relative abundance as well as the specific pathogens detected may help with clinical interpretation. In the study of Falsey et al, S. pneumoniae, H. influenzae, and M. catarrhalis—common respiratory pathogens—were associated with higher genomic copies per milliliter than organisms often considered colonizers, such as S. aureus, Group B Streptococcus, and Gram-negative rods.19 They also reported that high serum procalcitonin and bacterial adjudication were more often associated with sputa with 106 or 107 copies detected, suggesting an association with invasive disease. Buchan et al particularly found discordance at a lower level of quantification of 104 copies/mL by PN Panel, suggesting that targets quantified as 104 copies/mL by the PN Panel may frequently be quantified as 103 CFU/mL by routine culture. The clinical significance of additional culture-negative low-concentration (104 copies/mL) detections by the PN Panel requires further investigation, especially given that these detections have historically gone largely unreported based on current culture and reporting guidelines. Further, approximately 80% culture-negative/PN Panel–detected results are mostly attributed to empirical antibiotics prior to specimen collection. A negative culture may result in premature discontinuation of therapy, while correct identification may enable deescalation/targeted therapy.10 Murphy et al showed that the PN Panel had better performance at the limit of detection for monomicrobial cultures. In contrast, in polymicrobial cultures, the presence of low quantities of significant pathogens may be masked by large numbers of other pathogens or commensals.9 While culture remains the gold standard in the diagnosis of bacterial respiratory tract infections, it may be difficult to accurately recover all pathogens in clinical samples, as the organisms are in a complex matrix. In addition, culture results would be more affected by host immune response and prior antibiotic usage.

There remains, however, the potential to use semiquantitative bin results on nonfastidious growers. For example, S. aureus and P. aeruginosa are well-established nonfastidious bacteria that are known to cause both “colonization” in the respiratory tract and active pulmonary disease. Testing for S. aureus by PN Panel had a corresponding positive culture at a rate of 0% and 63.6% at semiquantitative bins 104 and 107, respectively. Testing for P. aeruginosa had a similar pattern of 0% and 86.7% at semiquantitative bins 104 and 107, respectively. Overall, in this study, only 4.2% of samples from the lowest quantitative bin had a corresponding positive culture result. These results raise the possibility that for positive bacterial results accompanied by quantitation, those from semiquantitative bin 104 may represent merely the presence of bacteria or residual bacterial DNA and not active infection. Determining definitively if this is the case would require additional clinical study. However, our findings are consistent with previous recommendations suggesting ≥105 CFUs as the quantitative cutoff suggesting active infection.17,18,20

The high sensitivity of the PN Panel in the setting of low organism abundance (i.e., colonization) could lead to overtreatment. If the 104 bin represents detection in an otherwise noninfected individual, clinicians responding to a “detected” result in this semiquantitative bin could cause unnecessary antibiotic exposure and antibiotic escalation. In the present study, antibiotic escalation occurred in 46 instances. In 9 of 46 instances, it was because of a positive 104 bin result from S. aureus and P. aeruginosa testing (three and six escalations, respectively). Thus, if 104 represents a noninfectious process for those two organisms, this observation could indicate possible inappropriate antibiotic escalation if used without proper clinical review.

This was a retrospective study examining the results of the PN Panel testing at a large quaternary referral center and included samples from a diverse study population that included both inpatients and outpatients, mixed sample types (sputum, tracheal aspirate, and BAL), and varying severity of disease (including extracorporeal membrane oxygenation patients). A weakness of the study was not evaluating the viral results of the PN Panel compared to other diagnostic techniques. We elected not to evaluate this measure as our facility does not have the laboratory methods readily available to routinely culture viruses or confirm via alternative PCR for all viruses in all samples due to financial and laboratory constraints. Furthermore, diagnostic methods that are used to evaluate viral infection vary widely, which means our results may not have represented real-life performance across different settings. In addition, at the time of the study, sputum samples were not routinely screened via Gram stain before being run on BioFire PN Panel. Overall, 1.1% of sputum samples retrospectively analyzed had Gram stains consistent with contamination, which may have affected performance.

In summary, this paper is a real-world application of the PN Panel with both inpatient and outpatient samples, the first reported to our knowledge. We recommend that if a clinician wants to use the PN Panel, a mandatory policy of concurrent respiratory culture should be applied, as there are many potential pathogens (Enterococcus spp., Stenotrophomonas maltophila, Nocardia spp., etc.) that are not targets in the PN Panel. Additionally, resistance patterns may be detected by culture that are not targets of the PN Panel. The PN Panel should remain a way to expedite organism identification and antibiotic adjustments in the setting of a physician’s clinical judgment, but largely serves in its current form as an adjunct to traditional culture methods.

Funding Statement

The study was funded by the Center for Advanced Heart and Lung Disease, Baylor University Medical Center, Dallas, Texas.

Disclosure statement/Funding

The authors report no conflicts of interest.

References

  • 1.National Center for Health Statistics . Pneumonia. Centers for Disease Control and Prevention; last reviewed July 31, 2024. https://www.cdc.gov/nchs/fastats/pneumonia.htm.
  • 2.Xu J, Murphy SL, Kochanek KD, Bastian BA.. Deaths: final data for 2013. Natl Vital Stat Rep. 2016;64(2):1–119. [PubMed] [Google Scholar]
  • 3.Metlay J, Waterer G, Long A, et al. Diagnosis and treatment of adults with community-acquired pneumonia. An official clinical practice guideline of the American Thoracic Society and Infectious Diseases Society of America. Am J Respir Crit Care Med. 2019;200(7):e45–e67. doi: 10.1164/rccm.201908-1581ST. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Jain S, Self WH, Wunderink RG, et al. Community-acquired pneumonia requiring hospitalization among U.S. adults. N Engl J Med. 2015;373(5):415–427. doi: 10.1056/NEJMoa1500245. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chalmers J, Taylor J, Singanayagam A, et al. Epidemiology, antiobiotic therapy and clinical outcomes in health care-associated pneumonia: A UK cohort study. Clin Inf Dis. 2011;53(2):107–113. doi: 10.1093/cid/cir274. [DOI] [PubMed] [Google Scholar]
  • 6.Musher D, Roig I, Cazares G, Stager C, Logan N, Safar H.. Can an etiologic agent be identified in adults who are hospitalized for community-acquired pneumonia: results of a one-year study. J Infect. 2013;67(1):11–18. doi: 10.1016/j.jinf.2013.03.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Templeton K, Scheltinga S, Sillekens P, et al. Development and clinical evaluation of an internally controlled single-tube multiplex real-time PCR assay for detection of Legionella pneumophila and other Legionella species. J Clin Microbiol. 2003;41(9):4016–4021. doi: 10.1128/JCM.41.9.4016-4021.2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Templeton K, Scheltinga S, Graffelman A, et al. Comparison and evaluation of real-time PCR, real-time nucleic acid sequence-based amplification, conventional PCR and serology for diagnosis of Mycoplasma pneumoniae. J Clin Microbiol. 2003;41(9):4366–4371. doi: 10.1128/JCM.41.9.4366-4371.2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Murphy C, Fowler R, Balada-Llasat J, et al. Multicenter evaluation of the BioFire FilmArray Pneumonia/Pneumonia Plus Panel for detection and quantification of agents of lower respiratory tract infection. J Clin Microbiol. 2020;58(7):e00128-00120. doi: 10.1128/JCM.00128-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Buchan B, Windham S, Balada-Llasat J, et al. Practical comparison of the BioFire FilmArray Pneumonia panel to routine diagnostic methods and potential impact on antimicrobial stewardship in adult hospitalized patients with lower respiratory tract infections. J Clin Microbiol. 2020;58(7):e00135–00120. doi: 10.1128/JCM.00135-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lee S, Ruan S, Pan S, Lee T, Chien J, Hsueh P.. Performance of a multiplex PCR pneumonia panel for the identification of respiratory pathogens and the main determinants of resistance from the lower respiratory tract speciments of adult patients in intensive care units. J Microbiol Immunol Infect. 2019;52(6):920–928. doi: 10.1016/j.jmii.2019.10.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Webber D, Wallace M, Burnham C, Anderson N.. Evaluation of the BioFire FilmArray pneumonia panel for detection of viral and bacterial pathogens in lower respiratory tract specimens in the setting of a tertiary care academic medical center. J Clin Microbiol. 2020;58(7):e00343-00320. doi: 10.1128/JCM.00343-20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.World Health Organization. The top 10 causes of death. August 7, 2024. https://www.who.int/news-room/fact-sheets/detail/the-top-10-causes-of-death.
  • 14.Mandell L, Wunderink R, Anzueto A, Infectious Diseases Society of America; American Thoracic Society , et al. Infectious Diseases Society of America/American Thoracic Society consensus guidelines on the management of community-acquired pneumonia in adults. Clin Infect Dis. 2007;44 (Suppl 2):S27–S72. doi: 10.1086/511159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Lim W, Woodhead M, British Thoracic Society . British Thoracic Society adult community acquired pneumonia audit 2009/10. Thorax. 2011;66(6):548–549. doi: 10.1136/thoraxjnl-2011-200081. [DOI] [PubMed] [Google Scholar]
  • 16.Olendorf D, Jeryan C, Boyden K, eds. Gale Encyclopedia of Medicine. Gale; 2008. [Google Scholar]
  • 17.Musher D, Jesudasen S, WBarwatt J, Cohen D, Moss B, Rodriguez-Barradas M.. Normal respiratory flora as a cause of community-acquired pneumonia. Open Forum Infect Dis. 2020;7(9):ofaa307. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Torres A, Lee N, Cilloniz C, Vila J, Van der Eerden M.. Laboratory diagnosis of pneumonia in the molecular age. Eur Respir J. 2016;48(6):1764–1778. doi: 10.1183/13993003.01144-2016. [DOI] [PubMed] [Google Scholar]
  • 19.Falsey AR, Branche AR, Croft DP, Formica MA, Peasley MR, Walsh EE.. Real-life assessment of biofire filmarray pneumonia panel in adults hospitalized with respiratory illness. J Infect Dis. 2024;229(1):214–222. doi: 10.1093/infdis/jiad221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Thorsteinsson S, Musher D, T F.. The diagnostic value of sputum culture in acute pneumonia. JAMA. 1975;233(8):894–895. doi: 10.1001/jama.1975.03260080056024. [DOI] [PubMed] [Google Scholar]

Articles from Proceedings (Baylor University. Medical Center) are provided here courtesy of Baylor University Medical Center

RESOURCES